| | Class 2 Device Recall Philips Expression Information Portal (IP5) |  |
| Date Initiated by Firm | September 14, 2014 |
| Create Date | September 21, 2016 |
| Recall Status1 |
Terminated 3 on December 14, 2016 |
| Recall Number | Z-2867-2016 |
| Recall Event ID |
74296 |
| 510(K)Number | K121424 |
| Product Classification |
Monitor,physiological,patient(without arrhythmia detection or alarms) - Product Code MWI
|
| Product | Philips Expression Information Portal (IP5) Model 865471. The IP5 is intended to be used as a secondary display control unit for the Expression MRI Patient Monitoring System and the Expression MR200 MRI Patient Monitoring System. |
| Code Information |
Serial numbers US23200121, US23200122, US23200123, US23200124, US23200125, US23200126, US23200127, US23200128, US23200129, US23200130, US23200131, US23200132, US23200133, US23200134, US23200135, US23200136, US23200137, US23200138, US23200139, US23200140, US23200141, US23200142, US23200143, US23200144, US23200145, US23200146, US23200147, US23200148, US23200149, US23200150, US23200151, US23200152, US23200153, US23200154, US23200155, US23200156, US23200157, US23200158, US23200159, US23200160, US23200161, US23200162, US23200163, US23200164, US23200165, US23200166, US23200167, US23200168, US23200169, US23200170, US23200172, US23200173, US23200176, US23200177, US23200178, US23200179, US23200180, US23200181, US23200182, US23200183, US23200184, US23200188, US23200189, US23200190, US23200191, US23200192, US23200193, US23200194, US23200195, US23200196, US23200197, US23200198, US23200199, US23200200, US23200201, US23200202, US23200203, US23200204, US23200205, US23200206, US23200207, US23200208, US23200209, US23200210, US23200211, US23200212, US23200213, US23200214, US23200215, US23200216, US23200217, US23200218, US23200219, US23200220, US23200221, US23200222, US23200223, US23200224, US23200225, US23200226, US23200227, US23200228, US23200229, US23200230, US23200231, US23200232, US23200233, US23200234, US23200235, US23200236, US23200237, US23200238, US23200239, US23200240, US23200241, US23200242, US23200243, US23200244, US23200245, US23200246, US23200247, US23200248, US23200249, US23200250, US23200251, US23200252, US23200253, US23200254, US23200255, US23200256, US23200258, US23200259, US23200260, US23200261, US23200262, US23200263, US23200264, US23200265, US23200266, US23200267, US23200268, US23200269, US23200270, US23200271, US23200272, US23200273, US23200274, US23200275, US23200276, US23200277, US23200278, US23200279, US23200280, US23200281, US23200282, US23200283, US23200284, US23200285, US23200286, US23200287, US23200288, US23200289, US23200290, US23200291, US23200292, US23200293, US23200294, US23200295, US23200296, US23200297, US23200298, US23200299, US23200300, US23200301, US23200302, US23200303, US23200304, US23200305, US23200306, US23200307, US23200308, US23200309, US23200310, US23200311, US23200312, US23200313, US23200314, US23200315, US23200316, US23200317, US23200318, US23200319, US23200320, US23200321, US23200322, US23200323, US23200324, US23200325, US23200326, US23200327, US23200328, US23200329, US23200330, US23200331, US23200332, US23200333, US23200334, US23200335, US23200336, US23200337, US23200338, US23200339, US23200340, US23200341, US23200342, US23200343, US23200344, US23200345, US23200346, US23200347, US23200349, US23200350, US23200351, US23200352, US23200353, US23200354, US23200356, US23200357, US23200358, US23200359, US23200360, US23200361, US23200362, US23200363, US23200364, US23200365, US23200366, US23200367, US23200368, US23200369, US23200370, US23200371, US23200372, US23200373, US23200374, US23200375, US23200376, US23200377, US23200378, US23200379, US23200380, US23200381, US23200382, US23200383, US23200385, US23200386, US23200387, US23200388, US23200389, US23200390, US23200391, US23200393, US23200394, US23200395, US23200397, US23200398, US23200399, US23200400, US23200401, US23200402, US23200403, US23200405, US23200406, US23200407, US23200408, US23200410, US23200413, US23200415, US23200416, US23200419, US23200421, US23200422, US23200423, US23200424, US23200425, US23200427, US23200428, US23200429, US23200431, US23200434, US34600510, US34600511, US34600512, US34600513, US34600514, US34600515, US34600516, US34600517, US34600518, US34600519, US34600520, US34600521, US34600522, US34600523, US34600524, US34600525, US34600526, US34600528, US34600529, US34600530, US34600531, US34600532, US34600533, US34600534, US34600535, US34600536, US34600537, US34600538, US34600539, US34600540, US34600541, US34600542, US34600543, US34600544, US34600545, US34600546, US34600547, US34600548, US34600549, US34600550, US34600551, US34600552, US34600553, US34600554, US34600555, US34600556, US34600557, US34600558, US34600559, US34600560