| Date Initiated by Firm | June 04, 2009 |
| Date Posted | August 27, 2012 |
| Recall Status1 |
Terminated 3 on July 16, 2014 |
| Recall Number | Z-2274-2012 |
| Recall Event ID |
62564 |
| 510(K)Number | K072426 |
| Product Classification |
Orthosis, spinal pedicle fixation, for degenerative disc disease - Product Code NKB
|
| Product | Apelo Modular Screwdriver labeled in part: "APELO PEDICLE SCREW SYSTEM***Modular Screwdriver***Atlas Spine, Inc.***www.AtlasSpine.com*** 1555 Jupiter Park Drive #4***Jupiter, Florida 33458***561-741-1108***Tissue Sleeve***MFG Date: XX/XX/XXXX***Lot #: XXXX***Weldment***MFG Date: 03/23/2009***Sleeve Assembly***MFG Date: XX/XX/XXXX***Lot #: XXXX***Qty: 1 Rx Only***Non-Sterile. Sterilize Prior to Use*** Rev. A***. The Modular Screwdriver is used to insert modular pedicle screw implants which are part of the Apelo System. |
| Code Information |
PN# 70175-001 CAT#2006-41-0000 Lot # 00686 |
| FEI Number |
3003855635
|
Recalling Firm/ Manufacturer |
Atlas Spine, Inc.
1555 Jupiter Park Dr Ste 4
Jupiter
FL
33458-8085
|
| For Additional Information Contact | Tom Smith 561-741-1108 Ext. 102 |
Manufacturer Reason for Recall | Atlas Spine, Inc. recalled their Apelo Modular Screwdriver, a component of the Apelo Pedicle Screw System Instrumentation, because of complaints received about the tip breaking. If the tip is fractured inside of the modular screw, the tip could become implanted in the body or if the entire screw is removed, it could cause a delay in the surgery. |
FDA Determined Cause 2 | Component design/selection |
| Action | Consignees were notified by Telephone and visit. All products will be returned, quarantined and then destroyed. |
| Quantity in Commerce | 4 pieces |
| Distribution | USA Distribution to the state of Florida only. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = NKB
|