| Date Initiated by Firm | June 30, 2011 |
| Date Posted | August 11, 2011 |
| Recall Status1 |
Terminated 3 on May 09, 2014 |
| Recall Number | Z-3002-2011 |
| Recall Event ID |
59378 |
| 510(K)Number | K073020 |
| Product Classification |
System,planning,radiation therapy treatment - Product Code MUJ
|
| Product | Eclipse Treatment Planning System;
Varian Medical Systems, Palo Alto, CA
Product Usage: Eclipse Treatment Planning System is used to plan radiotherapy treatments for patients with malignant or benign diseases. It is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (Brachytherapy) treatments. |
| Code Information |
Model # H48; Serial numbers: H480015 H480313 H480798 H481172 H481514 H481954 H482014 H482074 H482260 H482487 H482900 H483184 H483622 H480016 H480314 H480799 H481183 H481540 H481955 H482015 H482075 H482261 H482488 H482901 H483185 H483623 H480034 H480315 H480827 H481188 H481542 H481956 H482016 H482076 H482262 H482530 H482902 H483195 H483624 H480035 H480340 H480828 H481197 H481551 H481957 H482017 H482077 H482263 H482538 H482903 H483207 H483641 H480036 H480427 H480829 H481205 H481606 H481958 H482018 H482078 H482266 H482549 H482904 H483216 H483642 H480037 H480428 H480908 H481263 H481608 H481959 H482019 H482079 H482267 H482558 H482906 H483217 H483643 H480038 H480429 H480936 H481264 H481613 H481960 H482020 H482080 H482268 H482576 H482907 H483235 H483644 H480039 H480430 H480943 H481265 H481614 H481961 H482021 H482081 H482269 H482599 H482908 H483236 H483645 H480040 H480473 H480962 H481266 H481639 H481962 H482022 H482082 H482270 H482630 H482909 H483237 H483705 H480044 H480474 H480963 H481267 H481641 H481963 H482023 H482083 H482271 H482631 H482910 H483238 H483730 H480045 H480475 H480964 H481268 H481642 H481964 H482024 H482084 H482282 H482632 H482911 H483241 H483732 H480074 H480476 H480970 H481269 H481647 H481965 H482025 H482085 H482287 H482641 H482912 H483247 H483733 H480080 H480484 H480971 H481270 H481648 H481966 H482026 H482086 H482294 H482656 H482913 H483272 H483735 H480088 H480494 H480972 H481271 H481649 H481967 H482027 H482087 H482324 H482657 H482914 H483273 H483736 H480089 H480495 H480980 H481272 H481650 H481968 H482028 H482088 H482325 H482676 H482917 H483274 H483737 H480090 H480502 H480981 H481273 H481651 H481969 H482029 H482089 H482326 H482693 H482918 H483279 H483738 H480091 H480503 H480989 H481274 H481652 H481970 H482030 H482090 H482346 H482717 H482919 H483294 H483739 H480102 H480504 H480992 H481275 H481653 H481971 H482031 H482091 H482347 H482722 H482928 H483295 H483740 H480103 H480505 H480993 H481276 H481654 H481972 H482032 H482092 H482348 H482723 H482929 H483296 H483741 H480132 H480539 H480994 H481289 H481655 H481973 H482033 H482093 H482349 H482730 H482944 H483297 H483742 H480133 H480541 H481010 H481290 H481656 H481974 H482034 H482094 H482350 H482731 H482945 H483298 H483743 H480137 H480573 H481021 H481291 H481657 H481975 H482035 H482095 H482351 H482757 H483002 H483299 H483744 H480167 H480574 H481077 H481292 H481658 H481976 H482036 H482096 H482352 H482758 H483003 H483300 H483745 H480168 H480575 H481078 H481293 H481659 