Event occurred in (b)(6).Post-op (late), cerebral bleeding, consisting of left frontal lobe hemorrhage/ subcortical hemorrhage of left perinatal lobe, intracerebral hemorrhage with ventricular rupture, resulting in patient death.Anticoagulant-related.Inr at event = 11.96.On (b)(6) 2015, onxlti received a set of data from the sponsor of a (b)(6) post-market study, where-in they wanted on-x to do data analysis on a set of data containing adverse events.The valve on this mdr is one of the ae's.(b)(6).There are no statistically significant differences in these rates and this trial was run at inr of 1.5 to 2.5 versus the pma at 2.0 to 3.0.At lower inr lo rates are expected based on our research.So these events while mdr reportable as individual occurrences, they are overall in the range expected.
|