 Class 2 Device Recall Supreme Electrophysiology Extension Cablesee related information
Date Initiated by FirmJune 01, 2011
Date PostedAugust 05, 2011
Recall Status1 Terminated 3 on August 16, 2012
Recall NumberZ-2955-2011
Recall Event ID 59392
Product Classification Cable, electrode - Product Code IKD
ProductSupreme Electrophysiology Extension Cable, REF 401984, For use with Hexapolar Catheters, Sterile EO, Rx only, St Jude Medical, 14901 DeVeau Place, Minnetonka, MN 55345. Supreme Electrophysiology Extension Cable used to connect Electrophysiology catheters to external diagnostic or therapeutic equipment. The section indexing connector located a t the distal end of the cable will only connect to SMJ Electrophysiology catheters.
Code Information Batch # 3334664
FEI Number 2182269
Recalling Firm/
Manufacturer
St. Jude Medical
14901 DeVeau Place
Minnetonka MN 55345
For Additional Information Contact
952-933-4700
Manufacturer Reason
for Recall
St. Jude Medical is recalling a single lot (3334664) of Supreme Electrophysiology Extension Cable which were packaged with a Quadripolar cable instead of a hexapolar Cable. St. Jude Medical has determined that the incorrectly labeled product poses no safety risk since the extension cable can be safely used with a corresponding Quadripolar catheter but cannot be incorrectly connected to a Hexapolar catheter.
FDA Determined
Cause 2
Labeling mix-ups
ActionSt. Jude Medical sent an "IMPORTANT MEDICAL DEVICE FIELD ADVISORY NOTICE" letter dated June 8, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to locate the affected product and to complete and return an attached SJM AFD Reconciliation Form. The form will be used to replace product and as documentation that the customers received the notice. Fax the completed form to 877-497-031. Contact the Director of Product Marketing at 512-329-1054 for questions regarding this notice.
Quantity in Commerce32
DistributionNationwide Distribution-USA (nationwide) including the states of FL, NC, NY, and OR.