| | Class 2 Device Recall GE Healthcare Revolution EVO, Optima CT660, Optima CT670, Optima CT680 |  |
| Date Initiated by Firm | April 27, 2018 |
| Create Date | August 17, 2018 |
| Recall Status1 |
Terminated 3 on August 21, 2020 |
| Recall Number | Z-2870-2018 |
| Recall Event ID |
80563 |
| 510(K)Number | K131576 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | GE Healthcare Revolution EVO, Optima CT660, Optima CT670, Optima CT680. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Sold under the following product names:
CT GOLDSEAL LIGHTSPEED ULTRA, CT OPTIMA 660, CT OPTIMA 660 M40 3.5 LONG BJG, CT OPTIMA 660 M40 GT1700, CT OPTIMA 660 M40 GT1700 BEIJI, CT OPTIMA 660 M40 GT1700 HINO, CT OPTIMA 660 M40 GT2000 BEIJI, CT OPTIMA 660 M40 GT2000 HINO, CT OPTIMA CT 660, CT OPTIMA CT660, CT OPTIMA CT660 M4, CT OPTIMA CT660 M40, CT REVOLUTION EVO 3.5 LONG HIN, CT REVOLUTION EVO VT2000 HINO, CT660 MAR UPGRADE, MOBILE OPTIMA CT 660, OPTI660_A_G_2.0, OPTI660_M40_G, OPTI660_M40_L, OPTIMA 660, OPTIMA 660 128 FOR TW- DIRECT, OPTIMA 660 GT2000 M40 HINO MULTISLICE, OPTIMA 660 M40, OPTIMA 660 M40, OPTIMA 660 M40 3.5 HVY BJG, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 3.5 LONG HINO, OPTIMA 660 M40 3.5 MID, OPTIMA 660 M40 3.5 MID BJG, OPTIMA 660 M40 3.5 MID HINO, OPTIMA 660 M40 3.5/6, OPTIMA 660 M40 3.5/6 HVY BJG, OPTIMA 660 M40 3.5/6 HVY HINO, OPTIMA 660 M40 3.5/6 HVY WAUK, OPTIMA 660 M40 3.5/6 LONG, OPTIMA 660 M40 3.5/6 LONG B, OPTIMA 660 M40 3.5/6 LONG BJG, OPTIMA 660 M40 3.5/6 LONG HINO, OPTIMA 660 M40 3.5/6 LONG WAUK, OPTIMA 660 M40 3.5/6 MID BJ, OPTIMA 660 M40 3.5/6 MID BJG, OPTIMA 660 M40 3.5/6 MID HI, OPTIMA 660 M40 3.5/6 MID HINO, OPTIMA 660 M40 3.5/6 MID WA, OPTIMA 660 M40 3.5/6 MID WAUK, OPTIMA 660 M40 GT1700, OPTIMA 660 M40 GT1700 BEIJING, OPTIMA 660 M40 GT1700 HINO, OPTIMA 660 M40 GT1700 WAUK, OPTIMA 660 M40 GT2000 B, OPTIMA 660 M40 GT2000 BEIJI, OPTIMA 660 M40 GT2000 BEIJING, OPTIMA 660 M40 GT2000 HINO, OPTIMA 660 M40 GT2000 WAUK, OPTIMA 660 M40 MOBILE BEIJING, OPTIMA CT 660, OPTIMA CT660 ASIR, OPTIMA CT660 ASIR 128 M RU, OPTIMA CT660 FREEDOM MD HN2, OPTIMA CT660 M40 GT1700, OPTIMA CT660 M40 GT2000, OPTIMA CT660 M40 GT2000 BEIJIN, OPTIMA CT660 M40 GT2000 HINO, OPTIMA CT660 P2.5, OPTIMA CT660 P25 UPG KIT, OPTIMA CT660 PRO MID, OPTIMA CT660 SE, OPTIMA CT660 SE GT 1700, OPTIMA CT680, OPTIMA CT680 KL64 GT1700 BHW, REVOLUTION EVO, REVOLUTION EVO 3.5 HVY BJG, REVOLUTION EVO 3.5 HVY HINO, REVOLUTION EVO 3.5 HVY WAUK, REVOLUTION EVO 3.5 LONG BJG, REVOLUTION EVO 3.5 LONG HINO, REVOLUTION EVO 3.5 LONG WAUK, REVOLUTION EVO 3.5 MID BJG, REVOLUTION EVO 3.5 MID HINO, REVOLUTION EVO 3.5 MID WAUK, REVOLUTION EVO 3.5MID HINO CT, REVOLUTION EVO 3.6 CHN HINO, REVOLUTION EVO 3.6 HVY BJG, REVOLUTION EVO 3.6 HVY HINO, REVOLUTION EVO 3.6 HVY WAUK, REVOLUTION EVO 3.6 LONG, REVOLUTION EVO 3.6 LONG BJG, REVOLUTION EVO 3.6 LONG HIN, REVOLUTION EVO 3.6 LONG HINO, REVOLUTION EVO 3.6 LONG WAUK, REVOLUTION EVO 3.6 MID BJG, REVOLUTION EVO 3.6 MID HINO, REVOLUTION EVO 3.6 MID WAUK, REVOLUTION EVO 3.6B, REVOLUTION EVO 3.6B HVY HINO, REVOLUTION EVO 3.6B LONG, REVOLUTION EVO 3.6B LONG BJG, REVOLUTION EVO 3.6B LONG HINO, REVOLUTION EVO 3.6B MID BJG, REVOLUTION EVO 3.6B MID HINO, REVOLUTION EVO GT1700, REVOLUTION EVO GT2000, REVOLUTION EVO GT2000 HINO, REVOLUTION EVO GT2000X (HINO), REVOLUTION EVO GT2000X HINO, REVOLUTION EVO VT1700 BEIJING, REVOLUTION EVO VT1700 HINO, REVOLUTION EVO VT1700 WAUK, REVOLUTION EVO VT2000 BEIJING, REVOLUTION EVO VT2000 HINO, REVOLUTION EVO VT2000 WAUK, REVOLUTION EVO VT2000X BEIJING, REVOLUTION EVO VT2000X HINO, REVOLUTION EVO VT2000X WAUK, REVOLUTION EVO WITH GT1700V TA, REVOLUTION EVO WITH GT2000 TAB |
| Code Information |
Model Number, Serial Number (System ID): 5454001-260, CBCGG1600028HM (817261ULTRA); 5454001-240, 00000389996HM0 (83021861077017); 5454001-40, 00000074499YC5 (850210882); 5454001-40, 00000075588YC4 (850210905); 5454001-40, 00000069668YC2 (850210888); 5454001-40, 00000075080YC2 (850210904); 5454001-40, 00000071456YC8 (850210894); 5454001-40, 00000077687YC2 (850210910); 5454001-40, 00000073633YC0 (850210902); 5454001-40, 00000073638YC9 (850210900); 5454001-40, 40GAN1600013YC (850210908); 5454001-240, 00000424383HM8 (HU1113CT01); 5454001-240, 00000391751HM5 (DPN26338); 5454001-240, 00000387390HM8 (UE535A128); 5454001-240, 00000394185HM3 (NEV44709); 5454001-240, 00000394592HM0 (NEV44708); 5454001-240, 00000395002HM9 (NPV79344); 5454001-240, 00000415694HM9 (EP223318); 5454001-240, 00000414003HM4 (CS1346CT01); 5454001-240, 00000411639HM8 (CS1046CT01); 5454001-240, 00000385138HM3 (CS1335CT01); 5454001-240, 00000387391HM6 (H2021CT15); 5454001-240, 00000394590HM4 (H2049CT10); 5454001-240, 00000404914HM4 (H2016CT21); 5454001-240, 00000391756HM4 (H2017CT11); 5454001-240, 00000403534HM1 (HU1015CT03); 5454001-240, 00000411637HM2 (HU1158CT02); 5454001-240, 00000405579HM4 (RO1098CT02); 5454001-240, 