| | Class 2 Device Recall InTouch Critical Care Bed |  |
| Date Initiated by Firm | June 04, 2014 |
| Date Posted | July 28, 2014 |
| Recall Status1 |
Terminated 3 on March 14, 2016 |
| Recall Number | Z-2122-2014 |
| Recall Event ID |
68470 |
| Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
| Product | InTouch Critical Care Bed, Stryker Medical. Model Numbers 2130, 2131, 2140, 2141, 2151, 2152, 2155, and 2156. |
| Code Information |
Model 2130 T01168 T01723 T01331 T02213 T02554 T01544 T01167 T01722 T01330 T02212 T02553 T01609 T01166 T01721 T01329 T02211 T02552 T01608 T01165 T01310 T01328 T02210 T02551 T01543 T01164 T02089 T01327 T02209 T02550 T02515 T01163 T01967 T01326 T02208 T02549 T02514 T01162 T01966 T01325 T02207 T02173 T02513 T01161 T01965 T01324 T02206 T02172 T01613 T01160 T01964 T01323 T02205 T02171 T01612 T02644 T01963 T01322 T02204 T02170 T01611 T02068 T01962 T01321 T02203 T02169 T01610 T02067 T01961 T01699 T02202 T02168 T02174 T02066 T01960 T01698 T02201 T02167 T02180 T02512 T01959 T01697 T02200 T02166 T02179 T02511 T01958 T01696 T02199 T01931 T02178 T02510 T01957 T01695 T02198 T01930 T02177 T02509 T01956 T01665 T02197 T01929 T02176 T02508 T01955 T01664 T02196 T01928 T02175 T02507 T01954 T01663 T02152 T01927 T01572 T02506 T01953 T01662 T01369 T01926 T01548 T01388 T01952 T01661 T01368 T01925 T01547 T01387 T01951 T02237 T01377 T01924 T01526 T01386 T01950 T02236 T01376 T01923 T01525 T01385 T01949 T02235 T01375 T01922 T02660 T01384 T01948 T02234 T01374 T01921 T02659 T02077 T02239 T02233 T01373 T01920 T02090 T02076 T02256 T02232 T01372 T01919 T02088 T02075 T02255 T02231 T01371 T01918 T01971 T01720 T02087 T02230 T01370 T01917 T01970 T01719 T02086 T02229 T01689 T01916 T01933 T01718 T01348 T02228 T01688 T01915 T01419 T02065 T01347 T02227 T01687 T01914 T01418 T02064 T01346 T02226 T01686 T01913 T01417 T02063 T01345 T02225 T01685 T01912 T01416 T02062 T01344 T02224 T01684 T01911 T01415 T02061 T01343 T02223 T01683 T01910 T01414 T02060 T01342 T02222 T01682 T01909 T01413 T02059 T01341 T02221 T02564 T01908 T01412 T02058 T01340 T02220 T02563 T01907 T01411 T02016 T01339 T02219 T02090 T01906 T01410 T02015 T01338 T02218 T02088 T01905 T01409 T02014 T01337 T02217 T01555 T01551 T01408 T02013 T01336 T02216 T01554 T01550 T01407 T02012 T01335 T02215 T01553 T01546 T01406 T02011 T01332 T02214 T01552 T01545 T01405 T01401 T02566 T02565 T01404 Model 2131 U04413 U03329 U08829 U51367 U06561 U01013 U04412 U03328 U03872 U51366 U06560 U01012 U04411 U50318 U03871 U51365 U03270 U05809 U04410 U50317 U03870 U51364 U03269 U01621 U04409 U50316 U03869 U51363 U03268 U07592 U04408 U50315 U03868 U51362 U03267 U04982 U04407 U50314 U03867 U51361 U03266 U04981 U04406 U50313 U03866 U50581 U03265 U06925 U04405 U50312 U03865 U50580 U03264 U08990 U04404 U50311 U03074 U50579 U03263 U08602 U04403 U01739 U03010 U50578 U02613 U08601 U04402 U01738 U03009 U50577 U02612 U08600 U04401 U01737 U03008 U50576 U02611 U08599 U04400 U01736 U05467 U50419 U51206 U08598 U04399 U01735 U05466 U50418 U51205 U08597 U04398 U01734 U05465 U50417 U51204 U08596 U04397 U01733 U05464 U50416 U51203 U08595 U04396 U01732 U05463 U50415 U51202 U08594 U04395 U01731 U05462 U50414 U08737 U08593 U04394 U01730 U05461 U50413 U08736 U08592 U03062 U01729 U05460 U50412 U08735 U08591 U03061 U01728 U05459 U50411 U08734 U08590 U03060 U51160 U05458 U50410 U08733 U08589 U01259 U08722 U05457 U50409 U08732 U08588 U01258 U08721 U05456 U50404 U03660 U08587 U01257 U08720 U05455 U50403 U03659 U08586 U04577 U08719 U05454 U50402 U03658 U08585 U04422 U08718 U06914 U50401 U03657 U08584 U02179 U08717 U06913 U50400 U03656 U08583 U02178 U08716 U06912 U50399 U03655 U08582 U02177 U08715 U06911 U08096 U03654 U08581 U02176 U08714 U06910 U08095 U03653 U08580 U02175 U08713 U06909 U08094 U03652 U08579 U02174 U08712 U06908 U08093 U03651 U08578 U02173 U08711 U06907 U08085 U03650 U08577 U02172 U08710 U01401 U08084 U03649 U08576 U02171 U08709 U01400 U08083 U03648 U08575 U02861 U08708 U01399 U08082 U03647 U08574 U02860 U08707 U01398 U08081 U03646 U08573 U02859 U04949 U04372 U08080 U03645 U08572 U02858 U04948 U04371 U08079 U03644 U08571 U02857 U04947 U04370 U08078 U03643 U08570 U02856 U04946 U04369 U08077 U03642 U08569 U02855 U04945 U04368 U08076 U03641 U08568 U02854 U04944 U04367 U06733 U03640 U08567 U02853 U02921 U04366 U06732 U03639 U08566 U02852 U02920 U04365 U06731 U03638 U08565 U02851 U02919 U04364 U06730 U03637 U07420 U02850 U02918 U04363 U06729 U03636 U07239 U02849 U02917 U04362 U06728 U03635 U07238 U02848 U02916 U04361 U06727 U03634 U07237 U02847 U02915 U04360 U06726 U03633 U07236 U02846 U02914 U04359 U06725 U03632 U06820 U02845 U02913 U04358 U03366 U03631 U05021 U02844 U02912 U04357 U03365 U03630 U05020 U02843 U02911 U04356 U03364 U03629 U05019 U02842 U02910 U04355 U03363 U03628 U05018 U02841 U02909 U04354 U03362 U03627 U05017 U02840 U02908 U04353 U03361 U03626 U06800 U02839 U02907 U04352 U03360 U03625 U06799 U01331 U02906 U04351 U03359 U50656 U01712 U01330 U02905 U04350 U03358 U50655 U01711 U01329 U02904 U04349 U03357 U07957 U01710 U01328 U02903 U04348 U03356 U05780 U02816 U01327 U02902 U04347 U05506 U50088 U02806 U01326 U02901 U04346 U05505 U50087 U02425 U01325 U02900 U04345 U05504 U50086 U01391 U06519 U02899 U04344 U05503 U50085 U06017 U06518 U02898 U04343 U05502 U50084 U06016 U06517 U02897 U04342 U05501 U50083 U06015 U06516 U02896 U50575 U05500 U50082 U06014 U06515 U02895 U50190 U05499 U50081 U06013 U06514 U02894 U50189 U05498 U50080 U06012 U06513 U02893 U50188 U05497 U50079 U06011 U06512 U02892 U50187 U05204 U50078 U06010 U06511 U02891 U50186 U05203 U08539 U05920 U06510 U02890 U50185 U05202 U08538 U05919 U06509 U50650 U50184 U05201 U08537 U05918 U06508 U50649 U50183 U05200 U08536 U05917 U06507 U50648 U50182 U05199 U08535 U05916 U06506 U50647 U50181 U05198 U08534 U05915 U06505 U50646 U50180 U04701 U08533 U51464 U06504 U50645 U50179 U04700 U08532 U51463 U09003 U50644 U50178 U04699 U08531 U51462 U09002 U50643 U50177 U04698 U08530 U51461 U09001 U07025 U50176 U04697 U08529 U51460 U09000 U07024 U50175 U03083 U08528 U50805 U08999 U07023 U50174 U03082 U08527 U50804 U03327 U07022 U50173 U03081 U08526 U50803 U03326 U07021 U50172 U03080 U08525 U50802 U03325 U07020 U50171 U03079 U08329 U50801 U02992 U07019 U50169 U03078 U08328 U08937 U02991 U07018 U50168 U03077 U08327 U08936 U07419 U07017 U50167 U03076 U07583 U08935 U03832 U07016 U07928 U03075 U07582 U08934 U03831 U07015 U07927 U01699 U07581 U08933 U03830 U07014 U07926 U01698 U07580 U08932 U03829 U08888 U07925 U01697 U07579 U08849 U03828 U08887 U02741 U01696 U07578 U08848 U03827 U08886 U02740 U01695 U07577 U08847 U03826 U08885 U02739 U01694 U07576 U08117 U03825 U08884 U02738 U01693 U07575 U07464 U03824 U08883 U02737 U01692 U07574 U07463 U03823 U06641 U02736 U01691 U07573 U07462 U03822 U06640 U02735 U01690 U07572 U07461 U03821 U06639 U01514 U50392 U07571 U07460 U03820 U06638 U02415 U03042 U07570 U07459 U03819 U50807 U05447 U03041 U07569 U51712 U03818 U50806 U05446 U03040 U07568 U51437 U03817 U06793 U05445 U03039 U07567 U51436 U03816 U06792 U05444 U03038 U07566 U51435 U03815 U06791 U05443 U03037 U07565 U51434 U03814 U06790 U05442 U03036 U07564 U51433 U03813 U06789 U05441 U03035 U07563 U51432 U03812 U06788 U05440 U04005 U07562 U51431 U03811 U06787 U05439 U04004 U07561 U51430 U03810 U06786 U05438 U04003 U07560 U51429 U03809 U05210 U05437 U04002 U07559 U51428 U03808 U04254 U05436 U04001 U06154 U51427 U03807 U04253 U05435 U04000 U06153 U51426 U03806 U04252 U05434 U03999 U06152 U51425 U03805 U04251 U05433 U03998 U06151 U51424 U03804 U04250 U07640 U03997 U06150 U51423 U03803 U04249 U07639 U03996 U06149 U51422 U03802 U04248 U07638 U03995 U06148 U51421 U03801 U04247 U07637 U03994 U06147 U51420 U03800 U51208 U05064 U03993 U06146 U51419 U03799 U51207 U05063 U03992 U06145 U51418 U03798 U08487 U05424 U03991 U06144 U51417 U03797 U08486 U05423 U03990 U06143 U51416 U03796 U08485 U05422 U03989 U06142 U51415 U03795 U08484 U05421 U03988 U06141 U51414 U03794 U08483 U05420 U03987 U06140 U51413 U03793 U08473 U05419 U03986 U06139 U51412 U03792 U08472 U05418 U03985 U02838 U51411 U08743 U08471 U05417 U03984 U02837 U51410 U08742 U08470 U05416 U03983 U02836 U51409 U04936 U08469 U05415 U03982 U02835 U51408 U04935 U08468 U05414 U03981 U02834 U51407 U04934 U08467 U05413 U03980 U02833 U51406 U04933 U08466 U05412 U03979 U02832 U51405 U08524 U08465 U05411 U04187 U02831 U51404 U08523 U08464 U05410 U04186 U09007 U51403 U08522 U08425 U05409 U04185 U09006 U51402 U08521 U08424 U05408 U04184 U09005 U51401 U08520 U08423 U05407 U04183 U08701 U51400 U08519 U08422 U05406 U04182 U08700 U51399 U08518 U08421 U05516 U04181 U08699 U51398 U08517 U08420 U05515 U04180 U08698 U51397 U03139 U08419 U05514 U04179 U08697 U51396 U03138 U08418 U05513 U04178 U08696 U51395 U03137 U08417 U05512 U04177 U08695 U51394 U03136 U08416 U05511 U04176 U08694 U51393 U03135 U08414 U05510 U04175 U08693 U51392 U03134 U08413 U05509 U04174 U08692 U51391 U03133 U08412 U05508 U04173 U08691 U51390 U03132 U08411 U05507 U04172 U08690 U51389 U50800 U08410 U05786 U04171 U08689 U51388 U06378 U06122 U05785 U04170 U08688 U51387 U06377 U06121 U05784 U04169 U08687 U51386 U06376 U06120 U05783 U04168 U08686 U51385 U06375 U06119 U05782 U04167 U08683 U51102 U06374 U05373 U05781 U04166 U08682 U08613 U06373 U05372 U05493 U04165 U08681 U08612 U06372 U05371 U05492 U04164 U08680 U08611 U06371 U05370 U05491 U04163 U08679 U08610 U06370 U05369 U05490 U01254 U08678 U08609 U03892 U05368 U05489 U01253 U51744 U08608 U03891 U05367 U05488 U01252 U08816 U08495 U03890 U05366 U05487 U01251 U08815 U08494 U03889 U05365 U05486 U01250 U51964 U08493 U03058 U05364 U05485 U50444 U51893 U08492 U02768 U05363 U05484 U05590 U51892 U08491 U02256 U05362 U05483 U05589 U51891 U08490 U01534 U05361 U05482 U05588 U51890 U08489 U01513 U05360 U05481 U05587 U51889 U08488 U06647 U05359 U05480 U05586 U51888 U08119 U04393 U05358 U05479 U05585 U51887 U07477 U04392 U05357 U05478 U05584 U51886 U07476 U04391 U05356 U05477 U05583 U51885 U07475 U04390 U05355 U05476 U05582 U51884 U07474 U01374 U05354 U05475 U05581 U51883 U07473 U01373 U05353 U05474 U05580 U51882 U07472 U01372 U05352 U05473 U05579 U51881 U07471 U01371 U05351 U05472 U05578 U51645 U07470 U01370 U05350 U05471 U05577 U51644 U07469 U01369 U04722 U05470 U05576 U51643 U07468 U01368 U04721 U05469 U05575 U51642 U07467 U05215 U04720 U05468 U05574 U51476 U07466 U50583 U04719 