| Date Initiated by Firm | November 11, 2016 |
| Date Posted | December 14, 2016 |
| Recall Status1 |
Terminated 3 on May 05, 2017 |
| Recall Number | Z-0765-2017 |
| Recall Event ID |
75675 |
| 510(K)Number | K062497 |
| Product Classification |
Laparoscope, general & plastic surgery, reprocessed - Product Code NLM
|
| Product | Reprocessed Xcel Bladeless Trocar (H12LP)
Model # H12LP
Reprocessed Endoscopic Trocars are indicated for use to establish a port of entry for endoscopic instruments in patients requiring minimally invasive surgical procedures. |
| Code Information |
Model Lot Manufacture Date Expiration H12LP 3 5 3 6 585 3/3/2015 Mar-18 H12LP 3 5 3 6 880 3/3/2015 Mar-18 H12LP 3 5 3 9 579 3/3/2015 Mar-18 H12LP 3 5 4 0 175 3/5/2015 Mar-18 H12LP 3 5 4 2 016 3/5/2015 Mar-18 H12LP 3 5 4 3 606 3/5/2015 Mar-18 H12LP 3 5 4 3 610 3/9/2015 Mar-18 H12LP 3 5 4 6 115 3/9/2015 Mar-18 H12LP 3 5 4 6 751 3/9/2015 Mar-18 H12LP 3 5 4 6 767 3/9/2015 Mar-18 H12LP 3 5 4 8 518 3/11/2015 Mar-18 H12LP 3 5 4 8 368 3/13/2015 Mar-18 H12LP 3 5 4 9 294 3/13/2015 Mar-18 H12LP 3 5 4 9 583 3/13/2015 Mar-18 H12LP 3 5 5 7 373 3/16/2015 Mar-18 H12LP 3 5 5 9 424 3/16/2015 Mar-18 H12LP 3 5 5 9 453 3/16/2015 Mar-18 H12LP 3 5 6 0 044 3/16/2015 Mar-18 H12LP 3 5 6 0 160 3/16/2015 Mar-18 H12LP 3 5 6 1 254 3/16/2015 Mar-18 H12LP 3 5 6 1 368 3/16/2015 Mar-18 H12LP 3 5 6 1 797 3/16/2015 Mar-18 H12LP 3 5 6 2 147 3/16/2015 Mar-18 H12LP 3 5 6 2 759 3/16/2015 Mar-18 H12LP 3 5 6 2 771 3/16/2015 Mar-18 H12LP 3 5 6 4 013 3/16/2015 Mar-18 H12LP 3 5 6 7 588 3/20/2015 Mar-18 H12LP 3 5 6 9 977 3/20/2015 Mar-18 H12LP 3 5 7 2 483 3/20/2015 Mar-18 H12LP 3 5 7 3 055 3/20/2015 Mar-18 H12LP 3 5 7 4 237 3/23/2015 Mar-18 H12LP 3 5 7 4 758 3/23/2015 Mar-18 H12LP 3 5 7 6 241 3/26/2015 Mar-18 H12LP 3 5 7 6 256 3/26/2015 Mar-18 H12LP 3 5 7 8 827 3/26/2015 Mar-18 H12LP 3 5 7 9 519 3/27/2015 Mar-18 H12LP 3 5 8 1 796 3/27/2015 Mar-18 H12LP 3 5 8 4 936 3/27/2015 Mar-18 H12LP 3 5 8 5 050 3/27/2015 Mar-18 H12LP 3 5 8 5 045 3/31/2015 Mar-18 H12LP 3 5 9 0 552 3/31/2015 Mar-18 H12LP 3 5 9 1 204 3/31/2015 Mar-18 H12LP 3 5 9 5 638 4/6/2015 Apr-18 H12LP 3 6 0 0 132 4/6/2015 Apr-18 H12LP 3 5 9 5 640 4/9/2015 Apr-18 H12LP 3 5 9 8 881 4/9/2015 