U.S. Department of Health and Human Services

Class 1 Device Recall Curlin Infusion Administration Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Curlin Infusion Administration Setsee related information
Date Initiated by FirmMay 18, 2012
Date PostedMay 30, 2012
Recall Status1 Terminated 3 on February 22, 2013
Recall NumberZ-1659-2012
Recall Event ID 61895
510(K)NumberK981816 
Product Classification Tubing, fluid delivery - Product Code FPK
ProductNon-DEHP Microbore Administration Set with Non-Vented Spike Packaged Assembly, product code: 340-4114, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
Code Information CF1127990, CF1131190, CF1131191, CF1200290, CF1200291, CF1200292, CF1200490, CF1200491,  CF1201890, CF1202592, CF1204092, CF1204093, CF1205492, CF1205493, CF1205990, CF1206890, CF1206891.
FEI Number 1000117172
Recalling Firm/
Manufacturer
MOOG Medical Devices Group
4314 Zevex Park Ln
Salt Lake City UT 84123-7881
For Additional Information ContactMs. Laurie Brewer
801-264-1001 Ext. 224
Manufacturer Reason
for Recall
Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended.
FDA Determined
Cause 2
Process design
ActionThe firm, Moog Medical Device Group, sent an "Urgent Field Safety Notice" dated May 21, 2012 via certified letter and a press release direct to consignees/customers. The notice describes the product, problem and actions to be taken. The customers were instructed to do the following: If purchased by distributor -remove suspect product from inventory; contact your distributor to arrange replacement product and request a shipping return label as per the distributor's process and ship product to their location. If purchased by Moog-remove suspect product from inventory, contact Moog Customer Service at 1-800-970-2337, prompt #7 and Moog customer service will provide a call tag for convenient product return. For additional information or clarification, please contact the Post Market Surveillance Manager at (801) 264-1001, ext. 224 or via email at: lbrewer@moog.com.
Quantity in Commerce130,940 sets
DistributionNationwide distribution: USA including states of: AZ, CA, FL, IL, KS, MA, MI, MD, NC, ND, NY, OK, OH, and TX.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = FPK
-
-