• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 3 LEFT VENTRICULAR ASSIST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 3 LEFT VENTRICULAR ASSIST SYSTEM Back to Search Results
Model Number 106524
Device Problems Detachment Of Device Component (1104); Nonstandard Device (1420)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/11/2018
Event Type  malfunction  
Manufacturer Narrative
The heartmate 3 lvas was implanted during the (b)(6) clinical trial, ide (b)(4).Fda approval for heartmate 3 lvas was received on 23 august 2017.The same device is used commercially and in the ongoing (b)(6) trial.(b)(4).Approximate age of device - 11 months.Abbott has decided to initiate a voluntary field action for all lot numbers of heartmate iii distributed since september 2014.Internal investigation performed by abbott has determined there is a potential for an outflow graft occlusion due to twisting.Abbott performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification determined that bend relief rotation is inconsistently translated to the outflow graft hardware.A risk/benefit analysis was performed and it was determined that despite the potential for an outflow graft occlusion due to twisting, the associated residual risks are low and are considered acceptable.As a corrective action, consignees have been notified of the potential occlusion due to twisting.Additionally, abbott will continue to trend complaints related to this event based on original event description and/or results of product evaluation.The patient remains on lvad support.No additional information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
 
Event Description
The patient was implanted with a left ventricular assist device (lvad) on (b)(6) 2017.It was reported that review of recent chest x-rays revealed a possible outflow graft bend relief separation.There was no reported adverse patient impact due to the event.No additional information was provided.
 
Manufacturer Narrative
The review of the post-op x-rays of the lvad, revealed that the sealed outflow graft bend relief may not be fully engaged.These x-rays appear to show the bend relief hardware tilted out of alignment with respect to the sealed outflow graft hardware, indicating that the bend relief was not attached.The evaluation could not conclusively determine a root cause for the disconnected outflow graft bend relief.The patient was implanted after the corrective action implemented to address the heartmate 3 sealed outflow graft bend relief disconnection issue.The patient remains ongoing and no additional events have been reported at this time.A review of the device history records revealed the device met applicable specifications.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LEFT VENTRICULAR ASSIST SYSTEM
Type of Device
LEFT VENTRICULAR ASSIST SYSTEM
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 95488
MDR Report Key7657458
MDR Text Key113285061
Report Number2916596-2018-02598
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Remedial Action Recall
Type of Report Initial,Followup
Report Date 09/11/2018
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2020
Device Model Number106524
Device Catalogue Number106524
Device Lot Number6016077
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/24/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1774-2018
Patient Sequence Number1
Patient Age77 YR
Patient Weight100
-
-