| Date Initiated by Firm | August 10, 2026 |
| Date Posted | September 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3294-2026 |
| Recall Event ID |
99759 |
| Product Classification |
Hysterectomy kit - Product Code OJF
|
| Product | Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays
ABDOMINAL HYSTERECTOMY - 205989
Model No. MHAH55AM
LAVH-LAP COLON
Model No. SPLL54U
VAG HYST PACK
Model No. SPVH62P |
| Code Information |
MHAH55AM
UDI-DI 191072236487
Lots 42-8529911
42-8649211
42-8687711
42-8797811
42-8857011
42-8869711
SPLL54U
UDI-DI 191072239181
Lots 07-2516111
07-2528611
07-2545611
07-2547911
07-2561411
SPVH62P
UDI-DI 191072238771
Lots 07-2544611
12-7436711
|
| FEI Number |
3008492337
|
Recalling Firm/ Manufacturer |
American Contract Systems Inc
85 Shaffer Park Dr
Tiffin
OH
44883-9290
|
| For Additional Information Contact | Marlene Jones 419-455-2090 |
Manufacturer Reason for Recall | Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI). |
FDA Determined Cause 2 | Process change control |
| Action | Firm began notifying affected distributors via email on August 10, 2026. Distributors were instructed to notify end-users.
Customers were instructed to identify, segregate, and quarantine all impacted products. Customers are to add recall labels to affected kits to ensure the impacted component is discarded prior to use.
An amended customer letter was sent out on August 28, 2026, which added the following instruction:
Users should source alternative cleansing wipes for the impacted product lots and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters. |
| Quantity in Commerce | 1,804 kits total |
| Distribution | US Nationwide distribution in the states of Iowa, Minnesota, Wisconsin, and South Dakota. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|