H481977 H482037 H482097 H482353 H482759 H483004 H483301 H483746 H480169 H480585 H481079 H481294 H481669 H481978 H482038 H482098 H482354 H482764 H483005 H483302 H483747 H480170 H480586 H481080 H481295 H481670 H481979 H482039 H482099 H482355 H482767 H483006 H483303 H483748 H480191 H480596 H481081 H481300 H481711 H481980 H482040 H482100 H482356 H482770 H483007 H483306 H483836 H480192 H480610 H481082 H481301 H481712 H481981 H482041 H482101 H482357 H482771 H483008 H483340 H483837 H480193 H480645 H481083 H481306 H481722 H481982 H482042 H482102 H482358 H482772 H483009 H483356 H483838 H480194 H480652 H481084 H481307 H481723 H481983 H482043 H482103 H482359 H482775 H483012 H483357 H483839 H480195 H480670 H481085 H481309 H481757 H481984 H482044 H482104 H482360 H482776 H483013 H483373 H483845 H480196 H480676 H481086 H481310 H481766 H481985 H482045 H482105 H482361 H482777 H483018 H483374 H483860 H480197 H480677 H481087 H481315 H481767 H481986 H482046 H482106 H482362 H482778 H483019 H483375 H483874 H480198 H480678 H481099 H481316 H481789 H481987 H482047 H482107 H482397 H482779 H483020 H483376 H483925 H480239 H480682 H481100 H481334 H481795 H481988 H482048 H482108 H482398 H482781 H483021 H483382 H483933 H480240 H480683 H481101 H481335 H481796 H481989 H482049 H482109 H482399 H482782 H483022 H483386 H483943 H480241 H480685 H481102 H481374 H481805 H481990 H482050 H482110 H482425 H482783 H483023 H483390 H483944 H480242 H480686 H481103 H481375 H481806 H481991 H482051 H482111 H482426 H482784 H483047 H483391 H483945 H480244 H480687 H481104 H481376 H481807 H481992 H482052 H482112 H482428 H482785 H483053 H483392 H483946 H480250 H480699 H481105 H481381 H481815 H481993 H482053 H482113 H482429 H482786 H483054 H483393 H483947 H480251 H480700 H481119 H481382 H481816 H481994 H482054 H482138 H482430 H482787 H483055 H483394 H483948 H480252 H480701 H481120 H481383 H481839 H481995 H482055 H482139 H482431 H482788 H483120 H483401 H483949 H480253 H480702 H481122 H481393 H481850 H481996 H482056 H482140 H482432 H482789 H483123 H483403 H483950 H480254 H480712 H481130 H481394 H481851 H481997 H482057 H482141 H482433 H482820 H483124 H483404 H483951 H480255 H480713 H481131 H481395 H481852 H481998 H482058 H482142 H482434 H482824 H483125 H483416 H483952 H480258 H480719 H481133 H481423 H481853 H481999 H482059 H482143 H482435 H482825 H483126 H483417 H483953 H480259 H480720 H481135 H481424 H481854 H482000 H482060 H482166 H482439 H482827 H483127 H483418 H483957 H480260 H480722 H481138 H481428 H481855 H482001 H482061 H482167 H482450 H482835 H483128 H483419 H483960 H480261 H480723 H481145 H481432 H481856 H482002 H482062 H482168 H482451 H482836 H483129 H483420 H483980 H480262 H480724 H481149 H481444 H481901 H482003 H482063 H482169 H482452 H482848 H483130 H483421 H483981 H480263 H480727 H481150 H481455 H481902 H482004 H482064 H482170 H482464 H482859 H483131 H483422 H483982 H480264 H480734 H481151 H481456 H481903 H482005 H482065 H482171 H482465 H482860 H483132 H483448 H483983 H480285 H480735 H481152 H481457 H481904 H482006 H482066 H482172 H482466 H482861 H483133 H483450 H484010 H480293 H480745 H481154 