| | Class 2 Device Recall Stryker Orthopaedics |  |
| Date Initiated by Firm | December 29, 2011 |
| Date Posted | July 10, 2012 |
| Recall Status1 |
Terminated 3 on March 19, 2015 |
| Recall Number | Z-1982-2012 |
| Recall Event ID |
61923 |
| 510(K)Number | K936127 |
| Product Classification |
Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
|
| Product | Stryker Orthopaedics, Definition PM Femoral Hip Stem Catalog Numbers 6265-3-099 through 6265-3-107. Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah, NJ 07430. A subsidiary of Stryker Corp. Stryker France, ZAC Satorlas Green Pusignan, Av De Satolas Green 69881 Meyzeiu, Cedex France. Made in Ireland.
The Definition PM femoral hip stem is a forged, collared, straight stem with an integral premantle (PM) design. The Poly Methyl Methacrylate (PMMA) and Barium Sulphate (BaSO4) premantle is integral to the design as it aligns, centralizes and pressurizes the stem within the femoral canal. The forged stem is first grit blasted proximally and then the premantle is applied through an infection molding process. |
| Code Information |
Listed as Catalogue Number followed by Lot Number: 6265-3-099 8024501 6265-3-099 8024502 6265-3-099 11160601 6265-3-099 11650201 6265-3-099 11650202 6265-3-099 12915801 6265-3-099 12915802 6265-3-099 12915803 6265-3-099 12915804 6265-3-099 12915805 6265-3-099 14027701 6265-3-099 14027702 6265-3-099 14503501 6265-3-099 14525001 6265-3-099 14525002 6265-3-099 14525003 6265-3-099 19943601 6265-3-099 19943602 6265-3-099 19943603 6265-3-099 19943604 6265-3-099 21500101 6265-3-099 21500102 6265-3-099 22108201 6265-3-099 22108202 6265-3-099 22260101 6265-3-099 23164201 6265-3-099 23731701 6265-3-099 23731801 6265-3-099 23731802 6265-3-099 23731803 6265-3-099 23843302 6265-3-099 23843303 6265-3-099 23843304 6265-3-099 23843305 6265-3-099 23843306 6265-3-099 23844101 6265-3-099 24321501 6265-3-099 24321502 6265-3-099 24321503 6265-3-099 24589701 6265-3-099 24589702 6265-3-099 24589703 6265-3-099 24589704 6265-3-099 25025901 6265-3-099 25025902 6265-3-099 25025903 6265-3-099 25025904 6265-3-099 27951601 6265-3-099 27951602 6265-3-099 27951603 6265-3-099 27951604 6265-3-099 27951605 6265-3-099 28008201 6265-3-100 18010702 6265-3-100 18363101 6265-3-100 18363102 6265-3-100 18363103 6265-3-100 18363104 6265-3-100 18981801 6265-3-100 19253701 6265-3-100 19253702 6265-3-100 19601801 6265-3-100 19602301 6265-3-100 19602302 6265-3-100 19898201 6265-3-100 19898202 6265-3-100 19898203 6265-3-100 19898204 6265-3-100 20507301 6265-3-100 20507302 6265-3-100 20542501 6265-3-100 20542502 6265-3-100 20953101 6265-3-100 22260201 6265-3-100 22260202 6265-3-100 22349901 6265-3-100 22349902 6265-3-100 22441501 6265-3-100 22441502 6265-3-100 22499401 6265-3-100 32573902 6265-3-100 34145101 6265-3-100 34145102 6265-3-100 34145103 6265-3-100 34145104 6265-3-100 34145105 6265-3-100 34145106 6265-3-100 34748701 6265-3-100 34748702 6265-3-100 34748901 6265-3-100 34748903 6265-3-100 34748904 6265-3-100 35484501 6265-3-100 35484502 6265-3-100 35484503 6265-3-100 35484504 6265-3-100 36275202 6265-3-100 10212101R 6265-3-100 10316601R 6265-3-100 10316602R 6265-3-100 12037201R 6265-3-100 18718001R 6265-3-100 6645204R 6265-3-100 7898301R 6265-3-100 9718501R 6265-3-100 D3KTER 6265-3-100 D3KTJR 6265-3-101 7766001 6265-3-101 7766002 6265-3-101 7766003 6265-3-101 7836503 6265-3-101 7836504 6265-3-101 7851301 6265-3-101 7851302 6265-3-101 7851303 6265-3-101 8024701 6265-3-101 8024702 6265-3-101 8154301 6265-3-101 8154302 6265-3-101 8154303 6265-3-101 8220101 6265-3-101 8220102 6265-3-101 8367803 6265-3-101 8367804 6265-3-101 8908201 6265-3-101 8931701 6265-3-101 8931702 6265-3-101 8939001 6265-3-101 8965501 6265-3-101 8965502 6265-3-101 9027301 6265-3-101 9027302 6265-3-101 9051301 6265-3-101 9051302 6265-3-101 11335201 6265-3-101 11335202 6265-3-101 11335203 6265-3-101 11335204 6265-3-101 11495201 6265-3-101 11528501 6265-3-101 11528502 6265-3-101 11528503 6265-3-101 11790801 6265-3-101 11790802 6265-3-101 11790803 6265-3-101 11978401 6265-3-101 11978402 6265-3-101 11978403 6265-3-101 11978404 6265-3-101 11978405 6265-3-101 11995702 6265-3-101 11995703 6265-3-101 11995704 6265-3-101 12064201 6265-3-101 12064203 6265-3-101 12144701 6265-3-101 12144702 6265-3-101 12144703 6265-3-101 12144704 6265-3-101 12144705 6265-3-101 12725101 6265-3-101 14106902 6265-3-101 14150301 6265-3-101 14150302 6265-3-101 14150303 6265-3-101 14150304 6265-3-101 14422501 6265-3-101 14422502 6265-3-101 14422503 6265-3-101 14422504 6265-3-101 14422505 6265-3-101 14422506 6265-3-101 14468602 6265-3-101 14468604 6265-3-101 14468605 6265-3-101 14468606 6265-3-101 14485701 6265-3-101 14485702 6265-3-101 14485703 6265-3-101 14485704 6265-3-101 14485705 6265-3-101 14485706 6265-3-101 14504001 6265-3-101 14504002 6265-3-101 14504003 6265-3-101 14504004 6265-3-101 14504005 6265-3-101 14524901 6265-3-101 17674804 6265-3-101 17674805 6265-3-101 17674806 6265-3-101 17677701 6265-3-101 17878301 6265-3-101 17878302 6265-3-101 17878303 6265-3-101 17878304 6265-3-101 17878305 6265-3-101 17878306 6265-3-101 18036701 6265-3-101 18036702 6265-3-101 18036703 6265-3-101 18036704 6265-3-101 18036705 6265-3-101 18036706 6265-3-101 18982001 6265-3-101 18982002 6265-3-101 18998701 6265-3-101 18998702 6265-3-101 19013101 6265-3-101 19013102 6265-3-101 19866201 6265-3-101 19866202 6265-3-101 19898301 6265-3-101 19898302 6265-3-101 19898303 6265-3-101 22309504 6265-3-101 22441701 6265-3-101 22441702 6265-3-101 22441703 6265-3-101 22441704 6265-3-101 22461401 6265-3-101 22461402 6265-3-101 22499101 6265-3-101 22499102 6265-3-101 22499103 6265-3-101 22499104 6265-3-101 22499105 6265-3-101 22499106 6265-3-101 22499107 6265-3-101 22499108 6265-3-101 22602101 6265-3-101 22602102 6265-3-101 22602103 6265-3-101 22602104 6265-3-101 22614801 6265-3-101 22614802 6265-3-101 22614803 6265-3-101 22614804 6265-3-101 22924701 6265-3-101 22924702 6265-3-101 22924703 6265-3-101 22924704 6265-3-101 27364602 6265-3-101 27364603 6265-3-101 27364604 6265-3-101 27364605 6265-3-101 27561901 6265-3-101 27561902 6265-3-101 27561903 6265-3-101 27561904 6265-3-101 27562001 6265-3-101 27562002 6265-3-101 27813401 6265-3-101 27813402 6265-3-101 27813403 6265-3-101 27813404 6265-3-101 27813405 6265-3-101 27813406 6265-3-101 28384202 6265-3-101 28384203 6265-3-101 28384204 6265-3-101 28384205 6265-3-101 28528101 6265-3-101 28528102 6265-3-101 28528103 6265-3-101 28528104 6265-3-101 28732901 6265-3-101 28732902 6265-3-101 28732903 6265-3-101 32528304 6265-3-101 33091001 6265-3-101 33091002 6265-3-101 33091003 6265-3-101 33091004 6265-3-101 33746301 6265-3-101 33746302 6265-3-101 33746303 6265-3-101 33746304 6265-3-101 33746401 6265-3-101 33746402 6265-3-101 33746403 6265-3-101 33746404 6265-3-101 34275301 6265-3-101 34275302 6265-3-101 34275303 6265-3-101 34275304 6265-3-101 34617201 6265-3-101 34617202 6265-3-101 34617204 6265-3-101 34617205 6265-3-101 34617206 6265-3-101 34617207 6265-3-101 34617208 6265-3-101 34814301 6265-3-101 34814302 6265-3-101 34916401 6265-3-102 6154103 6265-3-102 6166001 6265-3-102 6166002 6265-3-102 6238201 6265-3-102 6238202 6265-3-102 6238203 6265-3-102 6238204 6265-3-102 6470601 6265-3-102 6470602 6265-3-102 6470603 6265-3-102 6470604 6265-3-102 6470605 6265-3-102 6470606 6265-3-102 6519101 6265-3-102 6519102 6265-3-102 6519103 6265-3-102 6519104 6265-3-102 6519105 6265-3-102 6519106 6265-3-102 6519201 6265-3-102 6519202 6265-3-102 6519203 6265-3-102 6519204 6265-3-102 6519205 6265-3-102 6519206 6265-3-102 6551501 6265-3-102 6551503 6265-3-102 7596903 6265-3-102 7596904 6265-3-102 7695101 6265-3-102 7695102 6265-3-102 7695103 6265-3-102 7695104 6265-3-102 7703601 6265-3-102 7703602 6265-3-102 7703603 6265-3-102 7703604 6265-3-102 7755501 6265-3-102 7755502 6265-3-102 7755503 6265-3-102 7755504 6265-3-102 7766101 6265-3-102 7766102 6265-3-102 7766103 6265-3-102 7766104 6265-3-102 7823702 6265-3-102 7823703 6265-3-102 7823704 6265-3-102 7836601 6265-3-102 7836602 6265-3-102 7836603 6265-3-102 7836604 6265-3-102 7851401 6265-3-102 7851402 6265-3-102 9524702 6265-3-102 9524703 6265-3-102 9541601 6265-3-102 9541602 6265-3-102 9567801 6265-3-102 9567802 6265-3-102 9640701 6265-3-102 9640702 6265-3-102 9640703 6265-3-102 9693301 6265-3-102 9693303 6265-3-102 10004401 6265-3-102 10004402 6265-3-102 10004403 6265-3-102 10004404 6265-3-102 10025502 6265-3-102 10025503 6265-3-102 10025504 6265-3-102 10026401 6265-3-102 10211601 6265-3-102 10211602 6265-3-102 10211603 6265-3-102 10211604 6265-3-102 10262001 6265-3-102 10262002 6265-3-102 10262003 6265-3-102 10262004 6265-3-102 11995803 6265-3-102 11995804 6265-3-102 12037301 6265-3-102 12037302 6265-3-102 12037303 6265-3-102 12037304 6265-3-102 12037305 6265-3-102 12064301 6265-3-102 12064302 6265-3-102 12064303 6265-3-102 12144802 6265-3-102 12144803 6265-3-102 12144804 6265-3-102 12337501 6265-3-102 12337502 6265-3-102 12337503 6265-3-102 12337504 6265-3-102 12549701 6265-3-102 12549702 6265-3-102 12549704 6265-3-102 12549705 6265-3-102 12549706 6265-3-102 12549707 6265-3-102 12549708 6265-3-102 12725205 6265-3-102 12725206 6265-3-102 12725207 6265-3-102 14150403 6265-3-102 14150404 6265-3-102 14172701 6265-3-102 14172702 6265-3-102 14172703 6265-3-102 14172704 6265-3-102 14195901 6265-3-102 14195902 6265-3-102 