Class 2 Device Recall DBS Lead Depth Stop

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall DBS Lead Depth Stopsee related information
Date Initiated by FirmMarch 17, 2017
Create DateAugust 23, 2017
Recall Status1 Terminated 3 on April 22, 2021
Recall NumberZ-3069-2017
Recall Event ID 77839
PMA NumberP960009 
Product Classification Stimulator, electrical, implanted, for parkinsonian tremor - Product Code MHY
ProductDBS Lead Depth Stop contained in Medtronic Deep Brain Stimulation (DBS) lead kits, models 3387/3387S, 3389/3389S, and 3391/3391S, The lead kit contains one DBS lead in addition to multiple accessories used in DBS lead implant procedures. Among these accessories is the DBS lead holder, also referred to as the lead depth stop. The lead holder affixes to the DBS lead during the lead implant procedure, marking the distance to the lead tip and providing control of DBS lead depth during implant by interfacing with the stereotactic system
Code Information Model 3387, Lot Serial #: 0206880294, 0206882647, 0206882779, 0206882780, 0206882781, 0206882782, 0206882783, 0206886184, 0206886323, 0206886328, 0206886330, 0206886331, 0206886332, 0206886333, 0206886603, 0206887813, 0206890443, 0206890444, 0206903488, 0206903489, 0206906792, 0206906793, 0206907159, 0206916391, 0206916392, 0206916393, 0206916394, 0206916395, 0206916396, 0206916397, 0206916398, 0206916399, 0206916797, 0206917120, 0206917121, 0206917122, 0206917123, 0206917124, 0206917125, 0206917126, 0206917127, 0206917128, 0206917129, 0206917130, 0206917131, 0206917132, 0206917302, 0206917303, 0206917304, 0206917306, 0206917307, 0206917308, 0206917310, 0206917312, 0206917560, 0206917561, 0206917563, 0206917564, 0206917565, 0206917566, 0206917568, 0206917569, 0206917570, 0206917572, 0206917573, 0206917574, 0206917575, 0206917576, 0206917577, 0206917578, 0206917579, 0206917620, 0206917621, 0206917622, 0206917623, 0206917624, 0206917625, 0206917626, 0206917627, 0206917628, 0206917629, 0206917630, 0206917631, 0206917632, 0206917633, 0206917634, 0206917635, 0206917636, 0206917637, 0206917638, 0206917639, 0206917660, 0206917661, 0206917662, 0206917663, 0206917664, 0206917665, 0206917666, 0206917667, 0206917668, 0206917669, 0206917670, 0206917671, 0206920443, 0206920444, 0206920445, 0206920447, 0206920448, 0206920449, 0206920450, 0206920451, 0206920452, 0206920453, 0206920454, 0206920455, 0206920456, 0206920457, 0206920458, 0206920459, 0206920460, 0206920461, 0206920462, 0206920463, 0206920663, 0206920664, 0206920665, 0206920666, 0206920667, 0206920668, 0206920669, 0206920670, 0206920671, 0206920672, 0206920673, 0206924158, 0206924492, 0206924493, 0206924494, 0206924604, 0206924605, 0206924606, 0206924607, 0206924608, 0206924609, 0206924610, 0206924611, 0206924612, 0206924613, 0206924614, 0206924615, 0206924616, 0206924617, 0206924618, 0206924619, 0206925140, 0206925141, 0206925142, 0206925143, 0206925245, 0206925290, 0206925291, 0206925292, 0206926111, 0206926112, 0206926414, 0206929827, 0206929828, 0206929829, 0206929830, 0206929849, 0206930146, 0206930736, 0206930737, 0206930941, 0206930942, 0206930943, 0206931910, 0206931911, 0206931927, 0206931928, 0206933408, 0206934345, 0206934346, 0206934347, 0206937919, 0206937920, 0206942506, 0206946796, 0206946798, 0206949034, 0206949035, 0206950164, 0206955498, 0206955499, 0206956649, 0206956650, 0206957515, 0206957516, 0206957517, 0206957518, 0206957519, 0206957801, 0206957802, 0206957803, 0206957804, 0206957805, 0206957806, 0206957807, 0206959205, 0206959206, 0206959207, 0206959208, 0206959209, 0206959210, 0206959211, 0206959212, 0206959213, 0206959214, 0206959215, 0206959216, 0206959217, 0206959218, 0206959220, 0206959221, 0206959222, 0206959223, 0206959224, 0206959501, 0206959502, 0206959503, 0206959682, 0206959684, 0206959686, 0206959687, 0206960104, 0206960105, 0206960106, 0206960107, 0206960108, 0206960109, 0206960147, 0206960148, 0206960149, 0206960655, 0206960658, 0206960666, 0206960667, 0206960668, 0206960669, 0206960670, 0206960671, 0206960676, 0206960677, 0206960678, 0206960679, 0206960680, 0206960681, 0206960694, 0206960695, 0206960696, 0206969559, 0206969560, 0206969561, 0206969562, 0206969563, 0206969564, 0206969565, 0206969566, 0206969567, 0206969568, 0206969569, 0206969570, 0206969571, 0206969572, 0206969573, 0206969574, 0206969575, 0206969576, 0206969577, 0206969578, 0206969691, 0206969692, 0206970927, 0206973200, 0206973201, 0206973202, 0206973203, 0206973204, 0206973205, 0206973206, 0206973800, 0206973801, 0206978789, 0206980541, 0206980542, 0206980543, 0206980970, 0206980971, 0206981705, 0206981706, 0206981707, 0206981714, 0206981715, 0206981754, 0206981771, 0206981772, 0206984645, 0206984646, 0206984650, 0206984675, 0206991945, 0206991946, 0206991947, 0206991948, 0206991949, 0206991950, 0206991951, 0206991952, 0206991953, 0206991954, 0206991955, 0206991956, 0206991957, 0206991958, 0206991959, 0206991960, 0206991961, 0206991962, 0206991963, 0206991964, 0206991965, 0206991966, 0206991967, 0206991968, 0206991969, 0206991970, 0206991971, 0206991972, 0206991973, 0206991974, 0206991975, 0206991976, 0206991977, 0206991978, 0206991979, 0206991984, 0206991985, 0206991986, 0206991987, 0206991988, 0206991989, 0206991990, 0206991991, 0206991992, 0206991993, 0206991994, 0206991995, 0206991996, 0206991997, 0206991998, 0206991999, 0206992000, 0206992001, 0206992002, 0206992003, 0206992004, 0206992005, 0206992006, 0206992007, 0206992008, 0206992009, 0206992010, 0206992011, 0206992012, 0206992013, 0206992014, 0206992015, 0206992016, 0206992017, 0206992018, 0206992019, 0206992020, 0206992021, 0206992022, 0206992023, 0206992024, 0206992025, 0206992026, 0206992027, 0206992028, 0206992029, 0206992030, 0206992031, 0206992032, 0206992033, 0206992034, 0206992035, 0206992036, 0206992037, 0206992038, 0206992039, 0206992040, 0206996540, 0206996541, 0206996542, 0206996543, 0206996544, 0206996545, 0206996546, 0206996547, 0206996548, 0206996549, 0206996550, 0206996551, 0206996552, 0206996553, 0206996554, 0206996555, 0206996556, 0206996557, 0206996558, 0206996559, 0206996560, 0206996561, 0206996562, 0206996563, 0207000225, 0207006520, 0207006777, 0207006778, 0207006779, 0207006780, 0207007919, 0207008220, 0207008221, 0207008222, 0207011212, 0207011213, 0207015958, 0207016168, 0207016796, 0207018882, 0207018883, 0207018884, 0207021792, 0207021793, 0207021794, 0207021795, 0207021796, 0207021797, 0207021798, 0207021799, 0207021800, 0207021801, 0207021802, 0207021803, 0207022120, 0207022121, 0207022122, 0207027753, 0207030645, 0207030646, 0207030647, 0207030648, 0207030649, 0207030650, 0207030651, 0207030652, 0207030653, 0207030654, 0207030655, 0207030656, 0207030657, 0207030658, 0207030659, 0207030780, 0207030781, 0207030980, 0207030981, 0207030982, 0207030983, 0207030985, 0207038190, 0207038192, 0207038195, 0207038423, 0207038424, 0207038425, 0207038426, 0207038427, 0207038434, 0207038435, 0207038436, 0207038437, 0207038438, 0207038439, 0207038440, 0207038441, 0207038442, 0207038443, 0207038457, 0207038458, 0207038459, 0207038460, 0207038461, 0207038462, 0207040184, 0207040185, 0207040187, 0207040188, 0207040189, 0207040190, 0207040191, 0207040239, 0207040240, 0207040241, 0207040242, 0207040243, 0207040244, 0207040573, 0207040574, 0207040575, 0207040576, 0207040712, 0207040713, 0207040714, 0207040715, 0207040718, 0207040719, 0207041016, 0207041811, 0207043628, 0207043629, 0207043630, 0207043631, 0207043632, 0207043658, 0207043659, 0207043660, 0207043871, 0207043872, 0207043874, 0207043875, 0207047746, 0207052711, 0207054947, 0207056625, 0207056626, 0207056628, 0207056629, 0207056630, 0207056631, 0207056807, 0207057256, 0207057257, 0207057427, 0207057428, 0207057429, 0207057430, 0207057431, 0207057432, 0207057433, 0207057434, 0207057435, 0207057436, 0207057437, 0207057438, 0207057439, 0207057440, 0207057441, 0207057442, 0207059495, 0207059565, 0207059566, 0207059567, 0207059956, 0207059959, 0207059963, 0207059985, 0207059986, 0207059987, 0207062240, 0207063038, 0207063039, 0207063040, 0207063133, 0207063134, 0207063135, 0207063136, 0207063137, 0207063138, 0207063139, 0207063140, 0207063141, 0207063142, 0207067134, 0207067135, 0207067136, 0207067137, 0207067138, 0207067139, 0207067140, 0207067141, 0207067142, 0207067143, 0207067146, 0207067147, 0207067148, 0207067847, 0207067848, 0207068325, 0207068326, 0207068327, 0207068449, 0207070300, 0207070321, 0207071704, 0207071705, 0207072960, 0207072961, 0207072962, 0207072963, 0207072964, 0207072965, 0207072966, 0207072967, 0207072968, 0207072969, 0207072970, 0207072971, 0207072972, 0207072973, 0207072974, 0207079003, 0207079004, 0207079339, 0207079340, 0207079341, 0207079342, 0207079343, 0207079414, 0207079415, 0207079416, 0207079417, 0207079418, 0207079419, 0207079435, 0207079436, 0207079437, 0207079438, 0207079439, 0207079440, 0207079441, 0207079467, 0207079468, 0207079469, 0207079470, 0207079471, 0207079472, 0207079473, 0207079474, 0207079475, 0207079476, 0207079477, 0207079478, 0207079479, 0207079480, 0207079481, 0207079482, 0207079483, 0207079484, 0207079485, 0207079486, 0207079487, 0207079488, 0207079489, 0207079490, 0207079491, 0207079492, 0207079493, 0207079494, 0207079495, 0207079496, 0207079497, 0207079498, 0207079641, 0207079642, 0207079643, 0207079645, 0207079680, 0207079681, 0207079701, 0207083910, 0207083911, 0207083912, 0207084903, 0207084904, 0207087901, 0207087902, 0207087903, 0207087904, 0207087905, 0207087906, 0207087907, 0207087908, 0207087909, 0207087910, 0207087951, 0207087952, 0207088003, 0207088004, 0207088005, 0207088431, 0207094140, 0207094142, 0207097722, 0207097723, 0207098161, 0207098162, 0207098163, 0207098164, 0207100512, 0207100514, 0207100515, 0207100516, 0207100517, 0207100810, 0207100848, 0207101008, 0207101009, 0207101010, 0207101023, 0207101031, 0207101053, 0207101120, 0207105162, 0207105163, 0207105164, 0207105165, 0207105166, 0207105167, 0207105168, 0207105169, 0207105170, 0207105171, 0207105172, 0207105176, 0207105321, 0207105322, 0207105323, 0207105324, 0207105325, 0207105326, 0207105327, 0207105328, 0207105329, 0207105330, 0207105331, 0207105332, 0207105333, 0207105334, 0207105335, 0207105336, 0207105352, 0207105356, 0207108624, 0207108625, 0207108635, 0207108636, 0207109103, 0207109704, 0207109705, 0207109706, 0207109708, 0207109709, 0207109710, 0207109711, 0207109712, 0207109713, 0207109714, 0207109715, 0207109716, 0207109717, 0207109718, 0207109719, 0207109720, 0207109725, 0207109726, 0207109728, 0207109729, 0207109733, 0207109734, 0207109737, 0207109815, 0207110541, 0207110542, 0207110543, 0207112954, 0207113814, 0207113815, 0207113816, 0207113817, 0207113818, 0207114006, 0207114007, 0207114746, 0207114852, 0207114853, 0207114854, 0207114855, 0207117588, 0207117589, 0207117590, 0207117741, 0207118073, 0207119315, 0207119316, 0207119317, 0207119318, 0207119319, 0207120110, 0207121868, 0207122061, 0207122062, 0207122063, 0207122064, 0207122065, 0207122066, 0207122067, 0207122068, 0207122069, 0207122070, 0207122071, 0207122072, 0207122099, 0207122100, 0207122101, 0207122127, 0207122128, 0207122129, 0207122183, 0207125639, 0207125641, 0207125642, 0207125643, 0207125754, 0207125851, 0207125876, 0207129909, 0207129910, 0207129912, 0207129913, 0207129917, 0207129920, 0207129921, 0207129922, 0207129923, 0207130018, 0207130019, 0207130020, 0207130021, 0207130022, 0207130023, 0207130024, 0207130025, 0207130026, 0207130027, 0207130028, 0207130029, 0207130081, 0207130299, 0207130300, 0207130301, 0207130302, 0207130303, 0207130304, 0207130305, 0207130306, 0207130307, 0207130308, 0207130724, 0207130786, 0207130787, 0207130788, 0207130789, 0207130790, 0207130845, 0207130846, 0207130847, 0207130848, 0207130849, 0207130850, 0207130851, 0207131732, 0207131734, 0207131757, 0207131758, 0207131759, 0207131780, 0207132032, 0207132033, 0207132098, 0207132677, 0207135004, 0207136742, 0207136743, 0207139918, 0207139919, 0207139920, 0207139921, 0207139922, 0207139923, 0207139924, 0207139925, 0207139926, 0207139927, 0207140714, 0207140715, 0207140716, 0207140717, 0207140718, 0207140719, 0207140720, 0207140724, 0207140725, 0207141032, 0207141033, 0207141034, 0207141035, 0207141036, 0207143663, 0207143678, 0207143679, 0207143680, 0207143681, 0207143682, 0207143683, 0207143697, 0207143698, 0207143700, 0207143701, 0207143707, 0207143708, 0207143712, 0207143713, 0207143714, 0207143715, 0207143716, 0207143717, 0207143718, 0207143719, 0207144809, 0207144817, 0207144818, 0207144819, 0207144840, 0207144841, 0207144842, 0207144843, 0207144844, 0207144846, 0207144847, 0207144848, 0207144883, 0207144884, 0207144885, 0207144886, 0207144887, 0207144888, 0207144889, 0207144890, 0207145866, 0207145867, 0207148304, 0207148305, 0207148306, 0207159258, 0207159259, 0207159260, 0207160611, 0207160612, 0207160614, 0207160615, 0207160678, 0207160679, 0207160680, 0207160708, 0207160709, 0207160710, 0207161115, 0207163879, 0207163881, 0207163882, 0207163883, 0207163884, 0207163887, 