H481458 H481905 H482007 H482067 H482178 H482467 H482862 H483146 H483487 H484011 H480297 H480780 H481155 H481459 H481907 H482008 H482068 H482189 H482473 H482863 H483147 H483507 H484058 H480298 H480785 H481156 H481482 H481908 H482009 H482069 H482190 H482474 H482865 H483151 H483517 H484059 H480305 H480786 H481164 H481492 H481912 H482010 H482070 H482201 H482475 H482881 H483152 H483521 H484060 H480306 H480787 H481165 H481497 H481913 H482011 H482071 H482203 H482478 H482897 H483159 H483617 H484061 H480311 H480788 H481166 H481500 H481914 H482012 H482072 H482258 H482479 H482898 H483182 H483620 H484077 H480312 H480795 H481171 H481513 H481915 H482013 H482073 H482259 H482480 H482899 H483183 H483621 H484082 H484113 H484360 H484807 H485222 H485690 H485967 H486358 H486860 H487167 H487925 H488505 H489234 H489756 H484114 H484361 H484808 H485233 H485691 H485968 H486359 H486861 H487168 H487926 H488506 H489235 H489757 H484149 H484367 H484809 H485234 H485712 H485977 H486360 H486862 H487169 H487933 H488509 H489236 H489758 H484151 H484368 H484810 H485242 H485713 H485978 H486361 H486863 H487170 H487934 H488510 H489237 H489788 H484153 H484384 H484811 H485246 H485714 H485979 H486362 H486864 H487171 H487957 H488517 H489238 H489789 H484154 H484394 H484812 H485250 H485731 H486007 H486363 H486865 H487172 H487958 H488518 H489239 H489790 H484158 H484395 H484813 H485253 H485738 H486016 H486364 H486918 H487173 H487974 H488529 H489240 H489805 H484159 H484396 H484814 H485254 H485740 H486024 H486365 H486919 H487181 H487975 H488530 H489241 H489806 H484200 H484398 H484815 H485255 H485741 H486025 H486366 H486934 H487229 H487976 H488547 H489271 H489845 H484201 H484407 H484816 H485256 H485742 H486037 H486367 H486937 H487242 H487999 H488548 H489272 H489876 H484202 H484446 H484817 H485259 H485749 H486057 H486368 H486946 H487266 H488001 H488593 H489273 H489908 H484203 H484447 H484818 H485260 H485750 H486058 H486390 H486947 H487267 H488002 H488594 H489274 H489909 H484204 H484448 H484819 H485261 H485756 H486059 H486391 H486948 H487268 H488003 H488595 H489275 H489910 H484205 H484449 H484820 H485283 H485757 H486060 H486392 H486962 H487269 H488004 H488697 H489276 H489933 H484206 H484462 H484830 H485284 H485758 H486068 H486393 H486984 H487270 H488005 H488724 H489277 H489934 H484207 H484463 H484831 H485297 H485759 H486075 H486394 H486985 H487271 H488009 H488734 H489278 H489962 H484208 H484549 H484832 H485301 H485760 H486076 H486401 H486986 H487272 H488010 H488735 H489279 H489963 H484209 H484550 H484833 H485305 H485761 H486081 H486402 H486987 H487273 H488011 H488763 H489280 H489984 H484210 H484551 H484841 H485306 H485762 H486114 H486414 H486988 H487274 H488012 H488764 H489281 H489985 H484211 H484578 H484859 H485319 H485763 H486115 H486415 H486989 H487275 H488013 H488775 H489282 H489991 H484212 H484579 H484892 H485320 H485764 H486116 H486435 H487016 H487276 H488014 H488776 H489283 H489992 H484222 H484580 H484893 H485321 H485765 H486122 H486446 H487017 H487282 H488015 H488777 H489284 H489993 H484224 H484581 H484894 H485322 H485766 H486138 H486447 H487071 H487380 H488016 H488789 H489285 H489994 H484225 