14195903 6265-3-102 14195904 6265-3-102 14255501 6265-3-102 14255502 6265-3-102 14255503 6265-3-102 14267501 6265-3-102 14267502 6265-3-102 14267503 6265-3-102 14468701 6265-3-102 14468702 6265-3-102 14504601 6265-3-102 14504603 6265-3-102 14504604 6265-3-102 14594701 6265-3-102 14594702 6265-3-102 14644601 6265-3-102 14644602 6265-3-102 14644603 6265-3-102 14644604 6265-3-102 16800701 6265-3-102 16800702 6265-3-102 16800703 6265-3-102 16935901 6265-3-102 16935902 6265-3-102 16935903 6265-3-102 16935904 6265-3-102 16935905 6265-3-102 16935906 6265-3-102 17091101 6265-3-102 17091102 6265-3-102 17136301 6265-3-102 17136302 6265-3-102 17136303 6265-3-102 17165801 6265-3-102 17540001 6265-3-102 17540002 6265-3-102 17540003 6265-3-102 17540004 6265-3-102 17540005 6265-3-102 17540006 6265-3-102 17592301 6265-3-102 17592302 6265-3-102 17592304 6265-3-102 17592305 6265-3-102 17592306 6265-3-102 17658701 6265-3-102 19344202 6265-3-102 19411402 6265-3-102 19411403 6265-3-102 19786301 6265-3-102 19786302 6265-3-102 19786303 6265-3-102 19786304 6265-3-102 19866301 6265-3-102 19866302 6265-3-102 19874901 6265-3-102 19874902 6265-3-102 19898401 6265-3-102 19898402 6265-3-102 19898403 6265-3-102 19909801 6265-3-102 19909802 6265-3-102 19909803 6265-3-102 19909804 6265-3-102 19909805 6265-3-102 19909806 6265-3-102 19959801 6265-3-102 19959802 6265-3-102 19959803 6265-3-102 19959804 6265-3-102 19959805 6265-3-102 19959806 6265-3-102 20058401 6265-3-102 22403006 6265-3-102 22499601 6265-3-102 22499602 6265-3-102 22615001 6265-3-102 22615002 6265-3-102 22658001 6265-3-102 22658002 6265-3-102 22707901 6265-3-102 22707902 6265-3-102 22901401 6265-3-102 22901402 6265-3-102 22901501 6265-3-102 22901502 6265-3-102 22901503 6265-3-102 22901504 6265-3-102 22921101 6265-3-102 22921102 6265-3-102 22921103 6265-3-102 22921104 6265-3-102 22999401 6265-3-102 22999402 6265-3-102 22999403 6265-3-102 22999404 6265-3-102 22999405 6265-3-102 23096101 6265-3-102 23096102 6265-3-102 23096103 6265-3-102 26043802 6265-3-102 26043804 6265-3-102 26349001 6265-3-102 26349002 6265-3-102 26349003 6265-3-102 26349004 6265-3-102 26349201 6265-3-102 26511001 6265-3-102 26511002 6265-3-102 26511003 6265-3-102 26511004 6265-3-102 26598701 6265-3-102 26633101 6265-3-102 26633102 6265-3-102 26742001 6265-3-102 26742002 6265-3-102 26742101 6265-3-102 26742102 6265-3-102 26742103 6265-3-102 26742104 6265-3-102 26967601 6265-3-102 26967602 6265-3-102 26967603 6265-3-102 26967604 6265-3-102 26967606 6265-3-102 27071701 6265-3-102 27071702 6265-3-102 29818606 6265-3-102 29818607 6265-3-102 29818608 6265-3-102 30139201 6265-3-102 30139401 6265-3-102 30139402 6265-3-102 30139403 6265-3-102 30139404 6265-3-102 30360501 6265-3-102 30360502 6265-3-102 30360503 6265-3-102 30360504 6265-3-102 30360505 6265-3-102 30614501 6265-3-102 30614502 6265-3-102 30614503 6265-3-102 30614601 6265-3-102 30614602 6265-3-102 30614603 6265-3-102 30614604 6265-3-102 30614605 6265-3-102 30836701 6265-3-102 30836702 6265-3-102 30836703 6265-3-102 30836704 6265-3-102 30836705 6265-3-102 30836706 6265-3-102 34932304 6265-3-102 35030901 6265-3-102 35030902 6265-3-102 35030903 6265-3-102 35030904 6265-3-102 35313301 6265-3-102 35313302 6265-3-102 35313303 6265-3-102 35313304 6265-3-102 35430301 6265-3-102 35430303 6265-3-102 10262005A 6265-3-102 11709704S 6265-3-102 14028603R 6265-3-102 15869606R 6265-3-102 16153406R 6265-3-102 18947304R 6265-3-102 19013201R 6265-3-102 19013203A 6265-3-102 19246302R 6265-3-102 19909806S 6265-3-102 19959802R 6265-3-102 19959806R 6265-3-102 20058401R 6265-3-102 20079101R 6265-3-102 21043302R 6265-3-102 30893103R 6265-3-103 6397905 6265-3-103 6414801 6265-3-103 6414802 6265-3-103 6414901 6265-3-103 6414902 6265-3-103 6414903 6265-3-103 6487101 6265-3-103 6487102 6265-3-103 6487103 6265-3-103 6487104 6265-3-103 6487105 6265-3-103 6609201 6265-3-103 6609202 6265-3-103 6609203 6265-3-103 6609204 6265-3-103 6609205 6265-3-103 6609207 6265-3-103 6679101 6265-3-103 6679102 6265-3-103 6679103 6265-3-103 6679104 6265-3-103 6679105 6265-3-103 6679106 6265-3-103 6679107 6265-3-103 6717401 6265-3-103 6717402 6265-3-103 6717403 6265-3-103 7535301 6265-3-103 7535302 6265-3-103 7535304 6265-3-103 7597101 6265-3-103 7597102 6265-3-103 7597103 6265-3-103 7597104 6265-3-103 7620301 6265-3-103 7620302 6265-3-103 7620303 6265-3-103 7620304 6265-3-103 7639801 6265-3-103 7639802 6265-3-103 7639803 6265-3-103 7639804 6265-3-103 7679201 6265-3-103 7679202 6265-3-103 7679203 6265-3-103 7679204 6265-3-103 7695001 6265-3-103 7695002 6265-3-103 7695003 6265-3-103 7695004 6265-3-103 7703801 6265-3-103 7703802 6265-3-103 7703803 6265-3-103 7703804 6265-3-103 9205302 6265-3-103 9205303 6265-3-103 9205304 6265-3-103 9205305 6265-3-103 9205306 6265-3-103 9277601 6265-3-103 9277602 6265-3-103 9277603 6265-3-103 9277605 6265-3-103 9277606 6265-3-103 9342101 6265-3-103 9342102 6265-3-103 9342103 6265-3-103 9342104 6265-3-103 9342105 6265-3-103 9342106 6265-3-103 9421601 6265-3-103 9421603 6265-3-103 9421604 6265-3-103 9421605 6265-3-103 9421901 6265-3-103 9421904 6265-3-103 9447601 6265-3-103 9447602 6265-3-103 9498601 6265-3-103 9498701 6265-3-103 9498801 6265-3-103 11060301 6265-3-103 11060302 6265-3-103 11060303 6265-3-103 11060304 6265-3-103 11078402 6265-3-103 11078403 6265-3-103 11078404 6265-3-103 11225601 6265-3-103 11253901 6265-3-103 11253902 6265-3-103 11253903 6265-3-103 11310001 6265-3-103 11310002 6265-3-103 11310003 6265-3-103 11310004 6265-3-103 11310005 6265-3-103 11310006 6265-3-103 11451201 6265-3-103 11451202 6265-3-103 11496602 6265-3-103 11496603 6265-3-103 11496604 6265-3-103 11503802 6265-3-103 11503803 6265-3-103 11718001 6265-3-103 11718002 6265-3-103 11718003 6265-3-103 12721504 6265-3-103 12721505 6265-3-103 12721506 6265-3-103 12721507 6265-3-103 12812501 6265-3-103 12812502 6265-3-103 12812503 6265-3-103 12812601 6265-3-103 12812701 6265-3-103 12812702 6265-3-103 12812901 6265-3-103 12812902 6265-3-103 12812903 6265-3-103 12812904 6265-3-103 12916501 6265-3-103 12916502 6265-3-103 12916503 6265-3-103 12916505 6265-3-103 12978901 6265-3-103 12978902 6265-3-103 12978903 6265-3-103 12978904 6265-3-103 12978905 6265-3-103 12978906 6265-3-103 12978907 6265-3-103 12978908 6265-3-103 13007801 6265-3-103 14107302 6265-3-103 14107303 6265-3-103 14150501 6265-3-103 14150502 6265-3-103 14150503 6265-3-103 14150504 6265-3-103 14172901 6265-3-103 14172902 6265-3-103 14172903 6265-3-103 14172904 6265-3-103 14196001 6265-3-103 14196003 6265-3-103 14254101 6265-3-103 14254102 6265-3-103 14254103 6265-3-103 14267601 6265-3-103 14267602 6265-3-103 14505201 6265-3-103 14505202 6265-3-103 14505203 6265-3-103 14505204 6265-3-103 14644701 6265-3-103 14644702 6265-3-103 14644703 6265-3-103 14644704 6265-3-103 14644705 6265-3-103 14644706 6265-3-103 17050101 6265-3-103 17050102 6265-3-103 17050103 6265-3-103 17050104 6265-3-103 17050105 6265-3-103 17091001 6265-3-103 17091002 6265-3-103 17091003 6265-3-103 17091004 6265-3-103 17322701 6265-3-103 17322702 6265-3-103 17322703
6265-3-103 17386201 6265-3-103 17386202 6265-3-103 17386203 6265-3-103 17386204 6265-3-103 17491301 6265-3-103 17491302 6265-3-103 17491303 6265-3-103 17711001 6265-3-103 17711002 6265-3-103 17711003 6265-3-103 17711004 6265-3-103 17842601 6265-3-103 17842602 6265-3-103 17842603 6265-3-103 17842604 6265-3-103 19832802 6265-3-103 19875001 6265-3-103 19875002 6265-3-103 19875003 6265-3-103 19875004 6265-3-103 19909701 6265-3-103 19909702 6265-3-103 19909703 6265-3-103 19909704 6265-3-103 19909705 6265-3-103 19909706 6265-3-103 19944101 6265-3-103 19944102 6265-3-103 19944103 6265-3-103 19944104 6265-3-103 19944105 6265-3-103 19944107 6265-3-103 19944108 6265-3-103 19944109 6265-3-103 19944110 6265-3-103 19944111 6265-3-103 20058501 6265-3-103 20058502 6265-3-103 20059101 6265-3-103 20096302 6265-3-103 20161401 6265-3-103 20161402 6265-3-103 22707603 6265-3-103 22707604 6265-3-103 22812901 6265-3-103 22812902 6265-3-103 22901601 6265-3-103 22901602 6265-3-103 22901603 6265-3-103 22901604 6265-3-103 22921201 6265-3-103 22921202 6265-3-103 22921203 6265-3-103 22921204 6265-3-103 22950801 6265-3-103 22950802 6265-3-103 22999501 6265-3-103 22999502 6265-3-103 22999503 6265-3-103 22999504 6265-3-103 23096201 6265-3-103 23096202 6265-3-103 23096203 6265-3-103 23096204 6265-3-103 23205701 6265-3-103 23410501 6265-3-103 23433601 6265-3-103 23455601 6265-3-103 23515801 6265-3-103 25357603 6265-3-103 25357604 6265-3-103 25357605 6265-3-103 25357606 6265-3-103 25357607 6265-3-103 25637301 6265-3-103 25637302 6265-3-103 25637303 6265-3-103 25637304 6265-3-103 25637305 6265-3-103 25637306 6265-3-103 25798401 6265-3-103 25798402 6265-3-103 25798403 6265-3-103 25798404 6265-3-103 25940301 6265-3-103 25940303 6265-3-103 25940304 6265-3-103 25940305 6265-3-103 26115501 6265-3-103 26115503 6265-3-103 26115504 6265-3-103 26289901 6265-3-103 26289902 6265-3-103 26289903 6265-3-103 26289904 6265-3-103 26369301 6265-3-103 28331606 6265-3-103 28331607 6265-3-103 28331608 6265-3-103 28570101 6265-3-103 28570102 6265-3-103 28570103 6265-3-103 28570104 6265-3-103 28570105 6265-3-103 28570106 6265-3-103 28584701 6265-3-103 28584702 6265-3-103 28584703 6265-3-103 28584704 6265-3-103 28584705 6265-3-103 28584706 6265-3-103 28890601 6265-3-103 28890602 6265-3-103 28890603 6265-3-103 