Apr-18 H12LP 3 6 0 0 653 4/9/2015 Apr-18 H12LP 3 6 1 8 260 4/23/2015 Apr-18 H12LP 3 6 0 0 885 4/28/2015 Apr-18 H12LP 3 6 0 1 108 4/28/2015 Apr-18 H12LP 3 6 2 7 407 4/28/2015 Apr-18 H12LP 3 6 0 0 956 4/29/2015 Apr-18 H12LP 3 6 1 8 738 5/4/2015 May-18 H12LP 3 6 1 8 740 5/4/2015 May-18 H12LP 3 6 2 1 258 5/4/2015 May-18 H12LP 3 6 2 1 270 5/4/2015 May-18 H12LP 3 6 2 1 465 5/4/2015 May-18 H12LP 3 6 2 2 983 5/4/2015 May-18 H12LP 3 6 2 3 066 5/4/2015 May-18 H12LP 3 6 2 5 266 5/4/2015 May-18 H12LP 3 6 2 5 278 5/4/2015 May-18 H12LP 3 6 2 6 334 5/4/2015 May-18 H12LP 3 6 2 6 368 5/4/2015 May-18 H12LP 3 6 2 6 967 5/4/2015 May-18 H12LP 3 6 2 9 293 5/4/2015 May-18 H12LP 3 6 3 0 383 5/4/2015 May-18 H12LP 3 6 2 7 257 5/6/2015 May-18 H12LP 3 6 2 7 411 5/6/2015 May-18 H12LP 3 6 3 0 752 5/6/2015 May-18 H12LP 3 6 3 0 768 5/6/2015 May-18 H12LP 3 6 3 2 514 5/6/2015 May-18 H12LP 3 6 3 3 784 5/6/2015 May-18 H12LP 3 6 3 1 516 5/8/2015 May-18 H12LP 3 6 3 3 922 5/8/2015 May-18 H12LP 3 6 4 2 785 5/8/2015 May-18 H12LP 3 6 4 2 841 5/8/2015 May-18 H12LP 3 6 4 2 650 5/11/2015 May-18 H12LP 3 6 4 3 660 5/11/2015 May-18 H12LP 3 6 4 3 694 5/11/2015 May-18 H12LP 3 6 4 4 701 5/11/2015 May-18 H12LP 3 6 4 6 364 5/11/2015 May-18 H12LP 3 6 4 7 952 5/11/2015 May-18 H12LP 3 6 4 7 968 5/11/2015 May-18 H12LP 3 6 4 8 893 5/11/2015 May-18 H12LP 3 6 4 8 903 5/11/2015 May-18 H12LP 3 6 3 3 933 5/13/2015 May-18 H12LP 3 6 5 1 572 5/13/2015 May-18 H12LP 3 6 5 1 586 5/13/2015 May-18 H12LP 3 6 4 4 729 5/15/2015 May-18 H12LP 3 6 5 4 744 5/15/2015 May-18 H12LP 3 6 5 5 084 5/15/2015 May-18 H12LP 3 6 5 5 097 5/15/2015 May-18 H12LP 3 6 5 6 718 5/15/2015 May-18 H12LP 3 6 5 6 725 5/15/2015 May-18 H12LP 3 6 6 0 049 5/18/2015 May-18 H12LP 3 6 6 2 085 5/18/2015 May-18 H12LP 3 5 9 1 178 5/21/2015 May-18 H12LP 3 5 9 4 312 5/21/2015 May-18 H12LP 3 6 6 2 403 5/21/2015 May-18 H12LP 3 6 6 9 729 5/22/2015 May-18 H12LP 3 6 7 7 361 5/22/2015 May-18 H12LP 3 6 7 8 137 5/22/2015 May-18 H12LP 3 6 7 8 171 5/22/2015 May-18 H12LP 3 6 7 9 918 5/22/2015 May-18 H12LP 3 6 7 9 939 5/27/2015 May-18 H12LP 3 6 8 0 287 5/27/2015 May-18 H12LP 3 6 8 0 959 5/27/2015 May-18 H12LP 3 6 7 2 428 5/28/2015 May-18 H12LP 3 6 8 0 577 5/28/2015 May-18 H12LP 3 6 9 2 959 