 Class 2 Device Recall BDsee related information
Date Initiated by FirmAugust 02, 2016
Create DateNovember 10, 2016
Recall Status1 Terminated 3 on April 22, 2019
Recall NumberZ-0394-2017
Recall Event ID 75150
Product Classification Anesthesia conduction kit - Product Code CAZ
Productcatalog # 405612 BD Spinal Anesthesia Tray 22G x 3.5 in. BD" Quincke Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog # 405622 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Quincke Spinal Needle Bupivacaine HCL (0.75%) with Dextrose (8.25%), 2 mL (Marcaine) Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog # 405623 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Quincke Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog #405624 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Quincke Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog # 405634 BD Spinal Anesthesia Tray 26G x 3.5 in. BD" Quincke Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog # 405636 BD Spinal Anesthesia Tray 26G x 3.5 in. BD" Quincke Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog #405641 BD Spinal Anesthesia Tray 24G x 3.5 in. Sprotte Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog #405648 BD Spinal Anesthesia Tray 22G x 3.5 in. BD" Whitacre Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog# 405649 BD Spinal Anesthesia Tray 22G x 3.5 in. BD" Whitacre Spinal Needle 25G x 3.5 in. BD" Quincke Spinal Needle Bupivacaine HCL (0.75%) with Dextrose (8.25%), 2 mL (Marcaine) Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog # 405682 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog #405684 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog #405685 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog#405694 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Quincke Spinal Needle 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Bupivacaine HCL (0.75%) with Dextrose (8.25%), 2 mL (Marcaine) Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog #405704 BD Spinal Anesthesia Tray 27G x 3.5 in. BD" Whitacre Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL catalog #405737 BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Bupivacaine HCL (0.75%) with Dextrose (8.25%), 2 mL (Marcaine) catalog# 405829 BD Spinal Anesthesia Tray 27G x 3.5 in. BD" Quincke Spinal Needle 25G x 3.5 in. BD" Whitacre Spinal Needle, TW
Code Information 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL B03H034D 1-May-2015 B04J262D 1-Oct-2016   25G x 3.5 in. BD" Whitacre Spinal Needle, TW Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL B01H034D 1-May-2015 B01H238D 1-Jan-2016 B01H259D 1-Jan-2016 B02J229D 1-Oct-2016 B03J364D 1-Oct-2016 B03K006D 1-Oct-2016 B03K048D 1-Oct-2016 B04H136D 1-Sep-2015  BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL B01H035D 1-May-2015 B01H036D 1-May-2015 B01H070D 1-Jun-2015 B01H091D 1-Jun-2015 B01H138D 1-Sep-2015 B02H031D 1-May-2015 B04H119D 1-Jun-2015  BD Spinal Anesthesia Tray 25G x 3.5 in. BD" Quincke Spinal Needle 25G x 3.5 in. BD" Whitacre Spinal Needle, TW Bupivacaine HCL (0.75%) with Dextrose (8.25%), 2 mL (Marcaine) Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL B01H191D 1-Oct-2016 B02H135D 1-Oct-2016 B03H050D 1-Jul-2016 B03H114D 1-Oct-2016   27G x 3.5 in. BD" Whitacre Spinal Needle Lidocaine HCL (5%) with Dextrose (7.5%), 2 mL B02H034D 1-May-2015   25G x 3.5 in. BD" Whitacre Spinal Needle, TW Bupivacaine HCL (0.75%) with Dextrose (8.25%), 2 mL (Marcaine) B03H153D 1-Sep-2015   27G x 3.5 in. BD" Quincke Spinal Needle 25G x 3.5 in. BD" Whitacre Spinal Needle, TW B02H077D 1-Jun-2015 B03H106D 1-Jun-2015
FEI Number 2243072
Recalling Firm/
Manufacturer
Becton Dickinson & Company
1 Becton Dr
Franklin Lakes NJ 07417-1815
For Additional Information Contact
201-847-6800
Manufacturer Reason
for Recall
A recall was initiated by Hospira, Inc. for several lots of 5% Lidocaine HCI and 7.5% Dextrose Injection, USP, 2mL Ampul, NDC # 0409-4712-01. Some BD" Spinal Trays contain the 5% Lidocaine Hydrochloride and 7.5% Dextrose Injection, USP, 2mL Ampul (NDC Number 0409-4712-01) being recalled by Hospira. As a result, BD is initiating a recall for the affected BD Spinal Trays.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionBD sent recall letters/business reply card was sent via UPS/email on 8/2/2016. Stericycle is handling returns.
Quantity in Commerce39,130 trays
DistributionNationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-