28890604 6265-3-103 29089802 6265-3-103 29089803 6265-3-103 29164901 6265-3-103 29164902 6265-3-103 29164903 6265-3-103 29164904 6265-3-103 29367701 6265-3-103 29387402 6265-3-103 31113004 6265-3-103 31113005 6265-3-103 31692201 6265-3-103 31692202 6265-3-103 31692203 6265-3-103 31692204 6265-3-103 31919001 6265-3-103 31919003 6265-3-103 31975401 6265-3-103 31975402 6265-3-103 31975403 6265-3-103 31975404 6265-3-103 32426201 6265-3-103 32426202 6265-3-103 32426203 6265-3-103 32426204 6265-3-103 32450901 6265-3-103 32450902 6265-3-103 32450903 6265-3-103 32738301 6265-3-103 32738302 6265-3-103 32738303 6265-3-103 32738304 6265-3-103 33374701 6265-3-103 33374702 6265-3-103 33374703 6265-3-103 33374704 6265-3-104 5351501 6265-3-104 5351502 6265-3-104 5471901 6265-3-104 5471902 6265-3-104 5657201 6265-3-104 5657202 6265-3-104 5657203 6265-3-104 5779502 6265-3-104 5779503 6265-3-104 5856602 6265-3-104 6030901 6265-3-104 6031001 6265-3-104 6154301 6265-3-104 6154302 6265-3-104 6154303 6265-3-104 6154304 6265-3-104 6275701 6265-3-104 6275702 6265-3-104 6275703 6265-3-104 6275704 6265-3-104 6275705 6265-3-104 6275706 6265-3-104 6292401 6265-3-104 6292403 6265-3-104 6292404 6265-3-104 6541901 6265-3-104 6541902 6265-3-104 9033401 6265-3-104 9033402 6265-3-104 9047901 6265-3-104 9047902 6265-3-104 9047903 6265-3-104 9047904 6265-3-104 9099701 6265-3-104 9099702 6265-3-104 9139301 6265-3-104 9139302 6265-3-104 9139303 6265-3-104 9205401 6265-3-104 9238901 6265-3-104 9277901 6265-3-104 9277902 6265-3-104 9341701 6265-3-104 9341702 6265-3-104 9341703 6265-3-104 9390801 6265-3-104 9390802 6265-3-104 9447701 6265-3-104 9499001 6265-3-104 9499101 6265-3-104 9499301 6265-3-104 9524801 6265-3-104 9524802 6265-3-104 9567601 6265-3-104 12725303 6265-3-104 12725304 6265-3-104 12725305 6265-3-104 12725306 6265-3-104 12725307 6265-3-104 13290801 6265-3-104 13290802 6265-3-104 13290803 6265-3-104 13290804 6265-3-104 13290805 6265-3-104 13290806 6265-3-104 13290807 6265-3-104 13290808 6265-3-104 13660401 6265-3-104 13660402 6265-3-104 13660403 6265-3-104 13660404 6265-3-104 13660405 6265-3-104 13660406 6265-3-104 13660407 6265-3-104 13660408 6265-3-104 14029201 6265-3-104 14029202 6265-3-104 14029203 6265-3-104 14029204 6265-3-104 14107401 6265-3-104 14107402 6265-3-104 17941801 6265-3-104 17941802 6265-3-104 17941803 6265-3-104 17941804 6265-3-104 18109601 6265-3-104 18109701 6265-3-104 18180301 6265-3-104 18180302 6265-3-104 18180303 6265-3-104 18180304 6265-3-104 18180305 6265-3-104 18180306 6265-3-104 18675001 6265-3-104 18675002 6265-3-104 18675003 6265-3-104 18675004 6265-3-104 19013401 6265-3-104 19013402 6265-3-104 19106801 6265-3-104 19106802 6265-3-104 19328301 6265-3-104 19328302 6265-3-104 19531301 6265-3-104 19531302 6265-3-104 19566001 6265-3-104 19566002 6265-3-104 19566003 6265-3-104 22977101 6265-3-104 22977102 6265-3-104 22977103 6265-3-104 22977104 6265-3-104 22977105 6265-3-104 23798001 6265-3-104 23798002 6265-3-104 23798003 6265-3-104 23798004 6265-3-104 23798005 6265-3-104 23798006 6265-3-104 23984701 6265-3-104 23984702 6265-3-104 23984703 6265-3-104 23984704 6265-3-104 23984705 6265-3-104 23984706 6265-3-104 24395101 6265-3-104 24395102 6265-3-104 24395103 6265-3-104 24395105 6265-3-104 24395106 6265-3-104 24395107 6265-3-104 24395108 6265-3-104 24395110 6265-3-104 25094101 6265-3-104 25094102 6265-3-104 29807602 6265-3-104 29807603 6265-3-104 29807801 6265-3-104 29807802 6265-3-104 29807803 6265-3-104 30214401 6265-3-104 30214402 6265-3-104 30214403 6265-3-104 30214404 6265-3-104 30214405 6265-3-104 30214406 6265-3-104 30214407 6265-3-104 30214408 6265-3-104 30944502 6265-3-104 30944503 6265-3-104 30944504 6265-3-104 31009402 6265-3-104 31009403 6265-3-104 31009404 6265-3-104 31704401 6265-3-104 31704402 6265-3-104 31704404 6265-3-104 31975601 6265-3-104 31975602 6265-3-104 31975603 6265-3-104 31975604 6265-3-104 32543401 6265-3-105 6470702 6265-3-105 6470703 6265-3-105 6470704 6265-3-105 6470705 6265-3-105 6470706 6265-3-105 8908601 6265-3-105 9139501 6265-3-105 9205501 6265-3-105 9277801 6265-3-105 9277802 6265-3-105 9390901 6265-3-105 9448201 6265-3-105 9524901 6265-3-105 9693501 6265-3-105 9693502 6265-3-105 9777801 6265-3-105 9815801 6265-3-105 9815802 6265-3-105 10082901 6265-3-105 10082902 6265-3-105 10163201 6265-3-105 10163202 6265-3-105 10212001 6265-3-105 10212002 6265-3-105 10262301 6265-3-105 10262302 6265-3-105 10368401 6265-3-105 17540103 6265-3-105 17592401 6265-3-105 17592402 6265-3-105 17658601 6265-3-105 17658602 6265-3-105 17658603 6265-3-105 18154401 6265-3-105 18154402 6265-3-105 18154403 6265-3-105 18180401 6265-3-105 18363301 6265-3-105 18363302 6265-3-105 18717801 6265-3-105 18717802 6265-3-105 18998901 6265-3-105 18998902 6265-3-105 19531801 6265-3-105 19531802 6265-3-105 19576101 6265-3-105 19576102 6265-3-105 19826301 6265-3-105 19826302 6265-3-105 19959901 6265-3-105 19959902 6265-3-105 19959903 6265-3-105 19959904 6265-3-105 19959905 6265-3-105 27795401 6265-3-105 27795402 6265-3-105 27795403 6265-3-105 27795404 6265-3-105 27964501 6265-3-105 27964502 6265-3-105 28008401 6265-3-105 29048701 6265-3-105 29048702 6265-3-105 29048703 6265-3-105 29048704 6265-3-105 31176301 6265-3-105 31176302 6265-3-105 31598401 6265-3-105 31598402 6265-3-105 31598403 6265-3-105 31598404 6265-3-105 31598405 6265-3-105 31598406 6265-3-105 32493201 6265-3-105 32493202 6265-3-105 32493203 6265-3-105 32493204 6265-3-105 33110201 6265-3-105 33110202 6265-3-105 33110203 6265-3-105 34204003 6265-3-106 13514208 6265-3-106 13846701 6265-3-106 14029601 6265-3-106 14029602 6265-3-106 14029603 6265-3-106 14618201 6265-3-106 14632701 6265-3-106 18075602 6265-3-106 18075603 6265-3-106 18592001 6265-3-106 18592002 6265-3-106 19833801 6265-3-106 19833802 6265-3-106 19898101 6265-3-106 19898102 6265-3-106 19898103 6265-3-106 20131501 6265-3-106 20131502 6265-3-106 21531101 6265-3-106 21572401 6265-3-106 21862401 6265-3-106 22351001 6265-3-106 22420401 6265-3-106 22420402 6265-3-106 22420403 6265-3-106 22420404 6265-3-106 22420405 6265-3-106 13514204R 6265-3-106 18075601A 6265-3-106 18075601R 6265-3-106 5736401A 6265-3-107 5921701 6265-3-107 6117101 6265-3-107 6117102 6265-3-107 7836701 6265-3-107 7836702 6265-3-107 7836703 6265-3-107 7836704 6265-3-107 8908801 6265-3-107 9500201 6265-3-107 10004601 6265-3-107 10004602 6265-3-107 10220201 6265-3-107 11275801 6265-3-107 11275802 6265-3-107 12037601 6265-3-107 12037602 6265-3-107 12064401 6265-3-107 12064402 6265-3-107 12725601 6265-3-107 12725602 6265-3-107 12725603 6265-3-107 12725604 6265-3-107 13190201 6265-3-107 31761506 6265-3-107 31761507 6265-3-107 31761508 6265-3-107 32560302 6265-3-099 28025701 6265-3-099 28025702 6265-3-099 31975802 6265-3-099 19943601S 6265-3-099 8024502R 6265-3-099 D1SABR 6265-3-099 D1SAER 6265-3-099 D2ACDR 6265-3-099 D2FRBR 6265-3-100 6373601 6265-3-100 6373602 6265-3-100 6373603 6265-3-100 6373604 6265-3-100 6373605 6265-3-100 6373606 6265-3-100 6645201 6265-3-100 6645202 6265-3-100 6645203 6265-3-100 6645204 6265-3-100 7075801 6265-3-100 7075802 6265-3-100 7075803 6265-3-100 7075804 6265-3-100 7152901 6265-3-100 7152902 6265-3-100 7152903 6265-3-100 7299301 6265-3-100 7299302 6265-3-100 7299303 6265-3-100 7299304 6265-3-100 7299305 6265-3-100 7299306 6265-3-100 7492601 6265-3-100 7492602 6265-3-100 7492603 6265-3-100 7639701 6265-3-100 7639702 6265-3-100 7898301 6265-3-100 7898701 6265-3-100 8024601 6265-3-100 8024602 6265-3-100 8024603 6265-3-100 9448101 6265-3-100 9525001 6265-3-100 9640801 6265-3-100 9718501 6265-3-100 9777401 6265-3-100 9850001 6265-3-100 9850002 6265-3-100 9850003 6265-3-100 10212101 6265-3-100 10316601 6265-3-100 10316602 6265-3-100 22499402 6265-3-100 22999101 6265-3-100 22999201 6265-3-100 22999202 6265-3-100 22999203 6265-3-100 22999204 6265-3-100 24027701 6265-3-100 24027702 6265-3-100 24027703 6265-3-100 24027704 6265-3-100 25547101 6265-3-100 25547102 6265-3-100 25547103 6265-3-100 25547104 6265-3-100 26598601 6265-3-100 26765701 6265-3-100 26765702 6265-3-100 26765703 6265-3-100 26765704 6265-3-100 26765705 6265-3-100 26765706 6265-3-100 27795201 6265-3-100 27795202 6265-3-100 27795203 6265-3-100 27795204 6265-3-100 27951701 6265-3-100 27951702 6265-3-101 5694301 6265-3-101 5694302 6265-3-101 5694501 6265-3-101 5750203 6265-3-101 5974601 6265-3-101 5974602 6265-3-101 5974603 6265-3-101 6031601 6265-3-101 6173601 6265-3-101 6173602 6265-3-101 6173603 6265-3-101 6173604 6265-3-101 6214101 6265-3-101 6214102 6265-3-101 6214103 6265-3-101 6345401 6265-3-101 6345402 6265-3-101 6345403 6265-3-101 6345404 6265-3-101 6345405 6265-3-101 6345406 6265-3-101 6345407 6265-3-101 6487301 6265-3-101 6487303 6265-3-101 6487304 6265-3-101 6487305 6265-3-101 6487306 6265-3-101 9061101 6265-3-101 9061102 6265-3-101 9099901 6265-3-101 9099902 6265-3-101 9138901 6265-3-101 9138902 6265-3-101 9138903 6265-3-101 9138904 6265-3-101 9205101 6265-3-101 9205102 6265-3-101 9205103 6265-3-101 9205104 6265-3-101 9238701 6265-3-101 9277701 6265-3-101 9277702 6265-3-101 9342301 6265-3-101 9390601 6265-3-101 9390602 6265-3-101 9421801 6265-3-101 9499501 6265-3-101 9541501 6265-3-101 9541502 6265-3-101 9541503 6265-3-101 9567901 6265-3-101 9567902 6265-3-101 9640901 6265-3-101 9678101 6265-3-101 12725102 6265-3-101 12725103 6265-3-101 12725104 6265-3-101 12725106 