| | Class 2 Device Recall CDI 101 Hematocrit/Oxygen Saturation Monitoring System |  |
| Date Initiated by Firm | November 01, 2010 |
| Date Posted | February 12, 2013 |
| Recall Status1 |
Terminated 3 on November 08, 2013 |
| Recall Number | Z-0802-2013 |
| Recall Event ID |
60695 |
| Product Classification |
Monitor, blood-gas, on-line, cardiopulmonary bypass - Product Code DRY
|
| Product | Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels. |
| Code Information |
Catalog number: 101 and Serial Numbers: 0503, 1000, 1017, 1070, 1071, 1073, 1074, 1076, 1077, 1080, 1082, 1083, 1107-1110, 1115-1117,1124,1126-1134, 1136, 1137, 1139-1146, 1148-1152, 1154-1156, 1159-1162, 1164-1166, 1168-1177, 1182, 1185-1190, 1192,1193, 1195, 1197, 1198, 1225, 1228, 1229, 1237, 1239-1241,1243-1245,1247, 1249-1326, 1332-1339, 1344-1347, 1349, 1350-1363, 1365-1409, 1411-1436,1438, 1440-1447, 1449-1453, 1455-1463,1465-1482,1484-1492,1494-1552,1555-1565, 1567-1600,1602,1605-1607,1609-1611,1613-1623,1625-1629,1632-1651,1653, 1655-1660, 1662-1664, 1666, 1667,1669-1679,1681-1684,1686,1689-1691, 1693-1710,1712-1715,1717-1750,1752-1757,1759-1771,1773-1785,1787,1788,1790-1809, 1811-1831,1833,1838,1839,1855-1858,1866, and TK001-TK011. |
| FEI Number |
1828100
|
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corporation
6200 Jackson Road
Ann Arbor
MI
48103-9586
|
| For Additional Information Contact | 734-741-6173 |
Manufacturer Reason for Recall | Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications. |
FDA Determined Cause 2 | Software design |
| Action | Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 1, 2010 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions or concerns. |
| Quantity in Commerce | 622 units |
| Distribution | Worldwide Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|