U05209 U05573 U51475 U07465 U50582 U01249 U05208 U05572 U51474 U07427 U06526 U01248 U05207 U05571 U51473 U07426 U06525 U01247 U05206 U05561 U51472 U07425 U03055 U01246 U01657 U05560 U51471 U07424 U03054 U01245 U01656 U05559 U51470 U07423 U03053 U01244 U01655 U05558 U51469 U07422 U03052 U01243 U01654 U05557 U51468 U07421 U03051 U01242 U01653 U05556 U51467 U07197 U03050 U01241 U01652 U05555 U51466 U07196 U03049 U01240 U01651 U05554 U51465 U07195 U03048 U01239 U01650 U05553 U51319 U07194 U03047 U01238 U01649 U05552 U50236 U07193 U03046 U01237 U01648 U05551 U50235 U07192 U03045 U01236 U01647 U05550 U50234 U07191 U03044 U01235 U01646 U05549 U50233 U07190 U03119 U01234 U01645 U05548 U50077 U07189 U02941 U01233 U01644 U05547 U50076 U07188 U02940 U01232 U01643 U05546 U08677 U07187 U02939 U01231 U01642 U05545 U08676 U07186 U02938 U01230 U01641 U05544 U08675 U07185 U02937 U01229 U01640 U05543 U08674 U07184 U02936 U01228 U01639 U05542 U08673 U07183 U02935 U01227 U01638 U05541 U08672 U07182 U02934 U01226 U01637 U05540 U08671 U07181 U02933 U01225 U04246 U05539 U08643 U07180 U02932 U01224 U04245 U05538 U08642 U07179 U50872 U01223 U04244 U05537 U08641 U07146 U50871 U01222 U04243 U05536 U07905 U05914 U50870 U01221 U04242 U05535 U07904 U05913 U50869 U01220 U04241 U05534 U07903 U05912 U50868 U01219 U04240 U05533 U07886 U04844 U50867 U01218 U04239 U05532 U07861 U04843 U50866 U01217 U50654 U05531 U07820 U04842 U50865 U01216 U06709 U05530 U07819 U04841 U50864 U01215 U06708 U05529 U07818 U04840 U50863 U01214 U05214 U05528 U07817 U04839 U50862 U01213 U05213 U05527 U07764 U04838 U50861 U01212 U05212 U05526 U07763 U04837 U50860 U01211 U05211 U05525 U07762 U04836 U50859 U01210 U05883 U05524 U07761 U04835 U50858 U01209 U05882 U05523 U07636 U04834 U50857 U01208 U05881 U05522 U07635 U04833 U50856 U01207 U05880 U07982 U07634 U04832 U50855 U01206 U05879 U07981 U07633 U04705 U50854 U01205 U03675 U06702 U07632 U04704 U50853 U01204 U03674 U06701 U07534 U04703 U50852 U01203 U03673 U06700 U07533 U04702 U50851 U01202 U03672 U06699 U07532 U04285 U07302 U06369 U03671 U06698 U07531 U03718 U07301 U06368 U03670 U06697 U07520 U03717 U07299 U06367 U03669 U06696 U07519 U03716 U07298 U06366 U03668 U06695 U06818 U03715 U07297 U06365 U03667 U06694 U06817 U03714 U07296 U06364 U03666 U06693 U06659 U03713 U07295 U06363 U03665 U06692 U06658 U03712 U07294 U06362 U03664 U06691 U06657 U03711 U07293 U06361 U06185 U06690 U06656 U03710 U07292 U06360 U06184 U06689 U06611 U03709 U07291 U06359 U06183 U06688 U06610 U03708 U07290 U06358 U06182 U06687 U06609 U03707 U50850 U06357 U06181 U06686 U06608 U03706 U50849 U06356 U06180 U06685 U06607 U03705 U50848 U06355 U06179 U06684 U06606 U03704 U50847 U06354 U06178 U06683 U06605 U03703 U50846 U06353 U08645 U06682 U06604 U03702 U50845 U06352 U08644 U06681 U06603 U03701 U50844 U06351 U07900 U06680 U06602 U03700 U50843 U06350 U07899 U06679 U06551 U03699 U50842 U06349 U06711 U06678 U06550 U03698 U50841 U06348 U02721 U06677 U06549 U03697 U50840 U06347 U02720 U06676 U06548 U03696 U50839 U06346 U02719 U06675 U06547 U03695 U50838 U06345 U02718 U06674 U06546 U03694 U50837 U06344 U02717 U06673 U06545 U03504 U50836 U06343 U02716 U06672 U06544 U50909 U50835 U06342 U02715 U06671 U06543 U50480 U50834 U06341 U02714 U06670 U06542 U50479 U50833 U06340 U02713 U06669 U06541 U50478 U50832 U08830 U02712 U06668 U06540 U50477 U50831 U07278 U02711 U06667 U06539 U50476 U50830 U07277 U02710 U06666 U06538 U50475 U50829 U07276 U02709 U06665 U06537 U50474 U50828 U07275 U02708 U06664 U06536 U08139 U50827 U07274 U02707 U06663 U06535 U08138 U03623 U07273 U02706 U06662 U06534 U08137 U03622 U07272 U02705 U06661 U06533 U08136 U03621 U07271 U02704 U06660 U06532 U08135 U03620 U07270 U02703 U07512 U06531 U08134 U03619 U07269 U02702 U07511 U06530 U08133 U03618 U06710 U02701 U07510 U06529 U08132 U02398 U50653 U02700 U50336 U06528 U08131 U02397 U50652 U02699 U50335 U06527 U08130 U02396 U50651 U02698 U50334 U05884 U08129 U02395 U08828 U02697 U50333 U05111 U08128 U02394 U07283 U02696 U50332 U05110 U08127 U02393 U07282 U02695 U50331 U04924 U05205 U02392 U07281 U02694 U50330 U04923 U05618 U02391 U07280 U02693 U50329 U04666 U05617 U02390 U02170 U02692 U50328 U04547 U05221 U02389 U51963 U02691 U50327 U04094 U05220 U02388 U51962 U02690 U50326 U04093 U05219 U02387 U51961 U06785 U50325 U04092 U05218 U02386 U51960 U06784 U50324 U04091 U05217 U02385 U51959 U04335 U50323 U04009 U05216 U02384 U51958 U04334 U50322 U04008 U04217 U01458 U51957 U04333 U50321 U04007 U04216 U01457 U51956 U04332 U50320 U04006 U04215 U01456 U51955 U04331 U50319 U03732 U04214 U01455 U51954 U04985 U03300 U03731 U04213 U01454 U07279 U04984 U03303 U03730 U04212 U01453 U01200 U04983 U03302 U03729 U04211 U08514 U01199 U04980 U07105 U03728 U04210 U05878 U01198 U04979 U07104 U03727 U04209 U05877 U01197 U04978 U07103 U03726 U04208 U05876 U01196 U04067 U07102 U03725 U04207 U05875 U01527 U04066 U07101 U03724 U04206 U05874 U01526 U04065 U07100 U03723 U03624 U05873 U01525 U07387 U07099 U03722 U01389 U05872 U07924 U07386 U07098 U03721 U01388 U05871 U07923 U07385 U07097 U03496 U50390 U05870 U04141 U50658 U07096 U03495 U50389 U03347 U08763 U50657 U01584 U03494 U01402 U03346 U08762 U51384 U01583 U03493 U03853 U03345 U08761 U51383 U01582 U03492 U03520 U03344 U08760 U51382 U01581 U03491 U03519 U03343 U08759 U51381 U01580 U03490 U03518 U03342 U08758 U51380 U01579 U03489 U03517 U03341 U08757 U51379 U01578 U03488 U03516 U03340 U08756 U51378 U07458 U03487 U03515 U03339 U08755 U51377 U04233 U03486 U03514 U03338 U08754 U51376 U04232 U03485 U03330 U03337 U08753 U51375 U04231 U03073 U08746 U03336 U08752 U51374 U04230 U02615 U51368 U03335 U08751 U51373 U04229 U02614 U06562 U03334 U08750 U51372 U04228 U01436 U01014 U03333 U08749 U51371 U06565 U01435 U01015 U03332 U08748 U51370 U06564 U01265 U06563 U03331 U08747 U51369 Model 2140 T01127 T02140 T02038 T01717 T01997 T01794 T01124 T01934 T02455 T01716 T01996 T01793 T01122 T02151 T02144 T01715 T01995 T01792 T01115 T02150 T02136 T01714 T01994 T01791 T01112 T02022 T01571 T01713 T01993 T01790 T01108 T02021 T01320 T01712 T01992 T01789 T01106 T01895 T01319 T01711 T01991 T01788 T01171 T01894 T01318 T01710 T01990 T01787 T01079 T01893 T01317 T01709 T01989 T01786 T01060 T01892 T01316 T01708 T01988 T01785 T01057 T01891 T01315 T01707 T01987 T01784 T01058 T01890 T01314 T01706 T01986 T01783 T01012 T01889 T01313 T01705 T01985 T01782 T01011 T01888 T01312 T01704 T01984 T01781 T01010 T01887 T01311 T01703 T01983 T01780 T01009 T01886 T02469 T01702 T01982 T01779 T01008 T01885 T02414 T01701 T01981 T01750 T01007 T01884 T02320 T01700 T01980 T01749 T01006 T01883 T02319 T01640 T01979 T01748 T01005 T01882 T02318 T01639 T01978 T01747 T01004 T01881 T02317 T01638 T01977 T01746 T01003 T01880 T02316 T01637 T01976 T01745 T01002 T01879 T02315 T01636 T01975 T01744 T01001 T01878 T02314 T01635 T01974 T01743 T01142 T01877 T02313 T01634 T01973 T01742 T01141 T01876 T02312 T01633 T01972 T01741 T01140 T01901 T02311 T01632 T02505 T02613 T01139 T01900 T02310 T01631 T01614 T02612 T01137 T01899 T02309 T02080 T01496 T02611 T01136 T01898 T02308 T01690 T01494 T02610 T01135 T01897 T02307 T01617 T01478 T02609 T01134 T01896 T02306 T02474 T01466 T02608 T01133 T01904 T02305 T02473 T01350 T02607 T01131 T01903 T02304 T02472 T02121 T02606 T01129 T01902 T02303 T02471 T02120 T02605 T01126 T01821 T02302 T02470 T02119 T02604 T01125 T01820 T02301 T02467 T02118 T02603 T01120 T01819 T02300 T02465 T02117 T02602 T01118 T01818 T02299 T01575 T02116 T02601 T01117 T01817 T02298 T02139 T02028 T02600 T01114 T01816 T02297 T02085 T02027 T02599 T01113 T01815 T01309 T02084 T02026 T02598 T01110 T01814 T01308 T02132 T02025 T02597 T01109 T01813 T01307 T02131 T02024 T02596 T01105 T01812 T01306 T02130 T02023 T02595 T01102 T01811 T01305 T02129 T01403 T02453 T01101 T01810 T01304 T02128 T01402 T02452 T01172 T01809 T01303 T02127 T02010 T02451 T01116 T01808 T01302 T02126 T02009 T02341 T01230 T01807 T01301 T02125 T02008 T02340 T01229 T01806 T01300 T01383 T02007 T02339 T01228 T01805 T01299 T01382 T02006 T02338 T01227 T01804 T01298 T01381 T02005 T02337 T01226 T01803 T01297 T01380 T02004 T02336 T01225 T01802 T01296 T01379 T02003 T02335 T01224 T01801 T01295 T02658 T02002 T02334 T01223 T01800 T01294 T02657 T02001 T02333 T01222 T01799 T01293 T02656 T02000 T02332 T01221 T01798 T01292 T02655 T02520 T02331 T01220 T01797 T01291 T02654 T02519 T02330 T01219 T01796 T01290 T02653 T02518 T02329 T01218 T02548 T01289 T02652 T02517 T02328 T01217 T02547 T01288 T01607 T02649 T02327 T01216 T02546 T01287 T01606 T02648 T02326 T01215 T02545 T01286 T01605 T02647 T02325 T01214 T02544 T01285 T01604 T02646 T02324 T01213 T02543 T01284 T01603 T02645 T02323 T01212 T02542 T01283 T01602 T02468 T02322 T01211 T02541 T01282 T01601 T02466 T02643 T01210 T02540 T01281 T01600 T02464 T01499 T01209 T02539 T01280 T01599 T02454 T01498 T01208 T02538 T01279 T01598 T01615 T01497 T01207 T02537 T01278 T01597 T02142 T01495 T01206 T02536 T01277 T01596 T02141 T01493 T01205 T02535 T01276 T01595 T02111 T01492 T01204 T02534 T01275 T01594 T02082 T01491 T01203 T02533 T01274 T01593 T02522 T01490 T01202 T02532 T01273 T01592 T02521 T01489 T01201 T02531 T01272 T01591 T02036 T01488 T01200 T02530 T01271 T01590 T01728 T01487 T01199 T02529 T01270 T01589 T01727 T01486 T01198 T02528 T01269 T01588 T01726 T01485 T01197 T02527 T01268 T01587 T01725 T01484 T01196 T02526 T01267 T01586 T01724 T01483 T01195 T02525 T01266 T01585 T02195 T01482 T01194 T02524 T01265 T01584 T02194 T01481 T01193 T02523 T01264 T01583 T02193 T01480 T01192 T02242 T01263 T01582 T02192 T01479 T01191 T02079 T01262 T01581 T02191 T01477 T01190 T02137 T01261 T01580 T02190 T01476 T01189 T01681 T01260 T01579 T02189 T01475 T01188 T01680 T01259 T01578 T02188 T01474 T01187 T01679 T01258 T01577 T02187 T01473 T01186 T01678 T01257 T01576 T02186 T01472 T01185 T01677 T01256 T02072 T02185 T01471 T01184 T01676 T01255 T02071 T02184 T01470 T01183 T01675 T01254 T02070 T02102 T01469 T01182 T01674 T01253 T02069 T02101 T01468 T01181 T01673 T01252 T02122 T02100 T01467 T01180 T01672 T01251 T02108 T02099 T01465 T01179 T01671 T01250 T02134 T02098 T01464 T01178 T01670 T01249 T02183 T02097 T01463 T01177 T01669 T01248 T02182 T02096 T01462 T01176 T01668 T01247 T002183 T02095 T01461 T01175 T01667 T01246 T002182 T02094 T01630 T01174 T01666 T01245 T02078 T02093 T01629 T01173 T02112 T01244 T02371 T02092 T01628 T01104 T02456 T01243 