| | Class 2 Device Recall VC150 Vital Signs Monitor |  |
| Date Initiated by Firm | October 22, 2015 |
| Date Posted | November 11, 2015 |
| Recall Status1 |
Terminated 3 on July 01, 2016 |
| Recall Number | Z-0264-2016 |
| Recall Event ID |
72475 |
| 510(K)Number | K133810 |
| Product Classification |
Monitor,physiological,patient(without arrhythmia detection or alarms) - Product Code MWI
|
| Product | CARESCAPE VC150 Vital Signs Monitor
Monitor vital signs in humans |
| Code Information |
Catalog numbers: 2067980-001, 2067980-002, 2067980-003, 2067980-004, 2067980-005, 2067980-006, 2067980-007, 2067980-008, 2067980-009, 2067980-010, 2067980-011, 2067980-012, 2067980-013, 2067980-014, 2067980-015, 2067980-016 Serial Numbers: In USA: SK514470004YA,SK514470005YA,SK515050001YA,SK515050002YA,SK515050003YA,SK515050004YA, SK515050005YA,SK515050006YA,SK515050007YA,SK515050008YA,SK515050009YA,SK515050010YA, SK515050011YA,SK515050012YA,SK515050013YA,SK515050014YA,SK515050015YA,SK515050026YA, SK515050027YA,SK515050028YA,SK515050029YA,SK515050030YA,SK515050031YA,SK515050032YA, SK515050033YA,SK515050034YA,SK515050035YA,SK515050036YA,SK515050037YA,SK515050038YA, SK515050039YA,SK515050040YA,SK515050041YA,SK515050042YA,SK515050043YA,SK515050044YA, SK515050045YA,SK515100051YA In Europe: SK514150001YA,SK514150004YA,SK514150005YA,SK514150007YA,SK514150008YA,SK514150009YA, SK514150010YA,SK514150011YA,SK514150019YA,SK514150020YA,SK514150023YA,SK514150026YA, SK514150027YA,SK514150028YA,SK514150029YA,SK514150030YA,SK514150031YA,SK514300002YA, SK514300003YA,SK514300004YA,SK514300005YA,SK514300006YA,SK514300007YA,SK514300008YA, SK514300009YA,SK514300010YA,SK514300011YA,SK514300014YA,SK514300015YA,SK514380001YA, SK514380002YA,SK514380003YA,SK514380004YA,SK514380005YA,SK514380006YA,SK514380007YA, SK514380008YA,SK514380009YA,SK514380010YA,SK514380011YA,SK514380012YA,SK514390001YA, SK514390005YA,SK514390006YA,SK514390007YA,SK514390008YA,SK514390009YA,SK514390010YA, SK514390011YA,SK514390012YA,SK514390013YA,SK514390014YA,SK514390016YA,SK514390017YA, SK514390018YA,SK514390019YA,SK514390020YA,SK514390021YA,SK514390023YA,SK514390024YA, SK514390025YA,SK514390026YA,SK514390027YA,SK514390028YA,SK514390029YA,SK514390030YA, SK514390031YA,SK514390032YA,SK514450001YA,SK514450002YA,SK514450003YA,SK514450004YA, SK514450005YA,SK514450006YA,SK514450008YA,SK514450011YA,SK514450018YA,SK514450021YA, SK514450022YA,SK514460003YA,SK514460004YA,SK514460005YA,SK514460006YA,SK514460007YA, SK514460008YA,SK514460009YA,SK514460010YA,SK514460011YA,SK514460012YA,SK514460013YA, SK514460014YA,SK514460015YA,SK514460016YA,SK514470006YA,SK514470007YA,SK514470008YA, SK514470009YA,SK514470010YA,SK514470011YA,SK514470012YA,SK514470013YA,SK514470014YA, SK514470015YA,SK514470016YA,SK514470018YA,SK514470020YA,SK514470025YA,SK514470026YA, SK514470027YA,SK514470028YA,SK514470029YA,SK514470030YA,SK514470031YA,SK514470032YA, SK514470033YA,SK514470034YA,SK514470035YA,SK514470036YA,SK514470037YA,SK514470038YA, SK514470039YA,SK514470040YA,SK514480003YA,SK514480004YA,SK514480005YA,SK514480006YA, SK514480007YA,SK514480008YA,SK514480009YA,SK514480010YA,SK514480011YA,SK514480012YA, SK514480013YA,SK514480014YA,SK514480015YA,SK514480016YA,SK514480017YA,SK514480018YA,SK514480019YA,SK514480020YA,SK514480021YA,SK514480022YA,SK514480023YA,SK514480024YA, SK514480025YA,SK514480026YA,SK514480027YA,SK514480028YA,SK514480029YA,SK514490018YA, SK514490019YA,SK514500001YA,SK514500002YA,SK514500003YA,SK514500004YA,SK514500005YA, SK514500006YA,SK514500007YA,SK514500008YA,SK514500009YA,SK514500011YA,SK515030069YA, SK515040001YA,SK515040002YA,SK515040003YA,SK515060004YA,SK515070014YA,SK515070015YA, SK515070016YA,SK515090008YA,SK515090009YA,SK515090010YA,SK515090011YA,SK515090012YA, SK515090013YA,SK515090014YA,SK515090015YA,SK515090016YA,SK515100035YA,SK515100036YA, SK515100037YA,SK515100038YA,SK515100039YA,SK515100040YA,SK515100041YA,SK515100042YA, SK515100043YA,SK515100044