0207163889, 0207164673, 0207164674, 0207164675, 0207164676, 0207164677, 0207164678, 0207164870, 0207164871, 0207164872, 0207168561, 0207168562, 0207168563, 0207168564, 0207168565, 0207168566, 0207168567, 0207168568, 0207168569, 0207168570, 0207168571, 0207168572, 0207168573, 0207168574, 0207168575, 0207168576, 0207168577, 0207168595, 0207168596, 0207168597, 0207168598, 0207168599, 0207168600, 0207168929, 0207168930, 0207168931, 0207168932, 0207168954, 0207168955, 0207168956, 0207168957, 0207168959, 0207168961, 0207168962, 0207168963, 0207168965, 0207168966, 0207168967, 0207168968, 0207168969, 0207168970, 0207168972, 0207168979, 0207168980, 0207168985, 0207168986, 0207168987, 0207169010, 0207174625, 0207174627, 0207174628, 0207174629, 0207174630, 0207174654, 0207175291, 0207175292, 0207175293, 0207175294, 0207175295, 0207175296, 0207176745, 0207177151, 0207177152, 0207177153, 0207177154, 0207180633, 0207180634, 0207180635, 0207180636, 0207180637, 0207180638, 0207180639, 0207180640, 0207180641, 0207180642, 0207180643, 0207180644, 0207180645, 0207180646, 0207180647, 0207180648, 0207180649, 0207180650, 0207180651, 0207181930, 0207181931, 0207181932, 0207181933, 0207181934, 0207181987, 0207181988, 0207181989, 0207181990, 0207181991, 0207181992, 0207181993, 0207181994, 0207181995, 0207181996, 0207181997, 0207182077, 0207182782, 0207182783, 0207182785, 0207182786, 0207182787, 0207182788, 0207182790, 0207182791, 0207182793, 0207182794, 0207182795, 0207182796, 0207182797, 0207182798, 0207182799, 0207182840, 0207182841, 0207182842, 0207182843, 0207182844, 0207182846, 0207185517, 0207185518, 0207185519, 0207185520, 0207185521, 0207185522, 0207185523, 0207185524, 0207185525, 0207185636, 0207185651, 0207192753, 0207192754, 0207192790, 0207192791, 0207192815, 0207192816, 0207192817, 0207192818, 0207192819, 0207192820, 0207192821, 0207192822, 0207192823, 0207192824, 0207192825, 0207192826, 0207192827, 0207192828, 0207192829, 0207192910, 0207192911, 0207202710, 0207202711, 0207202712, 0207202717, 0207202718, 0207202719, 0207202740, 0207202741, 0207202742, 0207202743, 0207202744, 0207202745, 0207202746, 0207202747, 0207202748, 0207202750, 0207202754, 0207202755, 0207202756, 0207202757, 0207202758, 0207202764, 0207202765, 0207202766, 0207202801, 0207202805, 0207202932, 0207202936, 0207202943, 0207202944, 0207202961, 0207203570, 0207203571, 0207208667, 0207208668, 0207208714, 0207208715, 0207208716, 0207208723, 0207220085, 0207220740, 0207220741, 0207220742, 0207220743, 0207220745, 0207220746, 0207225509, 0207225523, 0207225564, 0207225565, 0207225566, 0207225567, 0207225568, 0207225569, 0207225570, 0207225571, 0207226531, 0207226532, 0207226533, 0207226534, 0207226535, 0207226626, 0207226627, 0207226628, 0207226629, 0207226630, 0207226631, 0207226632, 0207226633, 0207226634, 0207226635, 0207226636, 0207226637, 0207230154, 0207232091, 0207232126, 0207232266, 0207234900, 0207234901, 0207234902, 0207240828, 0207240832, 0207240833, 0207240834, 0207240841, 0207240842, 0207240843, 0207240844, 0207240845, 0207240856, 0207240857, 0207240858, 0207241071, 0207241072, 0207241073, 0207241444, 0207241445, 0207241455, 0207241456, 0207241457, 0207241458, 0207241459, 0207241460, 0207241461, 0207241463, 0207246705, 0207247848, 0207247849, 0207247850, 0207250567, 0207250707, 0207250708, 0207250710, 0207250711, 0207250712, 0207250713, 0207250714, 0207250715, 0207250716, 0207250717, 0207250718, 0207250719, 0207250720, 0207250721, 0207250722, 0207250723, 0207250724, 0207250725, 0207250726, 0207250727, 0207250728, 0207250729, 0207250730, 0207250731, 0207250809, 0207250810, 0207250812, 0207250813, 0207250814, 0207250815, 0207250816, 0207250817, 0207250818, 0207250829, 0207254564, 0207254565, 0207254742, 0207255595, 0207257987, 0207257988, 0207257989, 0207257990, 0207257991, 0207258020, 0207258045, 0207258259, 0207258280, 0207258638, 0207258639, 0207258640, 0207258641, 0207259128, 0207259174, 0207259175, 0207260433, 0207260434, 0207261780, 0207261788, 0207261789, 0207261790, 0207266491, 0207266492, 0207266530, 0207266531, 0207266532, 0207268177, 0207268178, 0207268179, 0207268231, 0207268232, 0207270938, 0207270939, 0207270972, 0207270973, 0207275052, 0207275053, 0207275054, 0207275058, 0207275218, 0207275219, 0207278567, 0207278568, 0207278569, 0207278570, 0207278572, 0207278573, 0207278574, 0207278575, 0207278576, 0207278577, 0207278578, 0207278579, 0207278639, 0207279750, 0207280151, 0207280180, 0207280181, 0207280200, 0207280201, 0207280202, 0207280203, 0207280204, 0207280205, 0207280206, 0207280207, 0207280208, 0207280209, 0207280210, 0207280211, 0207280217, 0207280218, 0207280219, 0207280220, 0207280221, 0207280222, 0207280223, 0207280224, 0207280225, 0207280226, 0207280227, 0207280228, 0207280229, 0207280230, 0207280231, 0207280232, 0207280233, 0207280238, 0207280239, 0207283309, 0207283310, 0207283614, 0207283627, 0207283628, 0207283629, 0207283630, 0207283631, 0207283632, 0207283638, 0207283639, 0207283841, 0207283883, 0207283884, 0207283885, 0207283886, 0207283903, 0207283944, 0207283945, 0207283946, 0207283947, 0207283948, 0207283949, 0207283950, 0207283951, 0207283952, 0207283953, 0207283954, 0207283955, 0207283956, 0207283957, 0207288214, 0207288219, 0207289211, 0207289212, 0207289214, 0207289215, 0207289216, 0207291547, 0207291548, 0207291549, 0207291686, 0207291687, 0207291688, 0207294519, 0207298853, 0207298854, 0207298855, 0207298856, 0207298857, 0207298858, 0207298859, 0207298996, 0207299404, 0207299920, 0207299921, 0207299922, 0207299923, 0207299924, 0207299925, 0207299926, 0207299927, 0207299928, 0207300520, 0207300521, 0207300522, 0207300523, 0207300699, 0207301220, 0207301221, 0207303119, 0207303499, 0207303641, 0207303656, 0207303885, 0207303886, 0207304245, 0207304800, 0207304801, 0207305118, 0207305119, 0207306460, 0207308323, 0207308324, 0207308325, 0207308326, 0207308327, 0207308328, 0207309474, 0207309475, 0207309476, 0207309477, 0207309478, 0207309479, 0207309480, 0207309481, 0207309482, 0207309483, 0207309484, 0207309485, 0207309486, 0207309487, 0207309488, 0207309489, 0207309490, 0207309491, 0207309492, 0207309493, 0207309494, 0207309495, 0207309496, 0207309497, 0207309498, 0207309499, 0207309500, 0207309501, 0207309502, 0207309505, 0207309506, 0207309507, 0207309508, 0207309516, 0207309517, 0207309518, 0207309519, 0207309520, 0207309521, 0207309522, 0207309523, 0207309524, 0207309525, 0207309526, 0207309527, 0207309530, 0207309531, 0207309532, 0207309533, 0207309534, 0207309535, 0207309536, 0207319594, 0207324654, 0207324655, 0207324656, 0207324657, 0207324658, 0207324659, 0207324680, 0207324681, 0207324682, 0207324683, 0207324684, 0207324685, 0207324686, 0207324687, 0207324688, 0207324689, 0207324690, 0207324691, 0207324692, 0207324693, 0207324694, 0207324695, 0207324696, 0207324697, 0207324698, 0207324699, 0207324700, 0207326089, 0207326090, 0207326091, 0207326131, 0207326132, 0207327178, 0207337977, 0207342360, 0207348166, 0207348167, 0207348168, 0207348211, 0207348212, 0207348214, 0207348215, 0207348216, 0207348217, 0207348218, 0207348583, 0207350341, 0207353632, 0207353633, 0207353634, 0207353751, 0207353898, 0207359405, 0207359406, 0207359407, 0207359408, 0207359409, 0207359410, 0207359411, 0207359412, 0207359413, 0207359414, 0207359415, 0207359416, 0207359417, 0207359418, 0207359419, 0207360423, 0207360424, 0207360425, 0207360426, 0207360427, 0207360428, 0207360429, 0207360430, 0207360431, 0207360432, 0207360433, 0207360434, 0207360435, 0207360436, 0207360437, 0207360438, 0207360439, 0207361120, 0207361121, 0207361122, 0207361123, 0207361280, 0207361281, 0207361282, 0207361283, 0207361284, 0207361285, 0207361286, 0207361287, 0207361348, 0207368928, 0207368929, 0207368930, 0207368931, 0207368932, 0207368933, 0207368934, 0207368935, 0207372779, 0207372793, 0207372892, 0207372893, 0207372895, 0207372896, 0207379102, 0207379355, 0207379356, 0207379405, 0207379406, 0207380120, 0207380121, 0207380122, 0207380123, 0207385226, 0207396642, 0207396707, 0207396708, 0207397023, 0207397864, 0207397880, 0207397887, 0207397888, 0207397890, 0207397892, 0207397896, 0207397897, 0207397898, 0207397899, 0207397900, 0207397901, 0207397902, 0207397904, 0207397905, 0207397906, 0207397907, 0207397956, 0207401852, 0207401853, 0207404964, 0207407581, 0207409006, 0207409007, 0207409008, 0207409009, 0207413918, 0207413919, 0207414060, 0207414061, 0207414062, 0207414063, 0207414064, 0207414065, 0207414066, 0207414067, 0207419438, 0207422718, 0207424300, 0207424302, 0207424303, 0207424304, 0207424305, 0207424756, 0207428968, 0207428969, 0207428970, 0207428971, 0207428972, 0207428973, 0207430334, 0207430399, 0207430403, 0207430404, 0207430405, 0207432103, 0207432104, 0207432105, 0207432666, 0207437724, 0207441876, 0207444058, 0207444059, 0207444300, 0207446984, 0207448988, 0207448989, 0207448990, 0207448991, 0207448992, 0207448996, 0207449981, 0207449982, 0207449993, 0207449994, 0207449995, 0207449996, 0207449997, 0207449998, 0207454223, 0207454224, 0207454225, 0207454226, 0207454227, 0207454228, 0207454229, 0207454401, 0207454402, 0207454403, 0207454404, 0207454405, 0207454406, 0207454407, 0207454408, 0207454409, 0207456279, 0207457340, 0207457341, 0207457342, 0207457343, 0207457344, 0207457345, 0207457346, 0207457347, 0207457348, 0207457349, 0207457350, 0207457351, 0207457352, 0207457353, 0207457354, 0207457355, 0207457356, 0207457357, 0207457358, 0207457359, 0207457360, 0207457361, 0207457362, 0207457363, 0207457364, 0207457365, 0207457366, 0207457367, 0207457368, 0207457369, 0207457370, 0207457371, 0207457372, 0207457373, 0207457374, 0207457375, 0207457376, 0207457377, 0207457378, 0207457379, 0207457380, 0207457381, 0207457382, 0207457383, 0207457384, 0207457385, 0207457386, 0207457387, 0207457388, 0207457389, 0207457390, 0207457391, 0207457392, 0207457393, 0207457394, 0207457395, 0207457396, 0207457397, 0207457398, 0207457399, 0207457400, 0207457401, 0207457402, 0207457403, 0207457404, 0207457405, 0207457406, 0207457407, 0207457408, 0207457409, 0207457410, 0207457411, 0207457412, 0207457413, 0207457414, 0207457415, 0207457416, 0207457417, 0207457418, 0207457419, 0207457420, 0207457421, 0207457422, 0207460357, 0207460358, 0207460359, 0207460360, 0207460361, 0207460871, 0207460872, 0207466531, 0207466532, 0207466591, 0207466592, 0207466593, 0207466594, 0207466595, 0207466596, 0207466597, 0207466598, 0207466599, 0207466600, 0207466601, 0207466602, 0207466603, 0207466604, 0207466605, 0207466606, 0207466607, 0207466608, 0207466674, 0207469251, 0207469252, 0207471095, 0207471096, 0207474219, 0207498243, 0207498244, 0207498245, 0207498246, 0207498247, 0207498250, 0207498251, 0207507151, 0207507152, 0207507153, 0207507154, 0207507447, 0207507448, 0207507449, 0207520659, 0207523074, 0207526687, 0207531718, 0207531719, 0207531720, 0207534092, 0207534093, 0207534573, 0207534575, 0207536498, 0207536499, 0207536500, 0207536501, 0207536503, 0207536504, 0207536505, 0207536506, 0207538329, 0207538744, 0207538745, 0207541018, 0207546649, 0207546650, 0207546651, 0207546652, 0207546653, 0207546654, 0207546655, 0207546656, 0207546657, 0207546658, 0207547088, 0207547089, 0207547090, 0207547091, 0207547092, 0207547093, 0207547094, 0207547095, 0207547096, 0207547097, 0207547098, 0207547099, 0207547100, 0207547101, 0207547102, 0207547111, 0207547112, 0207547113, 0207547114, 0207547115, 0207547116, 0207547117, 0207547118, 0207547119, 0207547120, 0207547121, 0207547122, 0207547123, 0207547311, 0207548014, 0207548015, 0207548639, 0207548883, 0207548884, 0207548885, 0207548886, 0207549280, 0207550658, 0207554148, 0207555823, 0207561059, 0207561060, 0207567863, 0207574365, 0207585359, 0207588282, 0207588640, 0207588643, 0207588644, 0207588645, 0207588672, 0207588674, 0207588808, 0207588809, 0207588810, 0207588811, 0207588860, 0207588861, 0207588865, 0207588866, 0207592741, 0207595000, 0207595001, 0207595002, 0207595003, 0207595017, 0207598167, 0207598168, 0207598169, 0207600533, 0207600534, 0207600535, 0207601075, 0207602702, 0207604544, 0207605819, 0207606004, 0207609972, 0207609973, 0207609974, 0207609975, 0207609976, 0207609977, 0207609978, 0207611974, 0207611975, 0207612377, 0207615806, 0207615807, 0207615808, 0207615809, 0207615810, 0207615811, 0207615812, 0207615813, 0207615814, 0207615815, 0207615816, 0207615817, 0207615818, 0207615819, 0207615835, 0207615836, 0207620203, 0207620204, 0207620205, 0207620206, 0207620207, 0207620364, 0207620790, 0207620791, 0207620793, 0207620794, 0207620795, 0207620796, 0207621611, 0207621612, 0207621613, 0207621614, 0207621615, 0207621616, 0207621617, 0207621618, 0207621619, 0207621620, 0207621621, 0207621622, 0207621623, 0207621624, 0207621625, 0207621626, 0207621627, 0207621628, 0207621629, 0207621630, 0207621631, 0207621632, 0207621633, 0207621634, 0207621635, 0207621636, 0207621637, 0207621638, 0207621639, 0207621640, 0207621641, 0207621642, 0207621643, 0207621644, 0207621645, 0207621646, 0207621647, 0207621648, 0207621649, 