H484601 H484895 H485323 H485767 H486139 H486463 H487073 H487389 H488020 H488790 H489286 H48A011 H484226 H484602 H484896 H485334 H485768 H486140 H486481 H487079 H487390 H488023 H488791 H489287 H48A012 H484227 H484616 H484897 H485336 H485769 H486184 H486517 H487080 H487391 H488029 H488795 H489288 H48A013 H484233 H484635 H484898 H485359 H485770 H486185 H486518 H487081 H487410 H488030 H488796 H489289 H48A014 H484234 H484636 H484899 H485363 H485788 H486186 H486533 H487089 H487411 H488031 H488799 H489355 H48A015 H484238 H484637 H484900 H485364 H485790 H486209 H486543 H487130 H487412 H488034 H488829 H489356 H48A016 H484239 H484638 H484901 H485371 H485791 H486211 H486544 H487131 H487419 H488035 H488841 H489376 H48A017 H484240 H484663 H484942 H485386 H485835 H486212 H486545 H487132 H487420 H488036 H488887 H489377 H48A018 H484241 H484664 H484945 H485387 H485836 H486215 H486546 H487133 H487425 H488037 H488888 H489380 H48A072 H484242 H484674 H484946 H485393 H485837 H486216 H486605 H487134 H487488 H488038 H488892 H489381 H48A074 H484243 H484675 H484947 H485402 H485838 H486224 H486606 H487135 H487489 H488039 H488911 H489398 H48A075 H484244 H484677 H484966 H485403 H485839 H486225 H486607 H487136 H487490 H488040 H488920 H489419 H48A076 H484245 H484678 H484968 H485404 H485840 H486231 H486608 H487137 H487491 H488042 H488921 H489427 H48A077 H484246 H484679 H484987 H485421 H485841 H486232 H486609 H487138 H487523 H488074 H488922 H489428 H48A078 H484247 H484680 H485012 H485429 H485849 H486236 H486610 H487139 H487524 H488108 H488926 H489429 H48A079 H484248 H484681 H485013 H485456 H485856 H486237 H486611 H487140 H487529 H488134 H488979 H489430 H48A080 H484274 H484695 H485037 H485467 H485874 H486238 H486629 H487141 H487541 H488135 H488984 H489431 H48A081 H484275 H484696 H485081 H485474 H485890 H486239 H486630 H487142 H487554 H488186 H489015 H489432 H48A082 H484276 H484697 H485102 H485538 H485891 H486265 H486631 H487143 H487555 H488230 H489022 H489433 H48A083 H484279 H484699 H485103 H485551 H485892 H486266 H486632 H487144 H487556 H488231 H489027 H489510 H48A084 H484280 H484706 H485104 H485552 H485893 H486289 H486641 H487145 H487557 H488243 H489030 H489514 H48A085 H484281 H484745 H485105 H485553 H485894 H486290 H486651 H487146 H487561 H488244 H489031 H489528 H48A124 H484282 H484746 H485113 H485554 H485895 H486291 H486652 H487147 H487562 H488259 H489032 H489529 H48A126 H484285 H484747 H485114 H485555 H485896 H486292 H486657 H487148 H487563 H488275 H489033 H489555 H48A136 H484297 H484748 H485115 H485556 H485897 H486293 H486659 H487149 H487617 H488276 H489034 H489571 H48A157 H484298 H484749 H485116 H485586 H485898 H486308 H486665 H487150 H487639 H488291 H489045 H489597 H48A158 H484299 H484750 H485133 H485587 H485899 H486311 H486693 H487151 H487661 H488316 H489056 H489631 H48A202 H484300 H484751 H485134 H485589 H485900 H486312 H486694 H487152 H487676 H488371 H489062 H489663 H48A213 H484302 H484752 H485136 H485610 H485901 H486314 H486733 H487153 H487733 H488393 H489063 H489670 H48A214 H484303 H484753 H485143 H485617 H485902 H486315 H486805 H487154 H487734 H488396 H489095 H489671 