• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN CARDIO THORACIC UNKNOWN; SHUNT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN CARDIO THORACIC UNKNOWN; SHUNT Back to Search Results
Model Number UNKNOWN CARDIO THORACIC
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Death (1802)
Event Type  Death  
Manufacturer Narrative
Submit date (b)(6) 2016.An investigation is currently under way; upon completion the results will be forwarded.(b)(4).Several attempts to gather information from the customer were made.To date, no additional information has been received.If additional pertinent information becomes available, the report will be updated.
 
Event Description
It was reported to covidien on (b)(6) 2016 that a customer had an issue with an artery shunt.The investigator reported there was a patient death while the argyle shunt was in place.This was discovered during a literature review.
 
Manufacturer Narrative
This complaint was discovered during a literature review.The product and lot number involved in this event is unknown.This complaint has not been confirmed.The complaint sample was not returned to the manufacturing site for review.No lot number was provided, therefore, it is not possible to perform a device history record (dhr) review.All dhr¿s are reviewed for accuracy prior to a products release.Without the sample or additional information, it is not possible to establish if there is a failure mode related to manufacturing or the device.The source literature for this event was reviewed and the death could not be attributed to the shunt or a shunt complication.A brainstorming session was performed in order to identify the possible causes for the described event.The following potential causes were identified: non device related organ failure, non-device related complications during placement procedure, incorrect catheter size selection, or contaminated catheter.Based on the information in the mentioned literature, no causes could be confirmed to be linked to manufacturing activities or to the device.The complaint sample was not returned to the manufacturing site for review.Should the sample be returned in the future, this complaint will be re-opened for further investigation.Since the lot number was not provided and the sample was not returned, it is not possible to identify the specific products manufacturing date.Based on the reviewed literature related to this complaint, the device functioned as intended during the required time and it concludes shunting should be considered a viable treatment option.Based on the available information, the most probable causes are non-device related organ failure and non-device related complications during placement procedure.Per management discretion, an (b)(4) was initiated to perform a risk assessment linked to this complaint report.No further actions are required at this time.It must be noted that in-process controls, such as personnel training, incoming quality acceptance testing for raw material, in process visual inspection and visual acceptance sampling, and environmental and product biological monitoring, are in place to prevent nonconforming product from leaving the manufacturing operations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIO THORACIC UNKNOWN
Type of Device
SHUNT
Manufacturer (Section D)
COVIDIEN
covidien manufacturing solutions sa
edificio 820 calle #2 zona france coyol
alajuela 20101
CS  20101
Manufacturer (Section G)
COVIDIEN MANUFACTURING SOLUTIONS
edificio 820 calle #2
zona franca coyol
alajuela 20101
CS   20101
Manufacturer Contact
thom mcnamara
15 hampshire st.
mansfield, MA 02048
5084524811
MDR Report Key5402032
MDR Text Key37240653
Report Number3009211636-2016-00058
Device Sequence Number1
Product Code KDQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/11/2016
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received02/01/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUNKNOWN CARDIO THORACIC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-