BRACCO INJENNERING SA FASTLOAD CTA; INJECTOR SYSTEM Back to Search Results
Model Number 6730
Device Problems Break (1069); Collapse (1099); Detachment of Device or Device Component (2907); Torn Material (3024)
Patient Problem No Patient Involvement (2645)
Event Date 07/22/2015
Event Type  malfunction  
Event Description
Product defect in manufacturing of the fastload cta dual syringe pack used in the empower cta and empower cta injector systems.Two of three packages had the locking lugs on the syringe broken off.One of three had the low pressure connecting y-tube blue cap crushed and torn off.These all occurred in the same lot number.None were used on any patients, however, the broken locking lugs on the syringes made them unusable.None of the other dual packs in stock with this lot number had the same defect.Attempts were made to duplicate the error by dropping the packages from 5 feet to the floor.There was no damage.Fastload cta dual syringe pack with spikes contains: 2x 200ml fastload syringes, 1 low pressure connecting t tube, 1 small spike and 1 large spike.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FASTLOAD CTA
Type of Device
INJECTOR SYSTEM
Manufacturer (Section D)
BRACCO INJENNERING SA
lausanne
SZ 
MDR Report Key4949666
MDR Text Key6633926
Report NumberMW5044858
Device Sequence Number1
Product Code DXT
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/23/2015
3 Devices were Involved in the Event: 1   2   3  
1 Patient was Involved in the Event
Date FDA Received07/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2018
Device Model Number6730
Device Catalogue Number017347
Device Lot NumberIADA4-1505
Is the Reporter a Health Professional? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-