T02370 T02091 T01627 T01159 T01862 T01242 T02369 T02586 T01626 T01158 T01861 T01241 T02368 T02585 T01625 T01157 T01860 T01240 T02367 T02584 T01624 T01156 T01859 T01239 T02366 T02583 T01623 T01155 T01858 T01238 T02365 T02582 T01622 T01154 T01857 T01237 T02364 T02581 T01621 T01153 T01856 T01236 T02363 T02580 T01620 T01152 T01855 T01235 T02362 T02579 T01619 T01151 T01854 T01234 T02361 T02578 T02057 T01150 T01853 T01233 T02360 T02577 T02056 T01149 T01852 T01232 T02359 T02576 T01367 T01148 T01851 T01231 T02358 T02575 T01366 T01147 T01850 T01947 T02357 T01740 T01365 T01146 T01849 T01946 T02356 T01739 T01364 T01145 T01848 T01945 T02276 T01738 T01363 T01144 T01847 T01944 T02275 T01737 T01362 T01099 T01846 T01943 T02274 T01736 T01361 T01098 T01845 T01942 T02273 T01735 T01360 T01097 T01844 T01941 T02272 T01734 T01359 T01096 T01843 T01940 T02271 T01733 T01358 T01095 T01842 T01939 T02270 T01732 T01357 T01094 T01841 T01938 T02269 T01731 T01356 T01093 T01840 T01937 T02268 T01730 T01355 T01092 T01839 T01936 T02267 T01729 T01354 T01091 T01838 T01549 T02266 T01530 T01353 T01090 T01837 T02106 T02265 T01349 T01352 T01089 T01836 T02105 T02264 T02355 T01351 T01088 T01835 T02104 T02263 T02354 T02574 T01087 T01834 T02103 T02262 T02353 T02573 T01086 T01833 T02020 T02261 T02352 T02572 T01085 T01832 T02019 T02260 T02351 T02571 T01084 T01831 T02018 T02259 T02642 T02570 T01083 T01830 T02017 T02258 T02641 T02569 T01082 T01829 T02115 T02257 T02640 T02568 T01081 T01828 T02114 T02029 T02639 T02567 T01080 T01827 T02113 T01968 T02638 T02475 T01078 T01826 T01570 T01694 T02637 T02135 T01077 T01825 T01569 T02516 T02636 T02123 T01076 T01824 T01568 T02074 T02635 T02109 T01075 T01823 T01567 T02073 T02634 T01514 T01074 T01822 T01566 T01378 T02633 T01513 T01073 T02254 T01565 T02561 T02632 T01512 T01072 T02253 T01564 T02560 T02631 T01511 T01071 T02252 T01563 T02559 T02630 T01510 T01070 T02251 T01562 T02558 T02629 T01509 T01069 T02250 T01561 T02557 T02628 T01508 T01068 T02249 T01560 T02556 T02627 T01507 T01067 T02248 T01559 T02555 T02626 T01506 T01066 T01616 T02181 T01334 T02625 T01505 T01065 T02493 T02415 T01333 T02624 T01504 T01064 T02492 T02143 T02503 T02623 T01503 T01063 T02491 T02138 T02502 T02622 T01502 T01062 T02490 T01693 T02501 T02621 T01501 T01061 T02489 T01692 T02500 T02620 T01500 T01059 T02488 T01691 T02499 T02619 T01573 T01056 T02487 T01121 T02498 T02618 T01660 T01055 T02486 T01542 T02497 T02617 T01659 T01054 T02485 T01541 T02496 T02616 T01658 T01053 T02484 T01540 T02495 T02615 T01657 T01143 T02483 T01539 T01457 T02614 T01656 T01036 T02482 T01538 T01456 T02083 T01655 T01035 T02481 T01537 T01455 T01648 T01654 T01034 T02480 T01536 T01454 T01647 T01653 T01033 T02479 T01535 T01453 T01646 T01652 T01032 T02478 T02349 T01452 T01645 T01651 T01031 T02477 T02348 T01451 T01644 T01650 T01030 T02476 T02110 T01450 T01643 T01649 T01029 T01460 T01969 T01449 T01642 T01558 T01028 T01459 T01618 T01448 T01641 T01557 T01027 T01458 T02504 T01447 T02145 T01556 T01026 T02124 T02494 T01778 T02081 T02661 T01025 T01875 T02463 T01777 T02392 T071 T01024 T01874 T02462 T01776 T02391 T070 T01023 T01873 T02461 T01775 T02390 T069 T01022 T01872 T02460 T01774 T02389 T02562 T01021 T01871 T02459 T01773 T02388 T02450 T01020 T02147 T02458 T01772 T02387 T02449 T01019 T01446 T02457 T01771 T02386 T02448 T01018 T01445 T02350 T01770 T02385 T02447 T01017 T01444 T02347 T01769 T02384 T02446 T01016 T01443 T02346 T01768 T02383 T02445 T01015 T01442 T02345 T01767 T02382 T02444 T01014 T01441 T02344 T01766 T02381 T02443 T01138 T01440 T02343 T01765 T02380 T02442 T01132 T01434 T02342 T01764 T02379 T02441 T01130 T01433 T02133 T01763 T02378 T02440 T01128 T01432 T02651 T01762 T02377 T02439 T01123 T01431 T077 T01761 T02376 T02438 T01119 T01430 T076 T01760 T02375 T02437 T01111 T01429 T075 T01759 T02374 T02436 T01107 T01428 T074 T01758 T02373 T02435 T01103 T01427 T073 T01757 T02372 T02434 T01100 T01426 T072 T01756 T02296 T02433 T01052 T01425 T068 T01755 T02295 T02432 T01051 T01424 T067 T01754 T02294 T02431 T01050 T01423 T066 T01753 T02293 T02430 T01049 T01422 T065 T01752 T02292 T02429 T01048 T01421 T01534 T01751 T02291 T02428 T01047 T01420 T01533 T02413 T02290 T02427 T01046 T01870 T01532 T02412 T02289 T02426 T01045 T01869 T01531 T02411 T02288 T02425 T01044 T01868 T01529 T02410 T02287 T02424 T01043 T01867 T01528 T02409 T02286 T02423 T01042 T01866 T01527 T02408 T02285 T02422 T01041 T01865 T01524 T02407 T02284 T02421 T01040 T01864 T01523 T02406 T02283 T02420 T01039 T01863 T01522 T02405 T02282 T02419 T01038 T02397 T01521 T02404 T02281 T02418 T01037 T02396 T01520 T02403 T02280 T02417 T01229 T02395 T01519 T02402 T02279 T02416 T01170 T02394 T01518 T02401 T02278 T02321 T01169 T02393 T01517 T02400 T02277 T02247 T01012 T02055 T01516 T02399 T02165 T02246 T01011 T02054 T01515 T02398 T02164 T02245 T01010 T02053 T01439 T02241 T02163 T02244 T01009 T02052 T01438 T02037 T02162 T02243 T01008 T02051 T01437 T02035 T02161 T02238 T01007 T02050 T01436 T02034 T02160 T02153 T01006 T02049 T01435 T02033 T02159 T02107 T01005 T02048 T01398 T02032 T02158 T01935 T01004 T02047 T01397 T02031 T02157 T01400 T01003 T02046 T01396 T02030 T02156 T01399 T01002 T02045 T01395 T02591 T02155 T02036 T01001 T02044 T01394 T02590 T02154 T01932 T02650 T02043 T01393 T02589 T01574 T01795 T02240 T02042 T01392 T02588 T02594 T01998 T02146 T02041 T01391 T02587 T02593 T01389 T02149 T02040 T01390 T01999 T02592 T02039 T02148 Model 2141 U06231 U02047 U08108 U50198 U51306 U04966 U06230 U02046 U08107 U50197 U51447 U04965 U06229 U02045 U08106 U50196 U03978 U04964 U06228 U02044 U08105 U50195 U03977 U04963 U06227 U50269 U08104 U50194 U03976 U04962 U06226 U50268 U08103 U50193 U03975 U04961 U06225 U50267 U51656 U50192 U03974 U04960 U03033 U50266 U01904 U50191 U03973 U04959 U03240 U50265 U01903 U51038 U03972 U04958 U03239 U50264 U01902 U08786 U03971 U04957 U03238 U07333 U01901 U08785 U03970 U04956 U03237 U05566 U01900 U08784 U03969 U04955 U03236 U05349 U01150 U08783 U03968 U04954 U03235 U04995 U01149 U08782 U03967 U51648 U03234 U04994 U01148 U08781 U03966 U07608 U03233 U04992 U01147 U08780 U03965 U07607 U03232 U03513 U01146 U08779 U03964 U07606 U03231 U03512 U05649 U05610 U03963 U07605 U03230 U03503 U50750 U05603 U03962 U07604 U03229 U03502 U50749 U04993 U03961 U07603 U03882 U02258 U50748 U04829 U03960 U07558 U03881 U02257 U06242 U50510 U03959 U07557 U03880 U51749 U02603 U50509 U03958 U07556 U03879 U51748 U01182 U08921 U03957 U07555 U03878 U51747 U51331 U08920 U03956 U07554 U03877 U51746 U08435 U08919 U03955 U07553 U03876 U04626 U08434 U08918 U03954 U07552 U03875 U04625 U08409 U08917 U03953 U07551 U03874 U04624 U08408 U08916 U03952 U07550 U03873 U04623 U08407 U08915 U03951 U07549 U51117 U04622 U08406 U08914 U03950 U06040 U51116 U04621 U08405 U08913 U03949 U06039 U51115 U04620 U08404 U08912 U03948 U06038 U51114 U04619 U08403 U08911 U03947 U06037 U51113 U05609 U08402 U08910 U03946 U06036 U01743 U04930 U08387 U08909 U03945 U06035 U08639 U08899 U08386 U08908 U03944 U06034 U01720 U08647 U08385 U08907 U03943 U06033 U01719 U06177 U08384 U08906 U03942 U06032 U04928 U06176 U08383 U08905 U03941 U06031 U05451 U06175 U08382 U08904 U03940 U06030 U04925 U06174 U08381 U50947 U03939 U06029 U03499 U06173 U08380 U07691 U03938 U06028 U03498 U06172 U06993 U07690 U03937 U06027 U03497 U06171 U06992 U07689 U03936 U06026 U03092 U06170 U06991 U07688 U03935 U06025 U03091 U06169 U06990 U07687 U03934 U06024 U03090 U06168 U06989 U07686 U03933 U06023 U03089 U04515 U06988 U02723 U03932 U06022 U03088 U04514 U01130 U01384 U03931 U06021 U03087 U04513 U01129 U01377 U03930 U06020 U03086 U04512 U01128 U08447 U03929 U06019 U03085 U04511 U01127 U08446 U08304 U50346 U07116 U04510 U01126 U08445 U08303 U50345 U07115 U05758 U01125 U08444 U08561 U50344 U07114 U05757 U01124 U08443 U06164 U50343 U07113 U05756 U01123 U08442 U05907 U50342 U07112 U05755 U01122 U08441 U50492 U50341 U07111 U05754 U01121 U08440 U50249 U50340 U07110 U05753 U01120 U08439 U50096 U50339 U07109 U05752 U01119 U08438 U50094 U50338 U07095 U05751 U01118 U08437 U50073 U50337 U07094 U05750 U01117 U08436 U08895 U02888 U07093 U05749 U01116 U08389 U08638 U02887 U07092 U05748 U01115 U08388 U08614 U02886 U07091 U05747 U01114 U02601 U08564 U02885 U07090 U05746 U01113 U01926 U08562 U02884 U07089 U05745 U01112 U01925 U08314 U02883 U07088 U05744 U01111 U01924 U08313 U02882 U07087 U05743 U01110 U01923 U08312 U02881 U07086 U05742 U01109 U50984 U08311 U02880 U07085 U05741 U01108 U50983 U08310 U02879 U07084 U05740 U01107 U50982 U08309 U05763 U07083 U05739 U01106 U50981 U08308 U50506 U07082 U05738 U01105 U50980 U08307 U50505 U07081 U05737 U01104 U50979 U02411 U50504 U07080 U05736 U08116 U50978 U02410 U06779 U07079 U05735 U08115 U50977 U02409 U06524 U07078 U05734 U08114 U50976 U02408 U06069 U07077 U05733 U08113 U50975 U02407 U06068 U07076 U05732 U03191 U50974 U02406 U06067 U07075 U05731 U03190 U50973 U02405 U06066 U07074 U05730 U03189 U50972 U02404 U06065 U07073 U05729 U03188 U50971 U02403 U06064 U07072 U05728 U03187 U50970 U02402 U06063 U07071 U05727 U03186 U50969 U02401 U06062 U07070 U05726 U03185 U50968 U02400 U06061 U07069 U05725 U03184 U50967 U02399 U06060 U07068 U05724 U03183 U50966 U08640 U06059 U07067 U05723 U03182 U50965 U08100 U06058 U07066 U05722 U03181 U50964 U07440 U06057 U07065 U05721 U03180 U50963 U51916 U06056 U07064 U05720 U03179 U50962 U51915 U06055 U07063 U05719 U03178 U50961 U51914 U06050 U07062 U05718 U03177 U50960 U51913 U06049 U07061 U05717 U03176 U50959 U51912 U06048 U07060 U05716 U03175 U07614 U51911 U06047 U07059 U05715 U03174 U07613 U51910 U06046 U07058 U05714 U03164 U07612 U51909 U06045 U07057 U05713 U03163 U07611 U51908 U06044 U07056 U05712 U03162 U07610 U51907 U06043 U06491 U05711 U03161 U07609 U51906 U06042 U06490 U05710 U03160 U02446 U51905 U06041 U06489 U05709 U03159 U02445 U51904 U01718 U08956 U03590 U03158 U02444 U51903 U01717 U08955 U03589 U03157 U02443 U51902 U01716 U08954 U03588 U03156 U02442 U51901 U01715 U08953 U03587 U03155 U02441 U51900 U01714 U08952 U03586 U03154 U51654 U51899 U01713 U08951 U03585 U03153 U04138 U51898 U01709 U08950 U03584 U03152 U04137 U51797 U01708 U08949 U03583 U03151 U04136 U02617 U01707 U08948 U03582 U03150 U04135 U02616 U01706 U08947 U03581 U03149 U04134 U01755 U01705 U08946 U03580 U04338 U04133 U01510 U01704 U08945 U03579 U04337 U04132 U01383 U01703 U08944 U50610 U04336 U04131 U01381 U01702 U08943 U50609 U03032 U04130 