| | Class 2 Device Recall Inspire Medical Systems, Inc. |  |
| Date Initiated by Firm | May 11, 2026 |
| Date Posted | June 18, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2515-2026 |
| Recall Event ID |
98990 |
| PMA Number | P130008 |
| Product Classification |
Stimulator, hypoglossal nerve, implanted, apnea - Product Code MNQ
|
| Product | Brand Name: Inspire Medical Systems, Inc.
Product Name: Inspire Respiratory Sensing Lead
Model/Catalog Number: 4340
Software Version: N/A
Product Description: The Inspire Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG).
Component: N/A |
| Code Information |
Lot Code: Model No 4340 (USA only lead); UDI-DI [4340] 10855728005946; Inspire part number [4340] 900-014-003; serial numbers T18806, T45012, T45017, T49002, T67098, T67152, T69541, T69542, T69553, T69558, T69562
Model No 4340 (Global lead); UDI-DI [4340] 10810098650301; Inspire part number [4340] 900-014-005; serial numbers T91485, T91486, T91491, T91495, T91513, T91520, T91521, T91527, T91533, T91537, T91540, T91545, T96165, T96168, T96192, T96193, T96195, T96225, T96237, T98118, T98126, T98136, T98137, T98138, T98141, T98143, T98145, T98150, T98153, T98156, T98163, T98165, T98176, T98177, T98178, T98181, T98183, T98190, T96217
|
| FEI Number |
3007666314
|
Recalling Firm/ Manufacturer |
Inspire Medical Systems Inc.
5500 Wayzata Blvd Ste 1600
Golden Valley
MN
55416-1237
|
| For Additional Information Contact | 763-205-7974 |
Manufacturer Reason for Recall | Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery. |
FDA Determined Cause 2 | Process control |
| Action | Firm began notifying consignees on May 11, 2026. Letters were titled "URGENT Medical Device Correction."
The firm requests consignees' assistance in removing affected product from clinical inventory and returning them to Inspire. |
| Quantity in Commerce | 50 impacted units |
| Distribution | Worldwide distribution - US Nationwide and the country of Hong Kong. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = MNQ
|
|
|
|