Class 1 Device Recall Beacon Tip Torcon NB Advantage Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Beacon Tip Torcon NB Advantage Cathetersee related information
Date Initiated by FirmJuly 02, 2015
Date PostedAugust 05, 2015
Recall Status1 Terminated 3 on September 02, 2017
Recall NumberZ-2317-2015
Recall Event ID 71692
510(K)NumberK133130 
Product Classification Catheter, intravascular, diagnostic - Product Code DQO
ProductBeacon Tip Torcon NB Advantage Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.
Code Information All Lots of following Product numbers: HNBR4.0-35-100-P-NS-0, HNBR4.0-35-100-P-NS-C1, HNBR4.0-35-100-P-NS-C2, HNBR4.0-35-100-P-NS-CANADA-041480, HNBR4.0-35-100-P-NS-DAV, HNBR4.0-35-100-P-NS-H1, HNBR4.0-35-100-P-NS-JB1,  HNBR4.0-35-100-P-NS-JB2, HNBR4.0-35-100-P-NS-JB3,  HNBR4.0-35-100-P-NS-JIM,  HNBR4.0-35-100-P-NS-KMP,  HNBR4.0-35-100-P-NS-MPA, HNBR4.0-35-100-P-NS-O1,  HNBR4.0-35-100-P-NS-SIM1,  HNBR4.0-35-100-P-NS-SIM2,  HNBR4.0-35-100-P-NS-SIM3, HNBR4.0-35-100-P-NS-VERT,  HNBR4.0-35-100-P-NS-VTK,  HNBR4.0-35-100-P-NS-WNBG,  HNBR4.0-35-125-P-NS-0, HNBR4.0-35-125-P-NS-DAV,  HNBR4.0-35-125-P-NS-H1,  HNBR4.0-35-125-P-NS-JB1,  HNBR4.0-35-125-P-NS-JB2, HNBR4.0-35-125-P-NS-MPA,  HNBR4.0-35-125-P-NS-SIM2,  HNBR4.0-35-125-P-NS-VERT,  HNBR4.0-35-125-P-NS-VTK, HNBR4.0-35-125-P-NS-WNBG,  HNBR4.0-35-40-P-NS-KMP,  HNBR4.0-35-65-P-NS-0,  HNBR4.0-35-65-P-NS-C1, HNBR4.0-35-65-P-NS-C2,  HNBR4.0-35-65-P-NS-C3,  HNBR4.0-35-65-P-NS-DAV,  HNBR4.0-35-65-P-NS-H1,  HNBR4.0-35-65-P-NS-JB1, HNBR4.0-35-65-P-NS-KMP,  HNBR4.0-35-65-P-NS-MPA,  HNBR4.0-35-65-P-NS-RC1,  HNBR4.0-35-65-P-NS-RC2, HNBR4.0-35-65-P-NS-RDC,  HNBR4.0-35-65-P-NS-RIM,  HNBR4.0-35-65-P-NS-SIM1,  HNBR4.0-35-65-P-NS-SIM2, HNBR4.0-35-80-P-NS-C1,  HNBR4.0-35-80-P-NS-C2,  HNBR4.0-35-80-P-NS-JL2,  HNBR4.0-35-80-P-NS-JL2.5, HNBR4.0-35-80-P-NS-JL3,  HNBR4.0-35-80-P-NS-JR1,  HNBR4.0-35-80-P-NS-JR2,  HNBR4.0-35-80-P-NS-JR3.5, HNBR4.0-35-80-P-NS-RH,  HNBR4.0-35-80-P-NS-RLG,  HNBR4.0-35-80-P-NS-VS,  HNBR4.0-35-80-P-NS-VS1, HNBR4.0-35-80-P-NS-VS2,  HNBR4.0-35-80-P-NS-VS3,  HNBR4.0-NT-100-P-NS-H1,  HNBR4.0-NT-80-P-NS-RC1, HNBR4.0-NT-80-P-NS-RIM,  HNBR4.0-NT-80-P-NS-RLG, HNBR4.1-35-100-P-NS-0,  HNBR4.1-35-100-P-NS-C1, HNBR4.1-35-100-P-NS-C2,  HNBR4.1-35-100-P-NS-DAV,  HNBR4.1-35-100-P-NS-H1,  HNBR4.1-35-100-P-NS-JB1, HNBR4.1-35-100-P-NS-JB2,  HNBR4.1-35-100-P-NS-JB3,  HNBR4.1-35-100-P-NS-KMP,  HNBR4.1-35-100-P-NS-MPA, HNBR4.1-35-100-P-NS-O1,  HNBR4.1-35-100-P-NS-SIM1,  HNBR4.1-35-100-P-NS-SIM2,  HNBR4.1-35-100-P-NS-SIM3, HNBR4.1-35-100-P-NS-VERT,  HNBR4.1-35-100-P-NS-WNBG,  HNBR4.1-35-110-P-NS-JB1,  HNBR4.1-35-110-P-NS-JB2, HNBR4.1-35-110-P-NS-WNBG,  HNBR4.1-35-120-P-NS-JB2,  HNBR4.1-35-130-P-NS-DAV-T,  HNBR4.1-35-40-P-NS-KMP, HNBR4.1-35-65-P-NS-C1,  HNBR4.1-35-65-P-NS-C2,  HNBR4.1-35-65-P-NS-DAV,  HNBR4.1-35-65-P-NS-H1,  HNBR4.1-35-65-P-NS-JB1, HNBR4.1-35-65-P-NS-KMP,  HNBR4.1-35-65-P-NS-LEV1,  HNBR4.1-35-65-P-NS-MPA,  HNBR4.1-35-65-P-NS-RC1, HNBR4.1-35-65-P-NS-RC2,  HNBR4.1-35-65-P-NS-RDC,  HNBR4.1-35-65-P-NS-RIM,  HNBR4.1-35-65-P-NS-SHK1.0, HNBR4.1-35-65-P-NS-SIM1,  HNBR4.1-35-75-P-NS-CHG-D,  HNBR4.1-35-80-P-NS-C1,  HNBR4.1-35-80-P-NS-C2, HNBR4.1-35-80-P-NS-JL2,  HNBR4.1-35-80-P-NS-JL2.5,  HNBR4.1-35-80-P-NS-JL3,  HNBR4.1-35-80-P-NS-JR1, HNBR4.1-35-80-P-NS-JR2,  HNBR4.1-35-80-P-NS-JR3.5,  HNBR4.1-35-80-P-NS-RC1,  HNBR4.1-35-80-P-NS-RH, HNBR4.1-35-80-P-NS-RLG,  HNBR4.1-35-80-P-NS-VS,  HNBR4.1-35-80-P-NS-VS2,  HNBR4.1-NT-100-P-NS-H1, HNBR4.1-NT-80-P-NS-C1,  HNBR4.1-NT-80-P-NS-RC1,  HNBR4.1-NT-80-P-NS-RH,  HNBR4.1-NT-80-P-NS-RIM, HNBR4.1-NT-80-P-NS-RLG,  HNBR5.0-35-65-P-NS-RUC, HNBR5.0-35-90-P-NS-RUC
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Inc.
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations
812-339-2235
Manufacturer Reason
for Recall
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.
FDA Determined
Cause 2
Component design/selection
ActionCook Medical sent an Urgent Medical Device Recall letter dated July 2, 2015, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification and instructions for returning the affected product. All customers were notified via courier. Customers with questions were instructed to contact Cook Medical Customer Relations at (800) 457-4500 or (812) 339-2235. Firm issued press release 8/3/2015.
Quantity in Commerce451550
DistributionWorldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.