U01379 U01701 U08942 U50584 U01825 U04129 U50218 U01700 U08941 U02814 U01822 U04128 U50217 U03613 U08940 U01362 U50951 U04127 U50216 U03612 U08939 U01636 U50713 U04126 U50215 U03611 U04653 U01635 U50668 U03578 U50214 U03610 U01201 U01633 U50622 U03430 U50213 U03609 U51446 U01632 U50620 U03354 U04489 U03608 U02604 U01631 U50619 U03197 U04052 U03607 U01588 U01630 U50618 U02928 U01928 U03606 U51670 U01629 U50617 U08433 U01927 U03605 U07945 U01628 U50613 U08432 U01156 U03604 U04953 U01627 U50503 U08431 U01155 U03603 U04952 U01626 U50461 U08430 U01154 U03602 U04951 U01624 U50089 U08429 U01153 U03601 U04950 U01623 U08997 U08428 U01152 U03600 U07885 U01622 U08994 U08427 U01151 U03599 U07884 U01634 U08124 U08426 U01936 U03598 U07883 U01625 U06455 U07902 U01935 U03597 U07882 U50231 U04775 U07901 U01934 U03596 U07881 U50230 U04768 U02748 U01933 U03595 U07880 U50229 U04663 U02747 U01932 U03594 U07879 U50228 U04649 U02746 U01931 U03593 U07878 U50227 U04328 U02745 U01930 U03592 U07877 U50226 U03662 U02744 U01929 U03591 U07876 U50225 U03661 U02743 U05973 U06704 U07875 U50224 U07418 U02742 U05972 U06138 U07874 U50223 U02465 U06987 U05971 U06137 U50826 U50222 U02464 U06986 U05970 U07548 U50825 U50221 U02463 U06985 U01145 U07547 U50824 U50220 U02462 U06984 U01144 U07546 U50823 U50219 U02461 U06983 U01143 U07545 U50822 U05828 U04431 U06982 U01142 U07544 U03001 U05827 U04430 U06981 U01141 U07543 U03000 U05825 U08827 U06980 U01140 U07542 U02999 U05824 U08826 U06979 U01139 U07541 U02998 U05647 U08825 U06978 U03299 U07540 U02997 U05521 U08824 U06977 U08401 U07539 U02996 U05452 U08823 U06976 U08400 U07538 U02995 U04770 U08822 U06975 U08399 U07537 U02994 U08846 U08821 U06974 U08398 U07536 U02993 U08845 U08820 U06973 U08397 U07535 U02252 U06783 U08819 U06972 U08396 U50491 U51149 U51564 U08818 U06971 U08395 U50490 U51148 U51563 U03067 U06970 U08390 U50489 U51147 U51562 U03066 U06969 U01504 U50488 U51146 U51561 U03065 U06968 U50376 U50487 U51145 U51560 U03064 U06009 U50375 U50486 U51144 U51559 U03063 U06008 U50374 U50485 U51143 U51558 U02414 U06007 U50373 U50484 U51142 U51557 U02413 U06006 U50372 U50483 U51141 U51556 U02412 U06005 U50371 U50482 U51140 U51555 U02383 U06004 U50370 U50481 U51139 U51554 U02382 U06003 U50369 U04683 U51138 U51553 U02381 U06002 U50368 U51274 U51137 U51304 U02380 U06001 U50367 U51273 U51136 U51303 U02379 U06000 U50366 U51223 U51135 U51302 U02378 U05999 U50365 U51222 U51134 U51301 U02377 U05998 U50364 U51221 U51133 U51300 U02376 U05997 U50363 U51220 U51132 U51299 U02375 U05996 U50362 U51219 U51131 U51298 U02374 U05995 U50361 U51218 U51130 U51297 U02373 U05994 U50360 U51217 U51129 U51296 U02372 U05993 U50359 U51216 U51128 U51295 U02371 U05992 U50358 U51215 U51127 U51294 U02370 U05991 U50357 U51214 U51126 U08892 U02369 U05990 U50356 U51213 U51125 U06822 U02368 U05989 U50355 U51212 U51124 U04140 U02367 U05988 U50354 U51211 U51123 U04139 U02366 U05987 U50353 U51210 U51122 U08891 U50953 U05986 U50352 U51209 U51121 U06821 U05097 U05985 U50351 U51201 U51120 U51101 U05096 U05984 U50350 U51200 U51119 U51100 U05095 U05983 U50349 U51199 U51118 U05012 U05094 U05982 U50348 U51198 U04486 U05011 U05093 U50206 U01138 U51197 U03260 U01867 U05092 U50205 U01137 U51196 U08415 U01866 U05091 U50204 U01136 U51195 U08394 U01865 U05090 U50203 U01135 U51194 U08393 U01864 U05089 U06828 U01134 U51193 U08392 U06162 U05088 U06827 U01133 U50893 U08391 U06161 U05087 U06826 U01132 U50892 U06776 U06160 U05086 U06825 U01131 U50891 U06775 U06159 U02254 U06824 U01081 U50890 U06774 U06158 U02253 U06823 U01080 U50889 U06773 U06157 U08008 U06722 U01079 U50888 U06772 U06156 U08007 U06721 U01078 U50887 U06771 U06155 U08006 U06720 U01077 U50460 U06770 U02793 U08005 U06719 U01076 U50907 U06769 U02792 U08004 U06718 U01075 U50906 U50251 U02791 U08003 U06717 U01074 U50905 U07322 U02790 U08002 U04120 U01073 U50904 U07321 U02789 U08001 U04119 U01072 U50903 U07320 U02788 U08000 U04118 U01071 U50902 U07319 U02787 U07999 U04117 U01070 U50901 U07318 U02786 U07998 U04116 U50212 U50900 U07317 U02785 U07997 U04115 U51184 U50899 U07316 U02784 U07996 U04114 U51183 U50898 U07315 U02783 U07995 U04113 U51182 U50897 U07314 U02782 U04423 U04112 U51181 U50896 U07313 U02781 U04125 U04111 U51180 U50895 U07312 U02780 U04122 U04110 U51179 U50894 U07311 U02779 U04069 U04109 U51178 U50608 U07310 U02778 U04068 U04108 U51177 U50607 U07309 U02777 U03484 U04107 U51176 U50606 U07308 U02776 U03482 U04106 U51175 U50605 U07307 U07898 U03195 U04105 U51174 U50604 U07306 U07897 U03144 U04104 U51173 U50603 U07305 U07873 U03071 U04103 U04943 U50602 U07304 U07872 U02927 U04102 U04942 U50601 U07268 U07871 U04205 U04101 U04941 U50600 U07267 U07870 U04204 U04100 U04940 U50599 U07266 U07869 U04203 U04099 U01292 U50598 U07265 U07868 U04202 U04098 U01291 U50597 U07264 U07867 U04201 U04097 U01290 U50596 U07263 U07866 U04200 U04096 U01289 U04784 U07262 U07865 U04199 U04095 U01288 U04783 U07261 U07864 U04198 U02430 U01287 U04782 U07260 U07816 U04197 U02429 U01286 U04781 U07259 U07631 U04196 U02428 U01285 U04780 U07258 U07630 U04195 U51321 U01284 U04779 U07257 U07629 U04194 U51320 U01283 U04778 U07256 U07628 U04193 U50518 U01282 U02872 U07255 U07627 U04192 U50517 U01281 U02871 U07254 U07626 U04191 U50516 U01280 U02870 U07253 U07625 U04190 U50515 U01279 U02869 U07252 U07624 U04189 U50514 U01278 U02868 U07251 U04421 U04188 U50513 U01277 U02867 U07250 U04420 U02610 U50512 U05448 U02866 U07249 U04419 U02609 U50511 U04989 U02865 U07247 U04418 U02608 U06485 U04931 U02864 U07246 U04417 U02607 U06484 U04828 U02863 U07245 U04416 U51341 U06483 U04774 U02271 U07244 U04415 U51340 U06482 U04664 U02270 U07243 U04414 U51339 U06481 U04482 U02269 U07242 U03257 U51338 U06480 U08890 U02268 U07241 U03256 U51337 U06479 U08889 U01899 U51952 U03255 U50621 U06478 U08882 U01898 U51951 U03254 U03888 U06477 U08881 U01897 U51950 U07921 U03887 U06476 U08880 U01896 U51949 U02595 U03886 U06475 U08879 U01895 U51948 U01378 U03885 U06474 U08878 U01894 U51947 U01359 U03884 U06473 U08877 U01893 U51946 U51711 U03883 U06472 U08876 U01892 U51651 U51710 U07967 U04157 U08875 U01891 U08355 U51709 U07966 U04156 U08874 U01890 U08354 U51708 U07965 U04155 U08873 U01889 U08353 U51707 U07964 U04154 U08872 U01888 U08352 U04302 U07963 U04153 U08871 U01887 U08351 U04301 U07962 U04152 U08870 U01886 U08350 U04300 U07961 U04151 U08869 U01885 U08349 U04299 U07960 U04150 U08868 U01884 U08348 U04298 U07959 U04149 U08867 U01883 U08347 U04297 U07958 U04148 U50398 U01882 U08346 U04296 U04424 U04147 U50397 U01881 U08345 U04295 U03615 U04146 U50396 U01880 U08344 U04294 U03614 U04145 U50395 U01879 U08343 U04293 U02818 U04144 U50394 U01878 U08342 U04292 U02734 U04143 U50393 U01877 U08341 U04291 U02037 U04142 U05608 U01876 U08340 U04290 U01474 U01309 U04777 U01875 U08339 U04289 U01473 U01308 U02802 U01874 U08338 U04288 U01472 U01307 U02801 U01873 U08337 U04287 U01471 U01306 U02800 U01872 U08336 U04286 U01470 U01305 U02799 U01871 U08335 U04227 U01469 U01304 U02798 U01870 U08334 U04226 U01468 U01303 U02797 U01869 U08333 U04090 U01467 U01302 U02796 U01868 U08332 U04089 U01466 U01301 U02795 U05031 U08331 U04088 U01465 U01300 U01508 U05030 U08330 U04087 U01464 U01299 U01387 U05029 U08326 U04086 U01463 U01298 U01386 U05028 U08325 U04085 U01462 U01297 U01385 U05027 U08324 U04084 U01461 U01296 U01452 U05026 U08323 U04083 U01460 U01295 U01451 U05025 U08322 U04082 U01459 U01294 U01450 U05024 U08321 U04081 U06118 U51704 U01449 U05023 U08320 U04080 U06117 U51703 U01448 U05022 U08319 U04079 U06116 U51702 U01447 U02301 U08318 U04078 U06115 U51701 U01446 U02300 U08317 U04077 U06114 U51700 U01445 U02299 U08316 U04076 U06113 U51699 U01444 U02298 U08315 U04075 U06112 U51698 U01443 U02297 U01507 U04074 U06111 U51697 U01442 U02296 U01365 U04073 U06110 U51696 U01441 U02295 U04556 U02187 U06109 U51695 U01440 U02294 U04555 U02186 U06108 U51694 U01439 U02293 U04554 U02185 U06107 U51693 U01438 U02292 U04553 U02184 U06106 U51692 U01437 U02291 U04552 U02183 U06105 U51685 U51024 U02290 U04551 U02182 U06104 U51684 U51023 U02289 U04549 U02181 U06103 U08860 U51022 U02288 U04548 U02180 U06102 U08859 U51021 U02287 U07431 U01807 U06101 U08858 U51020 U02286 U07215 U01806 U06100 U08857 U51019 U02285 U06449 U01805 U06099 U08856 U51018 U02284 U06163 U01804 U06098 U08855 U51017 U02283 U51653 U01803 U06097 U08854 U51016 U02282 U51652 U01802 U06096 U08853 U51015 U02281 U08554 U01801 U06095 U08852 U51014 U02280 U08551 U01800 U06094 U08851 U51013 U02279 U08550 U01799 U06093 U08850 U51012 U02278 U08549 U01798 U06092 U06440 U51011 U02277 U08548 U01797 U06091 U05251 U51010 U02276 U08547 U01796 U06090 U05250 U51009 U02275 U08546 U01795 U06089 U05249 U51008 U02274 U08545 U01794 U06088 U05248 U51007 U02273 U08544 U01793 U06087 U05247 U51006 U02272 U08543 U01792 U06086 U05246 U51005 U51167 U08542 U01791 U06085 U05245 U51004 U51166 U08541 U01677 U06084 U05244 U51003 U51165 U08540 U01676 U01057 U05243 U51002 U03752 U06493 U01675 U01056 U05242 U51001 U03751 U06492 U01674 U01055 U05241 U51000 U03750 U02440 U01673 U01054 U05240 U50999 U03749 U02439 U01672 U01053 U05239 U50998 U03748 U02438 U01671 U01052 U05238 U50997 U03747 U02437 U01670 U01051 U05237 U50996 U02266 U02436 U01669 U01050 U05236 U50995 U02265 U02435 U01668 U01049 U05235 U50994 U02264 U02434 U01667 U01048 U05234 U50993 U02263 U02433 U01666 U01047 U05233 U50992 U02262 U02251 U01665 U01046 U05232 U50991 U02261 U02250 U01664 U01045 U03248 U50990 U02260 U02249 U01663 U01044 U03247 U50989 U02259 U02248 U01662 U01043 U03246 U50988 U01103 U02247 U01661 U01042 U03245 U50987 U01102 U02246 U01660 U01041 U03244 U50986 U01101 U02245 U01659 U04506 U03243 U50985 U01100 U02244 U01658 U04505 U03242 U50935 U01099 U02243 U01618 U04504 U03241 U06768 U01098 U02242 U01617 U04503 U03223 U06767 U01097 U02241 U01616 U04502 U03222 U06766 U01096 U02240 U01615 U04501 U03221 U06765 U01095 U02239 U01614 U04500 U03220 U06764 U01094 U02238 U01613 U04499 U03219 U06763 U01093 U02237 U01612 U04498 U03218 U06762 U01092 U02236 U01611 U04497 U03217 U06761 U01091 U01085 U01610 U04496 U03216 U06760 U01090 U01084 U01609 U04495 U03215 U06759 U01089 U01083 U01608 U04494 U03214 U06758 U01088 U01082 U01607 U04493 U03213 U06757 U01087 U51099 U01606 U04492 U03212 U06756 U01086 U51098 U01605 U04491 U03211 U06755 U04021 U51097 U01604 U04490 U03210 U06754 U04020 U51096 U01603 U03068 U03209 U06753 U04019 U51095 U01602 U02985 U03208 