Model Number ELITE 6500
Device Problems Thermal Decomposition of Device (1071); Smoking (1585)
Patient Problem No Information (3190)
Event Date 02/12/2014
Event Type  malfunction  
Event Description
During the start of a surgical case, while positioning the anesthetized patient, the anesthesiologist noticed an electrical burning smell.The doctor noted burning and charring of the alternating current (ac) line cord where it entered the base of the surgery table.The doctor pulled and released the ac power cord from the table and notified the surgeon.The case proceeded with the operative room (or) table in its powered off position.The table was removed from service after the case and the biomedical department was notified to investigate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SKYTRON ELITE 6500
Type of Device
TABLE, OPERATING-ROOM, AC-POWERED
Manufacturer (Section D)
SKYTRON
5085 corporate exchange blvd s
grand rapids MI 49512
MDR Report Key3676500
MDR Text Key4260250
Report Number3676500
Device Sequence Number1
Product Code FQO
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Type of Report Initial
Report Date 02/14/2014
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received02/14/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberELITE 6500
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/14/2014
Device Age20 YR
Event Location Hospital
Date Report to Manufacturer03/13/2014
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age25 YR
-
-