6/1/2015 Jun-18 H12LP 3 6 9 8 936 6/4/2015 Jun-18 H12LP 3 6 9 8 943 6/4/2015 Jun-18 H12LP 3 7 0 3 025 6/4/2015 Jun-18 H12LP 3 7 0 7 017 6/4/2015 Jun-18 H12LP 3 7 0 8 496 6/5/2015 Jun-18 H12LP 3 7 0 8 546 6/5/2015 Jun-18 H12LP 3 7 1 0 388 6/5/2015 Jun-18 H12LP 3 7 1 0 498 6/5/2015 Jun-18 H12LP 3 7 1 1 263 6/5/2015 Jun-18 H12LP 3 7 1 2 572 6/5/2015 Jun-18 H12LP 3 7 1 2 909 6/5/2015 Jun-18 H12LP 3 7 1 2 913 6/9/2015 Jun-18 H12LP 3 7 1 3 407 6/9/2015 Jun-18 H12LP 3 7 1 3 901 6/9/2015 Jun-18 H12LP 3 7 1 4 983 6/9/2015 Jun-18 H12LP 3 7 1 5 570 6/9/2015 Jun-18 H12LP 3 7 1 5 589 6/9/2015 Jun-18 H12LP 3 7 1 6 638 6/9/2015 Jun-18 H12LP 3 7 1 5 074 6/11/2015 Jun-18 H12LP 3 7 1 7 973 6/11/2015 Jun-18 H12LP 3 7 1 9 225 6/11/2015 Jun-18 H12LP 3 7 2 0 057 6/11/2015 Jun-18 H12LP 3 7 2 2 329 6/15/2015 Jun-18 H12LP 3 7 2 5 181 6/15/2015 Jun-18 H12LP 3 7 2 6 772 6/15/2015 Jun-18 H12LP 3 7 2 7 435 6/15/2015 Jun-18 H12LP 3 7 3 0 715 6/17/2015 Jun-18 H12LP 3 7 3 1 099 6/17/2015 Jun-18 H12LP 3 7 3 2 833 6/17/2015 Jun-18 H12LP 3 7 3 2 851 6/17/2015 Jun-18 H12LP 3 7 3 2 898 6/17/2015 Jun-18 H12LP 3 7 3 7 917 6/19/2015 Jun-18 H12LP 3 7 4 1 225 6/22/2015 Jun-18 H12LP 3 7 4 1 256 6/22/2015 Jun-18 H12LP 3 7 4 2 041 6/22/2015 Jun-18 H12LP 3 7 4 4 436 6/22/2015 Jun-18 H12LP 3 7 4 6 004 6/22/2015 Jun-18 H12LP 3 7 4 8 349 6/24/2015 Jun-18 H12LP 3 7 5 1 993 6/29/2015 Jun-18 H12LP 3 7 5 2 003 6/29/2015 Jun-18 H12LP 3 7 5 8 494 6/29/2015 Jun-18 H12LP 3 7 5 9 302 6/29/2015 Jun-18 H12LP 3 7 5 9 534 6/29/2015 Jun-18 H12LP 3 7 6 0 136 6/30/2015 Jun-18 H12LP 3 7 6 2 495 6/30/2015 Jun-18 H12LP 3 7 6 2 972 6/30/2015 Jun-18 H12LP 3 7 6 5 649 6/30/2015 Jun-18 H12LP 3 7 6 8 045 7/6/2015 Jul-18 H12LP 3 7 6 9 980 7/6/2015 Jul-18 H12LP 3 7 7 1 161 7/6/2015 Jul-18 H12LP 3 7 7 5 668 7/8/2015 Jul-18 H12LP 3 7 7 5 681 7/8/2015 Jul-18 H12LP 3 7 7 9 516 7/8/2015 Jul-18 H12LP 3 7 8 3 776 7/10/2015 Jul-18 H12LP 3 7 8 6 146 7/10/2015 Jul-18 H12LP 3 7 8 6 957 7/10/2015 Jul-18 H12LP 3 7 7 8 024 7/13/2015 Jul-18 H12LP 3 7 8 2 202 7/13/2015 Jul-18 H12LP 3 7 8 4 430 7/13/2015 Jul-18 H12LP 3 7 9 2 212 7/15/2015 Jul-18 H12LP 3 7 9 4 403 7/15/2015 Jul-18 H12LP 3 7 9 4 419 7/15/2015 Jul-18 H12LP 3 7 9 9 917 7/16/2015 Jul-18 H12LP 3 