• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall SYMBIQ OneChannel and TwoChannel Infusers

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall SYMBIQ OneChannel and TwoChannel Infuserssee related information
Date Initiated by FirmAugust 28, 2013
Date PostedNovember 20, 2013
Recall Status1 Terminated 3 on November 07, 2016
Recall NumberZ-0372-2014
Recall Event ID 66263
510(K)NumberK110901 
Product Classification Pump, infusion - Product Code FRN
Product*** 1) SYMBIQ One-Channel Infuser; *** 2) SYMBIQ Two-Channel Infuser; *** E09 batteries can be identified by the "E09" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery; *** E10 batteries can be identified by the "E10" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery. Symbiq is a general-purpose infuser designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intra-arterial, subcutaneous, or irrigation routes of administration.
Code Information *** 1) SYMBIQ One-Channel Infuser, List Number: 16026-04-01/02/03/04/25/26/43/44/45/46/51/52/53/54/55/56/57/58/65/66/69/70/77/78/79/80/81/82/83/84/87/88/89/90/93/94;  *** 2) SYMBIQ Two- Channel Infuser, List Number: 16027-04-01/02/03/04/25/26/43/44/45/46/51/52/53/54/55/56/57/58/65/66/69/70/77/78/79/80/81/82/83/84/87/88/89/90/ 93/94;  *** X) Battery Pack E10 Part Number: HSP3566-E10; *** X) Battery Pack E09 Part Number: HSP3566-E09
FEI Number 3013319212
Recalling Firm/
Manufacturer
Hospira Inc.
275 N Field Dr
Lake Forest IL 60045-2579
For Additional Information ContactHospira Global Complaint Management
800-441-4100
Manufacturer Reason
for Recall
Hospira shipped the incorrect replaceable battery pack (Part No: HSP3566-E10) instead of the correct E09 battery pack (Part No. HSP3566-E09) for the Symbiq Single and Dual Channel Infusers.
FDA Determined
Cause 2
Employee error
ActionThe firm, Hospira, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 28, 2013 to its customers. Hospira contracted with Stericycle, Inc., in Indianapolis, IN for customer notification, response, tracking, returns, effectiveness checks and destruction of returned product starting on August 28, 2013. The letter described the product, problem and actions to be taken. Customers were instructed to: 1) complete and return the attached reply form via fax to 1-866-470-3291 or e-mail to Hospira2921@stericycle.com, even if you do not currently have the affected batteries; 2) contact Stericycle at 1-877-779-7810 (M-F, 8am - 5pm ET) if they require additional response forms; and, 3) return any affected batteries upon receipt of the new (correct) replacement batteries to Stericycle. Customers are also instructed to forward the recall information to any accounts that they may have shipped the affected batteries to and have them contact Stericycle to recieve a reply form and return label. For further inquiries, please contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 6, availabel 24 hours a day/7 days per week.
Quantity in Commerce59 batteries
DistributionUS Distribution in states of: : KY, MN, ND, NJ, NY, PA, and SD.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = FRN
-
-