| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
065
|
Date of Entry 12/22/2025
|
|
FR Recognition Number
|
2-311
|
| Standard | (Included in ASCA) |
ISO 10993-4 Third edition 2017-04 [Including AMD1:2025] Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood [Including Amendment 1 (2025)] |
|
Scope/AbstractThis document specifies general requirements for evaluating the interactions of medical devices with blood. It describes a) a classification of medical devices that are intended for use in contact with blood, based on the intended use and duration of contact as defined in ISO 10993-1, b) the fundamental principles governing the evaluation of the interaction of devices with blood, c) the rationale for structured selection of tests according to specific categories, together with the principles and scientific basis of these tests. Detailed requirements for testing cannot be specified because of limitations in the knowledge and precision of tests for evaluating interactions of devices with blood. This document describes biological evaluation in general terms and may not necessarily provide sufficient guidance for test methods for a specific device. The changes in this document do not indicate that testing conducted according to prior versions of this document is invalid. For marketed devices with a history of safe clinical use, additional testing according to this revision is not recommended.
|
|
| Extent of Recognition
|
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Note: Amendment is recognized with the following corrections:
. Note a in Table 1 of Amendment 1 states "In vivo or ex vivo testing is not always unnecessary if clinically relevant in vitro thrombosis testing is performed." The word "unnecessary" is a typo, it should be "necessary" . Clause A1.1 of Amendment 1 states "Delete "of non-contact-, externally communicating and implant devices" from the sentence as follows: "and a list of tests for evaluating blood/device interactions."" Deleting the phrase "and a list of tests for evaluating blood/device interactions" from ISO 10993-4: 2017 version is an error. Clause A1.1 of Amendment 1 should state "Delete "of non-contact-, externally communicating and implant devices"" |
|
ASCA Notes
The following test methods are included in the ASCA pilot program: Complement Activation using a U.S. marketed ELISA kit; Direct and Indirect Hemolysis, in conjunction with ASTM F756-17 (recognition number: 2-250). |
|
Transition Period
| FDA recognition of ISO 10993-4 Third edition 2017-04 [Rec# 2-248] will be superseded by recognition of ISO 10993-4 Third edition 2017-04 [Rec# 2-311]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 2-248] until December 17, 2028. After this transition period, declarations of conformity to [Rec# 2-248] will not be accepted. |
|
Relevant FDA Guidance and/or Supportive Publications*
ISO 10993-12:2021 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials.
Use of International Standard ISO 10993-1, Biological evaluation of medical devices--Part 1: Evaluation and testing within a risk management process. Guidance for Industry and Food and Drug Administration Staff, Issued September 2023.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contacts
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
| *These are provided as examples and others may be applicable. |