6265-3-101 12725107 6265-3-101 12948501 6265-3-101 12948502 6265-3-101 12948601 6265-3-101 12948602 6265-3-101 12948701 6265-3-101 13120301 6265-3-101 13120303 6265-3-101 13120304 6265-3-101 13120305 6265-3-101 13120306 6265-3-101 13224501 6265-3-101 13224502 6265-3-101 13224503 6265-3-101 13224504 6265-3-101 13224601 6265-3-101 13224602 6265-3-101 13224603 6265-3-101 13224604 6265-3-101 13224701 6265-3-101 13224703 6265-3-101 13224704 6265-3-101 13429901 6265-3-101 14524902 6265-3-101 14524903 6265-3-101 14738801 6265-3-101 15236302 6265-3-101 15236303 6265-3-101 15236304 6265-3-101 15554601 6265-3-101 15554602 6265-3-101 15554603 6265-3-101 15554604 6265-3-101 15554605 6265-3-101 15554606 6265-3-101 15785201 6265-3-101 15785202 6265-3-101 15785203 6265-3-101 15785204 6265-3-101 15785205 6265-3-101 15785206 6265-3-101 15818801 6265-3-101 15818802 6265-3-101 15818803 6265-3-101 15818804 6265-3-101 16034001 6265-3-101 16034002 6265-3-101 16034003 6265-3-101 16059601 6265-3-101 16059602 6265-3-101 19898304 6265-3-101 19909901 6265-3-101 19909902 6265-3-101 19909903 6265-3-101 19909904 6265-3-101 19909905 6265-3-101 19909906 6265-3-101 19943701 6265-3-101 19943801 6265-3-101 19943901 6265-3-101 19943902 6265-3-101 19943903 6265-3-101 19943904 6265-3-101 20059201 6265-3-101 20059202 6265-3-101 20096101 6265-3-101 20096102 6265-3-101 20109001 6265-3-101 20109002 6265-3-101 20109003 6265-3-101 20109004 6265-3-101 20161201 6265-3-101 20161202 6265-3-101 20219301 6265-3-101 20219302 6265-3-101 20220401 6265-3-101 20317601 6265-3-101 22924705 6265-3-101 22999301 6265-3-101 22999302 6265-3-101 22999303 6265-3-101 22999304 6265-3-101 23096001 6265-3-101 23096002 6265-3-101 23096003 6265-3-101 23096004 6265-3-101 23205901 6265-3-101 23205902 6265-3-101 24342701 6265-3-101 24342702 6265-3-101 24342703 6265-3-101 24342704 6265-3-101 24342705 6265-3-101 24342706 6265-3-101 24678201 6265-3-101 24678202 6265-3-101 24678203 6265-3-101 24678204 6265-3-101 24678205 6265-3-101 24678206 6265-3-101 25081101 6265-3-101 25081102 6265-3-101 25081106 6265-3-101 25621901 6265-3-101 28732904 6265-3-101 29029801 6265-3-101 29029802 6265-3-101 29029803 6265-3-101 29029804 6265-3-101 29157301 6265-3-101 29157302 6265-3-101 29157303 6265-3-101 29157304 6265-3-101 29237001 6265-3-101 29237002 6265-3-101 29237003 6265-3-101 29237004 6265-3-101 29561201 6265-3-101 29561202 6265-3-101 29561203 6265-3-101 29561204 6265-3-101 29760701 6265-3-101 29760702 6265-3-101 29760703 6265-3-101 29760704 6265-3-101 29760705 6265-3-101 29760706 6265-3-101 29760707 6265-3-101 29760708 6265-3-101 30068401 6265-3-101 30068402 6265-3-101 34916402 6265-3-101 35144401 6265-3-101 35144402 6265-3-101 35144403 6265-3-101 35144404 6265-3-101 35144405 6265-3-101 35463201 6265-3-101 35463202 6265-3-101 35463204 6265-3-101 35471301 6265-3-101 35471303 6265-3-101 11790801R 6265-3-101 12725103R 6265-3-101 13224501R 6265-3-101 13596202R 6265-3-101 17674801R 6265-3-101 17674804R 6265-3-101 18982002R 6265-3-101 19013101R 6265-3-101 19909902R 6265-3-101 19943902R 6265-3-101 20161201S 6265-3-101 20161202R 6265-3-101 20366601R 6265-3-101 20527201S 6265-3-101 5471602B 6265-3-101 5694502B 6265-3-102 6551504 6265-3-102 6570501 6265-3-102 6570502 6265-3-102 6570503 6265-3-102 6570504 6265-3-102 6570505 6265-3-102 6570506 6265-3-102 6570507 6265-3-102 6645401 6265-3-102 6645402 6265-3-102 6645403 6265-3-102 6645404 6265-3-102 6679001 6265-3-102 6679002 6265-3-102 6679003 6265-3-102 6679004 6265-3-102 6679005 6265-3-102 6679006 6265-3-102 6679007 6265-3-102 6717301 6265-3-102 6717302 6265-3-102 6717303 6265-3-102 6717304 6265-3-102 6717305 6265-3-102 6717306 6265-3-102 6717307 6265-3-102 6828501 6265-3-102 7851403 6265-3-102 8025101 6265-3-102 8025102 6265-3-102 8112001 6265-3-102 8112003 6265-3-102 8154201 6265-3-102 8154202 6265-3-102 8154203 6265-3-102 8220201 6265-3-102 8220202 6265-3-102 8220203 6265-3-102 8908301 6265-3-102 8965401 6265-3-102 8965402 6265-3-102 9027401 6265-3-102 9027402 6265-3-102 9034301 6265-3-102 9050801 6265-3-102 9050802 6265-3-102 9050803 6265-3-102 9060901 6265-3-102 9060902 6265-3-102 9060903 6265-3-102 9099801 6265-3-102 9099802 6265-3-102 9099803 6265-3-102 9139101 6265-3-102 10262005 6265-3-102 10262006 6265-3-102 10321901 6265-3-102 10321902 6265-3-102 10367801 6265-3-102 10367802 6265-3-102 10367803 6265-3-102 10367804 6265-3-102 10999501 6265-3-102 10999503 6265-3-102 10999504 6265-3-102 10999505 6265-3-102 10999506 6265-3-102 11060001 6265-3-102 11060002 6265-3-102 11060003 6265-3-102 11060004 6265-3-102 11078301 6265-3-102 11078302 6265-3-102 11078303 6265-3-102 11078304 6265-3-102 11310101 6265-3-102 11310102 6265-3-102 11310103 6265-3-102 11310104 6265-3-102 11310105 6265-3-102 11335401 6265-3-102 12915901 6265-3-102 12915902 6265-3-102 12915903 6265-3-102 12916001 6265-3-102 12916101 6265-3-102 12916201 6265-3-102 12916202 6265-3-102 12916301 6265-3-102 12916302 6265-3-102 12978801 6265-3-102 12978802 6265-3-102 12978803 6265-3-102 12978804 6265-3-102 12978805 6265-3-102 12978806 6265-3-102 12978807 6265-3-102 12978808 6265-3-102 13120101 6265-3-102 13120102 6265-3-102 13120103 6265-3-102 13120104 6265-3-102 13120105 6265-3-102 13120106 6265-3-102 13120107 6265-3-102 13120108 6265-3-102 13120201 6265-3-102 13265101 6265-3-102 14644605 6265-3-102 14644606 6265-3-102 14792101 6265-3-102 14792103 6265-3-102 14792104 6265-3-102 14792105 6265-3-102 14792106 6265-3-102 14825301 6265-3-102 14825302 6265-3-102 14825303 6265-3-102 14825304 6265-3-102 14825305 6265-3-102 14825306 6265-3-102 15733201 6265-3-102 15733202 6265-3-102 15733203 6265-3-102 15733204 6265-3-102 15733205 6265-3-102 15733206 6265-3-102 15818901 6265-3-102 15818902 6265-3-102 15818903 6265-3-102 15818904 6265-3-102 15818905 6265-3-102 15818906 6265-3-102 15858001 6265-3-102 15858002 6265-3-102 17658702 6265-3-102 17658703 6265-3-102 17744801 6265-3-102 17744802 6265-3-102 17842501 6265-3-102 17842502 6265-3-102 17842503 6265-3-102 17842504 6265-3-102 17941601 6265-3-102 17941602 6265-3-102 17941603 6265-3-102 17941604 6265-3-102 17941605 6265-3-102 18131501 6265-3-102 18131502 6265-3-102 18131503 6265-3-102 18131504 6265-3-102 18131505 6265-3-102 18131506 6265-3-102 18154501 6265-3-102 18154502 6265-3-102 18154503 6265-3-102 18154504 6265-3-102 18154505 6265-3-102 18154506 6265-3-102 18180201 6265-3-102 18180202 6265-3-102 20058402 6265-3-102 20059001 6265-3-102 20079101 6265-3-102 20079102 6265-3-102 20096201 6265-3-102 20096202 6265-3-102 20109101 6265-3-102 20109102 6265-3-102 20161301 6265-3-102 20161302 6265-3-102 20232901 6265-3-102 20233001 6265-3-102 20233101 6265-3-102 20637901 6265-3-102 20804201 6265-3-102 20804202 6265-3-102 20847401 6265-3-102 20874901 6265-3-102 20906701 6265-3-102 20953201 6265-3-102 20975601 6265-3-102 20975602 6265-3-102 20977901 6265-3-102 21043301 6265-3-102 21209001 6265-3-102 21209002 6265-3-102 21405601 6265-3-102 23096104 6265-3-102 23096105 6265-3-102 23205801 6265-3-102 23205802 6265-3-102 24083701 6265-3-102 24083702 6265-3-102 24083703 6265-3-102 24083704 6265-3-102 24083705 6265-3-102 24083706 6265-3-102 24191601 6265-3-102 24191602 6265-3-102 24241201 6265-3-102 24241202 6265-3-102 24273201 6265-3-102 24310201 6265-3-102 24310202 6265-3-102 24310203 6265-3-102 24310204 6265-3-102 24310205 6265-3-102 24310206 6265-3-102 24533001 6265-3-102 24533002 6265-3-102 24533003 6265-3-102 24533004 6265-3-102 24533005 6265-3-102 24533006 6265-3-102 27071703 6265-3-102 27071704 6265-3-102 27071705 6265-3-102 27597901 6265-3-102 27597902 6265-3-102 27597903 6265-3-102 27597904 6265-3-102 27597905 6265-3-102 27813501 6265-3-102 27813502 6265-3-102 27813601 6265-3-102 27813602 6265-3-102 27813603 6265-3-102 27813604 6265-3-102 27823801 6265-3-102 27823802 6265-3-102 27823803 6265-3-102 27823804 6265-3-102 27823805 6265-3-102 28384401 6265-3-102 28384402 6265-3-102 28384403 6265-3-102 28384404 6265-3-102 28434001 6265-3-102 28434002 6265-3-102 28434003 6265-3-102 28434004 6265-3-102 30836801 6265-3-102 30836802 6265-3-102 30893102 6265-3-102 30893104 6265-3-102 30893105 6265-3-102 30893106 6265-3-102 30893107 6265-3-102 30893108 6265-3-102 30893109 6265-3-102 30893110 6265-3-102 32451001 6265-3-102 32451002 6265-3-102 32451003 6265-3-102 32451004 6265-3-102 32737501 6265-3-102 32737502 6265-3-102 32899101 6265-3-102 32899102 6265-3-102 32899103 6265-3-102 32899104 6265-3-102 32899105 6265-3-102 33374501 6265-3-102 33374502 6265-3-102 33374504 6265-3-102 33535801 6265-3-102 33535802 6265-3-102 33535803 6265-3-102 7004705R 6265-3-102 7823701R 6265-3-102 7823704R 6265-3-102 7836601R 6265-3-102 D6FRJ 6265-3-102 F1DWC 6265-3-103 5106401 6265-3-103 5340501 6265-3-103 5340502 6265-3-103 5340503 6265-3-103 5351402 6265-3-103 5471801 6265-3-103 5471802 6265-3-103 5707101 6265-3-103 5707102 6265-3-103 5707201 6265-3-103 5707302 6265-3-103 5750303 6265-3-103 5750304 6265-3-103 5853401 6265-3-103 5853402 6265-3-103 5853403 6265-3-103 5942301 6265-3-103 5942302 6265-3-103 5942303 6265-3-103 5974801 6265-3-103 5974802 6265-3-103 6717404 6265-3-103 6717405 6265-3-103 6801301 6265-3-103 6801302 6265-3-103 6801303 6265-3-103 6801304 6265-3-103 6828601 6265-3-103 6828602 6265-3-103 6828603 6265-3-103 6828604 6265-3-103 6946402 6265-3-103 6946403 6265-3-103 6946404 6265-3-103 6946405 6265-3-103 6946406 