U06752 U04018 U51094 U01601 U04696 U03207 U06751 U04017 U51093 U01600 U04695 U03206 U06750 U04016 U51092 U01599 U04694 U03205 U06749 U04015 U51091 U01598 U04693 U03204 U06748 U04014 U51090 U01597 U04692 U03203 U06747 U04013 U02639 U01596 U04691 U03202 U05981 U04012 U02638 U01595 U04690 U03201 U05980 U07163 U02637 U01594 U04689 U03200 U05979 U07162 U02636 U01593 U04688 U03140 U05978 U07161 U51750 U01592 U04687 U51089 U05977 U07160 U50500 U01591 U04686 U51088 U05976 U07159 U50070 U01590 U50508 U51087 U51158 U07158 U50069 U01589 U50454 U51086 U51157 U07157 U50068 U01574 U51164 U51085 U51156 U07156 U50067 U01573 U51163 U51084 U51155 U07155 U50066 U01572 U51162 U51083 U07822 U07154 U50065 U01571 U51161 U51082 U07821 U07153 U50064 U01570 U04665 U51081 U07815 U07152 U50063 U01569 U04439 U51080 U07814 U06224 U50062 U01568 U04435 U51079 U07813 U06223 U50061 U01567 U04027 U51078 U07812 U06222 U50060 U01566 U03927 U51077 U07811 U06221 U50059 U01565 U04736 U51076 U07810 U06220 U50058 U01564 U04735 U51075 U07809 U06219 U50057 U01563 U04734 U51074 U07808 U06218 U50056 U01562 U04733 U51073 U07807 U06217 U50055 U01561 U04732 U51072 U07806 U06216 U50054 U01560 U04731 U51071 U07785 U06215 U50053 U01559 U04730 U51070 U07784 U06214 U50052 U01558 U04729 U51069 U07783 U06213 U50051 U01557 U04718 U51068 U07782 U01481 U50047 U01556 U04717 U51067 U07781 U01480 U50046 U01555 U04716 U51066 U07780 U01479 U50045 U01554 U04715 U51065 U07779 U01478 U50044 U01553 U08032 U51064 U07778 U01477 U50043 U01552 U08031 U51063 U07777 U01476 U50042 U01551 U08030 U51062 U07776 U01475 U50041 U01550 U08029 U51061 U07685 U05035 U50040 U01549 U08028 U51060 U07684 U05034 U50039 U01548 U08027 U51059 U07683 U05033 U50038 U01547 U04929 U51058 U07682 U05032 U05759 U01546 U07231 U51057 U07681 U51444 U05520 U01545 U07230 U51056 U07680 U51706 U01746 U01544 U07229 U50876 U07679 U51705 U07439 U01543 U07228 U50875 U07678 U04526 U07142 U01542 U07227 U50874 U07677 U04525 U06781 U01541 U07226 U50873 U07676 U04524 U08463 U01540 U07225 U08636 U07657 U04523 U08462 U01539 U07224 U08635 U07656 U04522 U08461 U01538 U07223 U08634 U07655 U04521 U08460 U01537 U07222 U08633 U07654 U04520 U08459 U01536 U07221 U08632 U07653 U04051 U08458 U01535 U07220 U08631 U07652 U04050 U08457 U01496 U07219 U08630 U07651 U04049 U08456 U01495 U07218 U08629 U04238 U04048 U08455 U01494 U07217 U08628 U04237 U04047 U08454 U01493 U07216 U08627 U04236 U04046 U08453 U01492 U51792 U08626 U04235 U04045 U08452 U01491 U51791 U08625 U04234 U04044 U04875 U01490 U51790 U08624 U05646 U04043 U04874 U01489 U51788 U08623 U05645 U04042 U04873 U01488 U51232 U08622 U05644 U04041 U04872 U01487 U51231 U08621 U05643 U04040 U04871 U01486 U07513 U08620 U05642 U04039 U04870 U01485 U05274 U08619 U05641 U04038 U04869 U01484 U05273 U06871 U05640 U04037 U04868 U01483 U05272 U06870 U05639 U04036 U04867 U01482 U05271 U06869 U05638 U04035 U04866 U02817 U05270 U06868 U05637 U04034 U04865 U50709 U04846 U06867 U05636 U04033 U04855 U51360 U02574 U06866 U05635 U04032 U04854 U51359 U02573 U06865 U08122 U04031 U04853 U51358 U02572 U06864 U07623 U04030 U04852 U51357 U02571 U06863 U07621 U04029 U04851 U51356 U02570 U06862 U01185 U51604 U04850 U51355 U02569 U06861 U01184 U51603 U04849 U51354 U02568 U06860 U01183 U51602 U04848 U51353 U02567 U06859 U50633 U51601 U04847 U51352 U02566 U06858 U50632 U51600 U08026 U51351 U02565 U06857 U50631 U51599 U08025 U51350 U02564 U06856 U04064 U51598 U08024 U51349 U02563 U06855 U04063 U51597 U08023 U51348 U02562 U06854 U04062 U51789 U08022 U51347 U02561 U06853 U04061 U51787 U08021 U51346 U02560 U06852 U04060 U51786 U08020 U51345 U02559 U06851 U04059 U51785 U08019 U51344 U02558 U06850 U04058 U51784 U08018 U51343 U02557 U06849 U04057 U51783 U08017 U50666 U02556 U06803 U04056 U51782 U08016 U50611 U02555 U06802 U04055 U51781 U08015 U50498 U02554 U06801 U04054 U51780 U08014 U50248 U02553 U07710 U04053 U51779 U08013 U50232 U02552 U07709 U06956 U51778 U08012 U08101 U02551 U07708 U02365 U51777 U08011 U08099 U50240 U07707 U02364 U51776 U08010 U07615 U50239 U07706 U02363 U51775 U08009 U07429 U50238 U05833 U02362 U51774 U50880 U07177 U50237 U05832 U02361 U51773 U50711 U06520 U04618 U05831 U02360 U51772 U05073 U05651 U03353 U05830 U02359 U51771 U05072 U05252 U03196 U05829 U02358 U51770 U05071 U04880 U07128 U04860 U02357 U51769 U05070 U03253 U07127 U04859 U02356 U51768 U04772 U02508 U07126 U04858 U02355 U51767 U08555 U02507 U07125 U04857 U02354 U51766 U51818 U02506 U07124 U04856 U02353 U51765 U51817 U02505 U07123 U04028 U02352 U51764 U07351 U02504 U07122 U02432 U02351 U51763 U07350 U02503 U07121 U02431 U02350 U51762 U07349 U02502 U07120 U50642 U02349 U51761 U07348 U02501 U07119 U50641 U02348 U03481 U07347 U02500 U04826 U50640 U02347 U03480 U07346 U02499 U04825 U50639 U02346 U03479 U07345 U02498 U04824 U50638 U02345 U03478 U07344 U02497 U05969 U50637 U02344 U03477 U07332 U02496 U05968 U50636 U02343 U03476 U07331 U02495 U05967 U50635 U02342 U03475 U07330 U02494 U05966 U50634 U02311 U03474 U07329 U02493 U04902 U50037 U02310 U03473 U07328 U02492 U04901 U50036 U02309 U03472 U07327 U02491 U04900 U50035 U02308 U03471 U07326 U02490 U04899 U50034 U02307 U03470 U07325 U02489 U04898 U50033 U02306 U03469 U07324 U02488 U04897 U50032 U02305 U03468 U07323 U02487 U04896 U50031 U02304 U03467 U07289 U02486 U04895 U50030 U02303 U03466 U07288 U02485 U02990 U50029 U02302 U03465 U07287 U02484 U02633 U08154 U06815 U03464 U07286 U02483 U02632 U08153 U06814 U03463 U07285 U02482 U02631 U08149 U06813 U03462 U07284 U02481 U02630 U08148 U06812 U03461 U02984 U02480 U02629 U08147 U06811 U03460 U02983 U02479 U02628 U08146 U06810 U03459 U02982 U02478 U02627 U08144 U06809 U03458 U02981 U02477 U02626 U08143 U06808 U03457 U02980 U02476 U02625 U08141 U06807 U03456 U02979 U02475 U02624 U08140 U06806 U03455 U03754 U02474 U02623 U51923 U06805 U03454 U03753 U02473 U02622 U51922 U06804 U03453 U02599 U02472 U02621 U51921 U51333 U03452 U51459 U02471 U02620 U51920 U51026 U03451 U51458 U02470 U02619 U51919 U07956 U03450 U51457 U02469 U02618 U51918 U07954 U03449 U51456 U02468 U08558 U51917 U07953 U03448 U51455 U02467 U06334 U50941 U07952 U03447 U51454 U02466 U06130 U02767 U07951 U03446 U51453 U05602 U05802 U02766 U07950 U03445 U51452 U05601 U05801 U02765 U05432 U03444 U51451 U05600 U05800 U02764 U05431 U03443 U51450 U05599 U05799 U02763 U05430 U03442 U51449 U05598 U05798 U02762 U05429 U03441 U08618 U05597 U05797 U02761 U05428 U03440 U08617 U05596 U05796 U02760 U05427 U03439 U08616 U05595 U05795 U02759 U05426 U03438 U06413 U05594 U05794 U02758 U05425 U03437 U06412 U05593 U05793 U50938 U02034 U03436 U06411 U05592 U05792 U06951 U02029 U03435 U06410 U05591 U05791 U06950 U01366 U03434 U06409 U04816 U05790 U06949 U05060 U03433 U06408 U04815 U05789 U02204 U05059 U03432 U51596 U04814 U05788 U02203 U05058 U03423 U51595 U04813 U05787 U02202 U05057 U03422 U51594 U04812 U50138 U02201 U05056 U03421 U51593 U04811 U07415 U02200 U05055 U03420 U51592 U04810 U04485 U02199 U05054 U03419 U51591 U03854 U51112 U02198 U05053 U03418 U51590 U03431 U51111 U02197 U05052 U03417 U51589 U03199 U51110 U02196 U05051 U03416 U51588 U04809 U51109 U02195 U05050 U03415 U51587 U04808 U51108 U02194 U05049 U03414 U04681 U04807 U51107 U02193 U05048 U03413 U04072 U04806 U51106 U02192 U05047 U03412 U51586 U04805 U51105 U02169 U05046 U03411 U51585 U04804 U51104 U02168 U05045 U03410 U51584 U04803 U51103 U02167 U05044 U03409 U51583 U04802 U09004 U02166 U03198 U03408 U51582 U50137 U07860 U02165 U07693 U03407 U51581 U05843 U07859 U02164 U06874 U03406 U51580 U05842 U07858 U02163 U05519 U03405 U51579 U05841 U07857 U02162 U03031 U03404 U51578 U05840 U07856 U02161 U03030 U03403 U51577 U05839 U07855 U02160 U03029 U03402 U51576 U05838 U07854 U02159 U03028 U03401 U51575 U05837 U07853 U02158 U03027 U03400 U05570 U05836 U07852 U02157 U03026 U03399 U05568 U05835 U07851 U02156 U03025 U03398 U02550 U05834 U07850 U02155 U03024 U03397 U02549 U01726 U07849 U02154 U03023 U03396 U02548 U01725 U07797 U02153 U03022 U03395 U02547 U01724 U07796 U02152 U03021 U03394 U02546 U01723 U07794 U02151 U03020 U03393 U02545 U01722 U07793 U02150 U03019 U03392 U02544 U01721 U07792 U02149 U02830 U03391 U02543 U01790 U07791 U02148 U02829 U03390 U02542 U01789 U07790 U02147 U02828 U03389 U02541 U01788 U07789 U02146 U02827 U03388 U02540 U01787 U07788 U02145 U02826 U03387 U02539 U01786 U07787 U02144 U02825 U03386 U02538 U01785 U07786 U02143 U02824 U03385 U02537 U01784 U07750 U02142 U02823 U03384 U02536 U01783 U07749 U02141 U02822 U03383 U02535 U01782 U07748 U02140 U02821 U03382 U02534 U01781 U07747 U02139 U02820 U03381 U02533 U01780 U07746 U02138 U05496 U03380 U02532 U01779 U07745 U02137 U05495 U03379 U02531 U01778 U07744 U02136 U05494 U03378 U02530 U01777 U07743 U02135 U02427 U03377 U02529 U51332 U07742 U02134 U02426 U03376 U02528 U50247 U07741 U02133 U02191 U03375 U02527 U50246 U07740 U02132 U02190 U03374 U02526 U06396 U07739 U02131 U51037 U03259 U02525 U06395 U07620 U02130 U01678 U03258 U02524 U06394 U07619 U01989 U50280 U04785 U02523 U06393 U07618 U01988 U50279 U04426 U02522 U06392 U07617 U01987 U50278 U03262 U02521 U06391 U07616 U01986 U50277 U50948 U02520 U06390 U07518 U01985 U50276 U08516 U02519 U06389 U07517 U01984 U50275 U05762 U02518 U06388 U07516 U01983 U50274 U05604 U02517 U06387 U07515 U01982 U50273 U04830 U02516 U06386 U07514 U01981 U50272 U50347 U02515 U06385 U07055 U01980 U50271 U06967 U02514 U06384 U07054 U01979 U50263 U03373 U02513 U06383 U05013 U01978 U50262 U03372 U02512 U06382 U04546 U01977 U50261 U03371 U02511 U06381 U04545 U01976 U50260 U03370 U02510 U06380 U04544 U01975 U01524 U03369 U02509 U06379 U04543 U01974 U01523 U03368 U06494 U04644 U04542 U01973 U01522 U51641 U03093 U04325 U04541 U01972 U01521 U51640 U51658 U04324 U04540 U01971 U01520 U51639 U50949 U04323 U04539 U01970 U01519 U51638 U50028 U04322 U04538 U01969 U01518 U51637 U50027 U04321 U04537 U01968 U01517 U51636 U50026 U04320 U04536 U01967 U01516 U51635 U50025 U04319 U04535 U01966 U01515 U51634 U50024 U04318 U04534 U01965 U07728 U51633 U50023 U04317 U04533 U01964 U07727 U51632 U50022 U04316 U04532 U01963 U07726 U51631 U50021 U04315 U04531 U01962 U07725 U51630 U50020 U04314 U04530 U01961 U07724 U51629 U50019 U04313 U04529 U01960 U07723 U51628 U50018 U04312 U04528 U01959 U07722 U51627 U50017 U04311 U04527 U01958 U07721 U51626 U50016 U04310 U02653 U01957 U07667 U51625 U50015 U04309 U02652 U01956 