0207621650, 0207621651, 0207621652, 0207621653, 0207621654, 0207621830, 0207621831, 0207622044, 0207622045, 0207622169, 0207622176, 0207622177, 0207622178, 0207622179, 0207623055, 0207625459, 0207627667, 0207627668, 0207627669, 0207627690, 0207627691, 0207627692, 0207627693, 0207627694, 0207627695, 0207627696, 0207627949, 0207628172, 0207631094, 0207632116, 0207634225, 0207634233, 0207637637, 0207642804, 0207642807, 0207642813, 0207647090, 0207647249, 0207651194, 0207651195, 0207651196, 0207651197, 0207651352, 0207652932, 0207652933, 0207653543, 0207660129, 0207660130, 0207660131, 0207660132, 0207660133, 0207660134, 0207660135, 0207660136, 0207660171, 0207660878, 0207664645, 0207664646, 0207664647, 0207670273, 0207670274, 0207670278, 0207670279, 0207674827, 0207674828, 0207674829, 0207681428, 0207682857, 0207682886, 0207682888, 0207682890, 0207682892, 0207682903, 0207682904, 0207682906, 0207682908, 0207682909, 0207682913, 0207682917, 0207682918, 0207682940, 0207682942, 0207682944, 0207682946, 0207682948, 0207682950, 0207682951, 0207682953, 0207682955, 0207682957, 0207682959, 0207682960, 0207682962, 0207682964, 0207683075, 0207683076, 0207683077, 0207683078, 0207683079, 0207683100, 0207683101, 0207683103, 0207683104, 0207683105, 0207683106, 0207683108, 0207683109, 0207683110, 0207683111, 0207683112, 0207683113, 0207683114, 0207683115, 0207683116, 0207683117, 0207684419, 0207684662, 0207684663, 0207685023, 0207685138, 0207685139, 0207685454, 0207687157, 0207687571, 0207687572, 0207692231, 0207692232, 0207694357, 0207699502, 0207699503, 0207699504, 0207699507, 0207700963, 0207704751, 0207704753, 0207704959, 0207704960, 0207705029, 0207705030, 0207705031, 0207705032, 0207705033, 0207705034, 0207705035, 0207705036, 0207705041, 0207705042, 0207705075, 0207708402, 0207710979, 0207711007, 0207720013, 0207720602, 0207721774, 0207722086, 0207729498, 0207729499, 0207730540, 0207730541, 0207730542, 0207730543, 0207730544, 0207730545, 0207730546, 0207730547, 0207730548, 0207730549, 0207730550, 0207731594, 0207731595, 0207733668, 0207733678, 0207733695, 0207733696, 0207733697, 0207733698, 0207733699, 0207733819, 0207734053, 0207734799, 0207735690, 0207737625, 0207737626, 0207737627, 0207737628, 0207737629, 0207737630, 0207737838, 0207737839, 0207742414, 0207742415, 0207742416, 0207742417, 0207742418, 0207742419, 0207742420, 0207742421, 0207742422, 0207742423, 0207742452, 0207742453, 0207742454, 0207742455, 0207742456, 0207742457, 0207742458, 0207742459, 0207743128, 0207747304, 0207747305, 0207747306, 0207747307, 0207747308, 0207747309, 0207747417, 0207747418, 0207747419, 0207750080, 0207752765, 0207752766, 0207752767, 0207752768, 0207752769, 0207752770, 0207752771, 0207752772, 0207756643, 0207756644, 0207756645, 0207756646, 0207756648, 0207756716, 0207756717, 0207756718, 0207756719, 0207756720, 0207756721, 0207756722, 0207756757, 0207757244, 0207757245, 0207757246, 0207757247, 0207757352, 0207757353, 0207757354, 0207757355, 0207757356, 0207757357, 0207757358, 0207757359, 0207757361, 0207757362, 0207757451, 0207757452, 0207757453, 0207761819, 0207761900, 0207762185, 0207762186, 0207762187, 0207762188, 0207762189, 0207762190, 0207762191, 0207762192, 0207762193, 0207762194, 0207762195, 0207762196, 0207762197, 0207762198, 0207762199, 0207762200, 0207762201, 0207762202, 0207762203, 0207762204, 0207762205, 0207762206, 0207762207, 0207762208, 0207762209, 0207762210, 0207762211, 0207762212, 0207762213, 0207762214, 0207762215, 0207762216, 0207762217, 0207762218, 0207762681, 0207762682, 0207762683, 0207762684, 0207762722, 0207762723, 0207764452, 0207767622, 0207767623, 0207767624, 0207767625, 0207767626, 0207767627, 0207767628, 0207767629, 0207767928, 0207767929, 0207768006, 0207768007, 0207768008, 0207768009, 0207768010, 0207768011, 0207768012, 0207768013, 0207769247, 0207769248, 0207769249, 0207769250, 0207769251, 0207769252, 0207769253, 0207770721, 0207770722, 0207770723, 0207773147, 0207773148, 0207773149, 0207773150, 0207773151, 0207777434, 0207777435, 0207777436, 0207777437, 0207777438, 0207777439, 0207777440, 0207777441, 0207777442, 0207777443, 0207777444, 0207777445, 0207777446, 0207777447, 0207777878, 0207778540, 0207778541, 0207780669, 0207780670, 0207780671, 0207780672, 0207780673, 0207780761, 0207780762, 0207785216, 0207787723, 0207787724, 0207787725, 0207787726, 0207787727, 0207787728, 0207787729, 0207789155, 0207795322, 0207795323, 0207795324, 0207800432, 0207800433, 0207800481, 0207800482, 0207800483, 0207800484, 0207800485, 0207800486, 0207800487, 0207800494, 0207800495, 0207800496, 0207800497, 0207800498, 0207800500, 0207800509, 0207800510, 0207800511, 0207800512, 0207800655, 0207806145, 0207810062, 0207815195, 0207815196, 0207815197, 0207815198, 0207815211, 0207815225, 0207817620, 0207820647, 0207820648, 0207820726, 0207828400, 0207828401, 0207836043, 0207836044, 0207836119, 0207836120, 0207836121, 0207836122, 0207836123, 0207836201, 0207836212, 0207836213, 0207836214, 0207836215, 0207836247, 0207836248, 0207836249, 0207836250, 0207836251, 0207840219, 0207845496, 0207845535, 0207845537, 0207845549, 0207845550, 0207845689, 0207845691, 0207845692, 0207845706, 0207845707, 0207845708, 0207845709, 0207845710, 0207845711, 0207845712, 0207845713, 0207845726, 0207845748, 0207845749, 0207845750, 0207845882, 0207845883, 0207852426, 0207852427, 0207852428, 0207852429, 0207852430, 0207852431, 0207852432, 0207852433, 0207852434, 0207852435, 0207852436, 0207852437, 0207852868, 0207856675, 0207856676, 0207856677, 0207856678, 0207858516, 0207858517, 0207858538, 0207861043, 0207861044, 0207861705, 0207864175, 0207864176, 0207864177, 0207864361, 0207864362, 0207864363, 0207864364, 0207864365, 0207864366, 0207864367, 0207864368, 0207864369, 0207864370, 0207864371, 0207864372, 0207864383, 0207864384, 0207864385, 0207864398, 0207864828, 0207864829, 0207866171, 0207866172, 0207866173, 0207866174, 0207866326, 0207868463, 0207868464, 0207868465, 0207868466, 0207868467, 0207868468, 0207868469, 0207868470, 0207868471, 0207868472, 0207868473, 0207868474, 0207869975, 0207869976, 0207869977, 0207874443, 0207874444, 0207874863, 0207874864, 0207874865, 0207874866, 0207874867, 0207874868, 0207874869, 0207874870, 0207874871, 0207874872, 0207874873, 0207874874, 0207874875, 0207874876, 0207875002, 0207877664, 0207893315, 0207893316, 0207893619, 0207893620, 0207893621, 0207893622, 0207893623, 0207894148, 0207898073, 0207898074, 0207908591, 0207908592, 0207908593, 0207908669, 0207908670, 0207908671, 0207911576, 0207911942, 0207911943, 0207914091, 0207914191, 0207914192, 0207914193, 0207915278, 0207915279, 0207915400, 0207915401, 0207915402, 0207915403, 0207915404, 0207915405, 0207915406, 0207915407, 0207915408, 0207915409, 0207915410, 0207915411, 0207915412, 0207915413, 0207915414, 0207915415, 0207915416, 0207915417, 0207915418, 0207915419, 0207915420, 0207915421, 0207915422, 0207915423, 0207915424, 0207915425, 0207915426, 0207915427, 0207915428, 0207915429, 0207915430, 0207915431, 0207915432, 0207915433, 0207915434, 0207915718, 0207915719, 0207915751, 0207915752, 0207915753, 0207915754, 0207917203, 0207917204, 0207920752, 0207920849, 0207928098, 0207928099, 0207930499, 0207931724, 0207931725, 0207931726, 0207931727, 0207931728, 0207931729, 0207931730, 0207931731, 0207931732, 0207932178, 0207932179, 0207932420, 0207932422, 0207932423, 0207932425, 0207932922, 0207932931, 0207932932, 0207932933, 0207932936, 0207933359, 0207933620, 0207933621, 0207933622, 0207933623, 0207936872, 0207936873, 0207939381, 0207939382, 0207946682, 0207946683, 0207946685, 0207946687, 0207953913, 0207954846, 0207954854, 0207954855, 0207954856, 0207954857, 0207954858, 0207954859, 0207954960, 0207954961, 0207954962, 0207954963, 0207954964, 0207954965, 0207954966, 0207954967, 0207954968, 0207954969, 0207954970, 0207954975, 0207954976, 0207954977, 0207954978, 0207955290, 0207955291, 0207955292, 0207955293, 0207955294, 0207955295, 0207955296, 0207955377, 0207955378, 0207956234, 0207959140, 0207959141, 0207959790, 0207962083, 0207977529, 0207977531, 0207977554, 0207977559, 0207977561, 0207977563, 0207979721, 0207983410, 0207983411, 0207989763, 0207990251, 0207990252, 0207997018, 0207997019, 0207997020, 0207997021, 0207997022, 0207997023, 0207997024, 0207997025, 0207997026, 0207997027, 0207997028, 0207997029, 0207997030, 0207997031, 0207997032, 0207997033, 0207999254, 0208002305, 0208002306, 0208002406, 0208002445, 0208003966, 0208003967, 0208003968, 0208005256, 0208010165, 0208010166, 0208010167, 0208010765, 0208010766, 0208010767, 0208011663, 0208011777, 0208011778, 0208011779, 0208014584, 0208014585, 0208014690, 0208024166, 0208025332, 0208025333, 0208025334, 0208025953, 0208028973, 0208043346, 0208043347, 0208043348, 0208043349, 0208043356, 0208043357, 0208043522, 0208043523, 0208043524, 0208043911, 0208043920, 0208043921, 0208043922, 0208043936, 0208043937, 0208043961, 0208043962, 0208047388, 0208051576, 0208052427, 0208052428, 0208052429, 0208052663, 0208052664, 0208052839, 0208057180, 0208057831, 0208062216, 0208062217, 0208062218, 0208062219, 0208063420, 0208063422, 0208063423, 0208063426, 0208063427, 0208063428, 0208063429, 0208063430, 0208063431, 0208063432, 0208063433, 0208063434, 0208063435, 0208063436, 0208064011, 0208064012, 0208067770, 0208067771, 0208067772, 0208068436, 0208072524, 0208072525, 0208075029, 0208081024, 0208081025, 0208081026, 0208081027, 0208081028, 0208081029, 0208081030, 0208081031, 0208081033, 0208081034, 0208081035, 0208081036, 0208081126, 0208081127, 0208081128, 0208081129, 0208081130, 0208081459, 0208081460, 0208081466, 0208089190, 0208089191, 0208089192, 0208094688, 0208098151, 0208099997, 0208099998, 0208102070, 0208102336, 0208108170, 0208108171, 0208108172, 0208108173, 0208108174, 0208108175, 0208108176, 0208108177, 0208108178, 0208108366, 0208108367, 0208108368, 0208108369, 0208108370, 0208116602, 0208116603, 0208116604, 0208116605, 0208116606, 0208116666, 0208116667, 0208116668, 0208117530, 0208117531, 0208122468, 0208122469, 0208122470, 0208122471, 0208122472, 0208122622, 0208122623, 0208131550, 0208131551, 0208131741, 0208132220, 0208135681, 0208136763, 0208136764, 0208136765, 0208142174, 0208142175, 0208142176, 0208143008, 0208143009, 0208143053, 0208143054, 0208143055, 0208146369, 0208146370, 0208146371, 0208146515, 0208146610, 0208146611, 0208147136, 0208147137, 0208147532, 0208148901, 0208148902, 0208153989, 0208153990, 0208154524, 0208154533, 0208154556, 0208154557, 0208154558, 0208154571, 0208154572, 0208154574, 0208154575, 0208154576, 0208154577, 0208154645, 0208154647, 0208154648, 0208154932, 0208155696, 0208155697, 0208157689, 0208157690, 0208165718, 0208165719, 0208168147, 0208168148, 0208168202, 0208168203, 0208168204, 0208168205, 0208168265, 0208168266, 0208168267, 0208168373, 0208172598, 0208173010, 0208173011, 0208173012, 0208177529, 0208177530, 0208177593, 0208177594, 0208177595, 0208177596, 0208177661, 0208177662, 0208177663, 0208179236, 0208179237, 0208184327, 0208184328, 0208192606, 0208192923, 0208192926, 0208192927, 0208192928, 0208193015, 0208193017, 0208193080, 0208193083, 0208193085, 0208193087, 0208193089, 0208195084, 0208195085, 0208195086, 0208196288, 0208196289, 0208196290, 0208196291, 0208198855, 0208198856, 0208198915, 0208198916, 0208198917, 0208198919, 0208198980, 0208198981, 0208198982, 0208199153, 0208199154, 0208199155, 0208199170, 0208200284, 0208202772, 0208202773, 0208203643, 0208203867, 0208204728, 0208204729, 0208204730, 0208207836, 0208207837, 0208208698, 0208209147, 0208209148, 0208219894, 0208220884, 0208220885, 0208220886, 0208220887, 0208220921, 0208220923, 0208220924, 0208220925, 0208220926, 0208220927, 0208220928, 0208220929, 0208220930, 0208220931, 0208220932, 0208220933, 0208220934, 0208221259, 0208221300, 0208221313, 0208221314, 0208221315, 0208221316, 0208221382, 0208221736, 0208221737, 0208222358, 0208222710, 0208222879, 0208224281, 0208224381, 0208224382, 0208224383, 0208224384, 0208224385, 0208224386, 0208224387, 0208224388, 0208224390, 0208224391, 0208224392, 0208224393, 0208224394, 0208224395, 0208224396, 0208224397, 0208224398, 0208224399, 0208224440, 0208224441, 0208224442, 0208224443, 0208224444, 0208224445, 0208227573, 0208227574, 0208232213, 0208232214, 0208234582, 0208237314, 0208237358, 0208237359, 0208237541, 0208237542, 0208237663, 0208237664, 0208237710, 0208237711, 0208238379, 0208238398, 0208238399, 0208238400, 0208238401, 0208238402, 0208238420, 0208238428, 0208240463, 0208246974, 0208246975, 0208246976, 0208246977, 0208246978, 0208246979, 0208247040, 0208247041, 0208247042, 0208247043, 0208247044, 0208247045, 0208247046, 0208247047, 0208247048, 0208247049, 0208247050, 0208247051, 0208247121, 0208247122, 0208247123, 0208247124, 0208247125, 0208247127, 0208247128, 0208247129, 0208247130, 0208247131, 0208247132, 0208247133, 0208247134, 0208247135, 0208247136, 0208247137, 0208247138, 0208247139, 0208247205, 0208248478, 0208248479, 0208249070, 0208249071, 0208249072, 