 Class 2 Device Recall Sechrist Stryker Medical Stretchers Model 742see related information
Date Initiated by FirmJuly 19, 2011
Date PostedJuly 22, 2011
Recall Status1 Terminated 3 on August 29, 2013
Recall NumberZ-2861-2011
Recall Event ID 59201
Product Classification Stretcher, Wheeled - Product Code FPO
ProductSechrist Stryker Medical Stretchers Model 742 Intended to transport patients in a horizontal position.
Code Information Model: 742  Lots: 1102035194, 1102035195, 1102035196, 1102035197, 1102035198, 1102035199, 1102035200, 1102035201, 1102035202 and 1102035203.
FEI Number 1831750
Recalling Firm/
Manufacturer
Stryker Medical Division of Stryker Corporation
3800 East Centre Ave.
Portage MI 49002
For Additional Information ContactMatt Jordan
269-389-6604
Manufacturer Reason
for Recall
The firm has determined that the brake rod assemblies of various stretchers that were built with a 5th wheel were manufactured with roll pins that did not meet the materials specifications. Some roll pins may have fractured during the assembly process and could fall out of the brake rod assemblies, rendering the brake or steer systems inoperable from certain pedals or causing a reduction in overa
FDA Determined
Cause 2
Nonconforming Material/Component
ActionStryker Medical sent an "Urgent Medical Device Correction" dated July 19 ,2011 to all affected customers. The letter included; the name and model numbers of the stretchers, the reason for the correction and dates of manufacture. It contained the instructions to check the braking system and if found to be inoperable, to remove it from service, and if operable to check prior to each use. Also, information regarding contacting the recipients of any loaned or sold stretchers to relay the information in the letter and Medwatch reporting information. A self addressed postcard to confirm receipt of the letter was also included. For additional information call (269) 389-6604.
Quantity in Commerce10
DistributionWorldwide Distribution: Nationwide (USA) including the states of NY, MI, LA and CA ,and the countries of Canada, China, Spain, Japan, Korea, United Kingdom, South Africa.
Total Product Life CycleTPLC Device Report