00000405574HM5 (RO1303CT01); 5454001-40, 00000066732YC9 (RO1188CT01); 5454001-240, 00000420282HM6 (SK1042CT01); 5454001-240, 00000370111HM7 (MEX70100); 5454001-40, 00000063901YC3 (CS1085CT01); 5454001-40, 00000066209YC8 (RO1115CT03); 5454001-40, 00000061249YC9 (RU8891CT01); 5454001-240, 00000418621HM9 (CS1042CT02); 5454001-240, 00000419606HM9 (HU1012CT03); 5454001-240, 00000418622HM7 (600129CT01); 5454001-240, 00000389555HM4 (RO1042CT02); 5454001-240, 00000389550HM5 (RO1152CT02); 5454001-240, 00000385662HM2 (RO1271CT01); 5454001-40, 00000065926YC8 (RO1257CT01); 5454001-40, 00000067504YC1 (BG4693CT02); 5454001-40, 00000060793YC7 (DPN87723); 5454001-40, 00000068716YC0 (FPG39200); 5454001-41, 00000068076YC9 (82421290094); 5454001-230, 000000461669HM4 (82421160460); 5454001-41, 41GAN1600002YC (82421160466); 5454001-41, 00000074252YC8 (82421160438); 5454001-41, 00000073883YC1 (82421160436); 5454001-41, 00000074477YC1 (82421070440); 5454001-41, 00000074475YC5 (82421010361); 5454001-41, 00000068080YC1 (82421010346); 5454001-41, 00000073294YC1 (82421120546); 5454001-41, 00000073647YC0 (82421030444); 5454001-41, 00000068078YC5 (82421230363); 5454001-40, 00000060592YC3 (828210097); 5454001-41, 00000073645YC4 (82421111107); 5454001-41, 00000076283YC1 (82421111116); 5454001-41, 41GAN1600004YC (82421111133); 5454001-41, 00000076282YC3 (82421111114); 5454001-41, 41GAN1600005YC (82421111134); 5454001-41, 41GAN1700003YC (82421111140); 5454001-41, 00000076577YC6 (82421111121); 5454001-41, 00000076279YC9 (82421111111); 5454001-41, 41GAN1600008YC (82421040394); 5454001-41, 00000078206YC0 (82421290109); 5454001-41, 00000076783YC0 (82421290110); 5454001-41, 00000074478YC9 (82421160446); 5454001-41, 00000074255YC1 (82421160448); 5454001-41, 00000077952YC0 (82421160458); 5454001-41, 00000078205YC2 (82421180160); 5454001-41, 41GAN1600010YC (82421240200); 5454001-41, 00000074476YC3 (82421070441); 5454001-41, 41GAN1700008YC (82421070457); 5454001-41, 41GAN1600006YC (82421010379); 5454001-41, 41GAN1700001YC (82421010382); 5454001-41, 00000077954YC6 (82421090592); 5454001-41, 00000078651YC7 (82421090589); 5454001-41, 00000075364YC0 (82421120557); 5454001-41, 00000073646YC2 (82421120550); 5454001-41, 00000076280YC7 (82421120569); 5454001-41, 00000076782YC2 (82421200286); 5454001-41, 00000076781YC4 (82421020220); 5454001-41, 00000077562YC7 (82421020223); 5454001-41, 41GAN1600011YC (82421030463); 5454001-41, 00000075596YC7 (82421030448); 5454001-41, 41GAN1700007YC (82421100936); 5454001-41, 41GAN1600007YC (82421100908); 5454001-41, 00000073338YC6 (82421100856); 5454001-41, 00000074250YC2 (82421100859); 5454001-41, 00000073293YC3 (82421130182); 5454001-41, 00000075363YC2 (82421130183); 5454001-41, 41GAN1700006YC (82421130196); 5454001-41, 00000077953YC8 (82421310295); 5454001-41, 00000076281YC5 (82421230402); 5454001-41, 41GAN1700004YC (82421050123); 5454001-41, 00000074253YC6 (82421050116); 5454001-41, 00000074254YC4 (82421050117); 5454001-41, 41GAN1600009YC (82421050121); 5454001-41, 00000075857YC3 (82421270166); 5454001-41, 00000076576YC8 (82421270168); 5454001-41, 00000073885YC6 (82421260022); 5454001-41, 41GAN1600003YC (82421140419); 5454001-41, 00000070639YC0 (82421220132); 5454001-41, 00000069698YC9 (82421150165); 5454001-41, 00000068082YC7 (82421290100); 5454001-41, 00000071624YC1 (82421290103); 5454001-41, 00000068709YC5 (82421290098); 5454001-41, 00000072375YC9 (82421160431); 5454001-41, 00000068079YC3 (82421160416); 5454001-41, 00000071623YC3 (82421240182); 5454001-41, 00000072376YC7 (82421240174); 5454001-41, 00000070286YC0 (82421070428); 5454001-41, 00000072739YC6 (82421090564); 5454001-41, 00000070638YC2 (82421210345); 5454001-41, 00000072543YC2 (82421210352); 5454001-41, 00000068708YC7 (82421120507); 5454001-41, 00000068075YC1 (82421120499); 5454001-41, 00000072542YC4 (82421120533); 5454001-41, 00000068081YC9 (82421200262); 5454001-41, 00000075856YC5 (82421200273); 5454001-41, 00000072541YC6 (82421200274); 5454001-41, 00000072082YC1 (82421020214); 5454001-41, 00000072243YC9 (82421020215); 5454001-41, 00000074479YC7 (82421020213); 5454001-41, 00000070457YC7 (82421100818); 5454001-41, 00000073292YC5 (82421310246); 5454001-41, 00000074251YC0 (82421310284); 5454001-41, 00000072377YC5 (82421230380); 5454001-41, 00000070287YC8 (82421250176); 5454001-41, 00000071114YC3 (82421250181); 5454001-41, 00000068757YC4 (82421140358); 5454001-41, 00000072374YC2 (82421140370); 5454001-60, 00000069442YC2 (HU1098CT01); 5454001-60, 00000062894YC1 (HU1099CT06); 5454001-240, 00000406014HM1 (X868131011); 5454001-40, 00000066337YC7 (A1765916); 5454001-240, 00000436951HM8 (10583CTM04); 5454001-40, 00000070242YC3 (OH0167); 5454001-40, 00000060300YC1 (OJ0012); 5454001-40, 00000072692YC7 (OJ0034); 5454001-40, 00000064459YC1 (OJ0022); 5454001-40, 00000066656YC0 (OJ1023); 5454001-40, 00000067859YC9 (OJ1027); 5454001-40, 00000070845YC3 (OJ1040); 