0208251143, 0208252171, 0208252172, 0208256960, 0208256961, 0208256962, 0208256963, 0208256964, 0208256978, 0208256979, 0208257950, 0208257951, 0208257952, 0208257953, 0208257954, 0208258086, 0208262892, 0208262893, 0208262894, 0208262895, 0208262896, 0208262897, 0208262898, 0208262899, 0208264020, 0208264021, 0208264022, 0208264023, 0208264024, 0208264025, 0208264026, 0208264027, 0208264028, 0208264029, 0208264030, 0208264031, 0208264032, 0208264033, 0208264034, 0208264035, 0208264036, 0208264037, 0208264038, 0208264039, 0208264040, 0208264041, 0208264042, 0208264043, 0208264044, 0208264045, 0208264046, 0208266135, 0208266708, 0208266709, 0208266710, 0208266711, 0208266712, 0208266713, 0208266714, 0208266715, 0208266716, 0208270615, 0208270616, 0208270617, 0208270619, 0208274125, 0208274127, 0208289547, 0208289922, 0208290047, 0208290048, 0208290049, 0208290050, 0208290051, 0208290052, 0208290053, 0208290054, 0208290055, 0208290056, 0208290057, 0208290058, 0208290059, 0208290061, 0208290062, 0208290063, 0208290064, 0208290065, 0208290066, 0208290069, 0208290072, 0208290073, 0208290074, 0208290076, 0208290223, 0208290320, 0208290321, 0208290322, 0208290323, 0208290324, 0208290326, 0208290327, 0208290328, 0208290329, 0208290330, 0208290331, 0208290332, 0208290333, 0208290334, 0208290335, 0208290336, 0208290438, 0208290439, 0208290600, 0208290601, 0208290602, 0208290603, 0208290604, 0208290605, 0208290606, 0208290607, 0208290608, 0208290609, 0208290610, 0208290611, 0208290612, 0208290613, 0208290614, 0208290615, 0208290616, 0208290617, 0208290618, 0208290619, 0208290680, 0208290681, 0208292238, 0208292239, 0208301859, 0208302264, 0208302265, 0208302266, 0208302267, 0208302680, 0208302681, 0208303457, 0208303860, 0208303873, 0208316498, 0208316499, 0208320472, 0208323062, 0208323063, 0208323064, 0208323065, 0208327831, 0208327832, 0208327833, 0208327834, 0208327835, 0208327836, 0208328653, 0208328654, 0208328655, 0208328800, 0208328920, 0208328921, 0208328926, 0208328927, 0208328928, 0208328929, 0208328930, 0208328931, 0208328932, 0208328933, 0208329049, 0208338057, 0208338058, 0208338059, 0208338502, 0208338503, 0208347430, 0208347431, 0208347432, 0208347433, 0208347434, 0208347435, 0208347436, 0208347437, 0208347438, 0208347439, 0208347613, 0208347614, 0208347615, 0208347616, 0208347617, 0208347618, 0208347619, 0208347620, 0208348680, 0208348681, 0208348682, 0208348683, 0208348684, 0208348685, 0208348686, 0208348687, 0208348688, 0208348689, 0208348690, 0208348691, 0208348692, 0208348693, 0208348694, 0208348723, 0208358148, 0208358149, 0208358369, 0208358370, 0208358419, 0208362521, 0208362522, 0208362523, 0208362524, 0208362525, 0208362526, 0208362527, 0208362528, 0208362529, 0208374306, 0208374307, 0208374308, 0208374309, 0208374310, 0208374313, 0208374314, 0208374315, 0208374316, 0208374317, 0208374318, 0208374319, 0208374340, 0208374341, 0208374342, 0208374343, 0208374345, 0208374346, 0208374348, 0208374349, 0208374350, 0208374351, 0208374352, 0208374353, 0208374354, 0208374355, 0208374356, 0208374357, 0208374358, 0208374366, 0208374520, 0208374521, 0208374595, 0208374656, 0208374657, 0208374658, 0208374659, 0208374660, 0208378209, 0208378747, 0208378748, 0208378749, 0208378750, 0208378751, 0208378752, 0208378753, 0208378754, 0208378755, 0208382360, 0208382361, 0208390395, 0208390396, 0208390397, 0208390398, 0208390399, 0208390400, 0208390401, 0208390402, 0208390406, 0208390407, 0208390408, 0208390410, 0208390411, 0208390421, 0208391882, 0208393322, 0208393323, 0208394838, 0208394839, 0208412406, 0208412407, 0208412408, 0208412409, 0208412412, 0208412413, 0208412414, 0208412415, 0208412416, 0208412418, 0208412419, 0208412700, 0208412701, 0208412702, 0208412703, 0208412704, 0208412706, 0208412707, 0208412708, 0208413226, 0208425644, 0208425649, 0208425861, 0208430655, 0208431280, 0208433918, 0208433919, 0208433920, 0208435504, 0208438533, 0208438534, 0208438535, 0208438536, 0208438539, 0208439128, 0208439129, 0208439130, 0208439131, 0208439158, 0208439159, 0208439160, 0208439161, 0208439162, 0208439163, 0208439167, 0208439950, 0208439951, 0208452221, 0208452443, 0208452493, 0208452494, 0208452495, 0208452496, 0208455056, 0208455057, 0208455671, 0208455672, 0208456036, 0208456401, 0208456402, 0208456431, 0208456432, 0208456433, 0208456434, 0208456435, 0208456436, 0208456437, 0208456438, 0208456439, 0208456440, 0208456441, 0208456456, 0208456457, 0208456458, 0208456459, 0208456460, 0208456461, 0208456462, 0208457652, 0208457653, 0208457714, 0208457905, 0208457906, 0208457910, 0208457911, 0208459242, 0208459243, 0208460868, 0208461960, 0208461961, 0208466216, 0208466217, 0208466218, 0208466219, 0208466960, 0208466961, 0208466962, 0208466963, 0208466964, 0208466965, 0208466966, 0208466967, 0208466968, 0208466969, 0208466970, 0208466971, 0208466972, 0208466973, 0208466974, 0208466975, 0208471615, 0208475914, 0208475915, 0208476147, 0208476148, 0208477550, 0208477551, 0208477552, 0208477553, 0208477554, 0208477555, 0208477556, 0208477557, 0208477558, 0208477559, 0208477560, 0208477561, 0208477562, 0208477563, 0208477564, 0208477565, 0208477566, 0208477567, 0208477568, 0208477569, 0208479355, 0208479356, 0208479357, 0208479358, 0208479375, 0208479376, 0208482805, 0208482806, 0208484062, 0208487141, 0208487143, 0208488977, 0208490435, 0208490436, 0208490437, 0208490438, 0208492691, 0208492692, 0208497228, 0208497230, 0208498108, 0208498369, 0208498370, 0208498371, 0208498372, 0208498373, 0208498374, 0208498375, 0208498376, 0208498377, 0208498378, 0208498379, 0208498400, 0208498402, 0208498403, 0208498404, 0208498405, 0208498406, 0208498407, 0208498408, 0208498409, 0208498410, 0208498411, 0208498412, 0208498413, 0208498414, 0208498415, 0208498416, 0208498417, 0208498418, 0208500113, 0208501083, 0208501084, 0208501085, 0208501990, 0208501991, 0208501992, 0208501993, 0208502259, 0208502744, 0208503232, 0208506498, 0208506499, 0208506872, 0208506873, 0208511006, 0208511016, 0208511035, 0208511537, 0208513310, 0208521073, 0208521074, 0208521989, 0208521990, 0208522773, 0208522778, 0208522780, 0208533453, 0208533500, 0208535754, 0208540384, 0208544992, 0208545296, 0208545297, 0208546139, 0208549895, 0208551051, 0208551052, 0208551055, 0208551056, 0208551057, 0208551058, 0208551059, 0208551060, 0208551061, 0208551062, 0208551063, 0208551064, 0208551065, 0208551066, 0208551067, 0208551068, 0208551069, 0208551070, 0208551071, 0208551072, 0208551076, 0208551077, 0208551078, 0208551079, 0208559507, 0208559508, 0208559517, 0208564693, 0208564694, 0208564695, 0208564696, 0208564697, 0208570231, 0208570232, 0208570233, 0208570234, 0208570235, 0208570236, 0208587986, 0208587987, 0208587988, 0208587989, 0208587990, 0208587991, 0208587992, 0208587993, 0208587994, 0208588625, 0208595269, 0208595270, 0208612755, 0208612756, 0208612757, 0208612758, 0208612759, 0208613075, 0208613076, 0208613077, 0208616118, 0208616143, 0208616144, 0208616145, 0208616146, 0208616149, 0208616150, 0208616151, 0208616153, 0208616154, 0208616155, 0208616156, 0208616157, 0208616158, 0208616192, 0208616193, 0208616194, 0208616196, 0208616198, 0208616270, 0208616271, 0208616308, 0208616877, 0208616878, 0208616879, 0208616880, 0208616900, 0208616901, 0208616902, 0208625502, 0208626099, 0208627463, 0208628120, 0208628121, 0208628122, 0208628123, 0208628124, 0208628125, 0208628126, 0208628140, 0208632188, 0208632189, 0208632264, 0208632265, 0208632266, 0208632267, 0208632268, 0208638539, 0208638541, 0208638542, 0208638543, 0208638544, 0208652385, 0208652386, 0208652387, 0208652388, 0208655794, 0208655797, 0208655798, 0208655799, 0208655800, 0208655977, 0208655978, 0208655993, 0208655994, 0208659943, 0208659944, 0208669893, 0208678537, 0208678538, 0208678539, 0208678560, 0208678561, 0208678562, 0208678563, 0208678564, 0208678565, 0208678566, 0208678567, 0208679441, 0208679442, 0208679443, 0208679672, 0208679673, 0208679674, 0208679675, 0208679676, 0208679677, 0208679678, 0208679679, 0208679680, 0208679681, 0208679682, 0208679683, 0208679684, 0208679685, 0208679686, 0208679687, 0208679688, 0208679689, 0208679690, 0208679691, 0208680230, 0208688283, 0208688284, 0208688285, 0208688287, 0208688288, 0208688527, 0208688598, 0208688599, 0208688620, 0208688621, 0208688622, 0208688623, 0208688624, 0208688625, 0208688626, 0208688627, 0208688628, 0208688629, 0208688630, 0208689079, 0208689080, 0208689081, 0208689082, 0208689083, 0208689084, 0208689085, 0208689086, 0208689087, 0208689088, 0208689089, 0208689090, 0208689091, 0208689092, 0208689093, 0208689094, 0208689095, 0208689096, 0208689102, 0208689103, 0208689104, 0208689105, 0208689107, 0208692733, 0208692735, 0208692746, 0208694365, 0208694366, 0208700347, 0208700556, 0208702142, 0208703147, 0208703148, 0208703149, 0208703150, 0208703151, 0208703152, 0208703153, 0208703154, 0208703155, 0208703156, 0208703157, 0208703158, 0208703159, 0208703160, 0208703161, 0208703162, 0208703163, 0208703164, 0208703165, 0208703166, 0208703169, 0208703170, 0208705155, 0208705156, 0208705157, 0208705158, 0208705159, 0208705920, 0208705921, 0208705922, 0208705923, 0208705924, 0208705927, 0208705928, 0208705929, 0208705930, 0208705931, 0208705932, 0208705933, 0208705934, 0208708074, 0208709034, 0208714630, 0208714631, 0208715513, 0208715514, 0208715515, 0208715516, 0208715517, 0208715606, 0208715607, 0208717498, 0208717499, 0208717520, 0208717521, 0208717522, 0208718119, 0208718988, 0208718989, 0208718991, 0208718992, 0208718993, 0208718994, 0208718995, 0208719193, 0208719221, 0208719500, 0208719501, 0208719502, 0208719503, 0208719505, 0208719506, 0208719507, 0208719508, 0208719510, 0208720745, 0208720746, 0208720747, 0208732550, 0208732551, 0208732552, 0208732553, 0208732554, 0208732555, 0208732556, 0208732557, 0208732558, 0208732559, 0208732560, 0208732561, 0208732562, 0208732563, 0208732565, 0208732566, 0208732567, 0208732568, 0208732569, 0208732570, 0208749184, 0208749189, 0208749190, 0208749191, 0208749192, 0208749193, 0208749194, 0208754263, 0208754264, 0208754265, 0208754268, 0208754271, 0208754273, 0208754274, 0208754275, 0208754276, 0208754542, 0208754543, 0208757258, 0208757259, 0208758920, 0208759044, 0208759046, 0208759050, 0208759051, 0208759052, 0208759053, 0208759054, 0208759055, 0208759056, 0208759058, 0208759660, 0208759661, 0208759662, 0208759663, 0208759664, 0208760718, 0208760743, 0208763346, 0208763347, 0208763348, 0208763349, 0208763350, 0208763351, 0208763352, 0208763353, 0208763354, 0208763355, 0208763356, 0208763357, 0208763707, 0208763709, 0208764053, 0208764054, 0208764055, 0208764056, 0208764057, 0208764058, 0208764059, 0208764801, 0208766267, 0208766268, 0208766269, 0208766270, 0208766271, 0208766272, 0208766273, 0208766274, 0208766277, 0208766278, 0208766346, 0208766347, 0208766348, 0208766349, 0208766351, 0208766352, 0208766380, 0208766381, 0208766382, 0208766812, 0208774650, 0208774651, 0208774652, 0208774653, 0208774654, 0208774655, 0208774656, 0208774657, 0208774658, 0208774659, 0208774680, 0208774681, 0208774682, 0208774759, 0208775440, 0208775441, 0208785332, 0208794853, 0208794854, 0208794855, 0208794856, 0208794857, 0208794858, 0208794859, 0208794880, 0208794881, 0208801473, 0208801474, 0208801475, 0208803034, 0208803348, 0208803349, 0208803350, 0208803351, 0208803680, 0208804840, 0208804841, 0208804842, 0208804843, 0208804844, 0208804845, 0208808901, 0208808915, 0208808916, 0208808917, 0208808918, 0208808919, 0208808995, 0208813875, 0208813876, 0208813877, 0208813878, 0208813879, 0208814046, 0208814467, 0208814468, 0208814469, 0208816552, 0208817076, 0208817077, 0208817604, 0208817685, 0208817686, 0208817687, 0208817688, 0208817689, 0208817690, 0208817691, 0208821012, 0208821013, 0208821154, 0208821619, 0208821789, 0208821790, 0208821791, 0208821792, 0208821793, 0208821794, 0208821795, 0208821796, 0208821831, 0208821935, 0208821936, 0208821937, 0208821938, 0208821939, 0208821976, 0208821977, 0208821978, 0208821979, 0208823120, 0208823121, 0208826582, 0208826651, 0208826950, 0208826951, 0208826961, 0208826964, 0208829576, 0208832016, 0208832080, 0208833131, 0208833132, 0208833133, 0208833134, 0208833135, 0208833136, 0208833137, 0208833138, 0208833139, 0208833242, 0208833243, 0208833245, 0208833263, 0208833264, 0208833265, 0208833302, 0208833303, 0208833304, 0208833306, 0208833309, 0208833312, 0208833314, 0208833316, 0208833317, 0208833318, 0208833319, 0208833320, 0208833321, 0208833322, 0208835585, 0208837062, 0208837063, 0208837102, 0208837103, 0208837104, 0208837105, 0208837106, 0208837107, 0208837489, 0208837490, 0208837577, 0208839357, 0208839359, 0208839360, 0208839361, 0208839362, 0208839363, 0208839364, 0208839365, 0208839366, 0208839367, 0208839368, 0208839369, 0208839370, 0208839371, 0208839372, 0208839813, 0208839814, 0208843168, 0208843169, 0208843170, 0208843171, 0208843172, 0208843173, 0208843174, 0208843175, 0208843176, 0208843177, 0208843178, 0208843179, 0208843200, 0208843212, 0208843213, 