7 9 9 929 7/16/2015 Jul-18 H12LP 3 8 0 0 013 7/16/2015 Jul-18 H12LP 3 8 0 3 761 7/20/2015 Jul-18 H12LP 3 8 0 3 918 7/20/2015 Jul-18 H12LP 3 8 0 4 868 7/20/2015 Jul-18 H12LP 3 8 0 7 840 7/22/2015 Jul-18 H12LP 3 8 0 9 891 7/22/2015 Jul-18 H12LP 3 8 1 0 037 7/22/2015 Jul-18 H12LP 3 8 1 0 783 7/22/2015 Jul-18 H12LP 3 8 1 0 806 7/22/2015 Jul-18 H12LP 3 8 1 1 525 7/23/2015 Jul-18 H12LP 3 8 1 1 778 7/23/2015 Jul-18 H12LP 3 8 1 3 019 7/23/2015 Jul-18 H12LP 3 8 1 3 042 7/23/2015 Jul-18 H12LP 3 8 1 5 917 7/28/2015 Jul-18 H12LP 3 8 1 6 258 7/28/2015 Jul-18 H12LP 3 8 1 7 655 7/29/2015 Jul-18 H12LP 3 8 2 0 257 7/29/2015 Jul-18 H12LP 3 8 2 1 740 7/29/2015 Jul-18 H12LP 3 8 1 8 770 7/30/2015 Jul-18 H12LP 3 8 2 2 395 7/30/2015 Jul-18 H12LP 3 8 2 3 167 7/30/2015 Jul-18 H12LP 3 8 2 8 192 7/30/2015 Jul-18 H12LP 3 8 2 8 222 7/30/2015 Jul-18 H12LP 3 8 3 0 778 7/30/2015 Jul-18 H12LP 3 8 2 9 338 8/5/2015 Aug-18 H12LP 3 8 3 1 629 8/5/2015 Aug-18 H12LP 3 8 3 3 206 8/7/2015 Aug-18 H12LP 3 8 3 4 196 8/7/2015 Aug-18 H12LP 3 8 3 4 203 8/7/2015 Aug-18 H12LP 3 8 4 1 494 8/10/2015 Aug-18 H12LP 3 8 4 3 375 8/10/2015 Aug-18 H12LP 3 8 4 6 698 8/10/2015 Aug-18 H12LP 3 8 4 5 650 8/12/2015 Aug-18 H12LP 3 8 4 8 891 8/12/2015 Aug-18 H12LP 3 8 4 9 655 8/12/2015 Aug-18 H12LP 3 8 5 2 513 8/12/2015 Aug-18 H12LP 3 8 5 3 141 8/17/2015 Aug-18 H12LP 3 8 5 4 774 8/17/2015 Aug-18 H12LP 3 8 6 1 804 8/19/2015 Aug-18 H12LP 3 8 6 3 303 8/21/2015 Aug-18 H12LP 3 8 6 4 693 8/21/2015 Aug-18 H12LP 3 8 7 0 240 8/24/2015 Aug-18 H12LP 3 8 6 6 398 8/26/2015 Aug-18 H12LP 3 8 7 3 578 8/26/2015 Aug-18 H12LP 3 8 7 9 078 8/26/2015 Aug-18 H12LP 3 8 8 1 364 8/26/2015 Aug-18 H12LP 3 8 8 6 277 8/26/2015 Aug-18 H12LP 3 8 8 0 942 8/28/2015 Aug-18 H12LP 3 8 9 0 070 8/28/2015 Aug-18 H12LP 3 8 9 0 089 8/28/2015 Aug-18 H12LP 3 8 9 1 092 8/28/2015 Aug-18 H12LP 3 8 9 4 286 8/31/2015 Aug-18 H12LP 3 8 9 4 678 8/31/2015 Aug-18 H12LP 3 8 9 6 022 8/31/2015 Aug-18 H12LP 3 8 9 7 381 8/31/2015 Aug-18 H12LP 3 9 0 1 817 9/2/2015 Sep-18 H12LP 3 9 1 9 589 9/10/2015 Sep-18 H12LP 3 9 2 3 533 9/10/2015 Sep-18 H12LP 4 0 7 0 594 11/2/2015 Nov-18 H12LP 4 0 7 1 923 11/2/2015 Nov-18 H12LP 4 0 8 6 802 11/6/2015 Nov-18 H12LP 4 1 0 8 634 11/16/2015 Nov-18 H12LP 4 1 0 9 