U07666 U51624 U50014 U04487 U02651 U01955 U07665 U51623 U50013 U02592 U02650 U01950 U07664 U51622 U51325 U02591 U02649 U01949 U07663 U51621 U01158 U02590 U02648 U01948 U07662 U51620 U01157 U02589 U02647 U01947 U07661 U51619 U04728 U02588 U02646 U01946 U07660 U51618 U04727 U02587 U02645 U01945 U07659 U51617 U04726 U02586 U02644 U01944 U07658 U51616 U04725 U02585 U02643 U01943 U07602 U51615 U04724 U02584 U02642 U01942 U07601 U51614 U04723 U02583 U05041 U01941 U07600 U51613 U06212 U02582 U05040 U51055 U07599 U51612 U06211 U02581 U05039 U51054 U07598 U51611 U06210 U02580 U05038 U51053 U07597 U51610 U06209 U02579 U05037 U51052 U07596 U51609 U06208 U02578 U05036 U51051 U07595 U51608 U06207 U02577 U04643 U51050 U07594 U51607 U06206 U02576 U04642 U51049 U07593 U51606 U04279 U02575 U04641 U51048 U07591 U51605 U04278 U50681 U04640 U51047 U07590 U06467 U04277 U50680 U04639 U05565 U07589 U06466 U04276 U50679 U04638 U05564 U07588 U06465 U04275 U50678 U04637 U05563 U07587 U06464 U04274 U50677 U04636 U05562 U07586 U06463 U04273 U50676 U04635 U50936 U07585 U06462 U04272 U50675 U04634 U07000 U07584 U06461 U03228 U50674 U04633 U06999 U03147 U06460 U03227 U50673 U04632 U06998 U02961 U06459 U03226 U50672 U04631 U06997 U02960 U06458 U03225 U50671 U04630 U06996 U02959 U06457 U03224 U50670 U04629 U06995 U02958 U06456 U04305 U03141 U04628 U06994 U02957 U51314 U04306 U03069 U04627 U04444 U02956 U51313 U04307 U02989 U04589 U04443 U02955 U51312 U04308 U02809 U04588 U04442 U02954 U51311 U04304 U02751 U04587 U08962 U02953 U51310 U04303 U02750 U04586 U08961 U02952 U51309 U50706 U02749 U04585 U08960 U02951 U50211 U50705 U01829 U04584 U08959 U02950 U50210 U50704 U01824 U04583 U08958 U02949 U50209 U50703 U01363 U04582 U08957 U02948 U50208 U07986 U04517 U04581 U07039 U02947 U50207 U07985 U04516 U04580 U06193 U02946 U06926 U07984 U51335 U04579 U06192 U02945 U03740 U07983 U01195 U04578 U06191 U02944 U03739 U06339 U01194 U04576 U06190 U02943 U03738 U06338 U01193 U04575 U06189 U02942 U03737 U06337 U01192 U04574 U06188 U01754 U03736 U06336 U01191 U04573 U06187 U01753 U03735 U03018 U51801 U04572 U06186 U01752 U03734 U03017 U04488 U04571 U51152 U01751 U03733 U03016 U04481 U04570 U51151 U01750 U03571 U03015 U04438 U04569 U51150 U01749 U03570 U03014 U04330 U04568 U06943 U01748 U03569 U03013 U03194 U04567 U06942 U01747 U03568 U03012 U03002 U04566 U06941 U08684 U03567 U03011 U02028 U04565 U04679 U07035 U03566 U05965 U08897 U04564 U04678 U07034 U03565 U05964 U03925 U04563 U04677 U07033 U03564 U05963 U03924 U04562 U04676 U07032 U03563 U05962 U03923 U04561 U04675 U07031 U03562 U05961 U03922 U04560 U04674 U07030 U03561 U05960 U03921 U04559 U04673 U07029 U03560 U05959 U03920 U04558 U04672 U07028 U03559 U05958 U03919 U04557 U04671 U07027 U03558 U05957 U03918 U51793 U04670 U07026 U03557 U05956 U03917 U07417 U04669 U07013 U03556 U05955 U03916 U07175 U04668 U07012 U03555 U05954 U03915 U06954 U51029 U07011 U03554 U05953 U03914 U06782 U07053 U07010 U03553 U05952 U03913 U05779 U07052 U07009 U03552 U05951 U03912 U05778 U07051 U07008 U02255 U05950 U03911 U05777 U07050 U07007 U01742 U05949 U03910 U05776 U07049 U07006 U01741 U05948 U03909 U05775 U07048 U07005 U06471 U05947 U03908 U05774 U07047 U07004 U06470 U05946 U03907 U05773 U07046 U06966 U06469 U50697 U03906 U05772 U07045 U06965 U03719 U50696 U03905 U05771 U07044 U06964 U03575 U50695 U03904 U05770 U07043 U06963 U03261 U50694 U03903 U05769 U07042 U06962 U02925 U50693 U03902 U05768 U02926 U06961 U02819 U50692 U03901 U05767 U04881 U06746 U02593 U50691 U03900 U05766 U04878 U06745 U08123 U50690 U03899 U05765 U04876 U06744 U06927 U50689 U03898 U05764 U04771 U06743 U06329 U50688 U03897 U01367 U04518 U06742 U06328 U50687 U03896 U51646 U04484 U06741 U06327 U50686 U03895 U51324 U04446 U06740 U06326 U50685 U03894 U50522 U04329 U06739 U06325 U50684 U03893 U50521 U04121 U06738 U06324 U50683 U03864 U50520 U04011 U06737 U06323 U50682 U03863 U50519 U03143 U06736 U06322 U05906 U03862 U07942 U03043 U06735 U06321 U05905 U03861 U07941 U02924 U06734 U06320 U05904 U03860 U07940 U02889 U06655 U06317 U04161 U03859 U07939 U02722 U06654 U06316 U04160 U03858 U07938 U02602 U06653 U06315 U04159 U03857 U07937 U02597 U06652 U06314 U04158 U03856 U07936 U02042 U06651 U06313 U03301 U03855 U07935 U02039 U06650 U06312 U51168 U03852 U07934 U02035 U06649 U06311 U50245 U03851 U07933 U01828 U06648 U06310 U50244 U03850 U07896 U01756 U06637 U06309 U50243 U03849 U07895 U03131 U06636 U06308 U50242 U03848 U07894 U03130 U06635 U06307 U50141 U03847 U07893 U03129 U06634 U06306 U04384 U03846 U07892 U03128 U06633 U06305 U04383 U03845 U07891 U03127 U06632 U06304 U04382 U03844 U07890 U03126 U06631 U06303 U04381 U03843 U07889 U03125 U06630 U06302 U04380 U03842 U07888 U03124 U06629 U06301 U04379 U03841 U07887 U03123 U06628 U06300 U04378 U03840 U07775 U03122 U06627 U06299 U04377 U03839 U07760 U03121 U06626 U06298 U04376 U03838 U07759 U03120 U06625 U06297 U04375 U03837 U07758 U51230 U06624 U06296 U04374 U03836 U07757 U51229 U06623 U06294 U04373 U03835 U07756 U51228 U06622 U06293 U51046 U03834 U07755 U51227 U06621 U06292 U51045 U03833 U07754 U51226 U06620 U06291 U04341 U03786 U07753 U51225 U06619 U06290 U04340 U03785 U07752 U51224 U06618 U06289 U04339 U03784 U07751 U03686 U06617 U06288 U06295 U03783 U07738 U03685 U06616 U06287 U01357 U03782 U07737 U03684 U06615 U06286 U01356 U03781 U07736 U03683 U06614 U06285 U01355 U03780 U07735 U03682 U06613 U06284 U01354 U03779 U07734 U03681 U06612 U06283 U01353 U03778 U07733 U03680 U03294 U06282 U01352 U03777 U07732 U03679 U03293 U06281 U01351 U03776 U07731 U03678 U03292 U06280 U01350 U03775 U07730 U03677 U03291 U06279 U01349 U03774 U07729 U03676 U08894 U06278 U01348 U03773 U07720 U04845 U51293 U06277 U01347 U03772 U07719 U01586 U51292 U06276 U01346 U03771 U07718 U01585 U51291 U06275 U01345 U03770 U07717 U07210 U51290 U06274 U01344 U03769 U07716 U07209 U51289 U06273 U01343 U03768 U07715 U07208 U51288 U06272 U01342 U03767 U07714 U07207 U51287 U06271 U51798 U03766 U07713 U07206 U51286 U06270 U07980 U03765 U07712 U05010 U51285 U06269 U07979 U03764 U07711 U05009 U51284 U06268 U07978 U03763 U07705 U05008 U51283 U06267 U07457 U03762 U07704 U05007 U51282 U06266 U07456 U03761 U07703 U05006 U51281 U06265 U07455 U03760 U07702 U05005 U51280 U06264 U07454 U03759 U07701 U05004 U51279 U03173 U07453 U03758 U07700 U05003 U51278 U03172 U07452 U03757 U07699 U05002 U51277 U03171 U07451 U02027 U07698 U05001 U51276 U03170 U07450 U02026 U07697 U05000 U51275 U03169 U07449 U02025 U07696 U04999 U08515 U03168 U07448 U02024 U07675 U04998 U06445 U03167 U07447 U02023 U07674 U04997 U06444 U01689 U07446 U02022 U07673 U04996 U06443 U01688 U07445 U02021 U07672 U04801 U06442 U01687 U07435 U02020 U07671 U04800 U06441 U01686 U07434 U02019 U07670 U04799 U06439 U01685 U07433 U02018 U07669 U04798 U06438 U01684 U07432 U01256 U07668 U04797 U06437 U01683 U07407 U01255 U50140 U04796 U06436 U01682 U07406 U08646 U07139 U04795 U06435 U01681 U07405 U08451 U07138 U04794 U06434 U01680 U07404 U08126 U04550 U04793 U06433 U01679 U07403 U08098 U03193 U04792 U06432 U08817 U07402 U08097 U05085 U04791 U06431 U08102 U07401 U06955 U05084 U04790 U06430 U51318 U07400 U50391 U05083 U04789 U06429 U51317 U07399 U50388 U05082 U04788 U06428 U51316 U07398 U50387 U05081 U04787 U06427 U51315 U07397 U01314 U05080 U04786 U06426 U04864 U07396 U07235 U05079 U07973 U06425 U04863 U07395 U07234 U05078 U07972 U06424 U04862 U07394 U07233 U05077 U07971 U06423 U04861 U07393 U07232 U05076 U07970 U06422 U06960 U07392 U06643 U05075 U07969 U06421 U06959 U07391 U05648 U05074 U07968 U06420 U06958 U07390 U05606 U02635 U02038 U06419 U06957 U07389 U04445 U02634 U01835 U06418 U05277 U07388 U04436 U03192 U01834 U06417 U05276 U07384 U06468 U03165 U01833 U06416 U05275 U07383 U02775 U03146 U01832 U06415 U50799 U07382 U02774 U02930 U01831 U51154 U50798 U07381 U02773 U51272 U01830 U51153 U50797 U07380 U02772 U51271 U01745 U50092 U50796 U07379 U02771 U51270 U01744 U08685 U07303 U07378 U02770 U51269 U01361 U07036 U07248 U07377 U02769 U51268 U01358 U05975 U04480 U07376 U02041 U51267 U06777 U05974 U04479 U07375 U05347 U51266 U06559 U01171 U04478 U07374 U06944 U51265 U06558 U01418 U04477 U07373 U06875 U51264 U06557 U01417 U04476 U07372 U03743 U51263 U06556 U01416 U04475 U07371 U04509 U51262 U06555 U01415 U04474 U07370 U04508 U51261 U06554 U01414 U04473 U07369 U04507 U51260 U06553 U01413 U04472 U07368 U03145 U51259 U06552 U01412 U04471 U07367 U05891 U51258 U06407 U01411 U51305 U07366 U05890 U51257 U06406 U01410 U07622 U07365 U05889 U51256 U06405 U01409 U51342 U07364 U05888 U51255 U06404 U01408 U50630 U07363 U05887 U51254 U06403 U01407 U50629 U07358 U05886 U51253 U06402 U01406 U50628 U07357 U51897 U51252 U06401 U01405 U50627 U07356 U51896 U51251 U06400 U01404 U50626 U07355 U08903 U51250 U04916 U01403 U50625 U07354 U08902 U51249 U04915 U08212 U50624 U07353 U08901 U51248 U04914 U08211 U50623 U07352 U08900 U51247 U04913 U08210 U50450 U07343 U07165 U51246 U04912 U08209 U50449 U07342 U07164 U51245 U04911 U08208 U09016 U07341 U07041 U51244 U04910 U08207 U08938 U07340 U07038 U51243 U04909 U08206 U08866 U07339 U06952 U51242 U04908 U08205 U08865 U07338 U06946 U51241 U04907 U08204 U08864 U07337 U06780 U07240 U04906 U08203 U08863 U07336 U06705 U07108 U04905 U08202 U08862 U07335 U06646 U07107 U04904 U08201 U08861 U07334 U06644 U07106 U04903 U08200 U08778 U50495 U06453 U50707 U02338 U08199 U08606 U08125 U06452 U50421 U02337 U08198 U08605 U07922 U06451 U50072 U02336 U06241 U08604 U07214 U06448 U05069 U02335 U06240 U08603 U07174 U06447 U05068 U02334 U06239 U08302 U07151 U06446 U05067 U02333 U06238 U08301 U07147 U06331 U05066 U02332 U06237 U08300 U07143 U06330 U51027 U02331 U06236 U08299 U07140 U06167 U50420 U02330 U06235 U08273 U07037 U06166 U05826 U02329 U06234 U08272 U06872 U06136 U05449 U02328 U06233 U08271 U06134 U06135 U03574 U02327 U06232 U08270 U06133 U05611 U50712 U02326 U05634 U08269 U06131 U04682 U02689 U02325 U05633 U08268 U06129 U08292 U02688 U02324 U05632 U08267 U05910 U08291 U02687 U02323 U05631 U08266 U05909 U08290 U02686 U02322 U05630 U08265 U05902 U08289 U02685 U02321 U05629 U08264 U05612 U08288 U02684 U02320 U05628 U08263 U05605 U08287 U02683 U02319 U05627 U08262 U05567 U08286 U02682 U02235 U05626 U08261 U04883 U08285 U02681 U02234 U05625 U08260 U50473 U08284 U02680 U02233 U05624 U08259 U50472 U08283 U02679 U02232 U05623 U08258 U50471 U08282 U02678 U02231 