| Date Initiated by Firm | July 16, 2014 |
| Date Posted | August 26, 2014 |
| Recall Status1 |
Terminated 3 on December 16, 2014 |
| Recall Number | Z-2553-2014 |
| Recall Event ID |
68854 |
| 510(K)Number | K130276 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338. |
| Code Information |
All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338 |
| FEI Number |
2517506
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc.
500 Gbc Dr Ms 514
PO BOX 6101
Newark
DE
19702-2466
|
| For Additional Information Contact | Customer Support 800-441-9250 |
Manufacturer Reason for Recall | The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator. |
FDA Determined Cause 2 | Device Design |
| Action | Siemens notified end users of a potential safety issue with the Reagent Lid hinge on the Dimension EXL instrument via an Urgent Medical Device Correction letter, dated July 2014. The letter indicated Customer Service Engineers will inspect and adjust the tension on the reagent lid hinges during every visit and replace hinges which no longer properly adjust. Users were urged to use caution when the Reagent Lid is open and notify Siemens the Reagent Lid is shifting down. |
| Quantity in Commerce | 2757 |
| Distribution | Nationwide and internationally, including Canada and Mexico |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|