| Date Initiated by Firm | December 24, 2015 |
| Date Posted | February 12, 2016 |
| Recall Status1 |
Terminated 3 on February 22, 2017 |
| Recall Number | Z-0806-2016 |
| Recall Event ID |
73086 |
| 510(K)Number | K111862 |
| Product Classification |
Accelerator, linear, medical - Product Code IYE
|
| Product | MRIdian ViewRay Radiation Therapy System, VIEWRAY Inc. |
| Code Information |
Model #: 10000, Serial #'s: 100, 101, 102, 104 & 105 |
| FEI Number |
3007546534
|
Recalling Firm/ Manufacturer |
Viewray Incorporated
2 Thermo Fisher Way
Oakwood Village
OH
44146-6536
|
| For Additional Information Contact | Mr. Garth M. Nobis 650-252-0925 |
Manufacturer Reason for Recall | The system loaded a completion fraction in the incorrect order after a treatment interruption. |
FDA Determined Cause 2 | Software design |
| Action | ViewRay sent an Customer Advisory Notice dated December 24, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Workaround: If you want to move a completion plan to a later date in the Delivery Calendar then do not schedule it until right before you are ready to treat that completion plan.
While we will address this issue in the next release of software we recommend that in the event of a completion plan you always review the plan name loaded on the TPDS and verify the remaining time and beams are as expected with the completion plan.
If you have any questions or concerns regarding this issue, please do not hesitate to contact ViewRay Customer Support at support@viewray.com or call the ViewRay" Customer Support center at 855-286-
8875. |
| Quantity in Commerce | 5 units |
| Distribution | US Distribution to the states of : CA, FL, MO & WI., and Internationally to Republic of Korea. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = IYE
|