5454001-40, 00000077686YC4 (OJ1044); 5454001-40, 00000059194YC1 (OJ1009); 5454001-40, 00000069398YC6 (OJ1029); 5454001-40, 00000070038YC5 (OJ1030); 5454001-40, 00000078192YC2 (OJ1047); 5454001-40, 00000078193YC0 (OJ1046); 5454001-40, 40GAN1600011YC (OJ1051); 5454001-40, 00000061244YC0 (OJ0014); 5454001-40, 00000064505YC1 (OJ1018); 5454001-40, 00000072530YC9 (OJ1036); 5454001-40, 00000062560YC8 (OJ0017); 5454001-40, 00000066730YC3 (OJ1026); 5454001-40, 00000077688YC0 (OJ0039); 5454001-40, 00000077943YC9 (OJ0037); 5454001-40, 00000057152YC1 (OJ0001); 5454001-40, 00000070417YC1 (OJ0050); 5454001-40, 00000071458YC4 (OJ1060); 5454001-40, 00000057832YC8 (OJ1001); 5454001-40, 00000057833YC6 (OJ0004); 5454001-40, 00000058397YC1 (OJ0002); 5454001-40, 00000060939YC6 (OJ1015); 5454001-40, 00000070037YC7 (OJ1031); 5454001-40, 00000059383YC0 (OJ1011); 5454001-40, 00000070846YC1 (OJ0027); 5454001-40, 00000077282YC2 (OJ0036); 5454001-40, 00000060447YC0 (OJ0015); 5454001-40, 00000060593YC1 (OJ0020); 5454001-40, 00000062519YC4 (OJ0018); 5454001-40, 00000063152YC3 (OJ1020); 5454001-40, 00000069865YC4 (OJ0031); 5454001-40, 00000070849YC5 (OJ0033); 5454001-40, 00000070851YC1 (OJ0032); 5454001-40, 00000076548YC7 (OJ0035); 5454001-40, 00000057672YC8 (OJ0003); 5454001-40, 00000058706YC3 (OJ0005); 5454001-40, 00000058879YC8 (OJ1010); 5454001-40, 00000061774YC6 (OJ0016); 5454001-40, 00000071089YC7 (OJ1032); 5454001-40, 00000075587YC6 (OJ1042); 5454001-40, 00000067508YC2 (OJ0025); 5454001-40, 00000071453YC5 (OJ0028); 5454001-40, 00000060594YC9 (OJ0009); 5454001-40, 00000065206YC5 (OJ1017); 5454001-40, 00000067494YC5 (OJ1025); 5454001-40, 00000064210YC8 (OJ0030); 5454001-40, 00000065210YC7 (OJ1022); 5454001-40, 00000067887YC0 (OJ0026); 5454001-40, 00000059197YC4 (OJ0006); 5454001-40, 00000070416YC3 (OJ1033); 5454001-40, 00000071088YC9 (OJ1050); 5454001-40, 00000072532YC5 (OJ0055); 5454001-40, 00000074498YC7 (OJ1055); 5454001-40, 00000059193YC3 (OJ1007); 5454001-40, 00000059872YC2 (OJ1008); 5454001-40, 00000072235YC5 (OJ1035); 5454001-40, 00000063388YC3 (OJ1016); 5454001-40, 00000072531YC7 (OJ1037); 5454001-40, 00000073318YC8 (OJ1039); 5454001-40, 00000074649YC5 (OJ1045); 5454001-40, 00000065207YC3 (OJ1019); 5454001-40, 00000063902YC1 (OJ1013); 5454001-40, 00000064039YC1 (OJ0023); 5454001-40, 00000078634YC3 (OJ1049); 5454001-40, 00000061247YC3 (OJ0013); 5454001-40, 00000064634YC9 (OJ1021); 5454001-40, 00000072529YC1 (OJ0029); 5454001-40, 00000072534YC1 (OJ1038); 5454001-40, 00000075082YC8 (OJ1077); 5454001-40, 00000077283YC0 (OJ1043); 5454001-40, 00000059012YC5 (OJ1100); 5454001-40, 00000071087YC1 (OJ1088); 5454001-40, 00000062227YC4 (OJ1014); 5454001-40, 00000065204YC0 (OJ0024); 5454001-40, 00000058878YC0 (OJ1004); 5454001-40, 00000066935YC8 (OJ1024); 5454001-40, 00000073315YC4 (OJ1041); 5454001-40, 00000059381YC4 (OJ1005); 5454001-40, 00000059384YC8 (OJ1006); 5454001-40, 00000059548YC8 (OJ1012); 5454001-40, 00000067890YC4 (OJ1028); 5454001-40, 00000071457YC6 (OJ1034); 5454001-40, 00000073316YC2 (OJ0051); 5454001-40, 00000078404YC1 (OJ1048); 5454001-40, 00000064033YC4 (OJ0021); 5454001-40, 00000064502YC8 (OJ0177); 5454001-40, 00000065925YC0 (OJ1101); 5454001-40, 00000075357YC4 (OJ1102); 5454001-240, 00000381118HM9 (00269CTS03); 5454001-40, 00000071454YC3 (DZ1204CT02); 5454001-40, 64213YC2 (641754CT660); 5454001-40, 00000066087YC8 (815625OPT1); 5454001-40, 00000067555YC3 (410822EAS128); 5454001-40, 00000067177YC6 (828696CT02); 5454001-40, 00000063905YC4 (336719CT3); 5454001-40, 60084YC1 (732294OPT); 5454001-40, 62226YC6 (732294EROPT); 5454001-40, 00000060085YC8 (845561CTER); 5454001-40, 63539YC1 (614PMC680); 5454001-40, 63903YC9 (541677MOICT660); 5454001-40, 63387YC5 (814849CT); 5454001-40, 66933YC3 (615768CT660); 5454001-40, 00000066210YC6 (409899CT660); 5454001-40, 00000067557YC9 (361SBCT660); 5454001-40, 00000067174YC3 (METHCYPRESSCT); 5454001-40, 63150YC7 (817288MCICT660); 5454001-40, 63144YC0 (281523SJCT); 5454001-40, 00000065923YC5 (281BACT2); 5454001-40, 64630YC7 (206223CCUCT); 5454001-140, CJRAX1600038CN (8589); 5454001-140, 00000443704CN2 (8588); 5454001-240, 386834HM6 (8115); 5454001-240, 395530HM9 (609584OPT2); 5454001-140, CJRAX1700001CN (CJRAX1700001CN); 5454001-240, 00000359343HM1 (00216CTS10); 5454001-240, 00000368079HM0 (10698CTS01); 5454001-240, 00000379128HM2 (00224CTS05); 5454001-140, CJRAX1600006CN (619280O660); 5454001-240, 398384HM8 (303HICRCT1); 5454001-40, 69392YC9 (305682OPT); 5454001-40, 00000072687YC7 (812478CT660); 5454001-40, 70027YC8 (574267CT660); 5454001-140, CJRAX1600014CN (270793CT2); 5454001-240, 402444HM4 (504391660); 5454001-40, 00000068988YC5 (616394CT660); 5454001-140, Not Available (910VACT2); 5454001-240, 00000388470HM7 (585798660); 5454001-40, 69396YC0 (330757OP660); 5454001-140, Not Available (724773OPTCT); 5454001-140, 443127CN6 (615316660); 5454001-140, CJRAX1600009CN (804289CT660); 5454001-40, 