0208843214, 0208843215, 0208843216, 0208849131, 0208851526, 0208851527, 0208852878, 0208852879, 0208856443, 0208856452, 0208856453, 0208856454, 0208856455, 0208856456, 0208856457, 0208856458, 0208856459, 0208856748, 0208856749, 0208857815, 0208857816, 0208858247, 0208858248, 0208858249, 0208858250, 0208858251, 0208858252, 0208858253, 0208858254, 0208858255, 0208858256, 0208858257, 0208858258, 0208858259, 0208858260, 0208858261, 0208862451, 0208862452, 0208862453, 0208862454, 0208862455, 0208862456, 0208862457, 0208862458, 0208862459, 0208862874, 0208862875, 0208862876, 0208862877, 0208862878, 0208864099, 0208864300, 0208864301, 0208866453, 0208866880, 0208871343, 0208871785, 0208871786, 0208872016, 0208872017, 0208874284, 0208874285, 0208874286, 0208875667, 0208890875, 0208890876, 0208890877, 0208890878, 0208890879, 0208892980, 0208892981, 0208895229, 0208895230, 0208895231, 0208899494, 0208899495, 0208900447, 0208906207, 0208906208, 0208911113, 0208911114, 0208911115, 0208916297, 0208920983, 0208920984, 0208930674, 0208936586, 0208937076, 0208937077, 0208937078, 0208937079, 0208937292, 0208937293, 0208937294, 0208937320, 0208937321, 0208942476, 0208946533, 0208949856, 0208956920, 0208956921, 0208956954, 0208959468, 0208959469, 0208963119, 0208963140, 0208963141, 0208963142, 0208963143, 0208963144, 0208963145, 0208963146, 0208963147, 0208963148, 0208963149, 0208963150, 0208963226, 0208963227, 0208963299, 0208963340, 0208963342, 0208963343, 0208963344, 0208963345, 0208963346, 0208963347, 0208963348, 0208963349, 0208963350, 0208963351, 0208963352, 0208963353, 0208965395, 0208965397, 0208965398, 0208968003, 0208968004, 0208977539, 0208980186, 0208980738, 0208980881, 0208982966, 0208983462, 0208983466, 0208987949, 0208987951, 0208987953, 0208988021, 0208988022, 0208988023, 0208988024, 0208988025, 0208988026, 0208988027, 0208988028, 0208988029, 0208988030, 0208988031, 0208988032, 0208988660, 0208988661, 0208988680, 0208988681, 0208988682, 0208988896, 0208988898, 0208988899, 0208989353, 0208989355, 0208989356, 0208989357, 0208989358, 0208989581, 0208989582, 0208989583, 0208989584, 0208989585, 0208989586, 0208989587, 0208989588, 0208989589, 0208989590, 0208989591, 0208989600, 0208989601, 0208992702, 0208992740, 0208995176, 0208995177, 0208995178, 0208995179, 0208995480, 0208995695, 0208995696, 0208995697, 0208995698, 0208998252, 0208998253, 0208998254, 0208998255, 0208998256, 0208998257, 0208998258, 0208998259, 0208998663, 0208999040, 0208999041, 0208999340, 0208999341, 0209001628, 0209001629, 0209002689, 0209002690, 0209002870, 0209004658, 0209004659, 0209004890, 0209004891, 0209005400, 0209005401, 0209005402, 0209005403, 0209005404, 0209005405, 0209005406, 0209005407, 0209005408, 0209007210, 0209007969, 0209008355, 0209010079, 0209010080, 0209010081, 0209010082, 0209010083, 0209010084, 0209010085, 0209010086, 0209010087, 0209010088, 0209010089, 0209010090, 0209010091, 0209010092, 0209011131, 0209011132, 0209014817, 0209014818, 0209016419, 0209016782, 0209020292, 0209020365, 0209020366, 0209020367, 0209020368, 0209020369, 0209020371, 0209023646, 0209023647, 0209024727, 0209030016, 0209030018, 0209030651, 0209036766, 0209036767, 0209036768, 0209036769, 0209036770, 0209036771, 0209040869, 0209043588, 0209043589, 0209043590, 0209043591, 0209043592, 0209043593, 0209043594, 0209043595, 0209043596, 0209044568, 0209044569, 0209044570, 0209044571, 0209044572, 0209044573, 0209044574, 0209044575, 0209044576, 0209045614, 0209045615, 0209047234, 0209047235, 0209047236, 0209049558, 0209051167, 0209051168, 0209051863, 0209055925, 0209055926, 0209055927, 0209056904, 0209056913, 0209057407, 0209059668, 0209059669, 0209059670, 0209059671, 0209059820, 0209059821, 0209059822, 0209059823, 0209062209, 0209062210, 0209062211, 0209062212, 0209062213, 0209062214, 0209062215, 0209063310, 0209063314, 0209063315, 0209063944, 0209063945, 0209067183, 0209067288, 0209070795, 0209070796, 0209070797, 0209070798, 0209070799, 0209071331, 0209072100, 0209072105, 0209072125, 0209074745, 0209078525, 0209078527, 0209078528, 0209078529, 0209078531, 0209078534, 0209078535, 0209078536, 0209078537, 0209078538, 0209078539, 0209078586, 0209078869, 0209078870, 0209078910, 0209078911, 0209078912, 0209079407, 0209079408, 0209079414, 0209079415, 0209079416, 0209079417, 0209084185, 0209084187, 0209084606, 0209084607, 0209084608, 0209084609, 0209084610, 0209088279, 0209090800, 0209090809, 0209090810, 0209091631, 0209091825, 0209094299, 0209094426, 0209094427, 0209094731, 0209096359, 0209096360, 0209096361, 0209096362, 0209096363, 0209096364, 0209096365, 0209096366, 0209096409, 0209097704, 0209097705, 0209098476, 0209098477, 0209098478, 0209098479, 0209098600, 0209099389, 0209099390, 0209099391, 0209099407, 0209100767, 0209100773, 0209100774, 0209100775, 0209100776, 0209101209, 0209101264, 0209101265, 0209101266, 0209102091, 0209102813, 0209108856, 0209109042, 0209109043, 0209109044, 0209109045, 0209109046, 0209109047, 0209109048, 0209109220, 0209110084, 0209110085, 0209110086, 0209110087, 0209110088, 0209110089, 0209110090, 0209110091, 0209111473, 0209111474, 0209111475, 0209111476, 0209111477, 0209111478, 0209111479, 0209111620, 0209111621, 0209111622, 0209112372, 0209116351, 0209117711, 0209117712, 0209121365, 0209121383, 0209126365, 0209126366, 0209126711, 0209126712, 0209126713, 0209126714, 0209126715, 0209128354, 0209131124, 0209131125, 0209143537, 0209143538, 0209143539, 0209144400, 0209144401, 0209144402, 0209147706, 0209147707, 0209147772, 0209147805, 0209147806, 0209147903, 0209147904, 0209147905, 0209147906, 0209151504, 0209151507, 0209151509, 0209151511, 0209151525, 0209151528, 0209151530, 0209151532, 0209151544, 0209151546, 0209151549, 0209151551, 0209151554, 0209151556, 0209155101, 0209155102, 0209155103, 0209155152, 0209155153, 0209155159, 0209155309, 0209155312, 0209155314, 0209158640, 0209159134, 0209159135, 0209159136, 0209159425, 0209159426, 0209159427, 0209162965, 0209165873, 0209168925, 0209168926, 0209168927, 0209168928, 0209168933, 0209168951, 0209168952, 0209172775, 0209176810, 0209176813, 0209182982, 0209182983, 0209182984, 0209183710, 0209183887, 0209185594, 0209185595, 0209188298, 0209188311, 0209189906, 0209193858, 0209194116, 0209194381, 0209194400, 0209196099, 0209198720, 0209198775, 0209201184, 0209203836, 0209203911, 0209203972, 0209204355, 0209204383, 0209209011, 0209209012, 0209209021, 0209209022, 0209209219, 0209209249, 0209209250, 0209209251, 0209209252, 0209209253, 0209209340, 0209209341, 0209209342, 0209209606, 0209209607, 0209211975, 0209212086, 0209212087, 0209212088, 0209212089, 0209212092, 0209212093, 0209212094, 0209212095, 0209212521, 0209212524, 0209212526, 0209213180, 0209227966, 0209227967, 0209228560, 0209228561, 0209230530, 0209230531, 0209233324, 0209233325, 0209234597, 0209234600, 0209235186, 0209237216, 0209237329, 0209237330, 0209237331, 0209237332, 0209237333, 0209237382, 0209238245, 0209241170, 0209241171, 0209242116, 0209242502, 0209242503, 0209242504, 0209242505, 0209242506, 0209242656, 0209242659, 0209242765, 0209242774, 0209242776, 0209242777, 0209243837, 0209244038, 0209244039, 0209244340, 0209244341, 0209244342, 0209248176, 0209248640, 0209248641, 0209249650, 0209249655, 0209255341, 0209262280, 0209262579, 0209262752, 0209263003, 0209263054, 0209263152, 0209263154, 0209263155, 0209264521, 0209266947, 0209267840, 0209269830, 0209269831, 0209270377, 0209271361, 0209271373, 0209276622, 0209276623, 0209276637, 0209277465, 0209278531, 0209278532, 0209278533, 0209279645, 0209282515, 0209282517, 0209282526, 0209282527, 0209282761, 0209282762, 0209282871, 0209282872, 0209282873, 0209282874, 0209282875, 0209282876, 0209282877, 0209282878, 0209284384, 0209284385, 0209284386, 0209284387, 0209284398, 0209287415, 0209287416, 0209287417, 0209287476, 0209287612, 0209287613, 0209287614, 0209287615, 0209287616, 0209287617, 0209287618, 0209291524, 0209292226, 0209292637, 0209292638, 0209292671, 0209292672, 0209294875, 0209298384, 0209298385, 0209298386, 0209298387, 0209299181, 0209299184, 0209299185, 0209299243, 0209299368, 0209303220, 0209303221, 0209303222, 0209309155, 0209309156, 0209309157, 0209309158, 0209309390, 0209309391, 0209309817, 0209309833, 0209309835, 0209309836, 0209309837, 0209309838, 0209311102, 0209313548, 0209313549, 0209313550, 0209313551, 0209313552, 0209313554, 0209313555, 0209313556, 0209313557, 0209313558, 0209313560, 0209313921, 0209313922, 0209314037, 0209314039, 0209314251, 0209315470, 0209318049, 0209318050, 0209318051, 0209318121, 0209318176, 0209318177, 0209318806, 0209318822, 0209318823, 0209323066, 0209323067, 0209323126, 0209323127, 0209323812, 0209323815, 0209323820, 0209326070, 0209326071, 0209329314, 0209329316, 0209329499, 0209329837, 0209329838, 0209330799, 0209332095, 0209335451, 0209335452, 0209335675, 0209335676, 0209335677, 0209335812, 0209335813, 0209335814, 0209336661, 0209336662, 0209336663, 0209341127, 0209341128, 0209341129, 0209341130, 0209341131, 0209341614, 0209341615, 0209341707, 0209341708, 0209341723, 0209341724, 0209341725, 0209341728, 0209341729, 0209342956, 0209342958, 0209344014, 0209345134, 0209345135, 0209345136, 0209345204, 0209345206, 0209345456, 0209345457, 0209345764, 0209345765, 0209349506, 0209349507, 0209349952, 0209350553, 0209350825, 0209350887, 0209352788, 0209352789, 0209354999, 0209356412, 0209356413, 0209356422, 0209356423, 0209356424, 0209356480, 0209356481, 0209358450, 0209359857, 0209359858, 0209360593, 0209360595, 0209360596, 0209360663, 0209360664, 0209360665, 0209364436, 0209365941, 0209365942, 0209366209, 0209366210, 0209366211, 0209366212, 0209367098, 0209368403, 0209368404, 0209368413, 0209368525, 0209368526, 0209368801, 0209368804, 0209368811, 0209368812, 0209370879, 0209370909, 0209370921, 0209370922, 0209370923, 0209370975, 0209374057, 0209374058, 0209374529, 0209374530, 0209374531, 0209374586, 0209374587, 0209374589, 0209374590, 0209374591, 0209374592, 0209374593, 0209374594, 0209374999, 0209375750, 0209375751, 0209375825, 0209376393, 0209376395, 0209376814, 0209379222, 0209379223, 0209379810, 0209379811, 0209379815, 0209379817, 0209379818, 0209379822, 0209379823, 0209379869, 0209383041, 0209383042, 0209385080, 0209385081, 0209385082, 0209385083, 0209385084, 0209385086, 0209385087, 0209385125, 0209385126, 0209385127, 0209385128, 0209385647, 0209392669, 0209392701, 0209392702, 0209392784, 0209392887, 0209392893, 0209392894, 0209393773, 0209393779, 0209393781, 0209393782, 0209393784, 0209393786, 0209393787, 0209393788, 0209394992, 0209394995, 0209397031, 0209397547, 0209400052, 0209400111, 0209400112, 0209401840, 0209401845, 0209401846, 0209401847, 0209402316, 0209403257, 0209403964, 0209407833, 0209407839, 0209408655, 0209409049, 0209409050, 0209409051, 0209409414, 0209409463, 0209409901, 0209409902, 0209411898, 0209411899, 0209411900, 0209414792, 0209417875, 0209417876, 0209421884, 0209421886, 0209421887, 0209422108, 0209422109, 0209422110, 0209422111, 0209422115, 0209422116, 0209422117, 0209422118, 0209422119, 0209423235, 0209423236, 0209423237, 0209423238, 0209423239, 0209423241, 0209428178, 0209428179, 0209428180, 0209428181, 0209432579, 0209432760, 0209434846, 0209434847, 0209436004, 0209436005, 0209440146, 0209440147, 0209440148, 0209445531, 0209445532, 0209445533, 0209454050, 0209454051, 0209454096, 0209454097, 0209473049, 0209473050, 0209473051, 0209473193, 0209473260, 0209473261, 0209473262, 0209473263, 0209473264, 0209473265, 0209473266, 0209473267, 0209473268, 0209473269, 0209473270, 0209473271, 0209473272, 0209473273, 0209478562, 0209483652, 0209483653, 0209483654, 0209483655, 0209483656, 0209489057, 0209489322, 0209489326, 0209490302, 0209490303, 0209490307, 0209490308, 0209490310, 0209490311, 0209490312, 0209491500, 0209493378, 0209493659, 0209493780, 0209493781, 0209493782, 0209493951, 0209494175, 0209494324, 0209494336, 0209494337, 0209494338, 0209494339, 0209494520, 0209494521, 0209494522, 0209494523, 0209494524, 0209494560, 0209494561, 0209494624, 0209499045, 0209499046, 0209501841, 0209501842, 0209501843, 0209502559, 0209502560, 0209502561, 0209502562, 0209502868, 0209502869, 0209503204, 0209503205, 0209503206, 0209503207, 0209503208, 0209503209, 0209503210, 0209503211, 0209503213, 0209503214, 0209503215, 0209503216, 0209503217, 0209503218, 0209503219, 0209503460, 0209503461, 0209503462, 0209510570, 0209510575, 0209510582, 0209510583, 0209510584, 0209510585, 0209515212, 0209515219, 0209515600, 0209515601, 0209525433, 0209525746, 0209525747, 0209525748, 0209525749, 0209525750, 0209525751, 0209525752, 0209530143, 0209530144, 0209530145, 0209530146, 0209530147, 0209530148, 0209544587, 0209544588, 0209544805, 0209550294, 0209550295, 0209550789, 0209565166, 0209565167, 0209565168, 0209565169, 0209568827, 0209568828, 0209568829, 0209568830, 0209568831, 0209568832, 0209568833, 0209568834, 0209568835, 0209568836, 0209568837, 0209568838, 0209568839, 0209569215, 0209569216, 0209569217, 0209569239, 0209569253, 0209569254, 0209569255, 