| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
065
|
Date of Entry 12/22/2025
|
|
FR Recognition Number
|
2-308
|
| Standard | |
ASTM F1984-25 Standard Practice for Testing for Whole Complement Activation in Serum by Solid Materials |
|
Scope/Abstract1.1 This practice provides a protocol for rapid, in vitro screening for whole complement activating properties of solid materials used in the fabrication of medical devices that will contact blood. 1.2 This practice is intended to evaluate the acute in vitro whole complement activating properties of solid materials intended for use in contact with blood. For this practice, the words "serum" and "complement" are used interchangeably (most biological supply houses use these words synonymously in reference to serum used as a source of complement). 1.3 This practice consists of two procedural parts. Procedure A describes exposure of solid materials to a standard lot of human serum, using a 0.1 mL serum/13 x 100 mm disposable test tube. Cellulose acetate powders and fibers are used as examples of test materials. Procedure B describes assaying the exposed serum for significant functional whole complement depletion as compared to control samples. 1.4 This practice does not address function, elaboration, or depletion of individual complement components, nor does it address the use of plasma as a source of complement. 1.5 This practice is one of several developed for the assessment of the biocompatibility of materials. Practice F748 may provide guidance for the selection of appropriate methods for testing materials for other aspects of biocompatibility.
|
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ASTM F1984-99 (Reapproved 2018) [Rec# 2-141] will be superseded by recognition of ASTM F1984-25 [Rec# 2-308]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 2-141] until December 19, 2027. After this transition period, declarations of conformity to [Rec# 2-141] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| 21 CFR Part 58 Good Laboratory Practice for Nonclinical Laboratory Studies |
|
Relevant FDA Guidance and/or Supportive Publications*
ISO 10993-12 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials.
Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices--Part 1: Evaluation and testing within a risk management process", Issued September 2023.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contacts
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
| *These are provided as examples and others may be applicable. |