00000067559YC5 (509837SSCT660); 5454001-240, 00000445858HM4 (M4479021); 5454001-240, 00000436942HM7 (M4168143); 5454001-240, 00000444817HM1 (M2437606); 5454001-240, 00000444819HM7 (H1066CT35); 5454001-240, 00000439096HM9 (AMS14202); 5454001-240, 00000440558HM5 (10448CTS04); 5454001-240, 00000444703HM3 (CS1349CT01); 5454001-240, 00000439099HM3 (B1529611); 5454001-240, 00000437597HM8 (M4159716); 5454001-240, 00000428555HM7 (M4161851); 5454001-240, 00000441884HM4 (M187068304); 5454001-240, 00000430852HM4 (M4478903); 5454001-240, 00000432094HM1 (M4478428); 5454001-240, 00000421001HM9 (M4168431); 5454001-240, 00000452615HM8 (M4461518); 5454001-240, 00000443977HM4 (M2148261); 5454001-240, 00000443978HM2 (M4218721); 5454001-240, 00000428552HM4 (A1020801); 5454001-240, 00000458681HM4 (M4179506); 5454001-240, 00000448451HM5 (M4486928); 5454001-240, 00000443909HM7 (M4184875); 5454001-240, 00000452614HM1 (M9483447); 5454001-240, 00000421448HM2 (M4487918); 5454001-240, 00000435173HM0 (M1322129); 5454001-240, 00000436948HM4 (M9914223); 5454001-240, 00000439097HM7 (A190136611); 5454001-240, 00000445857HM6 (M4202316); 5454001-240, 00000441749HM9 (M2228042); 5454001-240, 00000044057HM7 (M4476121); 5454001-240, 00000420278HM4 (A5177237); 5454001-240, 00000437586HM1 (LLS01200); 5454001-240, 00000444816HM3 (PL3956CT01); 5454001-240, 00000461075HM4 (PL3794CT01); 5454001-240, 00000443979HM0 (PL3824CT01); 5454001-240, 00000436248HM9 (RO1334CT01); 5454001-240, 00000437598HM6 (RO1098CT04); 5454001-240, 00000430001HM8 (BD003CT01); 5454001-240, 00000439101HM7 (33044CTS01); 5454001-240, 00000430859HM9 (00187CTS04); 5454001-240, 00000439100HM9 (33044CTS03); 5454001-240, 00000439745HM1 (33044CTS02); 5454001-40, 00000074651YC1 (RU1231CT01); 5454001-40, 00000076549YC5 (RU7403CT01); 5454001-40, 00000078635YC0 (AE1065CT01); 5454001-240, 00000424385HM3 (83021001211615); 5454001-40, 00000074068YC8 (34433CTS01); 5454001-240, 00000429994HM7 (M2045017); 5454001-240, 00000436250HM5 (M6012202); 5454001-240, 00000445861HM8 (X191414804); 5454001-240, 00000453106HM7 (M4490018); 5454001-240, 00000440560HM1 (M2153107); 5454001-240, 00000438101HM8 (M4493024); 5454001-240, 00000437600HM0 (M176430612); 5454001-240, 00000436253HM9 (M4197019); 5454001-240, 00000441750HM7 (M2573414); 5454001-240, 00000445860HM0 (M215631207); 5454001-240, 00000441883HM6 (X751968030); 5454001-240, 00000425297HM9 (83021861016115); 5454001-240, 00000432093HM3 (83021005952416); 5454001-240, 00000423279HM9 (83021004500515); 5454001-240, 00000459671HM4 (83021606365816); 5454001-240, 00000422271HM7 (A171995612); 5454001-240, 00000463082HM8 (PL1560CT01); 5454001-240, 00000457516HM3 (RO1343CT01); 5454001-240, 00000453596HM9 (RO1150CT02); 5454001-240, 00000430857HM3 (RO1108CT03); 5454001-240, 00000453597HM7 (RO1337CT01); 5454001-240, 00000436252HM1 (RO1333CT01); 5454001-240, 00000435175HM5 (SK1015CT01); 5454001-240, 00000429002HM9 (ZA2816CT01); 5454001-240, 00000430854HM0 (34332CTS02); 5454001-240, 00000459774HM6 (34450CTS01); 5454001-40, 00000074885YC5 (DZ2092CT01); 5454001-40, 00000075081YC0 (83155336731); 5454001-40, 00000074278YC3 (83149051966); 5454001-40, 40GAN1700003YC (EG1023CT07); 5454001-40, 40GAN1600019YC (EG2601CT01); 5454001-40, 00000075847YC4 (EG2526CT01); 5454001-40, 00000073635YC5 (EG1245CT02); 5454001-40, 00000074279YC1 (EG1023CT06); 5454001-40, 00000075849YC0 (EG1065CT02); 5454001-40, 00000078407YC4 (EG2542CT01); 5454001-40, 00000076551YC1 (83021803092016); 5454001-40, 00000071227YC3 (5512CT0031); 5454001-40, 00000072236YC3 (5512CT0033); 5454001-40, 00000074647YC9 (5512CT0042); 5454001-40, 00000077039YC6 (5512CT0050); 5454001-40, 00000075359YC0 (KA2973CT01); 5454001-40, 00000078406YC6 (NG1280CT01); 5454001-40, 40GAN1600015YC (PL1687CT02); 5454001-40, 40GAN1700001YC (RU5899CT01); 5454001-40, 00000077689YC8 (RU7742CT01); 5454001-40, 00000072688YC5 (RU9478CT01); 5454001-40, 00000073851YC8 (SA1072CT01); 5454001-40, 00000073853YC4 (ZA2383CT02); 5454001-40, 00000077941YC3 (ZA2878CT01); 5454001-240, 00000437601HM8 (04008CTS01); 5454001-40, 40GAN1600014YC (34455CTS01); 5454001-40, 00000072686YC9 (06172CTS01); 5454001-40, 00000073285YC9 (5973CT0037); 5454001-40, 00000077558YC5 (5973CT0044); 5454001-240, 00000436945HM0 (TZ1052CT01); 5454001-240, 423282HM3 (309672CT660); 5454001-40, 72690YC1 (303SKYCT1); 5454001-40, 40GAN1600005YC (60001CTS02); 5454001-140, CJRAX1600007CN (302744OPTER); 5454001-140, CJRAX1600064CN (813844CT660); 5454001-140, CJRAX1600068CN (770848BCT); 5454001-140, CJRAX1600026CN (248661OPT1); 5454001-140, CJRAX1600060CN (717972CTER); 5454001-140, CJRAX1600050CN (864454OPT); 5454001-140, CJRAX1600034CN (253944CT660); 5454001-40, 00000072052YC4 (JO4093CT01); 5454001-240, 00000429001HM1 (PL1778CT01); 5454001-240, 421004HM3 (408972OPT660); 5454001-240, 00000439747HM7 (386719CT2); 5454001-240, 414948HM0 (808HRAECT660); 5454001-240, 420281HM8 (208359CT); 