| | Class 2 Device Recall VITROS Handheld Barcode Scanner |  |
| Date Initiated by Firm | December 29, 2014 |
| Date Posted | March 11, 2015 |
| Recall Status1 |
Terminated 3 on March 27, 2017 |
| Recall Number | Z-1263-2015 |
| Recall Event ID |
70264 |
| Product Classification |
barcode scanner used with in vitro diagnostic analyzer - Product Code N/A
|
| Product | VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System. This product is an accessory for use with VITROS Analyzer Systems. |
| Code Information |
Lot Numbers: 0914F11968, 0914F11969, 0914F11970, 0914F11971, 0914F11972, 0914F11973, 0914F11974, 0914F11975, 0914F11976, 0914F11977, 0914F11978, 0914F11979, 0914F11980, 0914F11981, 0914F11982, 0914F11983, 0914F11984, 0914F11985, 0914F11986, 0914F11987, 0914F11988, 0914F11989, 0914F11990, 0914F11997, 0914F12007, 0914F12016, 0914F12017, 0914F12020, 0914F12022, 0914F12028, 0914F12033, 0914F12037, 0914F12039, 0914F12040, 0914F12041, 0914F12046, 0914F12048, 0914F12049, 0914F12050, 0914F12053, 0914F12054, 0914F12055, 0914F12056, 0914F12057, 0914F12058, 0914F12059, 0914F12060, 0914F12061, 0914F12062, 0914F12063, 0914F12064, 0914F12065, 0914F12066, 0914F12067, 0914F12081, 0914F12082, 0914F12083, 0914F12084, 0914F12085, 0914F12086, 1014F12318, 1014F12319, 1014F12320, 1014F12321, 1014F12322, 1014F12323, 1014F12386, 1014F12387, 1014F12388, 1014F12389, 1014F12390, 1014F12391. |
| FEI Number |
1000136573
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics
100 Indigo Creek Dr
Rochester
NY
14626-5101
|
| For Additional Information Contact | Ms. Jennifer Paine 908-218-8776 |
Manufacturer Reason for Recall | In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended. |
FDA Determined Cause 2 | Software design |
| Action | On 12/29/2014, Customer letter (Ref. CL2014-315) was sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees (OCD Direct, Drop-Ship and US Federal Government) to inform them of the issue and provided with instructions for reconfiguring their Barcode Scanners. |
| Quantity in Commerce | Domestic: 62 units; Foreign: 10 units |
| Distribution | Worldwide Distribution-US (nationwide) and the countries of Canada, Australia and England. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|