| Date Initiated by Firm | January 05, 2017 |
| Date Posted | January 27, 2017 |
| Recall Status1 |
Terminated 3 on August 24, 2017 |
| Recall Number | Z-1107-2017 |
| Recall Event ID |
76165 |
| 510(K)Number | K090846 |
| Product Classification |
Fluorometer, for clinical use - Product Code KHO
|
| Product | Genetic Screening Processor (GSP); PerkinElmer Part Number= 2021-0010
Intended for in vitro quantitative and qualitative determination of analytes in body fluids. |
| Code Information |
Serial Numbers: 20210239 20210243 20210244 20210245 20210246 20210247 20210248 20210249 20210250 20210251 20210252 20210253 20210254 20210255 20210256 20210257 20210258 20210259 20210260 20210261 20210262 20210263 20210264 20210265 |
| FEI Number |
3002808498
|
Recalling Firm/ Manufacturer |
PerkinElmer Life and Analytical Sciences, Wallac, OY
Mustionkatu 6
P.O. Box 10
Turku
Finland
|
Manufacturer Reason for Recall | The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells. |
FDA Determined Cause 2 | Component change control |
| Action | The recall was initiated on January 5, 2017 via email sent through the PerkinElmer Technical Support group located in Akron Ohio.
Actions to be taken by user: 1. As instructed in the analyte specific kit inserts, the use of Kit Controls is required to assure the day-to-day validity of results. Patient results should only be reported if control results for the assay meet the laboratorys established criteria for acceptability. 2. If a movement error occurs customers will be directed to contact their local PerkinElmer field service representative. The movement error may indicate deterioration of the axle and therefore the use of the instrument should be avoided until field service personnel have inspected the status of the ball bushing axles or replaced affected ball bushing axles.
Actions to be taken by the PerkinElmer: As a corrective action, PerkinElmer will replace the affected ball bushing axles in all the GSP instruments identified in this FSCA/Recall. The replacement may take place either as part of the initial installation activity before the instrument is used, scheduled periodic maintenance, or during a dedicated service visit.
This field correction is considered as mandatory by the manufacturer. Customers will be contacted by PerkinElmer to arrange the field service visit. |
| Quantity in Commerce | 24 instruments and 3 spare disk remover modules |
| Distribution | MA, SC, TX, AL
OUS: China, Mexico, Peru, Morocco, Italy, Finland, Ecuador, Brazil |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = KHO
|