U05622 U08257 U50470 U08241 U02677 U02230 U05621 U08256 U04327 U08240 U02676 U02229 U05620 U08255 U04123 U08239 U02675 U02228 U05619 U08254 U03616 U08238 U02674 U02227 U05278 U08253 U03355 U08237 U02673 U02226 U04894 U08252 U03166 U08236 U02672 U02225 U04893 U08251 U08563 U08235 U02671 U02224 U04892 U08250 U08556 U08234 U02670 U02223 U04891 U08249 U08449 U08233 U02669 U02222 U04890 U08248 U07947 U08232 U02668 U02221 U04889 U08247 U07441 U08216 U02667 U02220 U04888 U08246 U07436 U08215 U02666 U02219 U04887 U08245 U07211 U08214 U02665 U02218 U04886 U08244 U07144 U08213 U02664 U02217 U04885 U08243 U06778 U08197 U02663 U02216 U04884 U08242 U05911 U08196 U50934 U02215 U02931 U08118 U03352 U08195 U50933 U02214 U01827 U08035 U03351 U08194 U50932 U02213 U01174 U08034 U03350 U08193 U50931 U02212 U01172 U08033 U03349 U08192 U50930 U02211 U01167 U07948 U03348 U08191 U50929 U02210 U01040 U07863 U50093 U08190 U50928 U02209 U01035 U07862 U02600 U08189 U50927 U02208 U01173 U07499 U01341 U08188 U50926 U02207 U04765 U07498 U01340 U08187 U50925 U02206 U04764 U07497 U01339 U08186 U50924 U50945 U04763 U07003 U01338 U08185 U50923 U50886 U04762 U07002 U01337 U08184 U50922 U50885 U04761 U07001 U01336 U08183 U50921 U50884 U04760 U06816 U01335 U08182 U50920 U08723 U04759 U06724 U01334 U08181 U50919 U08706 U04758 U06723 U01333 U08180 U50918 U08705 U04757 U06601 U01332 U08179 U50917 U08704 U04756 U06600 U01276 U08178 U50916 U08703 U04755 U06599 U01275 U08177 U50915 U08702 U04754 U06598 U01274 U08176 U50914 U08450 U04753 U06597 U01273 U08175 U50913 U07944 U04752 U06488 U01272 U08174 U50912 U07943 U04751 U06487 U01271 U08158 U50911 U06319 U04750 U06486 U01270 U08157 U50910 U06318 U04749 U06414 U01269 U08156 U06503 U04282 U04748 U05885 U01268 U08155 U06502 U04281 U04747 U05697 U01267 U03324 U06501 U04280 U04746 U05696 U50459 U03323 U06500 U08379 U04745 U05695 U01922 U03322 U06499 U08378 U04744 U05694 U01921 U03321 U06498 U08377 U04743 U05693 U01920 U03320 U06497 U08376 U04742 U05692 U01919 U03319 U06496 U08375 U04741 U05691 U01918 U51443 U06495 U08374 U04740 U05690 U01917 U51442 U05945 U08373 U04739 U05689 U01916 U51441 U05944 U08372 U04738 U05688 U01915 U51326 U05943 U08371 U01503 U05687 U01914 U51031 U05942 U08370 U01502 U05686 U01913 U51030 U05941 U08369 U01501 U05685 U01912 U51028 U05940 U08368 U01500 U05684 U01911 U50954 U05939 U08367 U01499 U05683 U01910 U04266 U05938 U08366 U01498 U05259 U01909 U04265 U05937 U08365 U01497 U05258 U01908 U04264 U05936 U08364 U06948 U05257 U51330 U04263 U05935 U08363 U04709 U05256 U50942 U04262 U05934 U08362 U04685 U05255 U05650 U04261 U05933 U08361 U04684 U05254 U05450 U04260 U05932 U08360 U04680 U05253 U08931 U04259 U05931 U08359 U03928 U05167 U08930 U04258 U05930 U08358 U03059 U05166 U08929 U04257 U05929 U08357 U06924 U05135 U08928 U04256 U05928 U08356 U06923 U05134 U08927 U04255 U05927 U02312 U06922 U05065 U08926 U51240 U05926 U01512 U06921 U05062 U08925 U51239 U05925 U01511 U06920 U05061 U08924 U51238 U05924 U08281 U06919 U05016 U03142 U51237 U05923 U08280 U06918 U05015 U01313 U51236 U05922 U08279 U06917 U05014 U51172 U51235 U05921 U08278 U06916 U04922 U51171 U51234 U05869 U08277 U06915 U04706 U51170 U51233 U05868 U08276 U06523 U04469 U51169 U08993 U05867 U08275 U04458 U04468 U04124 U08992 U05866 U08274 U04457 U04467 U50443 U08922 U05865 U04389 U04456 U04466 U50442 U05607 U05864 U04388 U04455 U04465 U50441 U03483 U05863 U04387 U04454 U04464 U50440 U08741 U05862 U04386 U04453 U04463 U50439 U08740 U05861 U04385 U04452 U04462 U50438 U08739 U05860 U03007 U04451 U04461 U50437 U08738 U05859 U03006 U04450 U04460 U50436 U07118 U05858 U03005 U04449 U04459 U50435 U07117 U05857 U03004 U05810 U04433 U50434 U06454 U05856 U03003 U50050 U04432 U50433 U04714 U05855 U51574 U50049 U04225 U50432 U04713 U05854 U51573 U50048 U04224 U50431 U04712 U05853 U51572 U07949 U04223 U50430 U04711 U05852 U51571 U06263 U04222 U50429 U02659 U05851 U51570 U06262 U04221 U50428 U02658 U05850 U51569 U06261 U04220 U50427 U02657 U05849 U51568 U06260 U04219 U50426 U02656 U05848 U51567 U06259 U04218 U50425 U02655 U05847 U51566 U06258 U04026 U50424 U02654 U05846 U51565 U06257 U04025 U50423 U01740 U05845 U50702 U06256 U04024 U50422 U51034 U05844 U50701 U06255 U04023 U05133 U50795 U05823 U50700 U06254 U04022 U05132 U50794 U05822 U50699 U06253 U03791 U05131 U50793 U05821 U50698 U06252 U03746 U05130 U50792 U05820 U50071 U06251 U03745 U05129 U50791 U05819 U08971 U06250 U03744 U05128 U50790 U05818 U08970 U06249 U03298 U05127 U50789 U05817 U08969 U02923 U03297 U05126 U50788 U05816 U08968 U01823 U03148 U05125 U50787 U05815 U08967 U01382 U03072 U05124 U50786 U05814 U08966 U01375 U02986 U05123 U50785 U05813 U07827 U01364 U02878 U05122 U50784 U05812 U07826 U01360 U02877 U05121 U50783 U05811 U07825 U01312 U02876 U05120 U50782 U05708 U07824 U05517 U02875 U05119 U50781 U05707 U07823 U07148 U02874 U05118 U50780 U05706 U07135 U06876 U02873 U05117 U50779 U05705 U07134 U06706 U02812 U05116 U50778 U05704 U07133 U51943 U02811 U05115 U50777 U05703 U07132 U51942 U02810 U05114 U50776 U05702 U07131 U51941 U02733 U05113 U50775 U05701 U03790 U51940 U02732 U05112 U50774 U05700 U03787 U51939 U02731 U03118 U50773 U05699 U03693 U51938 U02730 U03117 U50772 U05698 U51549 U51937 U02729 U03116 U50771 U05682 U51548 U51936 U02728 U03115 U50770 U05681 U51547 U51935 U02727 U03114 U50769 U05680 U51546 U51934 U02726 U03113 U50768 U05679 U51545 U51933 U02725 U03112 U50767 U05678 U51544 U51932 U02724 U03111 U50766 U05677 U51543 U51931 U02662 U03110 U50765 U05676 U51542 U51930 U02661 U03109 U50764 U05675 U51541 U51929 U02660 U03108 U50763 U05674 U51540 U51928 U02641 U03107 U50762 U05673 U51539 U51927 U02640 U03106 U50761 U05672 U51538 U51926 U02524 U03105 U50760 U04773 U51537 U51925 U02515 U03104 U50759 U03544 U51536 U51924 U02129 U03103 U50758 U03543 U51535 U50493 U02128 U03102 U50757 U03542 U51534 U06205 U02127 U03101 U50756 U03541 U51533 U06204 U02126 U03100 U50755 U03540 U51532 U06203 U02125 U03099 U50754 U03539 U51531 U06202 U02124 U03098 U50753 U03538 U51530 U06201 U02043 U03097 U50752 U03537 U51529 U06200 U01907 U03096 U50751 U03536 U51528 U06199 U01727 U03095 U50664 U03535 U51527 U06198 U01264 U03094 U50663 U03534 U51526 U06197 U01263 U06945 U50662 U03533 U51525 U06196 U01262 U05908 U50661 U03532 U51524 U06195 U01261 U04010 U50660 U03531 U51523 U06194 U01260 U03789 U50659 U03530 U51522 U02757 U01189 U06596 U50502 U03529 U51521 U02756 U01011 U06595 U50595 U03528 U51519 U02755 U08037 U06594 U50594 U03527 U51518 U02754 U08036 U06593 U50593 U03526 U51517 U02753 U05345 U06592 U50592 U03525 U51516 U02752 U05344 U06591 U50591 U03524 U51515 U03692 U05343 U06590 U50590 U03523 U51514 U03691 U05342 U06589 U50589 U03522 U51513 U03690 U05341 U06588 U50588 U03521 U51512 U03689 U05340 U06587 U50587 U51445 U51511 U03688 U05339 U06586 U50586 U05901 U51510 U03687 U05338 U06585 U50585 U05900 U51509 U03756 U01905 U06584 U08607 U05899 U51508 U03755 U01620 U06583 U05808 U05898 U51507 U01293 U01619 U06582 U05807 U05897 U51506 U02341 U01164 U06581 U05806 U05896 U51505 U02340 U08991 U06580 U05805 U05895 U51504 U02339 U06819 U06579 U05804 U05894 U51503 U03551 U03057 U06578 U05803 U05893 U51502 U03550 U01837 U06577 U05109 U05892 U51501 U03549 U02424 U06576 U05108 U07977 U51500 U03548 U02423 U06575 U05107 U07976 U51499 U03547 U02422 U06574 U05106 U07975 U51498 U03546 U02421 U06573 U05105 U07974 U51497 U03545 U02420 U06572 U05104 U02966 U51496 U50386 U02419 U06571 U05103 U02965 U51495 U50385 U02418 U06570 U05102 U02964 U51494 U50384 U02417 U06569 U05101 U02963 U51493 U50383 U02416 U06568 U05100 U02962 U51492 U50382 U01954 U06567 U05099 U02189 U51491 U50381 U01953 U04991 U05098 U02188 U51490 U50380 U01952 U04987 U04737 U51895 U51489 U50379 U01951 U04660 U04658 U51894 U51488 U50378 U01390 U04434 U04657 U51880 U51487 U50377 U50908 U50821 U04656 U51879 U51485 U02606 U03742 U50820 U04655 U51878 U51484 U02267 U03741 U50819 U04654 U51877 U51483 U02605 U08092 U50818 U07173 U51876 U51482 U51659 U08091 U50817 U07172 U51874 U51481 U07130 U08090 U50816 U07171 U51873 U51480 U07129 U08089 U50815 U07170 U51872 U51479 U01434 U08088 U50814 U07169 U51871 U51478 U01433 U08087 U50813 U07168 U51870 U51477 U01432 U08086 U50812 U07167 U51869 U08513 U01431 U01906 U50811 U07166 U51868 U08512 U01430 U51683 U50810 U06333 U51867 U08511 U01429 U51682 U50809 U06332 U51866 U08510 U01428 U51681 U50808 U04519 U50553 U08509 U01427 U51680 U05231 U04483 U50552 U08508 U01426 U51679 U05230 U01509 U50551 U08507 U01425 U51678 U05229 U07137 U50550 U08506 U01424 U51677 U05228 U07136 U50549 U08505 U01423 U51676 U05227 U02031 U50548 U08504 U01422 U51675 U05226 U50241 U50547 U08503 U01421 U51674 U05225 U08075 U50546 U08502 U01420 U51673 U05224 U08074 U50545 U08501 U01419 U51672 U05223 U08073 U50544 U08500 U06798 U51671 U05222 U08072 U50543 U08499 U06797 U51669 U05197 U08071 U50542 U08498 U06796 U51668 U05196 U08070 U50541 U08482 U06795 U51667 U05195 U08069 U50540 U08481 U06794 U51666 U05194 U08068 U50539 U08480 U50708 U51665 U05193 U08067 U50538 U08479 U50090 U51664 U05192 U08066 U50537 U08478 U08557 U51663 U05191 U08065 U50536 U08477 U07416 U51662 U05190 U08064 U08965 U08476 U06953 U51661 U05189 U08063 U08964 U08475 U06947 U51660 U05188 U08062 U08963 U08474 U06450 U51552 U05187 U08061 U04647 U51865 U06165 U51551 U05186 U08060 U04646 U51864 U05518 U51550 U05185 U08059 U04645 U51863 U08560 U51159 U05184 U08058 U02988 U51862 U07437 U50747 U05183 U08057 U02040 U51861 U07430 U50746 U05182 U08056 U01380 U51860 U07428 U50745 U05181 U08055 U07496 U51859 U07414 U50744 U05180 U08054 U07495 U51858 U07141 U50743 U05179 U08053 U07494 U51857 U05346 U50742 U05178 U08052 U07493 U51856 U03501 U50741 U05177 U08051 U07492 U51855 U03500 U50740 U05176 U08050 U07491 U51854 U02794 U50739 U05175 U08049 U07490 U51853 U02036 U50738 U05174 U08048 U07489 U51852 U02033 U50737 U05173 U08047 U07413 U51851 U02032 U50736 U05172 U08046 U07412 U51850 U02030 U50735 U05171 U08045 U07411 U51849 U01826 U50734 U05170 U08044 U07410 U51848 U50166 U50733 U05169 U08043 U07409 U51847 U50165 U50732 U05168 U08042 U07408 U51846 U50164 U50731 U05165 U08041 U07362 U51845 U50163 U50730 U05164 U08040 U07361 U51844 U50162 U50729 U05163 U08039 U07360 U51843 U50161 U50728 U05162 U08038 U07359 U51842 U50160 U50727 U05161 U51945 U07300 U51841 U50159 U50726 U05160 U51944 U04932 U51840 U50158 U50725 U05159 U08670 U04879 U51839 