, US34600561, US34600562, US34600563, US34600564, US34600565, US34600566, US34600567, US34600568, US34600569, US34600570, US34600571, US34600572, US34600573, US34600574, US34600575, US34600576, US34600577, US34600578, US34600579, US34600580, US34600581, US34600582, US34600583, US34600584, US34600585, US34600586, US34600587, US34600588, US34600589, US34600590, US34600591, US34600592, US34600592, US34600593, US34600594, US34600595, US34600596, US34600597, US34600598, US34600599, US34600600, US34600601, US34600602, US34600603, US34600604, US34600605, US34600606, US34600607, US34600608, US34600609, US34600610, US34600611, US34600612, US34600613, US34600614, US34600615, US34600616, US34600617, US34600618, US34600619, US34600620, US34600621, US34600622, US34600623, US34600624, US34600625, US34600626, US34600627, US34600628, US34600629, US34600630, US34600631, US34600632, US34600633, US34600634, US34600635, US34600636, US34600637, US34600638, US34600639, US34600640, US34600641, US34600642, US34600643, US34600644, US34600645, US34600646, US34600647, US34600648, US34600649, US34600650, US34600651, US34600652, US34600653, US34600654, US34600655, US34600656, US34600657, US34600658, US34600659, US34600660, US34600661, US34600662, US34600663, US34600664, US34600665, US34600666, US34600667, US34600668, US34600669, US34600670, US34600671, US34600672, US34600673, US34600674, US34600675, US34600676, US34600677, US34600678, US34600679, US34600680, US34600681, US34600682, US34600683, US34600684, US34600685, US34600686, US34600687, US34600688, US34600689, US34600690, US34600691, US34600692, US34600693, US34600694, US34600695, US34600696, US34600697, US34600698, US34600699, US34600700, US34600701, US34600702, US34600703, US34600704, US34600705, US34600706, US34600707, US34600708, US34600709, US34600710, US34600711, US34600712, US34600713, US34600714, US34600715, US34600716, US34600717, US34600718, US34600719, US34600720, US34600721, US34600722, US34600723, US34600724, US34600729, US34600730, US34600731, US34600732, US34600733, US34600734, US34600735, US34600736, US34600737, US34600738, US34600739, US34600740, US34600741, US34600742, US34600743, US34600744, US34600745, US34600746, US34600747, US34600748, US34600749, US34600750, US34600751, US34600752, US34600753, US34600754, US34600755, US34600756, US34600757, US34600758, US34600759, US34600760, US34600761, US34600762, US34600763, US34600764, US34600765, US34600766, US34600767, US34600768, US34600769, US34600770, US34600771, US34600772, US34600773, US34600774, US34600775, US34600776, US34600777, US34600778, US34600779, US34600780, US34600781, US34600782, US34600783, US34600784, US34600785, US34600786, US34600787, US34600788, US34600789, US34600790, US34600791, US34600792, US34600793, US34600794, US34600795, US34600796, US34600797, US34600798, US34600799, US34600800, US34600801, US34600802, US34600803, US34600804, US34600805, US41600814, US41600815, US41600816, US41600817, US41600818, US41600819, US41600820, US41600821, US41600822, US41600823, US41600824, US41600825, US41600826, US41600827, US41600828, US41600829, US41600830, US41600831, US41600832, US41600833, US41600834, US41600835, US41600836, US41600837, US41600838, US41600839, US41600840, US41600841, US41600842, US41600843, US41600844, US41600845, US41600846, US41600847, US41600848, US41600849, US41600850, US41600851, US41600852, US41600853, US41600854, US41600855, US41600856, US41600857, US41600858, US41600859, US41600860, US41600861, US41600862, US41600863, US41600865, US41600866, US41600867, US41600868, US41600869, US41600870, and US41600871. |
| FEI Number |
1051786
|
Recalling Firm/ Manufacturer |
Invivo Corporation
12151 Research Pkwy Ste 200
Orlando
FL
32826-3286
|
Manufacturer Reason for Recall | Frozen Display Numerics and Disabled Menu Keys after extended run time. This customer notification was sent August 22, 2014. |
FDA Determined Cause 2 | Software in the Use Environment |
| Action | A recall notification letter was sent to hospitals who purchased the recalled product via UPS March 2014. The letter instructed users to power cycle the IP5 unit every day. This is done by pressing the power button on the side of the unit. Customers were instructed to communicate the recall to all those subject and to attach an addendum to the IP5 Instructions for Use. The firm advised customers to call their Philips representative, option 1, then option 2, if they had any questions or concerns. |
| Quantity in Commerce | 639 units |
| Distribution | Worldwide distribution. US nationwide, Australia, Austria, Denmark, France, Germany, Iceland, Ireland, Italy, Netherlands, Norway, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, and England |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = MWI
|
|
|
|