934 11/16/2015 Nov-18 H12LP 4 1 1 3 667 11/20/2015 Nov-18 H12LP 4 1 1 8 604 11/20/2015 Nov-18 H12LP 4 1 1 7 718 11/25/2015 Nov-18 H12LP 4 1 3 3 825 11/25/2015 Nov-18 H12LP 4 1 3 8 769 11/25/2015 Nov-18 H12LP 4 1 4 7 801 11/30/2015 Nov-18 H12LP 3 8 4 9 330 12/3/2015 Dec-18 H12LP 4 1 5 4 685 12/4/2015 Dec-18 H12LP 4 1 5 8 429 12/4/2015 Dec-18 H12LP 4 1 7 0 275 12/9/2015 Dec-18 H12LP 4 1 7 3 170 12/9/2015 Dec-18 H12LP 4 1 7 8 006 12/9/2015 Dec-18 H12LP 4 1 7 9 279 12/9/2015 Dec-18 H12LP 4 1 8 7 693 12/11/2015 Dec-18 H12LP 4 1 9 8 853 12/16/2015 Dec-18 H12LP 4 2 0 2 332 12/16/2015 Dec-18 H12LP 4 2 3 5 633 1/5/2016 Jan-19 H12LP 4 2 4 5 630 1/8/2016 Jan-19 H12LP 4 2 7 4 018 1/18/2016 Jan-19 H12LP 4 2 8 2 444 1/22/2016 Jan-19 H12LP 4 3 8 6 368 3/3/2016 Mar-19 H12LP 4 4 0 1 698 3/9/2016 Mar-19 H12LP 4 5 1 2 285 4/29/2016 Apr-19 H12LP 4 5 1 3 266 4/29/2016 Apr-19 H12LP 4 5 1 5 596 5/2/2016 May-19 H12LP 4 5 1 8 648 5/2/2016 May-19 H12LP 4 5 1 8 607 5/4/2016 May-19 H12LP 4 5 2 4 457 5/4/2016 May-19 H12LP 4 5 2 8 655 5/6/2016 May-19 H12LP 4 5 2 9 566 5/6/2016 May-19 H12LP 4 5 3 0 457 5/6/2016 May-19 H12LP 4 5 3 1 659 5/6/2016 May-19 H12LP 4 5 3 3 116 5/6/2016 May-19 H12LP 4 5 3 9 867 5/9/2016 May-19 H12LP 4 5 4 1 625 5/11/2016 May-19 H12LP 4 5 4 5 105 5/11/2016 May-19 H12LP 4 5 4 5 612 5/11/2016 May-19 H12LP 4 5 4 5 198 5/13/2016 May-19 H12LP 4 5 4 8 729 5/13/2016 May-19 H12LP 4 5 4 8 738 5/13/2016 May-19 H12LP 4 5 5 1 779 5/13/2016 May-19 H12LP 4 5 5 1 780 5/13/2016 May-19 H12LP 4 5 5 7 026 5/16/2016 May-19 H12LP 4 5 6 1 287 5/18/2016 May-19 H12LP 4 5 6 2 934 5/18/2016 May-19 H12LP 4 4 5 8 288 5/20/2016 May-19 H12LP 4 4 5 9 232 5/20/2016 May-19 H12LP 4 4 6 0 302 5/20/2016 May-19 H12LP 4 4 6 6 103 5/20/2016 May-19 H12LP 4 4 6 6 644 5/20/2016 May-19 H12LP 4 4 6 7 337 5/20/2016 May-19 H12LP 4 4 6 9 802 5/20/2016 May-19 H12LP 4 4 7 0 971 5/20/2016 May-19 H12LP 4 4 7 3 956 5/20/2016 May-19 H12LP 4 4 7 3 960 5/20/2016 May-19 H12LP 4 4 7 4 932 5/20/2016 May-19 H12LP 4 4 7 4 945 5/20/2016 May-19 H12LP 4 4 7 6 349 5/20/2016 May-19 H12LP 4 4 7 7 829 5/20/2016 May-19 H12LP 4 4 7 9 578 5/20/2016 May-19 H12LP 4 5 6 6 912 5/20/2016 May-19 H12LP 4 5 6 7 856 5/20/2016 May-19 H12LP 4 5 7 2 487 