U50157 U50724 U05158 U08669 U04877 U51838 U50156 U50723 U05157 U03367 U04827 U51837 U50155 U50722 U05156 U02808 U04767 U51836 U50154 U50721 U05155 U02807 U08840 U51835 U09014 U50720 U05154 U51329 U08839 U51834 U09013 U50719 U05153 U50496 U08838 U51833 U09012 U50718 U05152 U51192 U08837 U51832 U09011 U50717 U05151 U51191 U08836 U51831 U09010 U50716 U05150 U51190 U08835 U51830 U09009 U50715 U05149 U51189 U08834 U51829 U09008 U50714 U05148 U51188 U08833 U51828 U08989 U50574 U05147 U51187 U08832 U51827 U08988 U50573 U05146 U51186 U08831 U51826 U08987 U50572 U05145 U51185 U07994 U51825 U08986 U50571 U05144 U07176 U07993 U51824 U08985 U50570 U05143 U07837 U07992 U51823 U08984 U50569 U05142 U07836 U07991 U51822 U08983 U50568 U05141 U07438 U07990 U51821 U08982 U50567 U05140 U50452 U07989 U51820 U50469 U50566 U05139 U50451 U07988 U51819 U50468 U50565 U05138 U50448 U07987 U51816 U50467 U50564 U05137 U50447 U07920 U51815 U50466 U50563 U05136 U50446 U07919 U51814 U50465 U50562 U03577 U50445 U07918 U51760 U50464 U50561 U03576 U03511 U07917 U51759 U50463 U50560 U04650 U08306 U07916 U51758 U50462 U50559 U04437 U51036 U07915 U51757 U01533 U50558 U04425 U50946 U07914 U51756 U01532 U50557 U04271 U50669 U07913 U51755 U01531 U50556 U04270 U50407 U07912 U51754 U01530 U50555 U04269 U07150 U07911 U08814 U01529 U50554 U04268 U51795 U07910 U08813 U01528 U08668 U04267 U51327 U07909 U08812 U01324 U08667 U03617 U50950 U07908 U08811 U01323 U08666 U02978 U50939 U07907 U08810 U01322 U08665 U02977 U06642 U07906 U08809 U01321 U08664 U02976 U50665 U07213 U08808 U01320 U08663 U02975 U07955 U04440 U08807 U01319 U08662 U02974 U50937 U01940 U08806 U01318 U08661 U02973 U50458 U01939 U08805 U01317 U08660 U02972 U04707 U01938 U08804 U01316 U08659 U02971 U04706 U01937 U08803 U01311 U08658 U02970 U01376 U51813 U08802 U01310 U08657 U02969 U51794 U51812 U08801 U01189 U08656 U02968 U51751 U51811 U08800 U08497 U08655 U02967 U51649 U51810 U08799 U08496 U08654 U08923 U51647 U51809 U08798 U50457 U08653 U07444 U51334 U51808 U08797 U50456 U08652 U07443 U51328 U51807 U08796 U50095 U08651 U07442 U51322 U51806 U08795 U06940 U08650 U06248 U51039 U51805 U08794 U06939 U08649 U06247 U51035 U51804 U08793 U06938 U08648 U06246 U51033 U51803 U08792 U06937 U08173 U06245 U51032 U51802 U08791 U06936 U08172 U06244 U51025 U51800 U08790 U06935 U08171 U06243 U50944 U51799 U08789 U06934 U08170 U04662 U50882 U51953 U08788 U06933 U08169 U08731 U50667 U02805 U08787 U06932 U08168 U08730 U50494 U02804 U06335 U06931 U08167 U08729 U08615 U02803 U08305 U06930 U08166 U08728 U08448 U05761 U04595 U06929 U08165 U08727 U08152 U05760 U04594 U06928 U08164 U08726 U08151 U05348 U04593 U06906 U08163 U08725 U08150 U03926 U04592 U06905 U08162 U08724 U08145 U02598 U04591 U06904 U08161 U50153 U08142 U07650 U04590 U06903 U08160 U50152 U06522 U07649 U05043 U06902 U08159 U50151 U05903 U07648 U05042 U06901 U08121 U50150 U05569 U07647 U51875 U06900 U08120 U50149 U05453 U07646 U08981 U06899 U07848 U50148 U04162 U07645 U08980 U06898 U07847 U50147 U03510 U07644 U08979 U06897 U07846 U50146 U03509 U07643 U08978 U06896 U07845 U50145 U03508 U07642 U08977 U06895 U07844 U50144 U03507 U07641 U08976 U06894 U07843 U50143 U03506 U07530 U08975 U06893 U07842 U50142 U03505 U07529 U08974 U06892 U07841 U04071 U02815 U07528 U08973 U06891 U07840 U04070 U51041 U07527 U08972 U06890 U07839 U04988 U51040 U07526 U06132 U06889 U07838 U04882 U50940 U07525 U01843 U06888 U07835 U04769 U50616 U07524 U01842 U06887 U07834 U04766 U07040 U07523 U01841 U06886 U07833 U04659 U04927 U07522 U01840 U06885 U07832 U04652 U02987 U07521 U01821 U06884 U07831 U04651 U51650 U07509 U01820 U06883 U07830 U04448 U51323 U07508 U01819 U06882 U07829 U04447 U51043 U07507 U01818 U06881 U07828 U04441 U51042 U07506 U01817 U06880 U06018 U03788 U50878 U07505 U01816 U06879 U05671 U06873 U50710 U07504 U01815 U06878 U05670 U06707 U50615 U07503 U01814 U06877 U05669 U04986 U50453 U07502 U01813 U06848 U05668 U51691 U50406 U07501 U01812 U06847 U05667 U51690 U08559 U07500 U01811 U06846 U05666 U51689 U07946 U07487 U01810 U06845 U05665 U51688 U04990 U07486 U01809 U06844 U05664 U51687 U04661 U07485 U01808 U06843 U05663 U51686 U51657 U07484 U01069 U06842 U05662 U50136 U50958 U07483 U01068 U06841 U05661 U50135 U50957 U07482 U01067 U06840 U05660 U50134 U50956 U07481 U01066 U06839 U05659 U50133 U50877 U07480 U01065 U06838 U05658 U50132 U50612 U07479 U01064 U06837 U05657 U50131 U50501 U07478 U01063 U06836 U05656 U50130 U50497 U07205 U01062 U06835 U05655 U50129 U50139 U07204 U01061 U06834 U05654 U50128 U08995 U07203 U01060 U06833 U05653 U50127 U06703 U07202 U01059 U06832 U05652 U50126 U06645 U07201 U01058 U06831 U05616 U50125 U04926 U07200 U02017 U06830 U05615 U50124 U04429 U07199 U02016 U06829 U05614 U50123 U04428 U07198 U02015 U05405 U05613 U50122 U04427 U50499 U02014 U05404 U04667 U50121 U03070 U50455 U02013 U05403 U04470 U50120 U50952 U06083 U02012 U05402 U03573 U50119 U50883 U06082 U02011 U05401 U03572 U50118 U50879 U06081 U02010 U05400 U03429 U50117 U02813 U06080 U02009 U05399 U03428 U50116 U01837 U06079 U02008 U05398 U03427 U50115 U01836 U06078 U02007 U05397 U03426 U50114 U01776 U06077 U02006 U05396 U03425 U50113 U01775 U06076 U02005 U05395 U03424 U50112 U01774 U06075 U02004 U05394 U03318 U50111 U01773 U06074 U02003 U05393 U03317 U50110 U01772 U06073 U02002 U05392 U03316 U50109 U01771 U06072 U02001 U05391 U03315 U50108 U01770 U06071 U02000 U05390 U03314 U50107 U01769 U06070 U01999 U05389 U03313 U50106 U01768 U06054 U01998 U05388 U03312 U50105 U01767 U06053 U01997 U05387 U03311 U50104 U01766 U06052 U01996 U05386 U03310 U50103 U01765 U06051 U01995 U05385 U03309 U50102 U01764 U04921 U01994 U05384 U03308 U50101 U01763 U04920 U01993 U05383 U03307 U50100 U01762 U04919 U01992 U05382 U03306 U50099 U01761 U04918 U01991 U05381 U03305 U50098 U01760 U04917 U01990 U05380 U03304 U50097 U01759 U51752 U50881 U05379 U03296 U02123 U01758 U51655 U07805 U05378 U03295 U02122 U01757 U50955 U07804 U05377 U03290 U02121 U01587 U50270 U07803 U05376 U03289 U02120 U01506 U50259 U07802 U05375 U03288 U02119 U01505 U50258 U07801 U05374 U03287 U02118 U01315 U50257 U07800 U05337 U03286 U02117 U01266 U50256 U07799 U05336 U03285 U02116 U01190 U50255 U07798 U05335 U03284 U02115 U01188 U50254 U08231 U05334 U03283 U02114 U01187 U50253 U08230 U05333 U03282 U02113 U01186 U50252 U08229 U05332 U03281 U02112 U01181 U08777 U08228 U05331 U03280 U02111 U01180 U08776 U08227 U05330 U03279 U02110 U01179 U08775 U08226 U05329 U03278 U02109 U01178 U08774 U08225 U05328 U03277 U02108 U01177 U08773 U08224 U05327 U03276 U02107 U01176 U08772 U08223 U05326 U03275 U02105 U01175 U08771 U08222 U05325 U03274 U02104 U01170 U08770 U08221 U05324 U03273 U02103 U01169 U08769 U08220 U05323 U03272 U02102 U01168 U08768 U08219 U05322 U03271 U02101 U01166 U08767 U08218 U05321 U02862 U02100 U01165 U08766 U08217 U05320 U02460 U02099 U01164 U08765 U50943 U05319 U02459 U02098 U01163 U08764 U03034 U05318 U02458 U02097 U01162 U08745 U01397 U05317 U02457 U02096 U01161 U08744 U01396 U05316 U02456 U02095 U01160 U50614 U01395 U05315 U02455 U02094 U01159 U50507 U01394 U05314 U02454 U02093 U01039 U50408 U01393 U05313 U02453 U02092 U01038 U50250 U01392 U05312 U02452 U02091 U01037 U50091 U51448 U05311 U02451 U02090 U01036 U50075 U08842 U05310 U02450 U02089 U01034 U08998 U08841 U05309 U02449 U02088 U01033 U07212 U06128 U05308 U02448 U02087 U01032 U07178 U06127 U05307 U02447 U02086 U01031 U07149 U06126 U05306 U02205 U02085 U01030 U04648 U06125 U05305 U01863 U02084 U01029 U04617 U06124 U05304 U01862 U02083 U01028 U04616 U06123 U05303 U01861 U02082 U01027 U04615 U03720 U05302 U01860 U02081 U01026 U04614 U02929 U05301 U01859 U02080 U01025 U04613 U02596 U05300 U01858 U02079 U01024 U04612 U02594 U05299 U01857 U02078 U01023 U04611 U51753 U05298 U01856 U02077 U01022 U04610 U04939 U05297 U01855 U02076 U01021 U04609 U04938 U05296 U01854 U02075 U01020 U04608 U04937 U05295 U01853 U02074 U01019 U04607 U04823 U05294 U01852 U02073 U01018 U04606 U04822 U05293 U01851 U02072 U01017 U04605 U04821 U05292 U01850 U02071 U01016 U04604 U04820 U05291 U01849 U02070 U01010 U04603 U04819 U05290 U01848 U02069 U01009 U04602 U04818 U05289 U01847 U02068 U01008 U04601 U04817 U05288 U01846 U02067 U01007 U04600 U04326 U05287 U01845 U02066 U01006 U04599 U51440 U05286 U01844 U02065 U01005 U04598 U51439 U05285 U01577 U02064 U01004 U04597 U51438 U05284 U01576 U02063 U01003 U04596 U06566 U05283 U01575 U02062 U01002 U51044 U03084 U05282 U07145 U02061 U01001 U08298 U01839 U05281 U05269 U02060 U07795 U08297 U01838 U05280 U05268 U02059 U02318 U08296 U50405 U05279 U05267 U02058 U02317 U08295 U06399 U04977 U05266 U02057 U02316 U08294 U06398 U04976 U05265 U02056 U02315 U08293 U06397 U04975 U05264 U02055 U02314 U04831 U04284 U04974 U05263 U02054 U02313 U04776 U04283 U04973 U05262 U02053 U08553 U04710 U03252 U04972 U05261 U02052 U08552 U51336 U03251 U04971 U05260 U02051 U08112 U50202 U03250 U04970 U04967 U02050 U08111 U50201 U03249 U04969 U51307 U02049 U08110 U50200 U51308 U04968 U50199 U02048 U08109 Model 2151 U50523 U50524 Model 2152 U50527 U50528 U50529 U50530 U50531 U50532 U50533 U50534 U50535 U51745 Model 2155 U50525 U50526 U51736 U51737 U51738 U51739 U51740 U51741 U51742 U51743 Model 2156 U51733 U51734 U51735 |
| FEI Number |
1831750
|
Recalling Firm/ Manufacturer |
Stryker Medical Division of Stryker Corporation
3800 E Centre Ave
Portage
MI
49002-5826
|
| For Additional Information Contact | 269-324-6609 |
Manufacturer Reason for Recall | Stryker has received complaints from the field alleging instances where the In Touch beds are hoisted or lifted from the floor and the casters have fallen off. The investigations into these casters have shown that in these cases the caster has broken on the top of the caster stem where the brake rod inserts into the caster. If more than one caster on a unit is damaged, brake functions can be affec |
FDA Determined Cause 2 | Device Design |
| Action | On 06/30/1014 the firm sent "Urgent Medical Device Notification" letters to their customers. |
| Quantity in Commerce | 11,632 units |
| Distribution | Worldwide Distribution -USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV, and the countries of Argentina, Brazil, Canada, Chile, China, EMEA, Greece, Italy, Japan, Korea, Latin America, Mexico, Poland, Singapore, South Africa, Taiwan, and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|