| | Class 1 Device Recall DORO LUCENT Base Unit |  |
| Date Initiated by Firm | July 05, 2018 |
| Date Posted | September 17, 2018 |
| Recall Status1 |
Terminated 3 on November 19, 2020 |
| Recall Number | Z-2970-2018 |
| Recall Event ID |
80628 |
| 510(K)Number | K032331 |
| Product Classification |
Holder, head, neurosurgical (skull clamp) - Product Code HBL
|
| Product | DORO LUCENT Base Unit, REF 1101.021 Headrest System used in open and percutaneous craniotomies and spinal surgeries |
| Code Information |
Serial Numbers: 1704002 1704004 1704005 1704007 1704011 1704041 1704010 1704019 1704012 1704014 1704016 1704017 1704021 1704024 1704025 1704036 1704038 1704018 1704008 1704034 1704026 1704027 1704028 1704029 1704030 1704032 1704035 1704006 1704037 1704039 |
| FEI Number |
3003923584
|
Recalling Firm/ Manufacturer |
Pro-Med Instruments Gmbh
Botzinger Str. 38
Freiburg Im Breisgau
Germany
|
Manufacturer Reason for Recall | The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system. |
FDA Determined Cause 2 | Component change control |
| Action | The firm initiated the recall by email on 07/05/2018.
This stage is only relevant for hospitals / users who are in possession with concerned products of two different revisions (mix-up):
a) Discontinue the use of the combination and quarantine the concerned products
b) Contact pro med instruments for further instructions relating to temporary replacement possibilities |
| Quantity in Commerce | 30 units |
| Distribution | Worldwide distribution to US states of CA, TX, TN, MS, GA, WV, and OH, Qatar, India, UK, Singapore, Australia, Germany, South Africa, Italy, and Japan |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = HBL
|
|
|
|