• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ConMed Hip Preservation System Signature Series

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ConMed Hip Preservation System Signature Seriessee related information
Date Initiated by FirmDecember 21, 2015
Date PostedMay 31, 2016
Recall Status1 Terminated 3 on May 24, 2018
Recall NumberZ-1842-2016
Recall Event ID 73123
Product Classification Bur, surgical, general & plastic surgery - Product Code GFF
ProductHip Preservation System Signature Series PreBent Burs under the following labels: 1) PreBent Spherical Bur, 5.5mm x 19cm Catalog Number HPS-HB11, 2) PreBent Polishing Bur, 5.5mm x 19cm Catalog Number HPS-HB12 and 3) PreBent Oval Bur, 6.0mm x 19cm Catalog Number HPS-HB13 The ConMed Hip Preservation System Signature Series is intended to be used for resection of osseous tissue (bone) in arthroscopic and endoscopic surgical procedures
Code Information Lot Numbers: 1) 569449, 577583, 577969, 579655, 580365, 581176, 581198, 582216, 587320, 594569, 658219, 669615, 673913 & 688136; 2) 569462, 577800, 577971 & 579652; 3) 569468, 577835, 577974, 579650, 583736, 583742, 587514, 627923, 659317 & 664751.
FEI Number 3001451571
Recalling Firm/
Manufacturer
ConMed Corporation
525 French Rd
Utica NY 13502-5945
Manufacturer Reason
for Recall
Metal shavings released from burs during use are due to contact between the bur and the hood of the PreBent Bur assembly. During the manufacturing process an assembly error occurred which caused inadvertent mixing of two sizes of bearings.
FDA Determined
Cause 2
Mixed-up of materials/components
ActionConMed sent an Urgent Medical Device Recall Notification letter dated December 17, 2015, to all affected customers. Customers are urged to stop use and return these devices immediately. Customers who have further distributed these items are urged to contact all of those organizations immediately and provide a copy of the recall notification letter. It is imperative that all end users of these devices receive the recall notice and respond immediately. Customers with questions can call 315-624-3237 or email largohpshb@conmed.com. The devices should be returned to ConMed Corporation NCR-47057 11155 Concept Blvd. Largo, FL 33773. For questions regarding this recall call 315- 624-3533.
Quantity in Commerce991 pieces
DistributionWorldwide Distribution - Nationwide to CA, CO, FL, IN, MD, MN, NC, NY, OH, TX, UT & VA; International: Canada, Wales
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-