5/23/2016 May-19 H12LP 4 5 7 4 907 5/23/2016 May-19 H12LP 4 5 7 5 693 5/23/2016 May-19 H12LP 4 5 7 5 705 5/23/2016 May-19 H12LP 4 5 7 1 871 5/25/2016 May-19 H12LP 4 5 7 4 911 5/25/2016 May-19 H12LP 4 5 7 6 086 5/25/2016 May-19 H12LP 4 5 7 6 443 5/25/2016 May-19 H12LP 4 5 7 6 462 5/25/2016 May-19 H12LP 4 5 7 7 392 5/25/2016 May-19 H12LP 4 5 7 7 407 5/25/2016 May-19 H12LP 4 5 8 1 177 5/25/2016 May-19 H12LP 4 5 8 0 431 5/27/2016 May-19 H12LP 4 5 8 1 725 5/27/2016 May-19 H12LP 4 5 8 3 479 5/27/2016 May-19 H12LP 4 5 8 3 936 5/27/2016 May-19 H12LP 4 5 8 6 615 5/27/2016 May-19 H12LP 4 5 8 5 685 6/1/2016 Jun-19 H12LP 4 5 8 9 373 6/1/2016 Jun-19 H12LP 4 5 8 9 677 6/1/2016 Jun-19 H12LP 4 5 9 0 455 6/1/2016 Jun-19 H12LP 4 5 8 9 665 6/3/2016 Jun-19 H12LP 4 5 9 6 066 6/3/2016 Jun-19 H12LP 4 5 9 0 464 6/6/2016 Jun-19 H12LP 4 5 9 3 587 6/6/2016 Jun-19 H12LP 4 5 9 4 649 6/6/2016 Jun-19 H12LP 4 5 9 8 005 6/6/2016 Jun-19 H12LP 4 6 0 0 043 6/6/2016 Jun-19 H12LP 4 6 0 2 957 6/6/2016 Jun-19 H12LP 4 6 0 3 588 6/6/2016 Jun-19 H12LP 4 6 0 3 590 6/6/2016 Jun-19 H12LP 4 5 9 7 997 6/8/2016 Jun-19 H12LP 4 5 9 8 476 6/8/2016 Jun-19 H12LP 4 6 0 7 573 6/8/2016 Jun-19 H12LP 4 6 0 8 496 6/8/2016 Jun-19 H12LP 4 6 0 8 514 6/8/2016 Jun-19 H12LP 4 6 0 9 747 6/8/2016 Jun-19 H12LP 4 6 1 0 175 6/8/2016 Jun-19 H12LP 4 6 1 0 839 6/8/2016 Jun-19 H12LP 4 6 1 1 167 6/8/2016 Jun-19 H12LP 4 6 1 1 179 6/8/2016 Jun-19 H12LP 4 6 1 2 670 6/8/2016 Jun-19 H12LP 4 6 1 0 894 6/10/2016 Jun-19 H12LP 4 6 1 3 359 6/10/2016 Jun-19 H12LP 4 6 1 3 889 6/10/2016 Jun-19 H12LP 4 6 1 4 603 6/10/2016 Jun-19 H12LP 4 6 1 4 912 6/10/2016 Jun-19 H12LP 4 6 1 5 536 6/10/2016 Jun-19 H12LP 4 6 1 8 135 6/10/2016 Jun-19 H12LP 4 6 1 5 246 6/13/2016 Jun-19 H12LP 4 6 1 6 746 6/15/2016 Jun-19 H12LP 4 6 5 8 690 6/29/2016 Jun-19 H12LP 4 6 5 9 478 7/1/2016 Jul-19 H12LP 4 6 5 9 552 7/1/2016 Jul-19 H12LP 4 6 5 9 568 7/1/2016 Jul-19 H12LP 4 6 6 0 191 7/1/2016 Jul-19 H12LP 4 6 6 2 392 7/1/2016 Jul-19 H12LP 4 6 6 2 668 7/1/2016 Jul-19 H12LP 4 6 6 3 276 7/1/2016 Jul-19 H12LP 4 6 6 4 081 7/1/2016 Jul-19 H12LP 4 6 6 4 364 7/5/2016 Jul-19 H12LP 4 6 6 6 763 7/5/2016 Jul-19 H12LP 4 6 6 8 273 7/5/2016 Jul-19 H12LP 4 6 7 2 293 7/5/2016 Jul-19 H12LP 4 6 6 7 846 