H48A225 H484308 H484798 H485144 H485618 H485903 H486319 H486819 H487155 H487794 H488406 H489096 H489672 H48A226 H484310 H484799 H485145 H485628 H485904 H486320 H486820 H487156 H487847 H488423 H489097 H489673 H48A227 H484334 H484800 H485146 H485656 H485905 H486327 H486842 H487157 H487866 H488446 H489156 H489674 H48A228 H484335 H484801 H485147 H485657 H485938 H486329 H486843 H487158 H487867 H488447 H489179 H489675 H48A242 H484354 H484802 H485152 H485667 H485939 H486331 H486844 H487159 H487875 H488448 H489180 H489676 H48A254 H484356 H484803 H485163 H485671 H485941 H486337 H486845 H487160 H487876 H488449 H489186 H489718 H48A292 H484357 H484804 H485191 H485675 H485944 H486354 H486846 H487164 H487887 H488450 H489194 H489740 H48A293 H484358 H484805 H485193 H485677 H485945 H486355 H486858 H487165 H487888 H488452 H489196 H489754 H48A294 H484359 H484806 H485208 H485689 H485962 H486357 H486859 H487166 H487908 H488464 H489232 H489755 H48A348 H48A369 H48A920 H48B294 H48B445 H48B600 H48B696 H48B789 H48B887 H48B969 H48C074 H48C160 H48C256 H48A370 H48A924 H48B295 H48B451 H48B602 H48B702 H48B790 H48B888 H48B971 H48C075 H48C161 H48C257 H48A398 H48A930 H48B298 H48B457 H48B604 H48B703 H48B791 H48B889 H48B973 H48C076 H48C162 H48C258 H48A407 H48A931 H48B299 H48B458 H48B607 H48B704 H48B792 H48B890 H48B975 H48C077 H48C163 H48C259 H48A408 H48A940 H48B300 H48B459 H48B608 H48B705 H48B793 H48B892 H48B976 H48C078 H48C168 H48C260 H48A415 H48A942 H48B302 H48B471 H48B609 H48B707 H48B794 H48B893 H48B977 H48C082 H48C170 H48C261 H48A416 H48A943 H48B309 H48B472 H48B610 H48B708 H48B795 H48B894 H48B978 H48C083 H48C171 H48C262 H48A417 H48A959 H48B312 H48B486 H48B611 H48B709 H48B797 H48B895 H48B979 H48C084 H48C173 H48C263 H48A418 H48A966 H48B331 H48B492 H48B612 H48B712 H48B798 H48B896 H48B994 H48C085 H48C174 H48C264 H48A453 H48A967 H48B332 H48B494 H48B613 H48B713 H48B799 H48B898 H48B995 H48C086 H48C175 H48C265 H48A454 H48A968 H48B333 H48B495 H48B614 H48B714 H48B800 H48B899 H48B996 H48C087 H48C176 H48C266 H48A455 H48A969 H48B334 H48B496 H48B615 H48B715 H48B801 H48B900 H48B997 H48C088 H48C177 H48C267 H48A456 H48A970 H48B335 H48B497 H48B616 H48B719 H48B802 H48B901 H48B998 H48C089 H48C178 H48C268 H48A462 H48A971 H48B336 H48B498 H48B617 H48B720 H48B803 H48B907 H48B999 H48C095 H48C179 H48C270 H48A474 H48A990 H48B340 H48B499 H48B618 H48B722 H48B804 H48B910 H48C000 H48C096 H48C180 H48C271 H48A478 H48A991 H48B341 H48B502 H48B619 H48B723 H48B805 H48B911 H48C001 H48C097 H48C183 H48C272 H48A484 H48A992 H48B342 H48B504 H48B622 H48B724 H48B806 H48B912 H48C002 H48C098 H48C184 H48C278 H48A485 H48B059 H48B343 H48B505 H48B630 H48B725 H48B807 H48B913 H48C003 H48C099 H48C185 H48C294 H48A487 H48B060 H48B344 H48B506 H48B633 H48B726 H48B808 H48B914 H48C004 H48C100 H48C186 H48C295 H48A506 H48B061 H48B345 H48B507 H48B634 H48B727 H48B809 H48B915 H48C005 H48C101 H48C187 H48C296 H48A507 H48B062 H48B349 H48B511 H48B635 H48B732 H48B810 H48B916 H48C006 H48C102 H48C188 H48C297 H48A509 H48B063 H48B350 H48B512 H48B636 H48B733 H48B812 H48B917 H48C007 H48C103 H48C194 H48C298 H48A510 H48B083 H48B351 H48B513 H48B637 