Coaguchek reported calibration problems with their int (pt) strips having inaccurate readings per letter (b)(6) dated (b)(6) 2018.Their solution was to send out more strips that were still not calibrated correctly.This resulted in several invalid readings of 7.5 and 7.9, well past the 4.5 allowable reading.This required a trip to the (b)(6) hospital, associated with (b)(6).This induced panic, emergency room visit to verify the results were valid or not, per the aforementioned letter from coaguchek.This problem should have been resolved in the beginning instead of sending more strips that had the same quality problem.This is inexcusable.They sent inr strips with a known problem to save their money, while passing the cost on to the (b)(6) hospital.
|