7/7/2016 Jul-19 H12LP 4 6 6 9 208 7/7/2016 Jul-19 H12LP 4 6 7 0 712 7/7/2016 Jul-19 H12LP 4 6 7 1 055 7/7/2016 Jul-19 H12LP 4 6 6 8 287 7/11/2016 Jul-19 H12LP 4 6 7 0 139 7/11/2016 Jul-19 H12LP 4 6 7 2 065 7/11/2016 Jul-19 H12LP 4 6 7 2 678 7/11/2016 Jul-19 H12LP 4 6 7 3 558 7/11/2016 Jul-19 H12LP 4 6 7 4 512 7/11/2016 Jul-19 H12LP 4 6 7 4 857 7/11/2016 Jul-19 H12LP 4 6 7 2 303 7/12/2016 Jul-19 H12LP 4 6 7 6 800 7/12/2016 Jul-19 H12LP 4 6 8 0 853 7/12/2016 Jul-19 H12LP 4 6 8 1 245 7/12/2016 Jul-19 H12LP 4 6 8 2 434 7/12/2016 Jul-19 H12LP 4 6 6 8 260 7/14/2016 Jul-19 H12LP 4 6 7 8 568 7/14/2016 Jul-19 H12LP 4 6 8 0 352 7/14/2016 Jul-19 H12LP 4 6 8 0 368 7/14/2016 Jul-19 H12LP 4 6 8 1 485 7/14/2016 Jul-19 H12LP 4 6 8 3 268 7/14/2016 Jul-19 H12LP 4 6 8 3 954 7/14/2016 Jul-19 H12LP 4 6 8 4 773 7/14/2016 Jul-19 H12LP 4 6 8 4 917 7/14/2016 Jul-19 H12LP 4 6 8 3 949 7/18/2016 Jul-19 H12LP 4 6 8 4 955 7/18/2016 Jul-19 H12LP 4 6 8 6 200 7/18/2016 Jul-19 H12LP 4 6 8 6 942 7/18/2016 Jul-19 H12LP 4 6 9 4 457 7/18/2016 Jul-19 H12LP 4 5 1 8 669 7/19/2016 Jul-19 H12LP 4 6 8 8 031 7/20/2016 Jul-19 H12LP 4 6 8 8 326 7/20/2016 Jul-19 H12LP 4 6 8 8 887 7/20/2016 Jul-19 H12LP 4 6 8 8 894 7/20/2016 Jul-19 H12LP 4 6 9 0 786 7/20/2016 Jul-19 H12LP 4 6 9 5 707 7/20/2016 Jul-19 H12LP 4 6 9 3 611 7/22/2016 Jul-19 H12LP 4 6 9 5 036 7/22/2016 Jul-19 H12LP 4 6 9 9 014 7/22/2016 Jul-19 H12LP 4 0 7 4 165 7/25/2016 Jul-19 H12LP 4 6 9 3 329 7/25/2016 Jul-19 H12LP 4 6 9 6 019 7/25/2016 Jul-19 H12LP 4 6 9 7 608 7/25/2016 Jul-19 H12LP 4 6 9 8 746 7/25/2016 Jul-19 H12LP 4 7 0 1 597 7/25/2016 Jul-19 H12LP 4 6 9 9 424 7/27/2016 Jul-19 H12LP 4 7 0 0 514 7/27/2016 Jul-19 H12LP 4 7 0 0 522 7/27/2016 Jul-19 H12LP 4 7 0 4 395 7/27/2016 Jul-19 H12LP 4 7 0 4 437 7/27/2016 Jul-19 H12LP 4 7 0 5 381 7/27/2016 Jul-19 H12LP 4 7 0 8 776 7/27/2016 Jul-19 H12LP 4 7 1 1 866 7/28/2016 Jul-19 H12LP 4 7 1 5 840 7/28/2016 Jul-19 H12LP 4 7 2 6 685 7/29/2016 Jul-19 H12LP 4 7 3 0 128 7/29/2016 Jul-19 H12LP 4 7 3 1 690 7/29/2016 Jul-19 H12LP 4 7 4 1 142 7/29/2016 Jul-19 H12LP 4 7 4 0 540 8/1/2016 Aug-19 H12LP 4 7 4 1 157 8/1/2016 Aug-19 H12LP 4 7 4 4 985 8/1/2016 Aug-19 H12LP 4 7 4 5 359 8/1/2016 Aug-19 H12LP 4 7 4 5 889 8/1/2016 Aug-19 