H48B734 H48B813 H48B918 H48C008 H48C104 H48C195 H48C299 H48A511 H48B084 H48B352 H48B516 H48B638 H48B735 H48B816 H48B919 H48C010 H48C105 H48C197 H48C300 H48A543 H48B085 H48B362 H48B517 H48B639 H48B736 H48B817 H48B926 H48C011 H48C106 H48C198 H48C301 H48A563 H48B087 H48B363 H48B521 H48B640 H48B737 H48B818 H48B927 H48C012 H48C112 H48C199 H48C302 H48A594 H48B142 H48B367 H48B522 H48B647 H48B738 H48B819 H48B928 H48C013 H48C115 H48C200 H48C303 H48A595 H48B150 H48B369 H48B523 H48B648 H48B739 H48B820 H48B929 H48C014 H48C116 H48C203 H48C304 H48A612 H48B156 H48B370 H48B524 H48B649 H48B740 H48B821 H48B930 H48C015 H48C117 H48C204 H48C305 H48A624 H48B187 H48B371 H48B525 H48B650 H48B741 H48B822 H48B934 H48C016 H48C118 H48C205 H48C306 H48A625 H48B188 H48B372 H48B526 H48B657 H48B742 H48B828 H48B935 H48C017 H48C119 H48C206 H48C307 H48A626 H48B189 H48B373 H48B527 H48B658 H48B743 H48B829 H48B936 H48C019 H48C120 H48C207 H48C308 H48A627 H48B190 H48B374 H48B528 H48B660 H48B744 H48B830 H48B937 H48C020 H48C121 H48C208 H48C311 H48A628 H48B198 H48B375 H48B531 H48B661 H48B745 H48B831 H48B938 H48C022 H48C122 H48C212 H48C312 H48A629 H48B199 H48B377 H48B532 H48B662 H48B746 H48B832 H48B939 H48C024 H48C123 H48C217 H48C313 H48A658 H48B200 H48B378 H48B537 H48B663 H48B747 H48B833 H48B941 H48C025 H48C124 H48C218 H48C315 H48A659 H48B208 H48B379 H48B539 H48B664 H48B748 H48B839 H48B942 H48C029 H48C125 H48C219 H48C316 H48A660 H48B209 H48B380 H48B542 H48B667 H48B749 H48B840 H48B943 H48C030 H48C126 H48C220 H48C317 H48A661 H48B213 H48B381 H48B543 H48B668 H48B754 H48B841 H48B944 H48C031 H48C127 H48C221 H48C318 H48A662 H48B214 H48B382 H48B544 H48B669 H48B755 H48B846 H48B945 H48C032 H48C128 H48C222 H48C319 H48A663 H48B215 H48B385 H48B545 H48B670 H48B756 H48B847 H48B946 H48C033 H48C129 H48C225 H48C320 H48A664 H48B216 H48B386 H48B546 H48B671 H48B757 H48B848 H48B947 H48C034 H48C130 H48C226 H48C323 H48A679 H48B219 H48B387 H48B547 H48B672 H48B760 H48B849 H48B948 H48C035 H48C131 H48C227 H48C324 H48A692 H48B229 H48B388 H48B553 H48B673 H48B762 H48B850 H48B949 H48C036 H48C132 H48C228 H48C325 H48A712 H48B232 H48B399 H48B554 H48B674 H48B763 H48B852 H48B950 H48C037 H48C133 H48C229 H48C326 H48A772 H48B233 H48B407 H48B555 H48B675 H48B764 H48B853 H48B951 H48C038 H48C134 H48C239 H48C327 H48A773 H48B234 H48B408 H48B556 H48B676 H48B765 H48B854 H48B952 H48C039 H48C135 H48C240 H48C328 H48A774 H48B235 H48B419 H48B557 H48B677 H48B766 H48B855 H48B953 H48C040 H48C136 H48C241 H48C329 H48A812 H48B236 H48B424 H48B577 H48B678 H48B767 H48B866 H48B954 H48C041 H48C137 H48C242 H48C335 H48A813 H48B237 H48B426 H48B579 H48B679 H48B768 H48B867 H48B955 H48C044 H48C138 H48C243 H48C336 H48A827 H48B238 H48B429 H48B580 H48B680 H48B774 H48B868 H48B956 H48C045 H48C139 H48C244 H48C337 H48A840 H48B243 H48B431 H48B583 H48B681 H48B775 H48B869 H48B957 H48C046 H48C140 H48C245 H48C338 H48A841 H48B244 H48B432 H48B586 H48B682 H48B776 H48B870 H48B958 H48C054 H48C141 H48C246 H48C339 H48A866 H48B245 H48B436 H48B587 H48B683 H48B777 H48B873 H48B959 H48C059 H48C142 H48C247 H48C340 H48A898 H48B246 H48B437 H48B588 H48B684 H48B778 H48B875 H48B960 H48C060 