 Class 2 Device Recall Regional Oximetry Sensorsee related information
Date Initiated by FirmNovember 25, 2014
Date PostedDecember 31, 2014
Recall Status1 Terminated 3 on September 01, 2015
Recall NumberZ-0901-2015
Recall Event ID 70045
510(K)NumberK132402 
Product Classification Oximeter - Product Code DQA
ProductSenSmart, Regional Oximetry EQUANOX Classic Plus Adult/Pediatric Sensor > 40 kg., Model 8003CA (REF 9824-001, 3-Wavelength), 8004CA (REF 9705-001, 4-Wavelength), with associated materials (INT-100 cable, operator's manual; and Quick Start Guide). Sensors are single-packed in a pouch, then multi-packed, 20 per shelf-box.
Code Information PRTNUM_72 PMDES1_01 LOTNUM_72 8825001 PKG ASSY, INT100 25605 8825001 PKG ASSY, INT100 414312 8825001 PKG ASSY, INT100 414353 8825001 PKG ASSY, INT100 414405 8825001 PKG ASSY, INT100 418061 9705001 PKG ASSY, 8004CA, 20PK 37245 9705001 PKG ASSY, 8004CA, 20PK 36349 9705001 PKG ASSY, 8004CA, 20PK 36475 9705001 PKG ASSY, 8004CA, 20PK 36476 9705001 PKG ASSY, 8004CA, 20PK 36477 9705001 PKG ASSY, 8004CA, 20PK 36593 9705001 PKG ASSY, 8004CA, 20PK 36693 9705001 PKG ASSY, 8004CA, 20PK 36694 9705001 PKG ASSY, 8004CA, 20PK 36695 9705001 PKG ASSY, 8004CA, 20PK 36696 9705001 PKG ASSY, 8004CA, 20PK 36697 9705001 PKG ASSY, 8004CA, 20PK 36698 9705001 PKG ASSY, 8004CA, 20PK 36886 9705001 PKG ASSY, 8004CA, 20PK 36887 9705001 PKG ASSY, 8004CA, 20PK 36989 9705001 PKG ASSY, 8004CA, 20PK 37066 9705001 PKG ASSY, 8004CA, 20PK 37094 9705001 PKG ASSY, 8004CA, 20PK 37151 9705001 PKG ASSY, 8004CA, 20PK 37245 9705001 PKG ASSY, 8004CA, 20PK 37374 9705001 PKG ASSY, 8004CA, 20PK 37410 9705001 PKG ASSY, 8004CA, 20PK 37514 9705001 PKG ASSY, 8004CA, 20PK 37515 9705001 PKG ASSY, 8004CA, 20PK 37591 9705001 PKG ASSY, 8004CA, 20PK 37607 9705001 PKG ASSY, 8004CA, 20PK 37669 9705001 PKG ASSY, 8004CA, 20PK 37670 9824001 PKG ASSY, 8003CA, 20 PK 36611 9824001 PKG ASSY, 8003CA, 20 PK 36888 9824001 PKG ASSY, 8003CA, 20 PK 36937 9824001 PKG ASSY, 8003CA, 20 PK 37090 9824001 PKG ASSY, 8003CA, 20 PK 37231 9824001 PKG ASSY, 8003CA, 20 PK 37575
FEI Number 2183646
Recalling Firm/
Manufacturer
Nonin Medical, Inc
13700 1st Ave N
Plymouth MN 55441-4595
For Additional Information ContactKim E. Aves
763-553-9968
Manufacturer Reason
for Recall
Nonin Medical is conducting a recall of certain regional oximetry sensors and associated products. They have received reports of signal loss during use of our shortcabled regional oximetry sensors with diaphoretic patients. No injury or illness have been reported.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionConsignees were sent a Nonin letter dated November 24, 2014. The letter described the problem and the product affected. They requested consignees to discontinue the use of the products and return it to Nonin. Consignees were asked to complete and return the "Medical Device Market Withdrawal Return Response Form". For questions and additional information they can contact James H. Russell at 763-577-2260 or e-mail jim.russell@nonin.com.
Quantity in Commerce5892 (5452 US, 440 OUS)
DistributionTo be provided later.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.