6265-3-103 6967401 6265-3-103 6967402 6265-3-103 6967501 6265-3-103 6967502 6265-3-103 7004401 6265-3-103 7004402 6265-3-103 7004403 6265-3-103 7004404 6265-3-103 7004406 6265-3-103 7004501 6265-3-103 7004502 6265-3-103 7004503 6265-3-103 7755401 6265-3-103 7755402 6265-3-103 7755403 6265-3-103 7755404 6265-3-103 7755405 6265-3-103 7755406 6265-3-103 7766202 6265-3-103 7766203 6265-3-103 7766204 6265-3-103 7823601 6265-3-103 7823603 6265-3-103 7851501 6265-3-103 7851503 6265-3-103 8908401 6265-3-103 8939201 6265-3-103 8939203 6265-3-103 8939204 6265-3-103 8939205 6265-3-103 8939206 6265-3-103 8965301 6265-3-103 8965302 6265-3-103 8965303 6265-3-103 8971801 6265-3-103 8971802 6265-3-103 8971804 6265-3-103 8971805 6265-3-103 8971806 6265-3-103 9498901 6265-3-103 9524501 6265-3-103 9524502 6265-3-103 9524503 6265-3-103 9524504 6265-3-103 9524505 6265-3-103 9524506 6265-3-103 9541701 6265-3-103 9541702 6265-3-103 9567502 6265-3-103 9567503 6265-3-103 9640501 6265-3-103 9640502 6265-3-103 9678001 6265-3-103 9678002 6265-3-103 9678003 6265-3-103 9718201 6265-3-103 9718202 6265-3-103 9718203 6265-3-103 9718204 6265-3-103 9777601 6265-3-103 9777602 6265-3-103 9815601 6265-3-103 9815602 6265-3-103 9815603 6265-3-103 9815604 6265-3-103 10004501 6265-3-103 11718004 6265-3-103 11718005 6265-3-103 11718006 6265-3-103 11718007 6265-3-103 11718008 6265-3-103 11718009 6265-3-103 11718010 6265-3-103 11718011 6265-3-103 11718012 6265-3-103 11718013 6265-3-103 11718015 6265-3-103 11718016 6265-3-103 11718018 6265-3-103 11718020 6265-3-103 11978501 6265-3-103 11978502 6265-3-103 11978503 6265-3-103 11978504 6265-3-103 11978505 6265-3-103 11995901 6265-3-103 11995902 6265-3-103 11995903 6265-3-103 11995904 6265-3-103 12037401 6265-3-103 12037402 6265-3-103 12037403 6265-3-103 12037404 6265-3-103 13007802 6265-3-103 13007803 6265-3-103 13007804 6265-3-103 13007805 6265-3-103 13265201 6265-3-103 13265202 6265-3-103 13265204 6265-3-103 13265205 6265-3-103 13265206 6265-3-103 13265207 6265-3-103 13265208 6265-3-103 13290601 6265-3-103 13290602 6265-3-103 13290603 6265-3-103 13290604 6265-3-103 13290605 6265-3-103 13290606 6265-3-103 13290607 6265-3-103 13290608 6265-3-103 13512401 6265-3-103 13512402 6265-3-103 13512403 6265-3-103 13512501 6265-3-103 13512502 6265-3-103 13512503 6265-3-103 13512504 6265-3-103 13512505 6265-3-103 14710902 6265-3-103 14710903 6265-3-103 14710904 6265-3-103 14710905 6265-3-103 14726401 6265-3-103 14726402 6265-3-103 14726403 6265-3-103 14792001 6265-3-103 14792002 6265-3-103 14792003 6265-3-103 14792004 6265-3-103 14792005 6265-3-103 14825401 6265-3-103 14825402 6265-3-103 14825403 6265-3-103 14825404 6265-3-103 14894201 6265-3-103 14894202 6265-3-103 14894203 6265-3-103 14894204 6265-3-103 14894205 6265-3-103 14894206 6265-3-103 15618602 6265-3-103 15618603 6265-3-103 15618604 6265-3-103 15618605 6265-3-103 15618606 6265-3-103 17842605 6265-3-103 17842606 6265-3-103 17878201 6265-3-103 17878202 6265-3-103 17878203 6265-3-103 17878204 6265-3-103 17878205 6265-3-103 17878206 6265-3-103 17941701 6265-3-103 17941702 6265-3-103 18189701 6265-3-103 18363501 6265-3-103 18363502 6265-3-103 18363503 6265-3-103 18378101 6265-3-103 18378102 6265-3-103 18378103 6265-3-103 18378104 6265-3-103 18378105 6265-3-103 18378106 6265-3-103 18429001 6265-3-103 18429002 6265-3-103 18429003 6265-3-103 18429004 6265-3-103 18429005 6265-3-103 18674901 6265-3-103 18674902 6265-3-103 20220101 6265-3-103 20220102 6265-3-103 20233201 6265-3-103 20233301 6265-3-103 20233401 6265-3-103 20507501 6265-3-103 20507502 6265-3-103 20514201 6265-3-103 20514202 6265-3-103 20560301 6265-3-103 20560302 6265-3-103 20638001 6265-3-103 20824501 6265-3-103 21209202 6265-3-103 21405701 6265-3-103 21484301 6265-3-103 21484302 6265-3-103 21531001 6265-3-103 21572301 6265-3-103 21662201 6265-3-103 21662202 6265-3-103 21717602 6265-3-103 21717603 6265-3-103 21759001 6265-3-103 21759002 6265-3-103 21790301 6265-3-103 21790302 6265-3-103 23515802 6265-3-103 23731601 6265-3-103 23731602 6265-3-103 23732001 6265-3-103 23732002 6265-3-103 23732003 6265-3-103 23732004 6265-3-103 23732005 6265-3-103 23732006 6265-3-103 23732007 6265-3-103 23732008 6265-3-103 23952501 6265-3-103 23952502 6265-3-103 23952503 6265-3-103 23952504 6265-3-103 23952505 6265-3-103 23952506 6265-3-103 23952507 6265-3-103 23952508 6265-3-103 24070101 6265-3-103 24070102 6265-3-103 24070103 6265-3-103 24070104 6265-3-103 24070105 6265-3-103 24070106 6265-3-103 24188801 6265-3-103 24188802 6265-3-103 26369302 6265-3-103 26369303 6265-3-103 26369304 6265-3-103 26369305 6265-3-103 26369306 6265-3-103 26633001 6265-3-103 26633002 6265-3-103 26684101 6265-3-103 26684102 6265-3-103 26684103 6265-3-103 26684104 6265-3-103 26684105 6265-3-103 26684106 6265-3-103 27003501 6265-3-103 27003503 6265-3-103 27003504 6265-3-103 27003505 6265-3-103 27086101 6265-3-103 27086102 6265-3-103 27086103 6265-3-103 27086104 6265-3-103 27086105 6265-3-103 27252301 6265-3-103 27252302 6265-3-103 27252303 6265-3-103 27252304 6265-3-103 27382604 6265-3-103 29400401 6265-3-103 29400402 6265-3-103 29400404 6265-3-103 29400405 6265-3-103 29400407 6265-3-103 29400408 6265-3-103 29561301 6265-3-103 29561302 6265-3-103 29561303 6265-3-103 29561304 6265-3-103 29839002 6265-3-103 29839004 6265-3-103 29839005 6265-3-103 30003002 6265-3-103 30003003 6265-3-103 30003004 6265-3-103 30003005 6265-3-103 30003006 6265-3-103 30003007 6265-3-103 30003008 6265-3-103 30165901 6265-3-103 30165902 6265-3-103 30165903 6265-3-103 30165904 6265-3-103 30165905 6265-3-103 30165906 6265-3-103 30165907 6265-3-103 33559701 6265-3-103 33559702 6265-3-103 33559703 6265-3-103 33559704 6265-3-103 33559801 6265-3-103 33559802 6265-3-103 33838302 6265-3-103 33838303 6265-3-103 33838304 6265-3-103 33838305 6265-3-103 33838306 6265-3-103 34145201 6265-3-103 34145202 6265-3-103 34145203 6265-3-103 34145204 6265-3-103 34145205 6265-3-103 34145206 6265-3-103 34668601 6265-3-103 34668602 6265-3-103 34668801 6265-3-103 34668802 6265-3-103 34827601 6265-3-103 34827602 6265-3-103 34963601 6265-3-103 34963602 6265-3-103 34963603 6265-3-103 34963604 6265-3-104 6541903 6265-3-104 6541904 6265-3-104 6541905 6265-3-104 6679801 6265-3-104 6679802 6265-3-104 6679803 6265-3-104 6679804 6265-3-104 6679805 6265-3-104 6679806 6265-3-104 6801201 6265-3-104 6801202 6265-3-104 6801203 6265-3-104 6801204 6265-3-104 6810801 6265-3-104 6810802 6265-3-104 6810803 6265-3-104 6810804 6265-3-104 7381701 6265-3-104 7381702 6265-3-104 7381703 6265-3-104 7381704 6265-3-104 7381705 6265-3-104 7381706 6265-3-104 7381707 6265-3-104 7434801 6265-3-104 7434802 6265-3-104 7434803 6265-3-104 9567602 6265-3-104 9567603 6265-3-104 9640601 6265-3-104 9640602 6265-3-104 9777701 6265-3-104 9777702 6265-3-104 9777703 6265-3-104 9815701 6265-3-104 10163301 6265-3-104 10163302 6265-3-104 10262201 6265-3-104 10316801 6265-3-104 10316901 6265-3-104 10368301 6265-3-104 10368302 6265-3-104 10403201 6265-3-104 10403202 6265-3-104 10999701 6265-3-104 10999702 6265-3-104 10999703 6265-3-104 10999704 6265-3-104 10999705 6265-3-104 11038701 6265-3-104 11038702 6265-3-104 11060401 6265-3-104 11060402 6265-3-104 11060403 6265-3-104 14107403 6265-3-104 14107404 6265-3-104 14196101 6265-3-104 14196102 6265-3-104 14255601 6265-3-104 14255602 6265-3-104 14468901 6265-3-104 14468902 6265-3-104 14505301 6265-3-104 14505302 6265-3-104 14617901 6265-3-104 14617902 6265-3-104 14710601 6265-3-104 14710602 6265-3-104 14710603 6265-3-104 14710604 6265-3-104 14710605 6265-3-104 14738501 6265-3-104 15442501 6265-3-104 15442502 6265-3-104 15554701 6265-3-104 15554702 6265-3-104 15554703 6265-3-104 15819001 6265-3-104 15819002 6265-3-104 15963601 6265-3-104 15963602 6265-3-104 19603301 6265-3-104 19603302 6265-3-104 19603303 6265-3-104 19603304 6265-3-104 19944401 6265-3-104 19944402 6265-3-104 19944403 6265-3-104 19944404 6265-3-104 19944405 6265-3-104 19944406 6265-3-104 20109201 6265-3-104 20109202 6265-3-104 20109203 6265-3-104 20109204 6265-3-104 20161501 6265-3-104 20161502 6265-3-104 20185401 6265-3-104 20185402 6265-3-104 20185403 6265-3-104 20185404 6265-3-104 20185405 6265-3-104 20185406 6265-3-104 20213701 6265-3-104 20213702 6265-3-104 20213703 6265-3-104 20213704 6265-3-104 20215501 6265-3-104 25094103 6265-3-104 25094104 6265-3-104 25094105 6265-3-104 25094106 6265-3-104 25764401 6265-3-104 25764402 6265-3-104 25764501 6265-3-104 25764502 6265-3-104 25764503 6265-3-104 25764504 6265-3-104 26115401 6265-3-104 26115402 6265-3-104 26115403 6265-3-104 26115404 6265-3-104 26115405 6265-3-104 26115406 6265-3-104 27099501 6265-3-104 27099502 6265-3-104 27099503 6265-3-104 27099504 6265-3-104 27099505 6265-3-104 27400601 6265-3-104 27400602 6265-3-104 27400603 6265-3-104 27400604 6265-3-104 27400605 6265-3-104 27646401 6265-3-104 32543402 6265-3-104 32543403 6265-3-104 32543404 6265-3-104 33450901 6265-3-104 33450902 6265-3-104 33450903 6265-3-104 33450904 6265-3-104 33450905 6265-3-104 33450906 6265-3-104 34220601 6265-3-104 34220602 6265-3-104 34220701 6265-3-104 34220702 6265-3-104 34220703 6265-3-104 34220704 6265-3-104 34791401 6265-3-104 34791402 6265-3-104 34791403 6265-3-104 34791404 6265-3-104 34827701 6265-3-104 34827702 6265-3-104 34916501 6265-3-104 34916502 6265-3-104 35362601 6265-3-104 35362602 6265-3-104 35362603 6265-3-105 6252402 6265-3-105 11052701 6265-3-105 11052702 6265-3-105 11052703 6265-3-105 11078501 6265-3-105 11078502 6265-3-105 11078503 6265-3-105 11450701 6265-3-105 11450702 6265-3-105 11504001 6265-3-105 11504002 6265-3-105 12145101
