| | Class 2 Device Recall OnBoard Imager (OBI); Trilogy Mx |  |
| Date Initiated by Firm | June 21, 2011 |
| Date Posted | August 31, 2011 |
| Recall Status1 |
Terminated 3 on March 15, 2012 |
| Recall Number | Z-3101-2011 |
| Recall Event ID |
59262 |
| 510(K)Number | K042720 K092871 |
| Product Classification |
Accelerator, linear, medical - Product Code IYE
|
| Product | On-Board Imager (OBI) 1.3, 1.4, 1.5; and Trilogy Mx, TrueBeam 1.0, Offline Review 1.0 - 2.0;
Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA
The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields relation to anatomical and/or fiducial landmarks. Trilogy Mx is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. |
| Code Information |
CODE(S): H080002 , H080065 , H080126 , H080186 , H080250 , H080313 , H080375 , H080437 , H080497 , H080558 , H080664 , H080740 , H082010 , H080003 , H080066 , H080127 , H080187 , H080251 , H080314 , H080376 , H080438 , H080498 , H080601 , H080665 , H080741 , H082011 , H080004 , H080068 , H080128 , H080188 , H080252 , H080315 , H080377 , H080439 , H080499 , H080602 , H080666 , H080742 , H082012 , H080005 , H080069 , H080129 , H080189 , H080254 , H080316 , H080378 , H080440 , H080500 , H080604 , H080667 , H080745 , H082013 , H080006 , H080070 , H080130 , H080190 , H080255 , H080317 , H080379 , H080441 , H080501 , H080605 , H080668 , H080746 , H082014 , H080008 , H080071 , H080131 , H080191 , H080256 , H080318 , H080380 , H080442 , H080502 , H080606 , H080669 , H080747 , H082015 , H080009 , H080072 , H080132 , H080192 , H080257 , H080319 , H080381 , H080443 , H080503 , H080607 , H080670 , H080748 , H082016 , H080010 , H080073 , H080133 , H080193 , H080258 , H080320 , H080382 , H080444 , H080504 , H080608 , H080672 , H080749 , H082017 , H080011 , H080074 , H080134 , H080194 , H080259 , H080321 , H080383 , H080445 , H080506 , H080609 , H080673 , H080753 , H082018 , H080012 , H080075 , H080136 , H080195 , H080260 , H080322 , H080384 , H080446 , H080507 , H080610 , H080674 , H080754 , H082020 , H080013 , H080076 , H080137 , H080196 , H080261 , H080324 , H080385 , H080447 , H080508 , H080611 , H080675 , H080755 , H082021 , H080014 , H080077 , H080138 , H080197 , H080262 , H080325 , H080386 , H080448 , H080509 , H080612 , H080676 , H080759 , H082022 , H080016 , H080078 , H080139 , H080198 , H080263 , H080326 , H080387 , H080449 , H080510 , H080613 , H080677 , H080760 , H082023 , H080017 , H080079 , H080140 , H080199 , H080264 , H080327 , H080389 , H080450 , H080511 , H080614 , H080678 , H080761 , H082024 , H080018 , H080080 , H080141 , H080200 , H080265 , H080328 , H080391 , H080451 , H080512 , H080615 , H080679 , H080762 , H082025 , H080019 , H080081 , H080142 , H080202 , H080266 , H080329 , H080392 , H080452 , H080513 , H080616 , H080680 , H080763 , H082026 , H080020 , H080082 , H080143 , H080203 , H080267 , H080330 , H080393 , H080453 , H080514 , H080617 , H080683 , H080764 , H082031 , H080021 , H080083 , H080144 , H080204 , H080268 , H080331 , H080394 , H080454 , H080515 , H080618 , H080684 , H080766 , H082032 , H080022 , H080084 , H080145 , H080205 , H080269 , H080332 , H080395 , H080455 , H080516 , H080619 , H080685 , H080768 , H082033 , H080024 , H080085 , H080147 , H080206 , H080270 , H080333 , H080396 , H080456 , H080517 , H080620 , H080686 , H080772 , H082034 , H080026 , H080086 , H080148 , H080207 , H080271 , H080334 , H080397 , H080457 , H080518 , H080621 , H080687 , H080773 , H082035 , H080027 , H080087 , H080149 , H080208 , H080272 , H080335 , H080398 , H080458 , H080519 , H080622 , H080688 , H080774 , H082036 , H080028 , H080088 , H080150 , H080209 , H080273 , H080336 , H080399 , H080459 , H080520 , H080623 , H080689 , H080775 , H082037 , H080029 , H080089 , H080151 , H080210 , H080274 , H080337 , H080400 , H080460 , H080521 , H080624 , H080690 , H080778 , H082038 , H080030 , H080090 , H080152 , H080211 , H080275 , H080338 , H080401 , H080461 , H080522 , H080625 , H080691 , H080783 , H082039 , H080031 , H080091 , H080153 , H080212 , H080276 , H080339 , H080402 , H080462 , H080523 , H080626 , H080692 , H080786 , H082040 , H080032 , H080092 , H080154 , H080213 , H080278 , H080340 , H080403 , H080463 , H080524 , H080627 , H080693 , H080788 , H082041 , H080033 , H080093 , H080155 , H080214 , H080279 , H080341 , H080404 , H080464 , H080525 , H080628 , H080695 , H080791 , H082042 , H080034 , H080094 , H080156 , H080215 , H080280 , H080342 , H080405 , H080465 , H080526 , H080629 , H080696 , H080798 , H082043 , H080035 , H080095 , H080157 , H080216 , H080281 , H080344 , H080406 , H080466 , H080527 , H080630 , H080697 , H080800 , H082044 , H080036 , H080096 , H080158 , H080217 , H080282 , H080345 .H080407 , H080467 , H080528 , H080631 , H080698 , H080801 , H082045 , H080037 , H080097 , H080159 , H080218 , H080283 , H080346 , H080408 , H080468 , H080529 , H080632 , H080699 , H080805 , H082046 , H080038 , H080099 , H080160 , H080219 , H080286 , H080347 , H080409 , H080469 , H080530 , H080633 , H080700 , H080812 , H082047 , H080039 , H080100 , H080161 , H080220 , H080287 , H080348 , H080410 , H080470 , H080531 , H080634 , H080701 , H080836 , H082048 , H080040 , H080101 , H080162 , H080221 , H080288 , H080349 , H080411 , H080471 , H080532 , H080635 , H080702 , H080840 , H082049 , H080041 , H080102 , H080163 , H080222 , H080289 , H080350 , H080412 , H080472 , H080534 , H080636 , H080703 , H080852 , H082050 , H080042 , H080103 , H080164 , H080224 , H080290 , H080351 , H080413 , H080473 , H080535 , H080637 , H080706 , H080857 , H082051 , H080043 , H080104 , H080165 , H080225 , H080291 , H080352 , H080414 , H080474 , H080536 , H080638 , H080707 , H080859 , H082052 , H080044 , H080105 , H080166 , H080226 , H080292 , H080353 , H080415 , H080476 , H080537 , H080641 , H080708 , H080860 , H082053 , H080045 , H080107 , H080167 , H080227 , H080294 , H080354 , H080416 , H080477 , H080538 , H080642 , H080709 , H080861 , H082054 , H080046 , H080108 , H080168 , H080228 , H080295 , H080355 , H080417 , H080478 , H080539 , H080643 , H080710 , H080864 , H082055 , H080047 , H080109 , H080169 , H080229 , H080296 , H080356 , H080418 , H080479 , H080540 , H080644 , H080711 , H080866 , H082056 , H080048 , H080110 , H080170 , H080230 , H080297 , H080357 , H080419 , H080480 , H080541 , H080645 , H080712 , H080868 , H082057 , H080049 , H080111 , H080171 , H080231 , H080298 , H080358 , H080420 , H080481 , H080542 , H080646 , H080713 , H080869 , H082058 , H080050 , H080112 , H080172 , H080232 , H080299 , H080359 , H080421 , H080482 , H080543 , H080647 , H080714 , H080870 , H082059 , H080051 , H080113 , H080173 , H080234 , H080300 , H080360 , H080423 , H080484 , H080544 , H080650 , H080716 , H080873 , H082060 , H080052 , H080114 , H080174 , H080235 , H080301 , H080361 , H080424 , H080485 , H080545 , H080651 , H080717 , H080881 , H082061 , H080053 , H080115 , H080175 , H080236 , H080302 , H080362 , H080425 , H080486 , H080546 , H080652 , H080718 , H081001 , H082062 , H080054 , H080116 , H080176 , H080237 , H080303 , H080363 , H080426 , H080487 , H080547 , H080653 , H080726 , H082000 , H082063 , H080055 , H080117 , H080177 , H080238 , H080304 , H080364 , H080427 , H080488 , H080548 , H080654 , H080727 , H082001 , H082064 , H080056 , H080118 , H080178 , H080239 , H080305 , H080365 , H080428 , H080489 , H080549 , H080655 , H080728 , H082002 , H082065 , H080057 , H080119 , H080179 , H080240 , H080306 , H080366 , H080429 , H080490 , H080550 , H080656 , H080729 , H082003 , H082066 , H080059 , H080120 , H080180 , H080243 , H080307 , H080368 , H080430 , H080491 , H080551 , H080657 , H080730 , H082004 , H082067 , H080060 , H080121 , H080181 , H080244 , H080308 , H080369 , H080431 , H080492 , H080552 , H080659 , H080733 , H082005 , H082068 , H080061 , H080122 , H080182 , H080245 , H080309 , H080370 , H080432 , H080493 , H080553 , H080660 , H080734 , H082006 , H082069 , H080062 , H080123 , H080183 , H080246 , H080310 , H080371 , H080433 , H080494 , H080554 , H080661 , H080736 , H082007 , H082070 , H080063 , H080124 , H080184 , H080247 , H080311 , H080373 , H080435 , H080495 , H080555 , H080662 , H080738 , H082008 , H082071 , H080064 , H080125 , H080185 , H080248 , H080312 , H080374 , H080436 , H080496 , H080557 , H080663 , H080739 , H082009 , H082072 , H082073 , H082137 , H082201 , H082265 , H082330 , H082393 , H082520 , H082584 , H082647 , H082712 , H082778 , H082842 , H082907 , H082074 , H082138 , H082202 , H082266 , H082331 , H082394 , H082521 , H082585 , H082649 , H082713 , H082779 , H082843 , H082910 , H082075 , H082139 , H082203 , H082267 , H082332 , H082395 , H082522 , H082586 , H082650 , H082714 , H082780 , H082844 , H082911 , H082076 , H082140 , H082204 , H082268 , H082333 , H082396 , H082523 , H082587 , H082651 , H082715 , H082781 , H082845 , H082912 , H082077 , H082141 , H082205 , H082269 , H082334 , H082397 , H082524 , H082588 , H082652 , H082716 , H082782 , H082846 , H082913 , H082078 , H082142 , H082206 , H082270 , H082335 , H082398 , H082525 , H082589 , H082653 , H082717 , H082783 , H082847 , H082914 , H082079 , H082143 , H082207 , H082271 , H082336 , H082399 , H082526 , H082590 , H082654 , H082718 , H082784 , H082848 , H082915 , H082080 , H082144 , H082208 , H082272 , H082337 , H082400 , H082527 , H082591 , H082655 , H082719 , H082785 , H082849 , H082916 , H082081 , H082145 , H082209 , H082273 , H082338 , H082401 , H082528 , H082592 , H082656 , H082720 , H082786 , H082850 , H082917 , H082082 , H082146 , H082210 , H082274 , H082339 , H082402 , H082529 , H082593 , H082657 , H082721 , H082787 , H082851 , H082919 , H082083 , H082147 , H082211 , H082275 , H082340 , H082403 , H082530 , H082594 , H082658 , H082722 , H082788 , H082852 , H082920 , H082084 , H082148 , H082212 , H082276 , H082341 , H082404 , H082531 , H082595 , H082659 , H082723 , H082789 , H082853 , H082921 , H082085 , H082149 , H082213 , H082277 , H082342 , H082406 , H082532 , H082596 , H082660 , H082724 , H082790 , H082854 , H082922 , H082086 , H082150 , H082214 , H082278 , H082343 , H082407 , H082533 , H082597 , H082661 , H082725 , H082791 , H082855 , H082923 , H082087 , H082151 , H082215 , H082279 , H082344 , H082408 , H082534 , H082598 , H082662 , H082726 , H082792 , H082856 , H082924 , H082088 , H082152 , H082216 , H082280 , H082346 , H082409 , H082535 , H082599 , H082663 , H082727 , H082794 , H082857 , H082925 , H082089 , H082153 , H082217 , H082282 , H082347 , H082410 , H082536 , H082600 , H082664 , H082730 , H082795 , H082858 , H082926 , H082090 , H082154 , H082218 , H082283 , H082348 , H082411 , H082537 , H082601 , H082666 , H082731 , H082796 , H082859 , H082927 , H082091 , H082156 , H082219 , H082284 , H082349 , H082412 , H082538 , H082602 , H082667 , H082732 , H082797 , H082860 , H082928 , H082092 , H082157 , H082220 , H082285 , H082350 , H082450 , H082539 , H082603 , H082668 , H082733 , H082798 , H082861 , H082929 , H082093 , H082158 , H082221 , H082286 , H082351 , H082451 , H082540 , H082604 , H082669 , H082734 , H082799 , H082862 , H082930 , H082094 , H082159 , H082222 , H082287 , H082352 , H082452 , H082541 , H082605 , H082670 , H082735 , H082800 , H082864 , H082931 , H082095 , H082160 , H082223 , H082288 , H082353 , H082453 , H082542 , H082606 , H082671 , H082736 , H082801 , H082865 , H082932 , H082096 , H082161 , H082224 , H082289 , H082354 , H082475 , H082543 , H082607 , H082672 , H082737 , H082802 , H082866 , H082933 , H082097 , H082162 , H082225 , H082290 , H082355 , H082476 , H082544 , H082608 , H082673 , H082738 , H082803 , H082868 , H082934 , H082098 , H082163 , H082226 , H082291 , H082356 , H082480 , H082545 , H082609 , H082674 , H082739 , H082804 , H082869 , H082935 , H082099 , H082164 , H082227 , H082292 , H082357 , H082482 , H082546 , H082610 , H082675 , H082740 , H082805 , H082870 , H082937 , H082100 , H082165 , H082228 , H082293 , H082358 , H082484 , H082547 , H082611 , H082676 , H082741 , H082806 , H082871 , H082938 , H082101 , H082166 , H082229 , H082294 , H082359 , H082485 , H082548 , H082613 , H082678 , H082743 , H082807 , H082872 , H082939 , H082102 , H082167 , H082230 , H082295 , H082360 , H082486 , H082549 , H082614 , H082679 , H082744 , H082808 , H082873 , H082940 , H082103 , H082168 , H082231 , H082296 , H082361 , H082487 , H082550 , H082615 , H082680 , H082745 , H082809 , H082874 , H082941 , H082104 , H082169 , H082232 , H082297 , H082362 , H082488 , H082551 , H082616 , H082681 , H082746 , H082810 , H082875 , H082942 , H082105 , H082170 , H082233 , H082298 , H082363 , H082489 , H082552 , H082617 , H082682 , H082747 , H082811 , H082876 , H082943 , H082106 , H082171 , H082234 , H082299 , H082364 , H082490 , H082553 , H082618 , H082683 , H082748 , H082812 , H082877 , H082944 , H082107 , H082172 , H082235 , H082300 , H082365 , H082491 , H082554 , H082619 , H082684 , H082749 , H082813 , H082878 , H082945 , H082108 , H082173 , H082236 , H082301 , H082366 , H082492 , H082555 , H082620 , H082685 , H082750 , H082814 , H082879 , H082946 , H082109 , H082174 , H082237 , H082302 , H082367 , H082493 , H082556 , H082621 , H082686 , H082751 , H082815 , H082880 , H082947 , H082110 , H082175 , H082238 , H082303 , H082368 , H082494 , H082557 , H082622 , H082687 , H082752 , H082816 , H082881 , H082948 , H082112 , H082176 , H082239 , H082304 , H082369 , H082495 , H082559 , H082623 , H082688 , H082753 , H082817 , H082882 , H082949 , H082113 , H082177 , H082241 , H082306 , H082370 , H082496 , H082560 , H082624 , H082689 , H082754 , H082818 , H082883 , H082950 , H082114 , H082179 , H082242 , H082307 , H082371 , H082498 , H082561 , H082625 , H082690 , H082755 , H082819 , H082884 , H082951 , H082115 , H082180 , H082243 , H082308 , H082372 , H082499 , H082562 , H082626 , H082691 , H082756 , H082820 , H082885 , H082952 , H082116 , H082181 , H082244 , H082309 , H082373 , H082500 , H082563 , H082627 , H082692 , H082757 , H082821 , H082886 , H082953 , H082117 , H082182 , H082245 , H082310 , H082374 , H082501 , H082564 , H082628 , H082693 , H082758 , H082822 , H082887 , H082954 , H082118 , H082183 , H082246 , H082311 , H082375 , H082502 , H082565 , H082629 , H082694 , H082759 , H082823 , H082888 , H082955 , H082119 , H082184 , H082247 , H082312 , H082376 , H082503 , H082566 , H082630 , H082695 , H082760 , H082824 , H082889 , H082957 , H082120 , H082185 , H082248 , H082313 , H082377 , H082504 , H082567 , H082631 , H082696 , H082761 , H082825 , H082890 , H082958 , H082121 , H082186 , H082249 , H082314 , H082378 , H082505 , H082568 , H082632 , H082697 , H082762 , H082826 , H082891 , H082959 , H082122 , H082187 , H082251 , H082315 , H082379 , H082506 , H082569 , H082633 , H082698 , H082763 , H082827 , H082892 , H082960 , H082123 , H082188 , H082252 , H082316 , H082380 , H082507 , H082570 , H082634 , H082699 , H082764 , H082828 , H082893 , H082961 , H082124 , H082189 , H082253 , H082317 , H082381 , H082508 , H082571 , H082635 , H082700 , H082765 , H082829 , H082894 , H082962 , H082125 , H082190 , H082254 , H082318 , H082382 , H082509 , H082572 , H082636 , H082701 , H082766 , H082830 , H082895 , H082963 , H082126 , H082191 , H082255 , H082319 , H082383 , H082510 , H082573 , H082637 , H082702 , H082767 , H082831 , H082896 , H082964 , H082127 , H082192 , H082256 , H082320 , H082384 , H082511 , H082574 , H082638 , H082703 , H082768 , H082832 , H082897 , H082965 , H082128 , H082193 , H082257 , H082321 , H082385 , H082512 , H082575 , H082639 , H082704 , H082769 , H082833 , H082898 , H082966 , H082129 , H082194 , H082258 , H082322 , H082386 , H082513 , H082576 , H082640 , H082705 , H082770 , H082834 , H082899 , H082967 , H082130 , H082195 , H082259 , H082323 , H082387 , H082514 , H082578 , H082641 , H082706 , H082771 , H082835 , H082900 , H082968 , H082131 , H082196 , H082260 , H082324 , H082388 , H082515 , H082579 , H082642 , H082707 , H082772 , H082836 , H082901 , H082969 , H082133 , H082197 , H082261 , H082325 , H082389 , H082516 , H082580 , H082643 , H082708 , H082773 , H082838 , H082902 , H082970 , H082134 , H082198 , H082262 , H082326 , H082390 , H082517 , H082581 , H082644 , H082709 , H082774 , H082839 , H082903 , H082971 , H082135 , H082199 , H082263 , H082327 , H082391 , H082518 , H082582 , H082645 , H082710 , H082776 , H082840 , H082904 , H082972 , H082136 , H082200 , H082264 , H082328 , H082392 , H082519 , H082583 , H082646 , H082711 , H082777 , H082841 , H082905 , H082973 , H082974 , H083041 , H083111 , H083176 , H083244 , H083311 , H083376 , H083442 , H083516 , H083590 , H083691 , H191070 , H082975 , H083042 , H083112 , H083177 , H083245 , H083312 , H083377 , H083443 , H083517 , H083591 , H084338 , H191071 , H082976 , H083043 , H083113 , H083178 , H083246 , H083313 , H083378 , H083445 , H083518 , H083592 , H191001 , H191072 , H082977 , H083044 , H083114 , H083179 , H083247 , H083314 , H083379 , H083446 , H083519 , H083593 , H191002 , H191073 , H082978 , H083045 , H083115 , H083180 , H083248 , H083315 , H083380 , H083448 , H083521 , H083594 , H191003 , H191076 , H082979 , H083046 , H083116 , H083181 , H083249 , H083316 , H083381 , H083449 , H083522 , H083595 , H191005 , H191077 , H082980 , H083047 , H083117 , H083182 , H083250 , H083317 , H083382 , H083450 , H083523 , H083596 , H191006 , H191078 , H082981 , H083048 , H083119 , H083183 , H083251 , H083318 , H083383 , H083451 , H083524 , H083599 , H191007 , H191079 , H082982 , H083049 , H083120 , H083184 , H083252 , H083319 , H083384 , H083452 , H083525 , H083600 , H191008 , H191080 , H082983 , H083050 , H083121 , H083185 , H083253 , H083321 , H083385 , H083453 , H083526 , H083601 , H191009 , H191082 , H082984 , H083051 , H083122 , H083186 , H083254 , H083322 , H083386 , H083454 , H083527 , H083602 , H191010 , H191083 , H082985 , H083052 , H083123 , H083187 , H083255 , H083323 , H083387 , H083455 , H083529 , H083604 , H191011 , H191085 , H082986 , H083053 , H083124 , H083188 , H083256 , H083324 , H083388 , H083456 , H083530 , H083607 , H191012 , H191086 , H082987 , H083054 , H083125 , H083189 , H083257 , H083325 , H083389 , H083457 , H083532 , H083608 , H191013 , H191087 , H082988 , H083055 , H083126 , H083190 , H083258 , H083326 , H083390 , H083458 , H083533 , H083609 , H191014 , H191088 , H082990 , H083056 , H083127 , H083191 , H083259 , H083327 , H083391 , H083459 , H083534 , H083610 , H191015 , H191089 , H082991 , H083057 , H083128 , H083192 , H083260 , H083328 , H083392 , H083461 , H083536 , H083611 , H191019 , H191090 , H082993 , H083058 , H083129 , H083193 , H083261 , H083330 , H083393 , H083462 , H083537 , H083612 , H191020 , H191091 , H082994 , H083059 , H083130 , H083194 , H083262 , H083331 , H083394 , H083463 , H083539 , H083616 , H191021 , H191092 , H082995 , H083060 , H083131 , H083195 , H083263 , H083332 , H083395 , H083464 , H083540 , H083617 , H191022 , H191094 , H082996 , H083061 , H083132 , H083196 , H083264 , H083333 , H083396 , H083465 , H083541 , H083618 , H191023 , H191095 , H082997 , H083062 , H083133 , H083197 , H083265 , H083334 , H083397 , H083466 , H083543 , H083620 , H191025 , H191098 , H082998 , H083063 , H083135 , H083198 , H083266 , H083335 , H083398 , H083467 , H083544 , H083621 , H191026 , H191100 , H082999 , H083064 , H083136 , H083199 , H083267 , H083336 , H083399 , H083469 , H083546 , H083622 , H191027 , H191101 , H083000 , H083065 , H083137 , H083200 , H083268 , H083337 , H083400 , H083470 , H083548 , H083623 , H191028 , H191102 , H083001 , H083066 , H083138 , H083202 , H083269 , H083338 , H083401 , H083471 , H083549 , H083624 , H191029 , H191106 , H083002 , H083067 , H083139 , H083204 , H083270 , H083339 , H083402 , H083473 , H083550 , H083625 , H191030 , H083003 , H083068 , H083140 , H083205 , H083271 , H083340 , H083403 , H083475 , H083551 , H083628 , H191031 , H083005 , H083069 , H083141 , H083206 , H083272 , H083341 , H083404 , H083476 , H083552 , H083629 , H191032 , H083006 , H083070 , H083142 , H083207 , H083273 , H083342 , H083405 , H083478 , H083553 , H083631 , H191033 , H083007 , H083071 , H083143 , H083208 , H083274 , H083343 , H083406 , H083480 , H083556 , H083632 , H191034 , H083008 , H083072 , H083144 , H083210 , H083275 , H083344 , H083407 , H083481 , H083557 , H083634 , H191035 , H083009 , H083073 , H083145 , H083211 , H083276 , H083345 , H083408 , H083482 , H083558 , H083635 , H191036 , H083010 , H083074 , H083146 , H083212 , H083277 , H083346 , H083409 , H083484 , H083559 , H083636 , H191038 , H083011 , H083075 , H083147 , H083213 , H083279 , H083347 , H083410 , H083485 , H083560 , H083637 , H191039 , H083012 , H083076 , H083148 , H083214 , H083280 , H083348 , H083412 , H083486 , H083561 , H083638 , H191040 , H083013 , H083077 , H083149 , H083216 , H083281 , H083349 , H083413 , H083487 , H083562 , H083639 , H191041 , H083014 , H083078 , H083150 , H083217 , H083282 , H083350 , H083414 , H083488 , H083563 , H083642 , H191042 , H083015 , H083079 , H083152 , H083218 , H083283 , H083351 , H083415 , H083489 , H083564 , H083643 , H191043 , H083016 , H083080 , H083153 , H083219 , H083284 , H083352 , H083416 , H083490 , H083565 , H083645 , H191044 , H083017 , H083082 , H083154 , H083220 , H083285 , H083353 , H083417 , H083491 , H083566 , H083646 , H191045 , H083018 , H083083 , H083155 , H083221 , H083287 , H083354 , H083418 , H083492 , H083567 , H083647 , H191046 , H083019 , H083084 , H083156 , H083222 , H083288 , H083355 , H083419 , H083494 , H083568 , H083648 , H191047 , H083020 , H083085 , H083157 , H083224 , H083289 , H083356 , H083420 , H083495 , H083569 , H083650 , H191048 , H083021 , H083086 , H083158 , H083225 , H083290 , H083357 , H083421 , H083496 , H083571 , H083651 , H191049 , H083022 , H083087 , H083159 , H083226 , H083291 , H083358 , H083422 , H083497 , H083572 , H083653 , H191050 , H083023 , H083088 , H083160 , H083227 , H083292 , H083359 , H083423 , H083499 , H083573 , H083654 , H191051 , H083024 , H083089 , H083161 , H083228 , H083293 , H083360 , H083424 , H083500 , H083574 , H083655 , H191052 , H083025 , H083090 , H083162 , H083229 , H083294 , H083361 , H083425 , H083502 , H083575 , H083659 , H191053 , H083026 , H083091 , H083163 , H083230 , H083295 , H083363 , H083426 , H083503 , H083576 , H083662 , H191054 , H083027 , H083092 , H083164 , H083231 , H083297 , H083364 , H083427 , H083504 , H083577 , H083663 , H191055 , H083029 , H083093 , H083165 , H083232 , H083298 , H083365 , H083428 , H083505 , H083578 , H083664 , H191056 , H083030 , H083094 , H083166 , H083233 , H083299 , H083366 , H083429 , H083506 , H083579 , H083665 , H191057 , H083031 , H083098 , H083167 , H083234 , H083300 , H083367 , H083430 , H083507 , H083580 , H083667 , H191058 , H083033 , H083103 , H083168 , H083235 , H083301 , H083368 , H083433 , H083508 , H083581 , H083668 , H191059 , H083034 , H083104 , H083169 , H083236 , H083302 , H083369 , H083434 , H083509 , H083582 , H083670 , H191060 , H083035 , H083105 , H083170 , H083237 , H083303 , H083370 , H083436 , H083510 , H083583 , H083671 , H191061 , H083036 , H083106 , H083171 , H083238 , H083304 , H083371 , H083437 , H083511 , H083584 , H083674 , H191063 , H083037 , H083107 , H083172 , H083239 , H083306 , H083372 , H083438 , H083512 , H083585 , H083682 , H191064 , H083038 , H083108 , H083173 , H083241 , H083308 , H083373 , H083439 , H083513 , H083586 , H083687 , H191066 , H083039 , H083109 , H083174 , H083242 , H083309 , H083374 , H083440 , H083514 , H083587 , H083689 , H191067 , H083040 , H083110 , H083175 , H083243 , H083310 , H083375 , H083441 , H083515 , H083588 , H083690 , H191069; ***12-20-12, AMENDED INFO ***: H083836, H081105, H083633, H081122, H083767, H083765 |
| FEI Number |
2916710
|
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc. Oncology Systems
911 Hansen Way
Palo Alto
CA
94304-1028
|
| For Additional Information Contact | Michael Pignataro 650-424-6471 |
Manufacturer Reason for Recall | Varian has identified a software anomaly in the 3D Match environment, regarding a slight mismatch between CT image and Structures in On-Board Imager w/s. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated June 21, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
The letter recommends several user actions. The firm is developing a correction for the issue and will contact customers when the correction is available.
Questions regarding this notice may be addressed by contacting the Varian Oncology Help Desk at 1-888-827-4265. |
| Quantity in Commerce | 2268 |
| Distribution | Worldwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|