0209569462, 0209569520, 0209569540, 0209569541, 0209570015, 0209570016, 0209570869, 0209570870, 0209570871, 0209573500, 0209573507, 0209574255, 0209574256, 0209574278, 0209574279, 0209574296, 0209574297, 0209574298, 0209574299, 0209574316, 0209574317, 0209574318, 0209574326, 0209574327, 0209574328, 0209574503, 0209574504, 0209577080, 0209577081, 0209577082, 0209577083, 0209577084, 0209579263, 0209580435, 0209581965, 0209581966, 0209581967, 0209581972, 0209581973, 0209581974, 0209581975, 0209581976, 0209582009, 0209582115, 0209582116, 0209582372, 0209586436, 0209586437, 0209586438, 0209586729, 0209586730, 0209586731, 0209586732, 0209586733, 0209588743, 0209588744, 0209588745, 0209588746, 0209588981, 0209588983, 0209590018, 0209591441, 0209591442, 0209591460, 0209591622, 0209592958, 0209592959, 0209593179, 0209593428, 0209593429, 0209593430, 0209593431, 0209593432, 0209593433, 0209593434, 0209595944, 0209595950, 0209595951, 0209596068, 0209596070, 0209596520, 0209596521, 0209596522, 0209596523, 0209596524, 0209596525, 0209596526, 0209596527, 0209596945, 0209598939, 0209606205, 0209606206, 0209606207, 0209606208, 0209606209, 0209606213, 0209607243, 0209607513, 0209607514, 0209607515, 0209607516, 0209607517, 0209607518, 0209607519, 0209610719, 0209612940, 0209612941, 0209612942, 0209612943, 0209612944, 0209612945, 0209612946, 0209612948, 0209612949, 0209612950, 0209612951, 0209612952, 0209612953, 0209612955, 0209612990, 0209612991, 0209612992, 0209612993, 0209612994, 0209612995, 0209612996, 0209612997, 0209612998, 0209612999, 0209613000, 0209613001, 0209613002, 0209613003, 0209613555, 0209613559, 0209614029, 0209614301, 0209615469, 0209615470, 0209615471, 0209615472, 0209617349, 0209617746, 0209617747, 0209617748, 0209627230, 0209627295, 0209627296, 0209627297, 0209636669, 0209639847, 0209639848, 0209639849, 0209639850, 0209639851, 0209639852, 0209639941, 0209643451, 0209647284, 0209647456, 0209647457, 0209647458, 0209647459, 0209647540, 0209648826, 0209651534, 0209651535, 0209651536, 0209651537, 0209652362, 0209652366, 0209652367, 0209652368, 0209658938, 0209658939, 0209659540, 0209659541, 0209659542, 0209659543, 0209664609, 0209664610, 0209664611, 0209664612, 0209664615, 0209664616, 0209664617, 0209664618, 0209664619, 0209665552, 0209665556, 0209665558, 0209665601, 0209665602, 0209665603, 0209667833, 0209669391, 0209669589, 0209669602, 0209669604, 0209669608, 0209669616, 0209669617, 0209669618, 0209670903, 0209671117, 0209671183, 0209671187, 0209671976, 0209671977, 0209671978, 0209671979, 0209674954, 0209674955, 0209675007, 0209675012, 0209675553, 0209675920, 0209676253, 0209676556, 0209680180, 0209680181, 0209680232, 0209680233, 0209680234, 0209680235, 0209680236, 0209680237, 0209680238, 0209680239, 0209680399, 0209680415, 0209680416, 0209680417, 0209680925, 0209681064, 0209681065, 0209686773, 0209699905, 0209699906, 0209700287, 0209704976, 0209704979, 0209705380, 0209708224, 0209710924, 0209710925, 0209711049, 0209711050, 0209711051, 0209711094, 0209711682, 0209711684, 0209714973, 0209714974, 0209714975, 0209717579, 0209718940, 0209718982, 0209719331, 0209719332, 0209719367, 0209724314, 0209724382, 0209724384, 0209724386, 0209728243, 0209728244, 0209728249, 0209728250, 0209730903, 0209730904, 0209730932, 0209730933, 0209730934, 0209730935, 0209730936, 0209730937, 0209730938, 0209730939, 0209731072, 0209731073, 0209731074, 0209731075, 0209731076, 0209731077, 0209731078, 0209731079, 0209731100, 0209731101, 0209731102, 0209732886, 0209732940, 0209732941, 0209732942, 0209732943, 0209732944, 0209732946, 0209732947, 0209732948, 0209733817, 0209733818, 0209734098, 0209734099, 0209734380, 0209736412, 0209738799, 0209738800, 0209747423, 0209747424, 0209747425, 0209747426, 0209747427, 0209747493, 0209747494, 0209747495, 0209747496, 0209747497, 0209747498, 0209747499, 0209747620, 0209747621, 0209750934, 0209750935, 0209750936, 0209750937, 0209750994, 0209751577, 0209751594, 0209756447, 0209756449, 0209756479, 0209756774, 0209756775, 0209756776, 0209756860, 0209761335, 0209766536, 0209766537, 0209766582, 0209767172, 0209767173, 0209772562, 0209774926, 0209774927, 0209774930, 0209774931, 0209775121, 0209775124, 0209781898, 0209782444, 0209782445, 0209782830, 0209783463, 0209783464, 0209783465, 0209783466, 0209783470, 0209783471, 0209783476, 0209785192, 0209785636, 0209792839, 0209795216, 0209796941, 0209796945, 0209798057, 0209798058, 0209798059, 0209798640, 0209798641, 0209798642, 0209801139, 0209806391, 0209806421, 0209806422, 0209806792, 0209806793, 0209806794, 0209806795, 0209806805, 0209806806, 0209807373, 0209807375, 0209807721, 0209807722, 0209807723, 0209807724, 0209811311, 0209811312, 0209830703, 0209830704, 0209830708, 0209830710, 0209830761, 0209830762, 0209830763, 0209831167, 0209831169, 0209831173, 0209831175, 0209831392, 0209831393, 0209831394, 0209831397, 0209831399, 0209831540, 0209831542, 0209831602, 0209835374, 0209835375, 0209835376, 0209835377, 0209842967, 0209848738, 0209848796, 0209849826, 0209857928, 0209857929, 0209859092, 0209859093, 0209859094, 0209859095, 0209859096, 0209859097, 0209859098, 0209859099, 0209859100, 0209859101, 0209864964, 0209864974, 0209865184, 0209865274, 0209865275, 0209865741, 0209865742, 0209871826, 0209871827, 0209871898, 0209871899, 0209871940, 0209871941, 0209871942, 0209871943, 0209871944, 0209877133, 0209877544, 0209877545, 0209877546, 0209877547, 0209877548, 0209877549, 0209878568, 0209878987, 0209878988, 0209878989, 0209879169, 0209879170, 0209881504, 0209881505, 0209881506, 0209882665, 0209882666, 0209882667, 0209893755, 0209893756, 0209901558, 0209901559, 0209901588, 0209901604, 0209901605, 0209901613, 0209901614, 0209917628, 0209917974, 0209923323, 0209923718, 0209923719, 0209924016, 0209928044, 0209928045, 0209928046, 0209929086, 0209929113, 0209932185, 0209933731, 0209933732, 0209933733, 0209933735, 0209933738, 0209935343, 0209935345, 0209935346, 0209935662, 0209935664, 0209935665, 0209935666, 0209935667, 0209935668, 0209935669, 0209935692, 0209935693, 0209935694, 0209935695, 0209935696, 0209948478, 0209949036, 0209962861, 0209967649, 0209967650, 0209967651, 0209967652, 0209967653, 0209967654, 0209967655, 0209967657, 0209967658, 0209967659, 0209981854, 0209981855, 0209983255, 0209983256, 0209983257, 0209983258, 0209983274, 0209983304, 0209983305, 0209995307, 0210006537, 0210016839, 0210019080, 0210019081, 0210026750, 0210032362, 0210032366, 0210032367, 0210032368, 0210032369, 0210032789, 0210032790, 0210033282, 0210033283, 0210037569, 0210037697, 0210037698, 0210037699, 0210038126, 0210038140, 0210038141, 0210038142, 0210038143, 0210038144, 0210038145, 0210038146, 0210038147, 0210038148, 0210038149, 0210038150, 0210038151, 0210038152, 0210038383, 0210038391, 0210038398, 0210038420, 0210038422, 0210038424, 0210038426, 0210038429, 0210038430, 0210047605, 0210047614, 0210048045, 0210051867, 0210052079, 0210058752, 0210061240, 0210061241, 0210068552, 0210069469, 0210069470, 0210074097, 0210074098, 0210074196, 0210074197, 0210074198, 0210074285, 0210074800, 0210074801, 0210074936, 0210074937, 0210079748, 0210081285, 0210086377, 0210093848, 0210093856, 0210099349, 0210099649, 0210099652, 0210100253, 0210100535, 0210100536, 0210100537, 0210104370, 0210107801, 0210107802, 0210107803, 0210107804, 0210112402, 0210113732, 0210113733, 0210116910, 0210116912, 0210116915, 0210116917, 0210123074, 0210134018, 0210144946, 0210155688, 0210164560, 0210164561, 0210164562, 0210164563, 0210165353, 0210165354, 0210165355, 0210165356, 0210165357, 0210169872, 0210170042, 0210173047, 0210173048, 0210174629, 0210174633, 0210174636, 0210174639, 0210175380, 0210175381, 0210175382, 0210176002, 0210178219, 0210178261, 0210178262, 0210178263, 0210178264, 0210178265, 0210178266, 0210178267, 0210178268, 0210178269, 0210178270, 0210178768, 0210187179, 0210187808, 0210187809, 0210187810, 0210187811, 0210187814, 0210187822, 0210187823, 0210187824, 0210187825, 0210187826, 0210187827, 0210187828, 0210187834, 0210187835, 0210187836, 0210187837, 0210187838, 0210194146, 0210194147, 0210194148, 0210194149, 0210194150, 0210194151, 0210194152, 0210194200, 0210194201, 0210194202, 0210194203, 0210194204, 0210194660, 0210195913, 0210195914, 0210195915, 0210195916, 0210195917, 0210195918, 0210195919, 0210197380, 0210197381, 0210197382, 0210200386, 0210201166, 0210205014, 0210206010, 0210206896, 0210209547, 0210209548, 0210209549, 0210212740, 0210218902, 0210218903, 0210219012, 0210219060, 0210221222, 0210226718, 0210226719, 0210226797, 0210226798, 0210231712, 0210240060, 0210240061, 0210241796, 0210245133, 0210245902, 0210245903, 0210245904, 0210245905, 0210245906, 0210245907, 0210245908, 0210245909, 0210250048, 0210250050, 0210250552, 0210250557, 0210250840, 0210254236, 0210262391, 0210262393, 0210263067, 0210263068, 0210263252, 0210263253, 0210263254, 0210263255, 0210267705, 0210267843, 0210275188, 0210276448, 0210279887, 0210279945, 0210279946, 0210279947, 0210279948, 0210282668, 0210282672, 0210282776, 0210287581, 0210287582, 0210287584, 0210289947, 0210289950, 0210289962, 0210289963, 0210289964, 0210289965, 0210289966, 0210294702, 0210294703, 0210294706, 0210294707, 0210294834, 0210294835, 0210294836, 0210294837, 0210294838, 0210294839, 0210294904, 0210294905, 0210294906, 0210294907, 0210294908, 0210294909, 0210294910, 0210294911, 0210295000, 0210295001, 0210295002, 0210295003, 0210295004, 0210295005, 0210295006, 0210296675, 0210296677, 0210296678, 0210296679, 0210296945, 0210296960, 0210296961, 0210296962, 0210297111, 0210297112, 0210298748, 0210298749, 0210298801, 0210300884, 0210300885, 0210300889, 0210306425, 0210306426, 0210306972, 0210306973, 0210308014, 0210308015, 0210308016, 0210308854, 0210312162, 0210312177, 0210312179, 0210312927, 0210322251, 0210322254, 0210322259, 0210322301, 0210323814, 0210323815, 0210328275, 0210328308, 0210328750, 0210328751, 0210328761, 0210328762, 0210328763, 0210328764, 0210328765, 0210328766, 0210336629, 0210336685, 0210336687, 0210337065, 0210337067, 0210337069, 0210337072, 0210337076, 0210337078, 0210337102, 0210337105, 0210337112, 0210337116, 0210337120, 0210337125, 0210337131, 0210337135, 0210337140, 0210337417, 0210337418, 0210344314, 0210349852, 0210354338, 0210364443, 0210364444, 0210364445, 0210364446, 0210364447, 0210364448, 0210364449, 0210364450, 0210364451, 0210364453, 0210364454, 0210364455, 0210364456, 0210364457, 0210364458, 0210364459, 0210364480, 0210364481, 0210364704, 0210364706, 0210364707, 0210364708, 0210364803, 0210365120, 0210365139, 0210365240, 0210365242, 0210365243, 0210365244, 0210365245, 0210365247, 0210365248, 0210365249, 0210365250, 0210365251, 0210365252, 0210365253, 0210365255, 0210365256, 0210365257, 0210366208, 0210366209, 0210366211, 0210366364, 0210367089, 0210367091, 0210367093, 0210367094, 0210367096, 0210368790, 0210371370, 0210371371, 0210371372, 0210371373, 0210371374, 0210371797, 0210371861, 0210372655, 0210373345, 0210373346, 0210373347, 0210373348, 0210373349, 0210373350, 0210373351, 0210373352, 0210373353, 0210373354, 0210373355, 0210373356, 0210373357, 0210373358, 0210373359, 0210373680, 0210373681, 0210373682, 0210373683, 0210373684, 0210373685, 0210373686, 0210373687, 0210373688, 0210373689, 0210373690, 0210373691, 0210373692, 0210373793, 0210373879, 0210378144, 0210378149, 0210378152, 0210378155, 0210378240, 0210378241, 0210378242, 0210378243, 0210378244, 0210378245, 0210378246, 0210378251, 0210378252, 0210378253, 0210378254, 0210378255, 0210378256, 0210378257, 0210378258, 0210378690, 0210378691, 0210378927, 0210383118, 0210384255, 0210384256, 0210384257, 0210384258, 0210384716, 0210387262, 0210390209, 0210390210, 0210399270, 0210399271, 0210400563, 0210400564, 0210400565, 0210400566, 0210400567, 0210400568, 0210405905, 0210406966, 0210406967, 0210411624, 0210411627, 0210412867, 0210417350, 0210417351, 0210417353, 0210417358, 0210420181, 0210420193, 0210420198, 0210420199, 0210420202, 0210420206, 0210422116, 0210422118, 0210422178, 0210422179, 0210426207, 0210426209, 0210426211, 0210426212, 0210427961, 0210427962, 0210427963, 0210434273, 0210434340, 0210438927, 0210438929, 0210438930, 0210438932, 0210438934, 0210438936, 0210438939, 0210439217, 0210439220, 0210439224, 0210439225, 0210439226, 0210439227, 0210439303, 0210439305, 0210439306, 0210439307, 0210439308, 0210439309, 0210439310, 0210439311, 0210439312, 0210439313, 0210439314, 0210439315, 0210439316, 0210439317, 0210439318, 0210439388, 0210444172, 0210444173, 0210444174, 0210444214, 0210444216, 0210444218, 0210444219, 0210444320, 0210444321, 0210444322, 0210444324, 0210450411, 0210450540, 0210467773, 0210467774, 0210467775, 0210467776, 0210467777, 0210467819, 0210473371, 0210473372, 0210473605, 0210473606, 0210473607, 0210473608, 0210473609, 0210473610, 0210473611, 0210473612, 0210473613, 0210473614, 0210473632, 0210473633, 0210473634, 0210473635, 0210473636, 0210473637, 0210473638, 0210473639, 0210473640, 0210473641, 0210473642, 0210474355, 0210474356, 0210474357, 