Class 2 Device Recall Medtronic Aptus(TM) HeliFX(TM) Thoracic EndoAnchor(TM) System HeliFX Guide 42 mm

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medtronic Aptus(TM) HeliFX(TM) Thoracic EndoAnchor(TM) System HeliFX Guide 42 mmsee related information
Date Initiated by FirmSeptember 13, 2017
Create DateSeptember 27, 2017
Recall Status1 Terminated 3 on May 18, 2020
Recall NumberZ-3227-2017
Recall Event ID 78127
510(K)NumberK171427 
Product Classification Endovascular suturing system - Product Code OTD
ProductMedtronic Aptus Heli-FX Thoracic EndoAnchor System Heli-FX Guide 42 mm, REF HG-18-90-42, STERILE EO, Rx Only The Heli-FX EndoAnchor System is intended to provide fixation and sealing between endovascular aortic grafts and the native artery. The Heli-FX System is indicated for use in patients who endovascular grafts have exhibited migration or endoleak, or are at risk for such complications, in whom augmented radial fixation and or sealing is required or regain or maintain adequate aneurysm exclusion.
Code Information UDI 00763000006679, Lot Number 0008674156 UDI 00763000006655, Lot Number 0008674157
FEI Number 3001452571
Recalling Firm/
Manufacturer
Medtronic Vascular, Inc.
3576 Unocal Pl
Santa Rosa CA 95403-1774
For Additional Information ContactKristen Hayward
508-261-8000
Manufacturer Reason
for Recall
It was determined that the deflection length indicated on the Guide catheter handle does not match the label on the box and sterile packaging for two lots.
FDA Determined
Cause 2
Labeling Change Control
ActionThe firm initiated their recall on September 13, 2017, by letter. The letter requested the consignee take the following actions: "1. Identify and quarantine all unused affected product as listed in your inventory. 2. Return all unused affected list product in your inventory to Medtronic. Contact Medtronic Customer Service at 1-800-854-3570 to initiate a product return and credit. Your local Medtronic Representative can assist you in the return and replacement of this product as necessary. 3. Complete the enclosed Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality or scan and email to RS.CFQFCA@medtronic .com. This notice needs to be passed on to all those who need to be aware within your organization where the potentially affected devices have been transferred." For further questions, please call (508) 261-8000.
Quantity in Commerce20 units
DistributionUS Distribution to the states of : AZ, CA, FL, GA, MA, MI, MN and NH.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = OTD
-
-