| | Class 2 Device Recall Secure |  |
| Date Initiated by Firm | June 23, 2016 |
| Create Date | July 22, 2016 |
| Recall Status1 |
Terminated 3 on April 11, 2023 |
| Recall Number | Z-2257-2016 |
| Recall Event ID |
74043 |
| Product Classification |
System/device, pharmacy compounding - Product Code NEP
|
| Product | AUTOCOMP6 XPS High Speed Compounder REF 58810 |
| Code Information |
000001, 003002, 003003, 003004, 003005, 003006, 003007, 003008, 003009, 003010, 003011, 003012, 003013, 003014, 003015, 003016, 003017, 003018, 003019, 003020, 003023, 003024, 003025, 003026, 003027, 003028, 003029, 003030, 003031, 003036, 003037, 003038, 003039, 003040, 003041, 003042, 003043, 003044, 003045, 003046, 003047, 003048, 003049, 003050, 003051, 003052, 003053, 003054, 003055, 003056, 003057 |
| FEI Number |
1000117446
|
Recalling Firm/ Manufacturer |
The Metrix Company
4400 Chavenelle Rd
Dubuque
IA
52002-2655
|
| For Additional Information Contact | Jennifer Clasen 563-556-8800 |
Manufacturer Reason for Recall | An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications. |
FDA Determined Cause 2 | Device Design |
| Action | On 6/23/2016, the recalling firm sent written recall notices to their distributors and end users notifying them of the recall. Customers and consignees were instructed to not use the devices, locate and quarantine all affected devices, and to complete the response form. |
| Quantity in Commerce | 16 devices |
| Distribution | Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|