0210474358, 0210474359, 0210474540, 0210474541, 0210474542, 0210474543, 0210474544, 0210474545, 0210474546, 0210474547, 0210474548, 0210474549, 0210474550, 0210474551, 0210478947, 0210478949, 0210478951, 0210478954, 0210478958, 0210478959, 0210479003, 0210491898, 0210491899, 0210492200, 0210492201, 0210492202, 0210492203, 0210492204, 0210492205, 0210492206, 0210492207, 0210492208, 0210492209, 0210492210, 0210492211, 0210492212, 0210492213, 0210492214, 0210492215, 0210492216, 0210492217, 0210492218, 0210492219, 0210492220, 0210492221, 0210492222, 0210492223, 0210492333, 0210492336, 0210492337, 0210492338, 0210492339, 0210492340, 0210492341, 0210492344, 0210492345, 0210492346, 0210492347, 0210492765, 0210495539, 0210495663, 0210495664, 0210495665, 0210495666, 0210495667, 0210495680, 0210495681, 0210495682, 0210496357, 0210496358, 0210496359, 0210496520, 0210496521, 0210496522, 0210496523, 0210496524, 0210496525, 0210496526, 0210496527, 0210497210, 0210497211, 0210497212, 0210502492, 0210502493, 0210502494, 0210502495, 0210502496, 0210502497, 0210502498, 0210502499, 0210505375, 0210505376, 0210505379, 0210505939, 0210506100, 0210506101, 0210506860, 0210506861, 0210508453, 0210512458, 0210512459, 0210512860, 0210512865, 0210512866, 0210512867, 0210512868, 0210512872, 0210512873, 0210512939, 0210516660, 0210516661, 0210528233, 0210528240, 0210528658, 0210528904, 0210528912, 0210529667, 0210529668, 0210529669, 0210540740, 0210540741, 0210540876, 0210540877, 0210541003, 0210541004, 0210550967, 0210550968, 0210550973, 0210551050, 0210551051, 0210551053, 0210553305, 0210553306, 0210553307, 0210557113, 0210557114, 0210557115, 0210563831, 0210566665, 0210566666, 0210566667, 0210566668, 0210566669, 0210566670, 0210566962, 0210566963, 0210566964, 0210566965, 0210566966, 0210566967, 0210566968, 0210566969, 0210566970, 0210566971, 0210566972, 0210566973, 0210566974, 0210566975, 0210566976, 0210566977, 0210566978, 0210566979, 0210566980, 0210566981, 0210566982, 0210566983, 0210566985, 0210566986, 0210566987, 0210566988, 0210566989, 0210566990, 0210566991, 0210566992, 0210566993, 0210566996, 0210566997, 0210567000, 0210567001, 0210567002, 0210567004, 0210567005, 0210567006, 0210570699, 0210571823, 0210571824, 0210571880, 0210574693, 0210577788, 0210581453, 0210581454, 0210581455, 0210581456, 0210581457, 0210581631, 0210581632, 0210586872, 0210587358, 0210587359, 0210592507, 0210592508, 0210592509, 0210605911, 0210606035, 0210606036, 0210606037, 0210606039, 0210606240, 0210606241, 0210606242, 0210606244, 0210606245, 0210606247, 0210613155, 0210613650, 0210613651, 0210615033, 0210615034, 0210615035, 0210615036, 0210615715, 0210615716, 0210615717, 0210615718, 0210615719, 0210616120, 0210616121, 0210616122, 0210616123, 0210616124, 0210616125, 0210616126, 0210616127, 0210616128, 0210616129, 0210616130, 0210616131, 0210617342, 0210617373, 0210617928, 0210617929, 0210617930, 0210617931, 0210617932, 0210617933, 0210617934, 0210617935, 0210617936, 0210617937, 0210617938, 0210617939, 0210617980, 0210617981, 0210617982, 0210617983, 0210617984, 0210617985, 0210617986, 0210617987, 0210617988, 0210617989, 0210617990, 0210617991, 0210618000, 0210618001, 0210618002, 0210618003, 0210618004, 0210618005, 0210618006, 0210618007, 0210618008, 0210618429, 0210618433, 0210618888, 0210618889, 0210619867, 0210625292, 0210625293, 0210625294, 0210625295, 0210626815, 0210626817, 0210626818, 0210626819, 0210626820, 0210626822, 0210626823, 0210626824, 0210626825, 0210626827, 0210626828, 0210626830, 0210626832, 0210632070, 0210632071, 0210634032, 0210640368, 0210640369, 0210640407, 0210640408, 0210643466, 0210643468, 0210643741, 0210644204, 0210644208, 0210644209, 0210644210, 0210644211, 0210648734, 0210648735, 0210649098, 0210654784, 0210658339, 0210658643, 0210664832, 0210664833, 0210664834, 0210664835, 0210664836, 0210664837, 0210664838, 0210664839, 0210664880, 0210664881, 0210664882, 0210666532, 0210671832, 0210680307, 0210680308, 0210680309, 0210680310, 0210680311, 0210680312, 0210680596, 0210680597, 0210680598, 0210680599, 0210680852, 0210680853, 0210680854, 0210680855, 0210680994, 0210680995, 0210680996, 0210681683, 0210681686, 0210681687, 0210682136, 0210682137, 0210682138, 0210682139, 0210686264, 0210686650, 0210686651, 0210686652, 0210686775, 0210686776, 0210686777, 0210686778, 0210686779, 0210687220, 0210687221, 0210687222, 0210687223, 0210687224, 0210687225, 0210692398, 0210693250, 0210693251, 0210693252, 0210693253, 0210693331, 0210693332, 0210693333, 0210693334, 0210697443, 0210697444, 0210697445, 0210697447, 0210697448, 0210697449, 0210697450, 0210697458, 0210699839, 0210699840, 0210700114, 0210700115, 0210700117, 0210704385, 0210704389, 0210704392, 0210704394, 0210704551, 0210704553, 0210704554, 0210704557, 0210705521, 0210705616, 0210705618, 0210707767, 0210707771, 0210707778, 0210707779, 0210710678, 0210711961, 0210711963, 0210711964, 0210711965, 0210718222, 0210718228, 0210718232, 0210718234, 0210718235, 0210721585, 0210722591, 0210722592, 0210722593, 0210722594, 0210722595, 0210728060, 0210728061, 0210728062, 0210728063, 0210728064, 0210728065, 0210728066, 0210728797, 0210728798, 0210733886, 0210733888, 0210733893, 0210733894, 0210733895, 0210733901, 0210733902, 0210733903, 0210745522, 0210745523, 0210745524, 0210745564, 0210759331, 0210759332, 0210759333, 0210759440, 0210759443, 0210759444, 0210762016, 0210763719, 0210763817, 0210767542, 0210767550, 0210767552, 0210767553, 0210767593, 0210774988, 0210779252, 0210779253, 0210779254, 0210779255, 0210779256, 0210779257, 0210779258, 0210779259, 0210779356, 0210779357, 0210780500, 0210780501, 0210780502, 0210780580, 0210780581, 0210780582, 0210780583, 0210780584, 0210780585, 0210780586, 0210780587, 0210780588, 0210780589, 0210780590, 0210780599, 0210780680, 0210780681, 0210780682, 0210780683, 0210780686, 0210780687, 0210780688, 0210780689, 0210780722, 0210780925, 0210780927, 0210780934, 0210792653, 0210794022, 0210817450, 0210817451, 0210817624, 0210817639, 0210817820, 0210817821, 0210828371, 0210828372, 0210844375, 0210844378, 0210844661, 0210844662, 0210844663, 0210844664, 0210844665, 0210844666, 0210844667, 0210844668, 0210844669, 0210844670, 0210852515, 0210852517, 0210852912, 0210852913, 0210852914, 0210853939, 0210854052, 0210854053, 0210854054, 0210854055, 0210854074, 0210854151, 0210854180, 0210854181, 0210854206, 0210854247, 0210859436, 0210859437, 0210859439, 0210859940, 0210859941, 0210863739, 0210863848, 0210863849, 0210863850, 0210863851, 0210863852, 0210863853, 0210863854, 0210863855, 0210863856, 0210863857, 0210863858, 0210863859, 0210863880, 0210863881, 0210863882, 0210863883, 0210863884, 0210863885, 0210863886, 0210863888, 0210863889, 0210863890, 0210863891, 0210863892, 0210863893, 0210863894, 0210864471, 0210864472, 0210864473, 0210866082, 0210866083, 0210868253, 0210868254, 0210874176, 0210874177, 0210875387, 0210875388, 0210880357, 0210880359, 0210880415, 0210880416, 0210880504, 0210880505, 0210882547, 0210882550, 0210887739, 0210889140, 0210889142, 0210889143, 0210889144, 0210889146, 0210889147, 0210889148, 0210889149, 0210889150, 0210889151, 0210889152, 0210889304, 0210890066, 0210895028, 0210895032, 0210895033, 0210895035, 0210897155, 0210897156, 0210897157, 0210897238, 0210897239, 0210897320, 0210897416, 0210897600, 0210897610, 0210897614, 0210897619, 0210897640, 0210897641, 0210897642, 0210897643, 0210897644, 0210897645, 0210897646, 0210897647, 0210897648, 0210897649, 0210897651, 0210897652, 0210897653, 0210897655, 0210897657, 0210897658, 0210897659, 0210897681, 0210897682, 0210897683, 0210897684, 0210897685, 0210897687, 0210897688, 0210897689, 0210897691, 0210897693, 0210897694, 0210904350, 0210904469, 0210904471, 0210904472, 0210904473, 0210904474, 0210904475, 0210904476, 0210904477, 0210904478, 0210904479, 0210905640, 0210905641, 0210905642, 0210905643, 0210905644, 0210905645, 0210905646, 0210905647, 0210905648, 0210905649, 0210905650, 0210905651, 0210905652, 0210905653, 0210905654, 0210905655, 0210905661, 0210908595, 0210920219, 0210925071, 0210925072, 0210925470, 0210926882, 0210926883, 0210926898, 0210926899, 0210926920, 0210926921, 0210926922, 0210926923, 0210932436, 0210932439, 0210932520, 0210932522, 0210932524, 0210932527, 0210945491, 0210945858, 0210945859, 0210946577, 0210946578, 0210946579, 0210946800, 0210947240, 0210947241, 0210947242, 0210947243, 0210952136, 0210953214, 0210953224, 0210953225, 0210953226, 0210953228, 0210953229, 0210953230, 0210953231, 0210956484, 0210956485, 0210956486, 0210956487, 0210960780, 0210960781, 0210960865, 0210966223, 0210966224, 0210966225, 0210966928, 0210966929, 0210966930, 0210967462, 0210967683, 0210976784, 0210981567, 0210981573, 0210981575, 0210981576, 0210987109, 0210989922, 0210989924, 0210990325, 0210990327, 0210990996, 0210995659, 0210995687, 0210999960, 0211003221, 0211003222, 0211003223, 0211003224, 0211003225, 0211003226, 0211003229, 0211003230, 0211003401, 0211003402, 0211003403, 0211003404, 0211005122, 0211005249, 0211005250, 0211005251, 0211006843, 0211007086, 0211007087, 0211007088, 0211007089, 0211013099, 0211013102, 0211013103, 0211013104, 0211013105, 0211017308, 0211017310, 0211017311, 0211017312, 0211017313, 0211017314, 0211017315, 0211017316, 0211017317, 0211017318, 0211017319, 0211017340, 0211017341, 0211017343, 0211017344, 0211017345, 0211017346, 0211017347, 0211017349, 0211017351, 0211021131, 0211021132, 0211021369, 0211021373, 0211043596, 0211047113, 0211047156, 0211047159, 0211047564, 0211047566, 0211047567, 0211051594, 0211051599, 0211051642, 0211051645, 0211051649, 0211051652, 0211051654, 0211051740, 0211051742, 0211051745, 0211051746, 0211051747, 0211058455, 0211063670, 0211063671, 0211069878, 0211084970, 0211084971, 0211084972, 0211084973, 0211084974, 0211084975, 0211084976, 0211084977, 0211084979, 0211085360, 0211085361, 0211085362, 0211085363, 0211085364, 0211085366, 0211085367, 0211085368, 0211085399, 0211085460, 0211085461, 0211089722, 0211089723, 0211089724, 0211096939, 0211101101, 0211101102, 0211101103, 0211101104, 0211105686, 0211105687, 0211105688, 0211105689, 0211105690, 0211105692, 0211105693, 0211105694, 0211105695, 0211105696, 0211105752, 0211105753, 0211105754, 0211105760, 0211105761, 0211105762, 0211105763, 0211105764, 0211105765, 0211105766, 0211105767, 0211105768, 0211105769, 0211105770, 0211105771, 0211105772, 0211105773, 0211106116, 0211106117, 0211106118, 0211106119, 0211106780, 0211106781, 0211106782, 0211106783, 0211106784, 0211106785, 0211106786, 0211106787, 0211106788, 0211106791, 0211106792, 0211106793, 0211106794, 0211106795, 0211106796, 0211106797, 0211106798, 0211106799, 0211113217, 0211115301, 0211115302, 0211115303, 0211115306, 0211115309, 0211115314, 0211115322, 0211115329, 0211115458, 0211115463, 0211115466, 0211115467, 0211115472, 0211115478, 0211115497, 0211115520, 0211115522, 0211115524, 0211115525, 0211115526, 0211115561, 0211115566, 0211115568, 0211121871, 0211128335, 0211128336, 0211128337, 0211128338, 0211128339, 0211128380, 0211128381, 0211128382, 0211128383, 0211128384, 0211130466, 0211136092, 0211136257, 0211144947, 0211157540, 0211157542, 0211157543, 0211162842, 0211164138, 0211164259, 0211164720, 0211173454, 0211173455, 0211173456, 0211173457, 0211173773, 0211173774, 0211173775, 0211173776, 0211173777, 0211173778, 0211173779, 0211173794, 0211173795, 0211173796, 0211173797, 0211173798, 0211173799, 0211173820, 0211173821, 0211173822, 0211173823, 0211173824, 0211173825, 0211173826, 0211173827, 0211173828, 0211173840, 0211173841, 0211173842, 0211173843, 0211173844, 0211173845, 0211173846, 0211173847, 0211173848, 0211184040, 0211184041, 0211184042, 0211184043, 0211184271, 0211194528, 0211194529, 0211194530, 0211194531, 0211194533, 0211194534, 0211196352, 0211201346, 0211201347, 0211201348, 0211201349, 0211201350, 0211201351, 0211227998, 0211227999, 0211229004, 0211229010, 0211229492, 0211229493, 0211229494, 0211229495, 0211229499, 0211231300, 0211231301, 0211251779, 0211331117, 0211331118, 0211338299, 0211338604, 0211338613, 0211338615, 0211338619, 0211338622, 0211338628, 0211338633, 0211338634, 0211338636, 0211338638, 0211338680, 0211338681, 0211338682, 0211338683, 0211338684, 0211338685, 0211338686, 0211338687, 0211338688, 0211338689, 0211338700, 0211338701, 0211338703, 0211338705, 0211338707, 0211338709, 0211338710, 0211338711, 0211338716, 0211338718, 0211338742, 0211338744, 0211338746, 0211338755, 0211338758, 0211338759, 0211338761, 0211338763, 0211338766, 0211338768, 0211338769, 0211338770, 0211338771, 0211338772, 0211338773, 0211338961, 0211343295, 0211343296, 0211343297, 0211343298, 0211343299, 0211376855, 0211390406, 0211405706, 0211405707, 0211405708, 0211413690, 0211418652, 0211418653, 0211496695, 0211496709, 0211496710, 0211496711, 0211496712, 0211496713, 0211496716, 0211496717, 0211496718, 0211496804, 0211496805, 0211496806, 0211496807, 0211496810, 0211496814, 0211496817, 0211496860, 0211496863, 0211496864, 0211496865, 0211496867, 