H12LP 4 7 4 6 386 8/1/2016 Aug-19 H12LP 4 7 4 9 418 8/1/2016 Aug-19 H12LP 4 7 4 9 402 8/3/2016 Aug-19 H12LP 4 7 5 0 299 8/3/2016 Aug-19 H12LP 4 7 5 1 118 8/3/2016 Aug-19 H12LP 4 7 5 3 896 8/5/2016 Aug-19 H12LP 4 7 5 3 906 8/5/2016 Aug-19 H12LP 4 7 5 7 161 8/5/2016 Aug-19 H12LP 4 7 6 0 422 8/5/2016 Aug-19 H12LP 4 7 6 2 156 8/8/2016 Aug-19 H12LP 4 7 6 6 617 8/8/2016 Aug-19 H12LP 4 7 6 6 751 8/8/2016 Aug-19 H12LP 4 7 4 8 478 8/9/2016 Aug-19 H12LP 4 7 7 2 854 8/11/2016 Aug-19 H12LP 4 7 7 6 931 8/11/2016 Aug-19 H12LP 4 7 7 7 716 8/11/2016 Aug-19 H12LP 4 7 8 6 543 8/12/2016 Aug-19 H12LP 4 7 8 9 098 8/12/2016 Aug-19 H12LP 4 7 9 3 711 8/15/2016 Aug-19 H12LP 4 7 9 9 173 8/17/2016 Aug-19 |
| FEI Number |
3009350690
|
Recalling Firm/ Manufacturer |
Stryker Sustainability Solutions
1810 W Drake Dr
Tempe
AZ
85283-4327
|
| For Additional Information Contact | 760-279-3180 |
Manufacturer Reason for Recall | An increase in reports indicating that the reprocessed H12LP Trocar adjustable plug may break in the packaging or during use. |
FDA Determined Cause 2 | No Marketing Application |
| Action | Stryker sent an Urgent Medical Device Recall letter to all affected customers on November 11, 2016.
Customers were instructed to discontinue use of the affected lots and were asked to follow the link below to complete the Recall Effectiveness Check Form and indicate if any devices from the lots referenced in Attachment 1 remain in their inventory. This form must be completed even if no affected product is found.
www.novasyte.com/sss142
Use the following reference number as you complete the form: [Ship to Account #].
If the form indicates that any affected devices remain in your inventory, a prepaid shipping label will be issued for the return of the product. Credit for all affected devices returned will be issued to your facility.
Address any questions regarding the recall to Novasyte by calling (760) 279-3180 or emailing sss142@novasyte.com. |
| Quantity in Commerce | 40,476 devices |
| Distribution | Worldwide Distribution - US (nationwide) and Internationally to Israel and Canada
Foreign distribution to the following; IL and CA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = NLM
|