6265-3-105 12145102 6265-3-105 12145103 6265-3-105 12145104 6265-3-105 13093101 6265-3-105 13093102 6265-3-105 13093103 6265-3-105 13093105 6265-3-105 13093106 6265-3-105 13093107 6265-3-105 13093108 6265-3-105 13513101 6265-3-105 13513201 6265-3-105 13513202 6265-3-105 13513203 6265-3-105 13513204 6265-3-105 13513205 6265-3-105 20514401 6265-3-105 20514402 6265-3-105 22261401 6265-3-105 22261402 6265-3-105 22409201 6265-3-105 22409202 6265-3-105 22559901 6265-3-105 22560001 6265-3-105 22560002 6265-3-105 22791901 6265-3-105 22950601 6265-3-105 22950602 6265-3-105 22976901 6265-3-105 22976902 6265-3-105 22976903 6265-3-105 22976904 6265-3-105 22976905 6265-3-105 22976906 6265-3-105 22976907 6265-3-105 23205601 6265-3-105 23205602 6265-3-105 23966901 6265-3-105 23966902 6265-3-105 23966903 6265-3-105 23966904 6265-3-105 23966905 6265-3-105 23966906 6265-3-105 34204004 6265-3-105 34204005 6265-3-105 34685401 6265-3-105 34685402 6265-3-105 34685403 6265-3-105 34685404 6265-3-105 34685405 6265-3-105 34685501 6265-3-105 34685502 6265-3-105 35653901 6265-3-105 35653902 6265-3-105 35653904 6265-3-105 195318013 6265-3-105 10212002R 6265-3-105 11052702R 6265-3-105 18717801R 6265-3-105 8908601RR 6265-3-106 5497201 6265-3-106 5497202 6265-3-106 5736401 6265-3-106 5736402 6265-3-106 6173901 6265-3-106 6173902 6265-3-106 6173903 6265-3-106 6252201 6265-3-106 6252202 6265-3-106 6263801 6265-3-106 22420406 6265-3-106 22559801 6265-3-106 22559802 6265-3-106 22657701 6265-3-106 22977001 6265-3-106 22977002 6265-3-106 22977003 6265-3-106 22977004 6265-3-106 22977005 6265-3-106 22977006 6265-3-106 22999703 6265-3-106 23939801 6265-3-106 23939802 6265-3-106 23939803 6265-3-106 23939804 6265-3-106 23939805 6265-3-106 23939806 6265-3-106 23939807 6265-3-106 24321601 6265-3-106 24321602 6265-3-106 24321603 6265-3-106 24321604 6265-3-106 24321605 6265-3-106 24502001 6265-3-106 24502002 6265-3-106 24502003 6265-3-106 24502004 6265-3-107 13190202 6265-3-107 13190203 6265-3-107 13190204 6265-3-107 13446501 6265-3-107 13446601 6265-3-107 13446801 6265-3-107 13447101 6265-3-107 13447201 6265-3-107 13447202 6265-3-107 13447203 6265-3-107 13899301 6265-3-107 13899302 6265-3-107 14080002 6265-3-107 14173001 6265-3-107 14173002 6265-3-107 14173003 6265-3-107 14173004 6265-3-107 14559601 6265-3-107 14559602 6265-3-107 14559603 6265-3-107 14738701 6265-3-107 20131601 6265-3-107 20131602 6265-3-107 22950501 6265-3-107 22950502 6265-3-107 23277401 6265-3-107 23277402 6265-3-107 32950603 6265-3-107 32950604 6265-3-107 33292601 6265-3-107 33292603 6265-3-107 33987802 6265-3-100 11038201 6265-3-100 11038202 6265-3-100 11038203 6265-3-100 11038204 6265-3-100 11224801 6265-3-100 11224802 6265-3-100 11224803 6265-3-100 11253601 6265-3-100 11253602 6265-3-100 11335801 6265-3-100 11335901 6265-3-100 11790701 6265-3-100 11790702 6265-3-100 11790703 6265-3-100 11995201 6265-3-100 11995202 6265-3-100 11995203 6265-3-100 12037201 6265-3-100 12037202 6265-3-100 12037203 6265-3-100 13784701 6265-3-100 13784702 6265-3-100 13784703 6265-3-100 13784704 6265-3-100 13784705 6265-3-100 13784706 6265-3-100 13784707 6265-3-100 13784708 6265-3-100 14027901 6265-3-100 14027902 6265-3-100 14485801 6265-3-100 15302801 6265-3-100 15302802 6265-3-100 15302803 6265-3-100 15302804 6265-3-100 15302805 6265-3-100 15302806 6265-3-100 15733101 6265-3-100 15733102 6265-3-100 16174701 6265-3-100 16174702 6265-3-100 16174703 6265-3-100 17091301 6265-3-100 17091302 6265-3-100 17604301 6265-3-100 17604302 6265-3-100 17604303 6265-3-100 17604304 6265-3-100 17842701 6265-3-100 17842702 6265-3-100 17842703 6265-3-100 17842704 6265-3-100 18010701 6265-3-100 27999501 6265-3-100 28557201 6265-3-100 28644401 6265-3-100 28644402 6265-3-100 28644403 6265-3-100 28646201 6265-3-100 28646202 6265-3-100 29838301 6265-3-100 29838302 6265-3-100 29838303 6265-3-100 29838304 6265-3-100 29838305 6265-3-100 29838306 6265-3-100 29988501 6265-3-100 29988502 6265-3-100 29988503 6265-3-100 29988504 6265-3-100 30423101 6265-3-100 30423102 6265-3-100 30423103 6265-3-100 30423104 6265-3-100 30423105 6265-3-100 30423106 6265-3-100 32573801 6265-3-100 32573802 6265-3-100 32573803 6265-3-100 32573901 6265-3-101 6487308 6265-3-101 6570401 6265-3-101 6570402 6265-3-101 6570403 6265-3-101 6570404 6265-3-101 6570405 6265-3-101 6570406 6265-3-101 6570407 6265-3-101 6645301 6265-3-101 6645302 6265-3-101 6645303 6265-3-101 6645304 6265-3-101 6678901 6265-3-101 6678902 6265-3-101 6678903 6265-3-101 6678904 6265-3-101 6678905 6265-3-101 7460901 6265-3-101 7460902 6265-3-101 7460903 6265-3-101 7460904 6265-3-101 7460905 6265-3-101 7460906 6265-3-101 7703101 6265-3-101 7703102 6265-3-101 7703103 6265-3-101 7703104 6265-3-101 9678102 6265-3-101 9693401 6265-3-101 9693402 6265-3-101 9718301 6265-3-101 9777501 6265-3-101 9777502 6265-3-101 9777504 6265-3-101 9849801 6265-3-101 10321801 6265-3-101 10367501 6265-3-101 10999401 6265-3-101 10999403 6265-3-101 10999404 6265-3-101 10999405 6265-3-101 10999406 6265-3-101 11038401 6265-3-101 11038402 6265-3-101 11038403 6265-3-101 11038404 6265-3-101 11078101 6265-3-101 11078102 6265-3-101 11078103 6265-3-101 11078104 6265-3-101 11224901 6265-3-101 11224902 6265-3-101 11253801 6265-3-101 11253802 6265-3-101 13429902 6265-3-101 13429903 6265-3-101 13429905 6265-3-101 13429906 6265-3-101 13596201 6265-3-101 13596202 6265-3-101 13596203 6265-3-101 13596205 6265-3-101 13596206 6265-3-101 13596207 6265-3-101 13596208 6265-3-101 13798501 6265-3-101 13798502 6265-3-101 13798503 6265-3-101 13798504 6265-3-101 13798505 6265-3-101 13798506 6265-3-101 13798507 6265-3-101 13798508 6265-3-101 14028101 6265-3-101 14028102 6265-3-101 14028103 6265-3-101 14028104 6265-3-101 14079801 6265-3-101 14079802 6265-3-101 14079803 6265-3-101 14106901 6265-3-101 16660401 6265-3-101 16660402 6265-3-101 16660403 6265-3-101 16660404 6265-3-101 16660405 6265-3-101 16660406 6265-3-101 16706001 6265-3-101 16706002 6265-3-101 16706003 6265-3-101 16800401 6265-3-101 16800402 6265-3-101 16800403 6265-3-101 16800404 6265-3-101 16800405 6265-3-101 16975501 6265-3-101 16975502 6265-3-101 16975503 6265-3-101 16975504 6265-3-101 17255801 6265-3-101 17255802 6265-3-101 17322601 6265-3-101 17322603 6265-3-101 17322605 6265-3-101 17491202 6265-3-101 17674801 6265-3-101 17674802 6265-3-101 17674803 6265-3-101 20317602 6265-3-101 20366601 6265-3-101 20366602 6265-3-101 20385101 6265-3-101 20385102 6265-3-101 20434301 6265-3-101 20434302 6265-3-101 20455101 6265-3-101 20455102 6265-3-101 20527202 6265-3-101 20560101 6265-3-101 20560102 6265-3-101 20581401 6265-3-101 20581402 6265-3-101 21862101 6265-3-101 21890201 6265-3-101 21947901 6265-3-101 21966701 6265-3-101 22189601 6265-3-101 22189602 6265-3-101 22260701 6265-3-101 22260801 6265-3-101 22287301 6265-3-101 22287302 6265-3-101 22309501 6265-3-101 22309502 6265-3-101 22309503 6265-3-101 25621902 6265-3-101 25621903 6265-3-101 25621904 6265-3-101 25741201 6265-3-101 25741202 6265-3-101 25741203 6265-3-101 25741501 6265-3-101 25741502 6265-3-101 25896701 6265-3-101 25896702 6265-3-101 25896703 6265-3-101 25896704 6265-3-101 25896705 6265-3-101 26454401 6265-3-101 26454402 6265-3-101 26454403 6265-3-101 26659301 6265-3-101 26659302 6265-3-101 26659303 6265-3-101 26659304 6265-3-101 26837601 6265-3-101 26837602 6265-3-101 26837603 6265-3-101 26837604 6265-3-101 26837605 6265-3-101 26837606 6265-3-101 27364601 6265-3-101 30068403 6265-3-101 30068404 6265-3-101 30068405 6265-3-101 30068406 6265-3-101 30068408 6265-3-101 30384601 6265-3-101 30384602 6265-3-101 30384603 6265-3-101 30384604 6265-3-101 30384605 6265-3-101 30384606 6265-3-101 30384607 6265-3-101 31256401 6265-3-101 31256402 6265-3-101 31256403 6265-3-101 31256404 6265-3-101 31256405 6265-3-101 31256406 6265-3-101 31256407 6265-3-101 32528201 6265-3-101 32528202 6265-3-101 32528203 6265-3-101 32528204 6265-3-101 32528205 6265-3-101 32528301 6265-3-101 32528302 6265-3-101 32528303 6265-3-101 5750201R 6265-3-102 5657101 6265-3-102 5657102 6265-3-102 5657103 6265-3-102 5657104 6265-3-102 5657105 6265-3-102 5707001 6265-3-102 5707002 6265-3-102 5707003 6265-3-102 5707004 6265-3-102 5779301 6265-3-102 5779302 6265-3-102 5779303 6265-3-102 5942101 6265-3-102 5942104 6265-3-102 5974701 6265-3-102 5974702 6265-3-102 5974703 6265-3-102 5974704 6265-3-102 6030501 6265-3-102 6030502 6265-3-102 6116801 6265-3-102 6116802 6265-3-102 6146601 6265-3-102 6146602 6265-3-102 6154101 6265-3-102 6154102 6265-3-102 6828502 6265-3-102 6828503 6265-3-102 6828504 6265-3-102 6828505 6265-3-102 6946302 6265-3-102 6946303 6265-3-102 7004601 6265-3-102 7004602 6265-3-102 7004603 6265-3-102 7004604 6265-3-102 7004605 6265-3-102 7004606 6265-3-102 7004701 6265-3-102 7004702 6265-3-102 7004703 6265-3-102 7004704 6265-3-102 7004705 6265-3-102 7434501 6265-3-102 7434502 6265-3-102 7434503 6265-3-102 7434504 6265-3-102 7534901 6265-3-102 7534902 6265-3-102 7534903 6265-3-102 7534904 6265-3-102 7596901 6265-3-102 7596902 6265-3-102 9139102 6265-3-102 9139103 6265-3-102 9139104 6265-3-102 9139105 6265-3-102 9205201 6265-3-102 9205202 6265-3-102 9205203 6265-3-102 