Model Number E SERIES |
Device Problems
Self-Activation or Keying (1557); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "pacer fault 122" message and restarted/rebooted on its own.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical canada evaluated the device and the reported malfunctions were not replicated or confirmed.The device was put through extensive testing without duplicating the malfunctions.Review of the device activity logs did not find evidence to support the reported event.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
|