H48C143 H48C248 H48C341 H48A899 H48B254 H48B438 H48B589 H48B685 H48B779 H48B876 H48B961 H48C062 H48C144 H48C249 H48C342 H48A900 H48B255 H48B439 H48B590 H48B686 H48B781 H48B877 H48B963 H48C063 H48C146 H48C250 H48C343 H48A901 H48B256 H48B440 H48B591 H48B687 H48B782 H48B879 H48B964 H48C064 H48C147 H48C251 H48C344 H48A902 H48B257 H48B441 H48B592 H48B688 H48B783 H48B880 H48B965 H48C065 H48C151 H48C252 H48C345 H48A903 H48B258 H48B442 H48B593 H48B689 H48B784 H48B881 H48B966 H48C066 H48C157 H48C253 H48C346 H48A911 H48B263 H48B443 H48B594 H48B690 H48B787 H48B885 H48B967 H48C068 H48C158 H48C254 H48C347 H48A918 H48B293 H48B444 H48B599 H48B695 H48B788 H48B886 H48B968 H48C073 H48C159 H48C255 H48C348 H48C349 H48C363 H48C383 H48C402 H48C419 H48C441 H48C471 H48C488 H48C517 H48C527 H48C552 H48C350 H48C364 H48C384 H48C403 H48C420 H48C442 H48C472 H48C489 H48C518 H48C528 H48C555 H48C351 H48C367 H48C388 H48C406 H48C421 H48C443 H48C473 H48C490 H48C519 H48C529 H48C556 H48C352 H48C368 H48C389 H48C407 H48C422 H48C446 H48C474 H48C491 H48C520 H48C530 H48C565 H48C353 H48C373 H48C390 H48C408 H48C423 H48C447 H48C475 H48C492 H48C521 H48C531 H48C588 H48C354 H48C375 H48C397 H48C412 H48C425 H48C458 H48C477 H48C504 H48C522 H48C532 H48C589 H48C355 H48C378 H48C398 H48C413 H48C427 H48C467 H48C478 H48C505 H48C523 H48C533 H48C358 H48C379 H48C399 H48C414 H48C428 H48C468 H48C484 H48C510 H48C524 H48C549 H48C359 H48C380 H48C400 H48C415 H48C429 H48C469 H48C486 H48C515 H48C525 H48C550 H48C362 H48C382 H48C401 H48C416 H48C434 H48C470 H48C487 H48C516 H48C526 H48C551 |
| FEI Number |
2916710
|
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc. Oncology Systems
911 Hansen Way
Palo Alto
CA
94304-1028
|
| For Additional Information Contact | Michael Pignataro 650-424-6471 |
Manufacturer Reason for Recall | In affected Eclipse versions, the treatment field energy may change from the current selection to a previous selection without use intervention and without invalidating the MU or calculated dose distribution. |
FDA Determined Cause 2 | Software design |
| Action | Varian sent an "Urgent Medical Device Correction/Urgent Field Safety Notice" letter dated June 30, 2011 to all affected consignees. The letter identifies the affected product, problem, and actions to be taken. The letter states that Varian will contact all affected customers to apply a database configuration script to the database and is preparing a correction for Eclipse Version 10 to address the issue. A Varian representative will contact each site to schedule correction when it is available. The letter also provides RECOMMENDED USER ACTIONS to follow. The Letters go on to reiterate caution and warnings from Eclipse instructions for use. All appropriate personnel should be notified of the correction notice. For future reference, this document will be posted to the Varian customer support website: http://www.MyVarian.com. For further clarification, contact your local Varian Customer Support District or Regional Manager. |
| Quantity in Commerce | 2436 |
| Distribution | Worldwide Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|