9205204 6265-3-102 9238801 6265-3-102 9238802 6265-3-102 9238803 6265-3-102 9342501 6265-3-102 9342502 6265-3-102 9342503 6265-3-102 9342504 6265-3-102 9342505 6265-3-102 9342506 6265-3-102 9390701 6265-3-102 9390702 6265-3-102 9421701 6265-3-102 9421702 6265-3-102 9448001 6265-3-102 9448002 6265-3-102 9448003 6265-3-102 9499701 6265-3-102 9499901 6265-3-102 9524701 6265-3-102 11335402 6265-3-102 11335403 6265-3-102 11335404 6265-3-102 11451502 6265-3-102 11496201 6265-3-102 11496202 6265-3-102 11496203 6265-3-102 11709701 6265-3-102 11709702 6265-3-102 11709703 6265-3-102 11709704 6265-3-102 11709705 6265-3-102 11709706 6265-3-102 11709707 6265-3-102 11709708 6265-3-102 11709709 6265-3-102 11709710 6265-3-102 11709711 6265-3-102 11709712 6265-3-102 11709713 6265-3-102 11709714 6265-3-102 11978601 6265-3-102 11978603 6265-3-102 11978604 6265-3-102 11978605 6265-3-102 11995801 6265-3-102 11995802 6265-3-102 13265102 6265-3-102 13265103 6265-3-102 13265104 6265-3-102 13265105 6265-3-102 13265107 6265-3-102 13265108 6265-3-102 13710001 6265-3-102 13710002 6265-3-102 13710003 6265-3-102 13710004 6265-3-102 13710006 6265-3-102 13710007 6265-3-102 13710008 6265-3-102 13759001 6265-3-102 13759002 6265-3-102 13759005 6265-3-102 13759006 6265-3-102 13759007 6265-3-102 13759008 6265-3-102 14028601 6265-3-102 14028602 6265-3-102 14028603 6265-3-102 14107201 6265-3-102 14107202 6265-3-102 14107203 6265-3-102 14150401 6265-3-102 14150402 6265-3-102 15858003 6265-3-102 15858004 6265-3-102 15869602 6265-3-102 15869603 6265-3-102 15869604 6265-3-102 15869605 6265-3-102 15869606 6265-3-102 16153401 6265-3-102 16153403 6265-3-102 16153404 6265-3-102 16153405 6265-3-102 16153406 6265-3-102 16174901 6265-3-102 16174902 6265-3-102 16174903 6265-3-102 16174904 6265-3-102 16705401 6265-3-102 16705403 6265-3-102 16705404 6265-3-102 16705405 6265-3-102 16705406 6265-3-102 16749201 6265-3-102 16749202 6265-3-102 16749203 6265-3-102 16749204 6265-3-102 16749205 6265-3-102 16749206 6265-3-102 18180203 6265-3-102 18428901 6265-3-102 18428902 6265-3-102 18561401 6265-3-102 18561402 6265-3-102 18561403 6265-3-102 18561404 6265-3-102 18591901 6265-3-102 18591902 6265-3-102 18869901 6265-3-102 18869902 6265-3-102 18869903 6265-3-102 18869904 6265-3-102 18869905 6265-3-102 18869906 6265-3-102 18947301 6265-3-102 18947302 6265-3-102 18947303 6265-3-102 19013202 6265-3-102 19013204 6265-3-102 19246301 6265-3-102 19246302 6265-3-102 19253901 6265-3-102 19253902 6265-3-102 19327701 6265-3-102 19327702 6265-3-102 19344201 6265-3-102 21405602 6265-3-102 21405603 6265-3-102 21483901 6265-3-102 21483902 6265-3-102 21530901 6265-3-102 21565001 6265-3-102 21662101 6265-3-102 21662102 6265-3-102 21790201 6265-3-102 21790202 6265-3-102 21862201 6265-3-102 21862202 6265-3-102 21948001 6265-3-102 21966801 6265-3-102 21982201 6265-3-102 21982202 6265-3-102 22108401 6265-3-102 22192501 6265-3-102 22192502 6265-3-102 22199401 6265-3-102 22350501 6265-3-102 22350502 6265-3-102 22403001 6265-3-102 22403002 6265-3-102 22403003 6265-3-102 22403004 6265-3-102 22403005 6265-3-102 24820301 6265-3-102 25190901 6265-3-102 25190902 6265-3-102 25190903 6265-3-102 25190904 6265-3-102 25190905 6265-3-102 25190906 6265-3-102 25190907 6265-3-102 25190908 6265-3-102 25458301 6265-3-102 25458302 6265-3-102 25458303 6265-3-102 25458304 6265-3-102 25458305 6265-3-102 25458306 6265-3-102 25622001 6265-3-102 25622101 6265-3-102 25622102 6265-3-102 25622103 6265-3-102 25622104 6265-3-102 25798201 6265-3-102 25798202 6265-3-102 25798203 6265-3-102 25845301 6265-3-102 25845302 6265-3-102 25845303 6265-3-102 25845304 6265-3-102 28434005 6265-3-102 28434006 6265-3-102 28557301 6265-3-102 28644601 6265-3-102 28644602 6265-3-102 28644603 6265-3-102 29022901 6265-3-102 29022902 6265-3-102 29022903 6265-3-102 29022904 6265-3-102 29205801 6265-3-102 29205802 6265-3-102 29205803 6265-3-102 29205804 6265-3-102 29205805 6265-3-102 29446601 6265-3-102 29446602 6265-3-102 29446603 6265-3-102 29446604 6265-3-102 29446701 6265-3-102 29446702 6265-3-102 29446703 6265-3-102 29818601 6265-3-102 29818602 6265-3-102 29818603 6265-3-102 29818604 6265-3-102 29818605 6265-3-102 33535804 6265-3-102 34082401 6265-3-102 34082402 6265-3-102 34082403 6265-3-102 34082404 6265-3-102 34082405 6265-3-102 34082406 6265-3-102 34082407 6265-3-102 34275401 6265-3-102 34275402 6265-3-102 34275403 6265-3-102 34275404 6265-3-102 34275405 6265-3-102 34572801 6265-3-102 34572802 6265-3-102 34572803 6265-3-102 34572804 6265-3-102 34572805 6265-3-102 34572806 6265-3-102 34572807 6265-3-102 34572808 6265-3-102 34572809 6265-3-102 34572810 6265-3-102 34572811 6265-3-102 34932301 6265-3-102 34932302 6265-3-102 34932303 6265-3-103 5974803 6265-3-103 6030801 6265-3-103 6030802 6265-3-103 6030803 6265-3-103 6030804 6265-3-103 6030805 6265-3-103 6116901 6265-3-103 6116902 6265-3-103 6117001 6265-3-103 6117002 6265-3-103 6154201 6265-3-103 6154202 6265-3-103 6154203 6265-3-103 6154204 6265-3-103 6166101 6265-3-103 6166102 6265-3-103 6214201 6265-3-103 6214202 6265-3-103 6214203 6265-3-103 6214204 6265-3-103 6214205 6265-3-103 6252001 6265-3-103 6252002 6265-3-103 6397901 6265-3-103 6397902 6265-3-103 6397903 6265-3-103 6397904 6265-3-103 7004504 6265-3-103 7004505 6265-3-103 7004506 6265-3-103 7075501 6265-3-103 7075502 6265-3-103 7075503 6265-3-103 7075504 6265-3-103 7075505 6265-3-103 7075506 6265-3-103 7075701 6265-3-103 7075704 6265-3-103 7075706 6265-3-103 7158701 6265-3-103 7158702 6265-3-103 7158703 6265-3-103 7158704 6265-3-103 7158705 6265-3-103 7158706 6265-3-103 7381801 6265-3-103 7381802 6265-3-103 7381803 6265-3-103 7381804 6265-3-103 7381805 6265-3-103 7434701 6265-3-103 7434702 6265-3-103 7434703 6265-3-103 7434704 6265-3-103 9027501 6265-3-103 9027502 6265-3-103 9027701 6265-3-103 9033501 6265-3-103 9033502 6265-3-103 9033503 6265-3-103 9033504 6265-3-103 9033505 6265-3-103 9051601 6265-3-103 9051602 6265-3-103 9051603 6265-3-103 9060501 6265-3-103 9060502 6265-3-103 9060503 6265-3-103 9060504 6265-3-103 9099601 6265-3-103 9100001 6265-3-103 9100002 6265-3-103 9100003 6265-3-103 9100004 6265-3-103 9100005 6265-3-103 9139201 6265-3-103 9139202 6265-3-103 9139203 6265-3-103 9139204 6265-3-103 9139205 6265-3-103 9205301 6265-3-103 10004502 6265-3-103 10004503 6265-3-103 10004504 6265-3-103 10004505 6265-3-103 10004506 6265-3-103 10048501 6265-3-103 10048502 6265-3-103 10048503 6265-3-103 10048504 6265-3-103 10048505 6265-3-103 10048506 6265-3-103 10082801 6265-3-103 10082802 6265-3-103 10082803 6265-3-103 10211901 6265-3-103 10211902 6265-3-103 10211903 6265-3-103 10211904 6265-3-103 10262101 6265-3-103 10316701 6265-3-103 10316702 6265-3-103 10368001 6265-3-103 10368002 6265-3-103 10999602 6265-3-103 10999603 6265-3-103 10999605 6265-3-103 10999606 6265-3-103 12037405 6265-3-103 12064101 6265-3-103 12064102 6265-3-103 12064103 6265-3-103 12064104 6265-3-103 12064105 6265-3-103 12064106 6265-3-103 12144901 6265-3-103 12144902 6265-3-103 12144903 6265-3-103 12144904 6265-3-103 12144905 6265-3-103 12337602 6265-3-103 12337603 6265-3-103 12337604 6265-3-103 12549601 6265-3-103 12549602 6265-3-103 12549603 6265-3-103 12549604 6265-3-103 12549605 6265-3-103 12549606 6265-3-103 12549607 6265-3-103 12549608 6265-3-103 12549609 6265-3-103 12549610 6265-3-103 12721501 6265-3-103 12721503 6265-3-103 13596401 6265-3-103 13596402 6265-3-103 13596403 6265-3-103 13596404 6265-3-103 13596405 6265-3-103 13596406 6265-3-103 13596407 6265-3-103 13596408 6265-3-103 13710101 6265-3-103 13710102 6265-3-103 13710104 6265-3-103 13710105 6265-3-103 13710106 6265-3-103 13710107 6265-3-103 13710108 6265-3-103 13823601 6265-3-103 13823602 6265-3-103 13823603 6265-3-103 13823604 6265-3-103 13823606 6265-3-103 13823607 6265-3-103 13823608 6265-3-103 14079901 6265-3-103 14079902 6265-3-103 14079903 6265-3-103 14079904 6265-3-103 14107301 6265-3-103 15963301 6265-3-103 15963302 6265-3-103 15963303 6265-3-103 15963304 6265-3-103 16059401 6265-3-103 16059402 6265-3-103 16059403 6265-3-103 16059404 6265-3-103 16585801 6265-3-103 16585802 6265-3-103 16585803 6265-3-103 16586001 6265-3-103 16586002 6265-3-103 16705501 6265-3-103 16705502 6265-3-103 16705503 6265-3-103 16705504 6265-3-103 16705505 6265-3-103 16705506 6265-3-103 16823201 6265-3-103 16823202 6265-3-103 16823203 6265-3-103 16823204 6265-3-103 16823205 6265-3-103 16823206 6265-3-103 16866701 6265-3-103 16866702 6265-3-103 18674903 6265-3-103 18870201 6265-3-103 18870202 6265-3-103 18870203 6265-3-103 18870204 6265-3-103 18870205 6265-3-103 18870206 6265-3-103 18982101 6265-3-103 18982102 6265-3-103 18998801 6265-3-103 18998802 6265-3-103 19160801 6265-3-103 19160802 6265-3-103 19246401 6265-3-103 19246402 6265-3-103 19295901 6265-3-103 19295902 6265-3-103 19344901 6265-3-103 19344902 6265-3-103 19400001 6265-3-103 19400002 6265-3-103 19400003 6265-3-103 19603601 6265-3-103 19603602 6265-3-103 19603603 6265-3-103 19603604 6265-3-103 19832801 6265-3-103 21862301 6265-3-103 21862302 6265-3-103 21862303 6265-3-103 21874001 6265-3-103 21890401 6265-3-103 21966901 6265-3-103 22108301 6265-3-103 22108302 6265-3-103 22130001 6265-3-103 22227501 6265-3-103 22261001 6265-3-103 22261101 6265-3-103 22261102 6265-3-103 22289001 6265-3-103 22289002 6265-3-103 22403101 6265-3-103 22403102 6265-3-103 22403103 6265-3-103 22403104 6265-3-103 