0211496869, 0211496871, 0211496873, 0211496874, 0211496878, 0211496884, 0211496885, 0211496886, 0211496891, 0211496892, 0211496893, 0211496894, 0211496895, 0211496896, 0211496897, 0211496898, 0211496899, 0211496940, 0211496942, 0211496944, 0211496948, 0211496949, 0211496951, 0211496952, 0211496953, 0211496960, 0211496968, 0211496969, 0211496970, 0211496972, 0211496973, 0211496974, 0211496975, 0211496976, 0211496977, 0211496978, 0211496980, 0211496981, 0211496982, 0211496983, 0211496984, 0211496985, 0211496986, 0211496987, 0211496988, 0211496989, 0211497000, 0211497001, 0211497003, 0211497005, 0211497006, 0211497007, 0211497008, 0211497009, 0211497011, 0211497065, 0211497066, 0211497392, 0211497395, 0211497397, 0211497398, 0211497560, 0211497563, 0211497567, 0211497570, 0211497571, 0211502588, 0211518958, 0211519420, 0211519442, 0211519444, 0211522457, 0211522458, 0211522459, 0211523422, 0211523434, 0211523436, 0211523438, 0211523439, 0211523560, 0211523561, 0211523600, 0211523601, 0211523603, 0211523604, 0211523605, 0211523608, 0211523610, 0211523612, 0211523613, 0211523614, 0211523615, 0211528900, 0211528901, 0211528902, 0211528903, 0211528904, 0211528905, 0211529219, 0211531662, 0211531664, 0211531667, 0211531767, 0211536556, 0211536724, 0211536727, 0211536730, 0211536732, 0211536733, 0211536736, 0211536739, 0211536763, 0211536766, 0211536768, 0211544557, 0211545625, 0211552121, 0211552122, 0211552123, 0211688941, 0211688944, 0211688945, 0211695120, 0211695121, 0211695122, 0211695123, 0211695124, 0211695125, 0211701573, 0211701574, 0211701575, 0211701576, 0211725977, 0211748831, 0211748832, 0211748833, 0211748834, 0211748835, 0211748836, 0211748837, 0211748838, 0211748839, 0211750961, 0211750962, 0211750963, 0211750964, 0211750965, 0211750966, 0211750972, 0211750973, 0211750974, 0211751021, 0211751022, 0211751023, 0211751024, 0211751025, 0211751026, 0211751027, 0211751028, 0211751029, 0211751030, 0211751031, 0211751032, 0211751033, 0211751040, 0211751041, 0211751042, 0211751043, 0211751044, 0211751060, 0211751080, 0211751081, 0211751082, 0211751083, 0211751116, 0211751117, 0211751118, 0211751119, 0211751130, 0211751131, 0211751132, 0211751133, 0211751134, 0211751135, 0211751136, 0211751137, 0211751138, 0211751139, 0211751360, 0211751361, 0211751362, 0211751363, 0211751364, 0211751365, 0211751366, 0211751367, 0211751368, 0211751369, 0211751370, 0211751371, 0211751372, 0211751373, 0211751374, 0211751520, 0211751521, 0211751522, 0211751523, 0211754907, 0211754908, 0211754909, 0211754910, 0211754911, 0211754912, 0211754913, 0211754914, 0211754915, 0211754916, 0211754917, 0211765977, 0211766330, 0211766331, 0211766332, 0211766333, 0211766334, 0211766335, 0211766336, 0211766337, 0211766338, 0211766339, 0211766463, 0211766466, 0211766468, 0211766469, 0211766472, 0211766475, 0211766476, 0211766478, 0211766479, 0211766480, 0211766481, 0211766482, 0211766486, 0211766487, 0211766488, 0211766489, 0211766490, 0211766500, 0211766501, 0211766502, 0211766510, 0211766512, 0211766513, 0211766514, 0211766515, 0211766516, 0211766517, 0211766518, 0211766540, 0211766541, 0211782283, 0211782284, 0211782285, 0211782286, 0211782287, 0211782392, 0211817536, 0211817537, 0211817538, 0211817539, 0211818620, 0211818621, 0211818622, 0211818623, 0211818624, 0211818625, 0211818626, 0211818627, 0211818628, 0211818629, 0211818630, 0211818631, 0211818632, 0211818633, 0211818634, 0211818635, 0211818636, 0211818637, 0211818638, 0211818639, 0211818680, 0211818681, 0211818682, 0211818683, 0211818684, 0211818685, 0211818686, 0211818687, 0211818688, 0211818689, 0211818690, 0211818691, 0211818692, 0211818693, 0211818694, 0211818695, 0211818696, 0211818697, 0211818698, 0211824073, 0211824074, 0211824075, 0211824076, 0211824274, 0211824275, 0211824276, 0211824277, 0211824278, 0211824279, 0211825093, 0211825094, 0211825095, 0211825096, 0211848465, 0211848466, 0211889349, 0211889350, 0211889351, 0211889352, 0211889353, 0211889355, 0211889356, 0211889357, 0211889358, 0211889359, 0211889660, 0211889661, 0211889662, 0211889664, 0211889665, 0211889666, 0211889667, 0211889668, 0211889669, 0211889670, 0211889671, 0211889672, 0211889673, 0211889674, 0211889675, 0211889676, 0211889677, 0211889678, 0211889679, 0211889680, 0211889681, 0211889682, 0211889683, 0211889684, 0211889690, 0211889691, 0211889692, 0211889693, 0211889694, 0211889695, 0211889696, 0211889697, 0211889698, 0211889699, 0211889700, 0211889701, 0211889702, 0211889703, 0211889704, 0211889705, 0211889708, 0211889709, 0211889710, 0211889711, 0211889712, 0211889713, 0211889714, 0211889715, 0211889716, 0211889717, 0211889718, 0211889719, 0211889720, 0211889721, 0211889722, 0211889723, 0211889724, 0211889725, 0211889726, 0211889727, 0211889728, 0211889729, 0211889730, 0211889731, 0211889733, 0211889734, 0211889735, 0211889736, 0211889737, 0211889738, 0211889739, 0211902813, 0211902816, 0211902817, 0211906783, 0211907140, 0211907142, 0211907144, 0211907145, 0211907146, 0211907814, 0211907815, 0211912148, 0211912149, 0211912150, 0211912151, 0211912152, 0211912153, 0211912154, 0211912155, 0211912156, 0211912157, 0211912158, 0211912159, 0211919855, 0211919858, 0211919859, 0211926970, 0211927057, 0211933649, 0211933651, 0211933652, 0211933653, 0211933655, 0211933656, 0211933657, 0211933658, 0211933659, 0211935680, 0211935695, 0211943841, 0211943846, 0211951937, 0211951938, 0211953023, 0211953424, 0211958807, 0211965804, 0211965886, 0211965888, 0211965889, 0211977044, 0211977055, 0211977057, 0211977100, 0211977101, 0211977103, 0211977105, 0211977139, 0211977162, 0211986392, 0212002931, 0212002932, 0212002933, 0212005951, 0212005956, 0212005999, 0212010103, 0212010140, 0212010141, 0212010142, 0212010143, 0212010145, 0212010146, 0212010147, 0212010148, 0212010150, 0212010151, 0212010153, 0212010154, 0212010155, 0212010156, 0212010157, 0212010158, 0212010159, 0212017010, 0212017608, 0212017963, 0212035680, 0212035683, 0212046316, 0212046317, 0212046319, 0212047360, 0212047362, 0212047368, 0212047371, 0212047375, 0212047379, 0212047408, 0212047410, 0212047414, 0212047418, 0212047422, 0212047425, 0212047429, 0212047433, 0212047439, 0212047443, 0212047445, 0212047451, 0212069285, 0212069286, 0212069287, 0212069288, 0212069289, 0212069290, 0212069291, 0212069292, 0212069293, 0212069294, 0212069295, 0212069296, 0212069297, 0212069298, 0212069299, 0212069720, 0212069721, 0212069722, 0212069723, 0212069724, 0212069725, 0212096739, 0212097296, 0212097297, 0212097360, 0212097361, 0212109948, 0212109949, 0212117007, 0212119848, 0212122682, 0212122683, 0212131168, 0212145541, 0212145542, 0212145543, 0212145544, 0212145545, 0212155823, 0212155824, 0212155825, 0212160252, 0212160253, 0212160254, 0212160256, 0212160257, 0212160258, 0212163451, 0212165301, 0212176424, 0212182782, 0212182783, 0212182827, 0212182828, 0212182830, 0212182837, 0212182838, 0212182839, 0212183100, 0212183101, 0212183102, 0212183103, 0212183104, 0212188937, 0212194130, 0212194640, 0212194642, 0212194643, 0212194647, 0212194649, 0212194651, 0212194652, 0212194656, 0212194658, 0212194700, 0212194703, 0212194704, 0212194705, 0212194706, 0212194707, 0212194708, 0212198744, 0212224183, 0212224558, 0212224701, 0212224704, 0212224707, 0212230714, 0212233991, 0212233993, 0212233994, 0212233995, 0212242852, 0212243205, 0212243206, 0212243207, 0212243209, 0212243210, 0212263057, 0212263116, 0212263118, 0212263119, 0212263181, 0212263182, 0212263183, 0212263184, 0212263185, 0212263696, 0212287387, 0212287388, 0212289537, 0212289538, 0212289539, 0212289720, 0212291473, 0212296918, 0212302723, 0212304702, 0212304703, 0212310143, 0212310147, 0212323653, 0212324003, 0212345751, 0212345752, 0212345753, 0212345754, 0212345756, 0212349303, 0212356380, 0212356381, 0212356382, 0212358961, 0212358962, 0212358963, 0212363742, 0212363743, 0212365932, 0212370808, 0212370811, 0212373369, 0212373370, 0212373373, 0212374780, 0212374781, 0212374782, 0212379063, 0212379064, 0212382454, 0212382455, 0212387370, 0212387371, 0212406943, 0212415450, 0212415451, 0212415452, 0212415453, 0212415454, 0212415477, 0212415770, 0212415771, 0212415772, 0212415773, 0212425555, 0212425556, 0212425557, 0212425558, 0212461288, 0212461289, 0212461290, 0212473934, 0212473935, 0212473936, 0212473937, 0212473938, 0212473939, 0212474776, 0212474777, 0212474778, 0212474779, 0212479080, 0212479083, 0212479084, 0212479085, 0212479086, 0212479087, 0212479089, 0212479090, 0212479091, 0212479092, 0212479093, 0212479094, 0212501660, 0212501662, 0212501664, 0212501668, 0212511955, 0212511956, 0212531176, 0212532545, 0212532546, 0212532877, 0212533725, 0212533856, 0212533857, 0212534210, 0212534211, 0212534212, 0212534213, 0212534214, 0212534215, 0212534216, 0212534649, 0212534650, 0212535040, 0212546590, 0212546591, 0212546592, 0212546593, 0212546594, 0212546595, 0212546596, 0212553862, 0212570047, 0212574149, 0212574150, 0212621427, 0212621428, 0212621429, 0212625777, 0212625778, 0212625779, 0212629517, 0212633964, 0212633965, 0212633966, 0212633967, 0212633968, 0212633970, 0212633971, 0212633972, 0212633974, 0212642491, 0212648053, 0212648294, 0212649252, 0212649253, 0212649254, 0212649255, 0212649256, 0212656407, 0212656408, 0212656409, 0212656410, 0212656411, 0212656412, 0212659326, 0212659327, 0212659328, 0212659331, 0212659332, 0212659333, 0212659335, 0212659337, 0212666088, 0212666089, 0212666090, 0212668632, 0212672719, 0212693924, 0212693925, 0212693926, 0212693927, 0212693928, 0212693929, 0212693930, 0212695373, 0212702903, 0212711320, 0212717800, 0212717801, 0212717802, 0212717803, 0212717804, 0212717805, 0212717806, 0212717807, 0212717808, 0212717809, 0212717810, 0212717811, 0212717812, 0212717813, 0212717814, 0212717815, 0212717816, 0212717817, 0212717818, 0212717819, 0212717900, 0212717901, 0212717902, 0212717903, 0212717904, 0212717905, 0212717906, 0212726757, 0212726758, 0212726759, 0212727360, 0212727361, 0212727362, 0212727363, 0212727365, 0212727366, 0212727367, 0212727368, 0212741333, 0212741334, 0212741335, 0212741336, 0212741337, 0212741338, 0212741339, 0212741460, 0212741461, 0212741462, 0212741463, 0212741464, 0212741465, 0212741466, 0212741467, 0212741468, 0212741469, 0212741470, 0212741471, 0212741472, 0212741473, 0212741474, 0212741475, 0212741476, 0212741477, 0212741478, 0212741479, 0212741520, 0212741521, 0212741522, 0212741523, 0212741524, 0212741525, 0212741526, 0212741527, 0212742323, 0212742324, 0212742325, 0212742326, 0212742327, 0212742328, 0212742329, 0212742330, 0212754196, 0212754197, 0212754198, 0212754199, 0212754320, 0212754322, 0212754323, 0212754324, 0212754325, 0212754397, 0212754561, 0212760990, 0212760992, 0212774143, 0212774144, 0212776602, 0212782497, 0212782642, 0212782643, 0212782644, 0212782645, 0212782646, 0212782648, 0212782649, 0212782650, 0212782651, 0212797473, 0212797479, 0212797500, 0212804760, 0212804764, 0212804767, 0212823605, 0212823606, 0212823607, 0212829143, 0212833423, 0212923529, 0212923530, VA129W9, VA13DAB, VA13JYA, VA14DTP, VA14DZ2, VA14F4D, VA14P8T, VA14P8TV01, VA159T2, VA15K33, VA15U9R, VA15U9RV01, VA15U9RV02, VA17NQ2, VA17T17, VA17ZER, VA19VVP, VA1A5UJ, VA1ABDD, VA1AXXE, VA1B6M8, VA1B7XW, VA1BL22, VA1BXGF, VA1C4BK, VA1C8E2, VA1CCYS, VA1CCZ9, VA1CN9M, VA1CNA0, VA1CPXN, VA1CU14, VA1CWLH, VA1DBVE, VA1DCHZ, VA1ES10.
FEI Number 2182207
Recalling Firm/
Manufacturer
Medtronic Neuromodulation
7000 Central Ave
Minneapolis MN 55421-1241
For Additional Information ContactMedtronic Neuromodulation Technical Services
800-707-0933
Manufacturer Reason
for Recall
Medtronic received reports that the DBS depth stop did not adequately secure to the lead, which can result in DBS lead placement beyond the intended target.
FDA Determined
Cause 2
Process control
ActionMedtronic sent an Urgent Medical Device Correction letter dated March 2017, to all affected consignees with important safety information on the use of the device. No product retrieval is involved at this time. The letter was addressed to Neurosurgeon. The letter described the product involved in the recall, provided information on the "Background", "Potential Risk to Patients", "Actions", & "Additional Information". For questions, contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 weekdays 7 am - pm CDT.
Quantity in Commerce86,024
DistributionWorldwide Distribution - US including AL, AR, AZ, CA,CO, CT, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN,M MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. Foreign Distribution: India, ANZ, Taiwan, Hong Kong, Canada, China, Japan, Korea, Malaysia, Philippines, Singapore, Thailand, Indonesia, Vietnam, Albania, Algeria, Austria, Belarus, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, India, Iran, Iraq, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, United Kingdom, Argentina, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Mexico, Panama, Peru, Puerto Rico, Uruguay, and Venezuela.
Total Product Life CycleTPLC Device Report