YA,SK515100045YA,SK515100046YA,SK515100047YA,SK515100048YA, SK515100049YA,SK515100050YA,SK515100054YA,SK515100056YA,SK515100057YA,SK515100058YA, SK515100059YA,SK515100060YA,SK515100061YA,SK515100063YA,SK515100064YA,SK515100066YA, SK515100067YA,SK515100068YA,SK515100069YA,SK515100070YA,SK515100071YA,SK515100072YA, SK515100073YA,SK515100074YA,SK515100075YA,SK515100077YA,SK515100078YA,SK515100079YA, SK515100080YA,SK515100082YA,SK515100083YA,SK515100085YA,SK515100087YA,SK515100088YA, SK515100094YA,SK515100096YA,SK515110001YA,SK515110004YA,SK515110005YA,SK515110006YA, SK515110008YA,SK515110009YA,SK515110010YA,SK515110011YA,SK515110013YA,SK515110019YA, SK515110021YA,SK515110022YA,SK515110023YA,SK515110024YA,SK515110025YA,SK515110026YA, SK515110027YA,SK515110029YA,SK515110031YA,SK515110032YA,SK515110036YA,SK515110037YA, SK515110038YA,SK515110039YA,SK515110040YA,SK515110042YA,SK515110044YA,SK515110046YA, SK515110048YA,SK515110049YA,SK515110054YA,SK515110055YA,SK515110056YA,SK515110057YA, SK515110058YA,SK515110061YA,SK515110067YA,SK515110068YA,SK515110075YA,SK515110076YA, SK515110077YA,SK515110078YA,SK515110079YA,SK515110080YA,SK515110081YA,SK515110082YA, SK515110083YA,SK515110084YA,SK515110085YA,SK515110086YA,SK515110087YA,SK515110088YA, SK515110089YA,SK515110090YA,SK515110091YA,SK515110092YA,SK515110116YA,SK515110117YA, SK515110118YA,SK515110119YA,SK515110120YA,SK515110122YA,SK515110123YA,SK515110124YA, SK515110125YA,SK515110126YA,SK515110127YA,SK515110128YA,SK515110129YA,SK515110130YA, SK515110131YA,SK515110132YA,SK515110133YA,SK515110134YA,SK515110135YA,SK515110136YA, SK515110138YA,SK515110139YA,SK515110140YA,SK515110141YA,SK515110142YA,SK515110143YA, SK515110144YA,SK515110145YA,SK515110146YA,SK515110147YA,SK515110148YA,SK515110149YA, SK515110150YA,SK515110151YA,SK515110152YA,SK515110153YA,SK515110154YA,SK515110155YA, SK515110156YA,SK515110157YA,SK515110158YA,SK515110159YA,SK515110160YA,SK515110161YA, SK515110162YA,SK515110163YA,SK515110164YA,SK515110165YA,SK515110166YA,SK515110168YA, SK515110169YA,SK515110170YA,SK515110171YA,SK515110172YA,SK515110175YA,SK515110176YA, SK515110177YA,SK515110178YA,SK515110181YA,SK515110197YA,SK515110198YA,SK515110199YA, SK515110200YA,SK515110202YA,SK515110203YA,SK515110204YA,SK515110205YA,SK515110206YA, SK515110207YA,SK515110208YA,SK515110214YA,SK515110219YA,SK515110222YA,SK515120010YA, SK515120018YA,SK515120028YA,SK515120030YA,SK515120031YA,SK515120033YA,SK515120034YA, SK515120035YA,SK515130004YA,SK515130005YA,SK515130006YA,SK515170019YA,SK515170022YA, SK514040029YA,SK514100003YA,SK514100005YA,SK514100006YA,SK514100007YA,SK514100010YA, SK514100011YA,SK514100013YA,SK514100014YA,SK514100015YA,SK514100030YA,SK514100031YA, SK514130002YA,SK514130004YA,SK514130005YA,SK514130010YA,SK514130016YA,SK514480032YA, SK514480033YA,SK514480034YA |
| FEI Number |
3007615597
|
Recalling Firm/ Manufacturer |
Innokas Medical Oy
Vihikari 10
Kempele
Finland
|
Manufacturer Reason for Recall | Monitor may shut down unintentionally without restarting. |
FDA Determined Cause 2 | Software design |
| Action | The firm, Innokas Medical, sent a "Urgent Field Safety Notice(s)" dated October 22, 2015 and October 23, 2015 by e-mail to their consignees. The notice informed the distributor of the device, about the deficiency, and correction of the device. The notice included the instructions regarding what to do with the product and information that the devices will be updated. The consignees were instructed to complete and return the Medical Device FSCA/Recall Return Response Acknowledgement and Receipt Form via fax to +358 8 562 3151, mail to: Innokas Medical Oy, Vihikari 10, FI-90440-Kempele, Finland or email to: feedback@innokasmedical.fi.
If you have any questions, contact the QA/RA Director at +358 44 562 3131. |
| Quantity in Commerce | 424 units (38 in US) |
| Distribution | Worldwide Distribution: US (nationwide) and Europe. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|