5454001-240, 426337HM2 (815464CT660); 5454001-240, 426336HM4 (765649CT660); 5454001-240, 425942HM0 (812537ERCT); 5454001-240, 429999HM6 (660562SFCT); 5454001-240, 420283HM4 (662624CT660); 5454001-240, 429003HM7 (NYP559904); 5454001-240, 420279HM2 (631548CTED); 5454001-240, 422274HM1 (419891CT); 5454001-240, 425941HM2 (937SRMC661); 5454001-240, 422276HM6 (419251ERCT); 5454001-240, 00000437585HM3 (817556CT660); 5454001-240, 425300HM1 (713363CT1); 5454001-240, 432091HM7 (METHSPRCT); 5454001-240, 429008HM6 (936271CT660); 5454001-240, 421449HM0 (435691CT); 5454001-240, 00000433594HM9 (H4478CT07); 5454001-240, 00000436947HM6 (C5125004); 5454001-240, 00000435176HM3 (A6184805); 5454001-240, 00000443980HM8 (A5304515); 5454001-240, 00000447695HM8 (A5422030); 5454001-240, 00000438131HM5 (RE1044CT02); 5454001-240, 00000467228HM3 (NEV33214); 5454001-240, 00000443320HM7 (BD001CT01); 5454001-240, 00000443321HM5 (BD004CT01); 5454001-40, 00000067893YC8 (205725OPT); 5454001-40, 74650YC3 (765282CT660); 5454001-40, 00000072365YC0 (410337ER660); 5454001-40, 70629YC1 (908685OPT); 5454001-140, CJRAX1500005CN (503KINCT); 5454001-40, 70237YC3. (412777OPTCT); 5454001-40, 74280YC9 (724861NORCT); 5454001-40, 00000077940YC5 (PK1490CT01); 5454001-40, 00000077038YC8 (PK1367CT01); 5454001-40, 00000075358YC2 (RU1251CT01); 5454001-40, 00000075589YC2 (RW1041CT01); 5454001-40, 00000074069YC6 (6073CT0017); 5454001-40, 40GAN1600006YC (UA4316CT01); 5454001-40, 00000073313YC9 (AE1557CT01); 5454001-140, CJRAX1600069CN (530SDCT); 5454001-140, 443482CN5 (303AUCT1); 5454001-140, CJRAX1600018CN (970641CT); 5454001-140, CJRAX1600057CN (219796CT660); 5454001-140, CJRAX1600028CN (913541M660); 5454001-140, CJRAX1600065CN (225SJCT); 5454001-140, CJRAX1600048CN (718245ERCT); 5454001-140, 442643CN3 (419251CT); 5454001-140, CJRAX1600032CN (724837MVCT); 5454001-140, CJRAX1600020CN (361SKCT660); 5454001-140, CJRAX1600066CN (254965CT660); 5454001-140, CJRAX1500008CN (804WATCT1); 5454001-240, 00000424384HM6 (CT8086CO); 5454001-240, 00000436943HM5 (NO4433CT01); 5454001-240, 420999HM5 (858657O660); 5454001-40, 71090YC5 (719767OPTIMA); 5454001-240, 428556HM5 (303SKYPCT); 5454001-240, 439746HM9 (303HIPCT); 5454001-240, 425299HM5 (772HFCT1); 5454001-240, 422275HM8 (912369LCT); 5454001-240, 420997HM9 (706437CT); 5454001-240, 425940HM4 (515295CT1); 5454001-240, 429004HM5 (815802CT660); 5454001-240, 432090HM9 (618643CT2); 5454001-240, 428553HM2 (765659CT660); 5454001-240, 426338HM0 (225927OCT); 5454001-240, 430858HM1 (248652OPT); 5454001-240, 428554HM0 (636332WCT); 5454001-240, 433285HM4 (828765BRCT); 5454001-240, 426339HM8 (505662CT660); 5454001-240, 427218HM3 (865687CT); 5454001-240, 421447HM4 (512SNWCT); 5454001-240, 00000433596HM4 (SHICT49); 5454001-240, 427217HM5 (972420L660); 5454001-40, 00000072691YC9 (210CSRNW660ER); 5454001-240, 421446HM6 (801773CT3); 5454001-240, 423278HM1 (703689OP660); 5454001-240, 423281HM5 (757507CT1); 5454001-240, 421000HM1 (757395AICCT); 5454001-240, 427216HM7 (276228OPT660); 5454001-240, 00000433284HM7 (HR4218CT01); 5454001-240, 00000448450HM7 (83021849468016); 5454001-240, 00000436946HM8 (83021800203516); 5454001-240, 00000467693HM8 (H153303401); 5454001-240, 00000452616HM6 (A5632441); 5454001-240, 00000433595HM6 (B5129930); 5454001-240, 00000434403HM2 (A5240506); 5454001-240, 00000433593HM1 (PPR58105); 5454001-240, 00000436251HM3 (MP842031); 5454001-240, 00000436944HM3 (MPX84723); 5454001-240, 00000462639HM6 (MPX52200); 5454001-240, 00000459775HM3 (MPX04219); 5454001-240, 00000435172HM2 (DEL26482); 5454001-240, 00000457425HM7 (FPA99429); 5454001-240, 00000454678HM4 (NE761114); 5454001-240, 00000454679HM2 (NE651516); 5454001-240, 00000454677HM6 (FP167510); 5454001-240, 00000459773HM8 (34377CTS02); 5454001-40, 00000077945YC4 (CT4931HN); 5454001-40, 40GAN1700002YC (5512CT0051); 5454001-40, 00000073637YC1 (JO4102CT01); 5454001-40, 00000073632YC2 (PK1300CT02); 5454001-40, 00000076301YC1 (1557CT0036); 5454001-40, 00000074648YC7 (6073CT0025); 5454001-40, 00000073287YC5 (1573CT0019); 5454001-40, 00000069669YC0 (251639PODCT2); 5454001-40, 70035YC1 (205487CT); 5454001-40, 72685YC1 (949557OPTM); 5454001-40, 71459YC2 (303FCBHCT1); 5454001-40, 72684YC4 (719598OPTIMA); 5454001-40, 71460YCO (904202TCCT); 5454001-40, 72366YC8 (904702CT1); 5454001-40, 73854YC2 (401CT2); 5454001-40, 00000071455YC0 (641664CT66); 5454001-40, 00000068989YC3 (318259OPT); 5454001-140, CJRAX1500003CN (337521OP660); 5454001-40, 69394YC5 (508486CT); 5454001-40, 00000068720YC2 (248524OPT660); 5454001-40, 72367YC6 (507275CT); 5454001-40, 73849YC2 (MC532CT); 5454001-40, 00000072368YC4 (308425FCT); 5454001-40, 69391YC1 (609267OPT2); 5454001-40, 70852YC9 (505368O660); 5454001-40, 72689YC3 (631726CT660); 5454001-140, CJRAX1600039CN (541274CT2); 5454001-40, 70040YC1 (717242CT); 5454001-40, 73312YC1 (570853660CT); 5454001-40, 00000073284YC2 (423855CT2); 5454001-40, 69666YC6 (931245PCT1); 