22403105 6265-3-103 22403106 6265-3-103 22601101 6265-3-103 22601102 6265-3-103 22657901 6265-3-103 22657902 6265-3-103 22707601 6265-3-103 22707602 6265-3-103 24191701 6265-3-103 24191702 6265-3-103 24273101 6265-3-103 24273102 6265-3-103 24273103 6265-3-103 24273104 6265-3-103 24321701 6265-3-103 24321801 6265-3-103 24321802 6265-3-103 24321803 6265-3-103 24321804 6265-3-103 24321805 6265-3-103 24321806 6265-3-103 24546501 6265-3-103 24546502 6265-3-103 24546503 6265-3-103 24546504 6265-3-103 24546505 6265-3-103 24795101 6265-3-103 24795103 6265-3-103 25026001 6265-3-103 25026002 6265-3-103 25026003 6265-3-103 25026004 6265-3-103 25026005 6265-3-103 25357601 6265-3-103 25357602 6265-3-103 27382605 6265-3-103 27382606 6265-3-103 27581201 6265-3-103 27581202 6265-3-103 27581203 6265-3-103 27581301 6265-3-103 27581302 6265-3-103 27581303 6265-3-103 27581304 6265-3-103 27704601 6265-3-103 27704602 6265-3-103 27704603 6265-3-103 27704604 6265-3-103 27704605 6265-3-103 27823901 6265-3-103 27823902 6265-3-103 28097201 6265-3-103 28097202 6265-3-103 28097203 6265-3-103 28097204 6265-3-103 28097205 6265-3-103 28097206 6265-3-103 28331601 6265-3-103 28331602 6265-3-103 28331603 6265-3-103 28331604 6265-3-103 28331605 6265-3-103 30165908 6265-3-103 30372801 6265-3-103 30372802 6265-3-103 30372805 6265-3-103 30448101 6265-3-103 30448102 6265-3-103 30448103 6265-3-103 30448104 6265-3-103 30601601 6265-3-103 30601602 6265-3-103 30601603 6265-3-103 30601604 6265-3-103 30601605 6265-3-103 30601606 6265-3-103 30601607 6265-3-103 30601608 6265-3-103 31071001 6265-3-103 31071002 6265-3-103 31071003 6265-3-103 31071004 6265-3-103 31071005 6265-3-103 31071006 6265-3-103 31071007 6265-3-103 31071008 6265-3-103 31113001 6265-3-103 31113002 6265-3-103 31113003 6265-3-103 34963605 6265-3-103 35212501 6265-3-103 35212502 6265-3-103 35212503 6265-3-103 35212504 6265-3-103 35212505 6265-3-103 35212506 6265-3-103 35743201 6265-3-103 35743202 6265-3-103 35743302 6265-3-103 35743304 6265-3-103 12144901R 6265-3-103 17842604R 6265-3-103 18982101R 6265-3-103 19160801R 6265-3-103 19944106S 6265-3-103 5707101A 6265-3-103 5707201A 6265-3-103 5707202A 6265-3-103 5750302A 6265-3-103 9060501A 6265-3-103 9139206R 6265-3-103 9342102R 6265-3-104 1125401 6265-3-104 1125402 6265-3-104 5106601 6265-3-104 5106603 6265-3-104 7679301 6265-3-104 7679302 6265-3-104 7679303 6265-3-104 7679304 6265-3-104 7766301 6265-3-104 7766302 6265-3-104 7766303 6265-3-104 7766304 6265-3-104 7851201 6265-3-104 7851202 6265-3-104 7851203 6265-3-104 7851204 6265-3-104 8153901 6265-3-104 8153902 6265-3-104 8220301 6265-3-104 8632101 6265-3-104 8632102 6265-3-104 8632103 6265-3-104 8632104 6265-3-104 8659901 6265-3-104 8659902 6265-3-104 8659903 6265-3-104 8790601 6265-3-104 8790602 6265-3-104 8908501 6265-3-104 8939401 6265-3-104 8965202 6265-3-104 11060404 6265-3-104 11225001 6265-3-104 11225002 6265-3-104 11276001 6265-3-104 11276002 6265-3-104 11276003 6265-3-104 11335501 6265-3-104 11335502 6265-3-104 11335503 6265-3-104 11451001 6265-3-104 11451002 6265-3-104 11451003 6265-3-104 11496801 6265-3-104 11496802 6265-3-104 11503901 6265-3-104 11503902 6265-3-104 11503903 6265-3-104 12037501 6265-3-104 12037503 6265-3-104 12037504 6265-3-104 12037505 6265-3-104 12145001 6265-3-104 12145002 6265-3-104 12145003 6265-3-104 12145004 6265-3-104 12725301 6265-3-104 12725302 6265-3-104 16059501 6265-3-104 16059502 6265-3-104 16175001 6265-3-104 16175002 6265-3-104 16552801 6265-3-104 16552802 6265-3-104 16552803 6265-3-104 16552804 6265-3-104 16705701 6265-3-104 16705702 6265-3-104 16705703 6265-3-104 16705704 6265-3-104 16705705 6265-3-104 16705706 6265-3-104 16866901 6265-3-104 16866902 6265-3-104 17091201 6265-3-104 17136401 6265-3-104 17136402 6265-3-104 17322801 6265-3-104 17322802 6265-3-104 17386501 6265-3-104 17386502 6265-3-104 17540201 6265-3-104 17540202 6265-3-104 17677801 6265-3-104 17677802 6265-3-104 20215502 6265-3-104 20233501 6265-3-104 20233502 6265-3-104 20317701 6265-3-104 20317702 6265-3-104 20390901 6265-3-104 20390902 6265-3-104 20824601 6265-3-104 22130101 6265-3-104 22261201 6265-3-104 22261301 6265-3-104 22403201 6265-3-104 22403203 6265-3-104 22461302 6265-3-104 22461304 6265-3-104 22602201 6265-3-104 22602202 6265-3-104 22602203 6265-3-104 22602204 6265-3-104 22657801 6265-3-104 22657802 6265-3-104 22708001 6265-3-104 22708002 6265-3-104 22950701 6265-3-104 22950702 6265-3-104 22950703 6265-3-104 22950704 6265-3-104 27646402 6265-3-104 27646501 6265-3-104 27646502 6265-3-104 27646503 6265-3-104 27964401 6265-3-104 27964402 6265-3-104 27999001 6265-3-104 28008301 6265-3-104 28427701 6265-3-104 28427702 6265-3-104 28427703 6265-3-104 28427704 6265-3-104 28890701 6265-3-104 28890702 6265-3-104 28890703 6265-3-104 28890704 6265-3-104 29165001 6265-3-104 29165002 6265-3-104 29165003 6265-3-104 29165004 6265-3-104 29367601 6265-3-104 29474401 6265-3-104 29474402 6265-3-104 29474403 6265-3-104 29474404 6265-3-104 29474405 6265-3-104 29807601 6265-3-104 36006901 6265-3-104 36006902 6265-3-104 36006903 6265-3-104 36139201 6265-3-104 36139202 6265-3-104 36139203 6265-3-104 11496802R 6265-3-104 19944401R 6265-3-104 20185404R 6265-3-104 29807802R 6265-3-104 8659903R 6265-3-104 8790601R 6265-3-104 8908501R 6265-3-104 9033402R 6265-3-104 9341702R 6265-3-105 5779701 6265-3-105 5779702 6265-3-105 5779703 6265-3-105 6063601 6265-3-105 6063602 6265-3-105 6063603 6265-3-105 6063604 6265-3-105 6166201 6265-3-105 6166202 6265-3-105 6166203 6265-3-105 6252401 6265-3-105 6470701 6265-3-105 13513901 6265-3-105 13846901 6265-3-105 13846902 6265-3-105 14029501 6265-3-105 14029502 6265-3-105 14029503 6265-3-105 14107501 6265-3-105 14107502 6265-3-105 14107503 6265-3-105 14644801 6265-3-105 14644802 6265-3-105 14644803 6265-3-105 14738601 6265-3-105 14738602 6265-3-105 16429401 6265-3-105 16429402 6265-3-105 16429403 6265-3-105 16429404 6265-3-105 16749701 6265-3-105 16749702 6265-3-105 16749703 6265-3-105 16867201 6265-3-105 16867202 6265-3-105 17386301 6265-3-105 17386302 6265-3-105 17540101 6265-3-105 17540102 6265-3-105 24188901 6265-3-105 24188902 6265-3-105 24241301 6265-3-105 24241302 6265-3-105 24562601 6265-3-105 24562603 6265-3-105 24562604 6265-3-105 24562605 6265-3-105 25740901 6265-3-105 25740902 6265-3-105 25740903 6265-3-105 25740904 6265-3-105 26044001 6265-3-105 26044002 6265-3-105 26052301 6265-3-105 26723901 6265-3-105 26723902 6265-3-105 26723903 6265-3-105 26723904 6265-3-105 26723905 6265-3-105 27436201 6265-3-105 27436202 6265-3-105 27436203 6265-3-105 27436204 6265-3-105 27795301 6265-3-105 27795302 6265-3-105 27795303 6265-3-106 6263802 6265-3-106 6263803 6265-3-106 6263804 6265-3-106 6263805 6265-3-106 8908701 6265-3-106 10394701 6265-3-106 11078901 6265-3-106 11078902 6265-3-106 11078903 6265-3-106 11254101 6265-3-106 11275901 6265-3-106 11275902 6265-3-106 11335701 6265-3-106 11335702 6265-3-106 12145201 6265-3-106 12145202 6265-3-106 12725501 6265-3-106 12725502 6265-3-106 12725503 6265-3-106 12725504 6265-3-106 13514201 6265-3-106 13514202 6265-3-106 13514203 6265-3-106 13514204 6265-3-106 13514205 6265-3-106 13514206 6265-3-106 13514207 6265-3-106 24502006 6265-3-106 24502007 6265-3-106 24502008 6265-3-106 24502009 6265-3-106 24502010 6265-3-106 25940501 6265-3-106 25940502 6265-3-106 25940503 6265-3-106 25940504 6265-3-106 27935701 6265-3-106 27935702 6265-3-106 27935703 6265-3-106 27935704 6265-3-106 27935705 6265-3-106 28026901 6265-3-106 30193701 6265-3-106 30193702 6265-3-106 30193704 6265-3-106 30193706 6265-3-106 30193707 6265-3-106 32916802 6265-3-106 11078901R 6265-3-106 11078902R 6265-3-106 11078903R 6265-3-106 11254101R 6265-3-106 11275901R 6265-3-106 11275902R 6265-3-107 23996401 6265-3-107 23996402 6265-3-107 23996403 6265-3-107 24070301 6265-3-107 24070302 6265-3-107 25458201 6265-3-107 25458202 6265-3-107 25458203 6265-3-107 25458204 6265-3-107 25458205 6265-3-107 25847001 6265-3-107 25847002 6265-3-107 25847003 6265-3-107 25847004 6265-3-107 26404501 6265-3-107 26404502 6265-3-107 26404503 6265-3-107 26404504 6265-3-107 26558501 6265-3-107 26558502 6265-3-107 26558503 6265-3-107 26558504 6265-3-107 31761501 6265-3-107 31761502 6265-3-107 31761503 6265-3-107 31761504 6265-3-107 31761505 6265-3-107 32560307 6265-3-107 32950601 6265-3-107 32950602 6265-3-107 3376104BA 6265-3-107 4356902B |
| FEI Number |
3003070421
|
Recalling Firm/ Manufacturer |
Stryker Howmedica Osteonics Corp.
325 Corporate Drive
Mahwah
NJ
07430
|
| For Additional Information Contact | Ms. Christie Samsa 201-831-6365 |
Manufacturer Reason for Recall | Stryker Orthopaedics has become aware of the potential for cracking of the PMMA coating on their PM COCR Hip Stem product. |
FDA Determined Cause 2 | Employee error |
| Action | Stryker Orthopaedics sent an Urgent Product Recall letter dated January 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their Stryker Sales Representative to arrange for return of any affected product. Customers were also asked to fax back the attached Product Recall Acknowledgement Form as soon as possible to 201-831-6069. For any questions customers should call 201-972-2100.
For any questions regarding this recall please call 201-831-5970. |
| Quantity in Commerce | 16,913 units |
| Distribution | Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = JDI
|
|
|
|