Model Number ISOLINE 2CR6 |
Device Problems
Nonstandard Device (1420); Failure to Analyze Signal (1539)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/04/2018 |
Event Type
Injury
|
Event Description
|
It was reported that several episodes with ventricular noise oversensing probably related to a lack of lead insulation were observed in the device memory.The internal cardioverter defibrillation (icd) replacement is imminent, the physician would like to know if the ventricular lead must be replaced.Recommendation to replace the ventricular lead were provided on (b)(6) 2018.
|
|
Event Description
|
It was reported that several episodes with ventricular noise oversensing probably related to a lack of lead insulation were observed in the device memory.The internal cardioverter defibrillation (icd) replacement is imminent, the physician would like to know if the ventricular lead must be replaced.Recommendation to replace the ventricular lead were provided on (b)(6) 2018.
|
|
Search Alerts/Recalls
|