5454001-40, 00000069863YC9 (615328CT01); 5454001-40, 74281YC7 (817551CT660); 5454001-40, 71620YC9 (915595CT660); 5454001-40, 71315YC6 (972403THCD); 5454001-40, 72852YC7 (920849CT1); 5454001-160, CJRBX1600041CN (608647CTEVO); 5454001-40, 00000077284YC8 (BY4033CT03); 5454001-40, 00000078636YC8 (BY4183CT01); 5454001-40, 00000072849YC3 (CI1087CT01); 5454001-40, 00000076550YC3 (5512CT0054); 5454001-40, 00000077559YC3 (5512CT0062); 5454001-40, 00000077684YC9 (5512CT0061); 5454001-40, 40GAN1600007YC (5512CT0058); 5454001-40, 40GAN1600009YC (5512CT0073); 5454001-40, 00000072693YC5 (KA2938CT01); 5454001-40, 40GAN1600017YC (1553CT0007); 5454001-40, 40GAN1700006YC (MA1650CT01); 5454001-40, 00000077944YC7 (1557CT0043); 5454001-40, 40GAN1600002YC (1557CT0039); 5454001-40, 00000077281YC4 (RU1695CT01); 5454001-40, 00000078405YC8 (RU3330CT02); 5454001-40, 40GAN1600016YC (1573CT0027); 5454001-40, 00000072053YC2 (1573CT0018); 5454001-40, 00000073634YC8 (1573CT0020); 5454001-40, 00000077685YC6 (1573CT0021); 5454001-40, 40GAN1600008YC (1573CT0030); 5454001-40, 00000072054YC0 (1573CT0017); 5454001-40, 40GAN1600001YC (1573CT0023); 5454001-40, 40GAN1600018YC (35073CTS01); 5454001-40, 00000075590YC0 (UZ1031CT01); 5454001-40, 40GAN1700005YC (5973CT0070); 454001-140, CJRMX1600015CN (CT1520AR); 5454001-140, CJRAX1600062CN (334288OPT660); 5454001-140, CJRAX1600052CN (334365OPT660); 5454001-140, CJRAX1600059CN (205345OPT660); 5454001-140, CJRAX1600024CN (480323CT1); 5454001-140, CJRAX1600067CN (209223OPT); 5454001-140, CJRAX1600036CN-DUP (760379CT); 5454001-140, CJRAX1600030CN (951486OPT660); 5454001-140, CJRAX1500009CN (303PSLCT1); 5454001-140, CJRAX1600045CN (904202DNCT); 5454001-140, CJRAX1600001CN (912739ECT); 5454001-140, CJRAX1500013CN (641743CT660); 5454001-240, 429007HM8 (812442CT660); 5454001-140, CJRAX1600013CN (502226OPT2); 5454001-140, CJRAX1600011CN (617296CT1); 5454001-140, CJRAX1600041CN (508427GSCT); 5454001-140, Not Available (573729OCT); 5454001-140, CJRAX1600051CN (BRUNSWICKCT1); 5454001-140, CJRAX1500006CN (SHICT01405); 5454001-140, CJRAX1600027CN (308282CT660); 5454001-140, CJRAX1600033CN (718544CT660); 5454001-160, Not Available (513870ERCT); 5454001-140, CJRAX1500011CN (419891VCT); 5454001-140, CJRAX1600063CN (937342OPT); 5454001-140, Not Available (419882CT); 5454001-140, CJRAX1600054CN (405307O660); 5454001-140, CJRAX1600056CN (918485W660); 5454001-140, CJRAX1600025CN (717867CT660); 5454001-140, CJRAX1600044CN (864841OPT); 5454001-140, CJRAX1600005CN (214BDCT660); 5454001-140, CJRAX1600029CN (915LPWEST); 5454001-140, CJRAX1600042CN (817UNIVCT660); 5454001-140, CJRAX1600058CN (940567CT660); 5454001-140, CJRAX1600037CN (830FRHCT660); 5454001-140, CJRAX1600002CN (804SCCT660); 5454001-140, CJRAX1600053CN (715644CT); 5454001-140, CJRAX1600072CN (CJRAX1600072CN); 5454001-240, 00000391752HM3 (83021026147514); 5454001-240, 00000391327HM4 (83021864863814); 5454001-240, 00000398387HM1 (83021827233015); 5454001-240, 389998HM6 (334CT660); 5454001-240, 398385HM5 (205820OPTIMA); 5454001-240, 00000389989HM5 (256429CT660); 5454001-240, 389549HM7 (251435MICT1); 5454001-240, 394594HM6 (251660CTHC); 5454001-240, 397584HM4 (479452C660); 5454001-240, 402442HM8 (954436OP64); 5454001-240, 399109HM8 (727843CT660); 5454001-240, 00000403050HM8 (309852CT1); 5454001-240, 402447HM7 (630873CT); 5454001-240, 403048HM2 (270527LCT); 5454001-240, 406019HM0 (859744OPT660); 5454001-240, 402969HM0 (985875CT1); 5454001-240, 388472HM3 (617754OPT2); 5454001-240, 419608HM5 (410822DIC64); 5454001-240, 389384HM9 (269659CT660); 5454001-240, 387793HM3 (601855OPCT); 5454001-240, 391750HM7 (406873CT2); 5454001-240, 390619HM5 (980487KMCT); 5454001-240, 00000391748HM1 (910CFHOKE660); 5454001-240, 395528HM3 (252641660CT); 5454001-40, 00000066085YC2 (973450OPT660); 5454001-240, 401670HM5 (9738956619CT660); 5454001-240, 389378HM1 (609267OPTM); 5454001-240, 00000389993HM7 (973579CTER); 5454001-240, 389995HM2 (732499OPT); 5454001-240, 391330HM8 (505983CT); 5454001-240, 403531HM7 (631DOLAN660); 5454001-240, Not Available (631743OPTMA); 5454001-240, 389380HM7 (518483CT660); 5454001-240, 409899HM2 (740439OPT); 5454001-240, 394591HM2 (419783CT); 5454001-240, 395527HM5 (330841OP660); 5454001-240, 382217HM8 (401348CT660); 5454001-240, 404913HM6 (615446FSED); 5454001-240, 420292HM5 (615342EDCT); 5454001-240, 405577HM8 (817408LGMC660); 5454001-240, 396124HM0 (972932K660); 5454001-240, 00000388473HM1 (972569OP66); 5454001-240, 00000388476HM4 (903TSJ660); 5454001-240, 380395HM4 (703645OP660); 5454001-240, 389554HM7 (608890DHCT); 5454001-240, 00000394596HM1 (DZ1771CT01); 5454001-240, 00000399114HM8 (X208617404); 5454001-240, 00000379125HM8 (M4400427); 5454001-240, 00000413610HM7 (A1112220); 5454001-240, 00000423280HM7 (M4482315); 5454001-240, 00000408407HM5 (D6008809); 5454001-240, 00000385659HM8 (D4191722); 5454001-240, 00000410650HM6 (M4491938); 5454001-240, 00000418620HM1 (M2036012); 5454001-240, 00000430855HM7 (M4484751); 5454001-240, 00000409894HM3 (M227137504); 5454001-240, 00000397412HM8 (H2073CT17); 5454001-240, 00000410652HM2 (GP1084CT01); 5454001-240, 00000421445HM8 (83021844488715); 5454001-240, 00000448452HM3 (83021601020316); 5454001-240, 00000395529HM1 (83021207242515); 5454001-240, 00000433286HM2 (83021248697916); 5454001-240, 00000415692HM3 (83021863020315); 5454001-240, 00000415693HM1 (83021861873715); 5454001-240, 00000414006HM7 (83021869996615); 5454001-240, 00000425298HM7 (83021861960015); 5454001-240, 00000428551HM6 (83021863396715); 5454001-240, 00000386836HM1 (83021284643214); 5454001-240, 00000399112HM2 (83021287156415); 5454001-240, 00000421002HM7 (83021284476115); 5454001-240, 00000432931HM4 (83021287156916); 5454001-240, 00000414002HM6 (83021209468215); 5454001-240, 00000399111HM4 (83021208438915); 5454001-240, 00000413444HM1 (83021201827115); 5454001-240, 00000421003HM5 (83021204475615); 5454001-240, 00000407422HM5 (83021200808215); 5454001-240, 00000407423HM3 (83021106947915); 5454001-240, 00000434404HM0 (83021049412216); 5454001-240, 00000395531HM7 (83021805881615); 5454001-240, 00000433287HM0 (83021808929616); 5454001-240, 00000403049HM0 (83021809699815); 5454001-240, 00000421444HM1 (83021804503215); 5454001-240, 00000440556HM9 (83021028867116); 5454001-240, 00000439098HM5 (83021028862616); 5454001-240, 00000440559HM3 (83021028788616); 5454001-240, 00000445859HM2 (83021021066616); 5454001-240, 00000452374HM2 (83021021066016); 5454001-240, 00000445862HM6 (83021021059416); 5454001-240, 00000399110HM6 (83021027245315); 5454001-240, 00000389382HM3 (A156941410); 5454001-240, 00000369367HM8 (A5144921); 5454001-240, 00000358590HM8 (KW1086CT02); 5454001-240, 00000403536HM6 (NG1190CT01); 5454001-240, 00000369366HM0 (NG1193CT01); 5454001-240, 00000419477HM5 (RO1314CT01); 5454001-240, 00000414004HM2 (DPB15863); 5454001-240, 00000398389HM7 (00082CTS02); 5454001-240, 00000394182HM0 (00216CTS11); 5454001-240, 00000435156HM5 (10583CTM03); 5454001-40, 66462YC3 (334727VATCT); 5454001-40, 00000067888YC8 (870414CT660); 5454001-40, 00000066735YC2 (623848CT); 5454001-40, 00000067500YC9 (860678CT660); 5454001-40, 63142YC4 (954HCAEROPT); 5454001-40, 00000064032YC6 (954518OPT); 5454001-40, 00000066574YC5 (863688CT660); 5454001-40, 00000067506YC6 (671CT1); 5454001-40, 00000064211YC6 (772SMCOPCT1); 5454001-40, 65927YC6 (515386CT660); 5454001-40, 60796YC0 (765331CT660); 5454001-40, 00000063900YC5 (913227FSED); 5454001-40, 63145YC7 (620356CT660); 5454001-40, 00000067864YC9 (CAPEERCT2); 5454001-40, 00000062520YC2 (586573OPT2); 5454001-40, 00000067865YC6 (704597OPT6601); 5454001-40, 00000067889YC6 (704597OPT660); 5454001-40, 00000067861YC5 (336564CT1); 5454001-40, 00000066655YC2 (716372CT660); 5454001-40, 64504YC4 (419436CT); 5454001-40, 67863YC1 (580481O660); 5454001-40, 00000060794YC5 (803461CT1); 5454001-40, 00000062057YC5 (864487CT1); 5454001-40, 00000067173YC5 (843761CT660); 5454001-40, 00000067501YC7 (864560ECT2); 5454001-40, Not Available (615446H660); 5454001-40, 00000064632YC3 (915ZERVCT); 5454001-40, Not Available (830CSR660128); 5454001-40, 67501YC7 (801313CT3); 5454001-40, 00000060942YC0 (608339CT); 5454001-40, 00000067172YC7 (DZ1853CT01); 5454001-40, 00000063143YC2 (DZ1723CT01); 5454001-40, 00000064631YC5 (AM1019CT01); 5454001-40, 00000070844YC6 (83127404555); 5454001-40, 00000063726YC4 (83135271130); 5454001-40, 00000060941YC2 (83119337881); 5454001-40, 00000065922YC7 (83132274380); 5454001-40, 00000064811YC3 (83131015593); 5454001-40, 00000078195YC5 (83151202561); 5454001-40, 00000063728YC0 (83129071457); 5454001-240, 00000402445HM1 (83126481210); 5454001-40, 00000067365YC7 (83138055398); 5454001-40, 00000068715YC2 (83141847794); 5454001-40, 00000062056YC7 (83125486755); 5454001-40, 00000069399YC4 (83142729121); 5454001-40, 00000062753YC9 (83121450295); 5454001-40, 00000070236YC5 (83142729095); 5454001-40, 00000067891YC2 (83141848312); 5454001-40, 00000069395YC2 (83142307182); 5454001-40, 00000063386YC7 (83122705534); 5454001-240, 00000381703HM8 (83115291141); 5454001-41, 72244YC7 (082421230390); 5454001-41, 41GAN1700002YC (82421250199); 5454001-40, 00000071228YC1 (EG2222CT01); 5454001-40, 00000063389YC1 (EG1736CT02); 5454001-40, 00000065209YC9 (ET1000CT02); 5454001-40, 00000067554YC6 (ET1031CT01); 5454001-240, 00000389990HM3 (M4161531); 5454001-240, 00000407424HM1 (X255095204); 5454001-40, 00000058707YC1 (M2748742); 5454001-40, 00000058880YC6 (C1874506); 5454001-40, 00000063382YC6 (A1279212); 5454001-40, 00000059195YC8 (X365212301); 5454001-40, 00000062518YC6 (83021101751614); 5454001-40, 00000062051YC8 (5512CT0002); 5454001-40, 00000061245YC7 (835210267); 5454001-40, 00000061772YC0 (835210268); 5454001-40, 00000059385YC5 (835210257); 5454001-40, 00000062054YC2 (835210263); 5454001-40, 00000067858YC1 (5512CT0017); 5454001-40, 00000069397YC8 (5512CT0022); 5454001-40, 00000070030YC2 (5512CT0020); 5454001-40, 00000070039YC3 (5512CT0025); 5454001-40, 00000066934YC1 (KA2960CT01); 5454001-40, 00000071091YC3 (NG1244CT01); 5454001-40, 00000064812YC1 (1557CT0010); 5454001-40, 00000072237YC1 (1557CT0030); 5454001-40, 00000070418YC9 (RU5897CT01); 5454001-40, 00000070239YC9 (RU0001CT01); 5454001-40, 00000070243YC1 (RU8065CT02); 5454001-40, 00000068596YC6 (RU7000CT01); 5454001-40, 00000070241YC5 (RU9522CT01); 5454001-40, 00000064635YC6 (RU1044CT02); 5454001-40, 00000070240YC7 (RU5891CT01); 5454001-40, 00000070031YC0 (RU4325CT01); 5454001-40, 00000070036YC9 (RU2039CT03); 5454001-40, 00000070032YC8 (RU6795CT01); 5454001-40, 00000070238YC1 (RU9369CT01); 5454001-40, 00000070414YC8 (RU1471CT01); 5454001-40, 00000059869YC8 (SA2420CT01); 5454001-40, 00000070034YC4 (SA2468CT01); 5454001-40, 00000071316YC4 (6073CT0014); 5454001-40, 00000059196YC6 (ZA1006CT04); 5454001-40, 00000069861YC3 (ZA1127CT03); 5454001-40, 00000067179YC2 (ZA1046CT01); 5454001-40, 00000067176YC8 (TZ1041CT01); 5454001-40, 00000063390YC9 (1573CT0002); 5454001-40, 00000064460YC9 (1573CT0003); 5454001-40, 00000071226YC5 (856210188); 5454001-40, 00000060083YC3 (06159CTS01); 5454001-40, 00000063383YC4 (06162CTS02); 5454001-40, 00000066576YC0 (06167CTS01); 5454001-40, 00000061250YC7 (16044CTS01); 5454001-40, 00000063541YC7 (34405CTS01); 5454001-40, 00000070847YC9 (34185CTS02); 5454001-40, 00000064463YC3 (63024CTS01); 5454001-40, 00000067178YC4 (UK5001CT02); 5454001-40, 00000064810YC5 (UZ1041CT01); 5454001-40, 00000063147YC3 (UZ1017CT01); 5454001-40, 00000073850YC0 (5973CT0024); 5454001-40, 00000063724YC9 (5973CT0006); 5454001-40, 00000063725YC6 (5973CT0007); 5454001-40, 00000063898YC1 (5973CT0008); 5454001-40, 00000064034YC2 (5973CT0009); 5454001-40, 00000064913YC7 (5973CT0013); 5454001-40, 00000065211YC5 (5973CT0012); 5454001-40, 00000064912YC9 (5973CT0015); 5454001-40, 00000072850YC1 (5973CT0021); 5454001-40, 00000062052YC6 (859210153); 5454001-40, 00000070028YC6 (DZ1714CT01); 5454001-140, 443449CN4 (205759CT11); 5454001-140, 438880CN7 (623584CT); 5454001-140, 443840CN4 (530CHICT660); 5454001-140, Not Available (415677CHOP); 5454001-140, 442993CN2 (318443OP660); 5454001-140, 440033CN9 (989497OPT); 5454001-140, 442037CN8 (732363OPT); 5454001-240, 00000389383HM1 (702877CT); 5454001-40, Not Available (775VACT660); 5454001-140, CJRMX1600008CN (718818ER660); 5454001-140, 438723CN9 262696OPT660); 5454001-140, CJRMX1500007CN (CJRMX1500007CN); 5454001-140, CJRMX1600017CN (CJRMX1600017CN); 5454001-140, CJRMX1600014CN (CJRMX1600014CN); 5454001-140, CJRMX1600019CN (CJRMX1600019CN); 5454001-140, CJRMX1600005CN (CJRMX1600005CN); 5454001-140, CJRMX1600012CN (CJRMX1600012CN); 5454001-140, CJRAX1700002CN (CJRAX1700002CN); 5454001-140, CJRMX1600011CN (CJRMX1600011CN); 5454001-140, CJRMX1600013CN (CJRMX1600013CN); 5454001-140, CJRMX1600018CN (CJRMX1600018CN); 5454001-140, CJRMX1600016CN (CJRMX1600016CN); 5454001-140, CJRAX1600055CN (CJRAX1600055CN); 5454001-140, CJRAX1600074CN (CT4003PA); 5454001-240, 00000376508HM8 (C4166934); 5454001-240, 00000370717HM1 (M2312312); 5454001-240, 00000385139HM1 (YE1172CT01); 5454001-140, 442011CN3 (205971OPT3); 5454001-240, 403052HM4 (205971OPT2); 5454001-240, 402446HM9 (205971OPT1); 5454001-240, 00000397413HM6 (520GVHCT); 5454001-240, 418623HM5 (303CENGCT); 5454001-240, 401669HM7 (302645CT660); 5454001-40, 72364YC3 (815300CT660); 454001-240, 00000370716HM3 (316689CT660); 5454001-240, 391328HM2 (601376OPCT); 5454001-140, 440262CN4 (828213CT1); 5454001-240, 00000395004HM5 (PORTERCT1); 5454001-240, 414005HM9 (308SIDNEYCT); 5454001-240, 402443HM6 (603433OPTA); 5454001-240, 412422HM8 (914273CT660); 5454001-240, 401667HM1 (845348OPT66064); 5454001-240, 409898HM4 (845348OPT660128); 5454001-240, 403533HM3 (614RMHN660); 5454001-240, 00000391753HM1 (503674OPT); 5454001-240, 406018HM2 (541MC660); 5454001-240, 419609HM3 (615342CT02); 5454001-240, 00000389381HM5 (936NMCS660); 5454001-240, 405581HM0 (713STL660); 5454001-240, 395533HM3 (METHCRCT660); 5454001-240, 391749HM9 (540536OP660); 5454001-240, 00000402440HM2 (83121450715); 5454001-240, 00000412420HM2 (BA4073CT03); 5454001-240, 00000417824HM0 (M4143637); 5454001-240, 00000373269HM0 (M4135324); 5454001-240, 00000405580HM2 (M2645710); 5454001-240, 00000376507HM0 (M4163780); 5454001-240, 00000367026HM2 (M4437829); 5454001-240, 00000386517HM7 (M4166662); 5454001-240, 00000385658HM0 (M4167714); 5454001-240, 00000412419HM4 (A8375420); 5454001-240, 00000381704HM6 (M4173826); 5454001-240, 00000385140HM9 (M4173821); 5454001-240, 00000435174HM8 (M4176538); 5454001-240, 00000412421HM0 (M44827199); 5454001-240, 00000403537HM4 (M4548539); 5454001-240, 00000394589HM6 (A4182039); 5454001-240, 00000411635HM6 (A4182041); 5454001-240, 00000411634HM9 (M6021216); 5454001-240, 00000383054HM4 (M4187025); 5454001-240, 00000367027HM0 (C2195291); 5454001-240, 00000403535HM8 (M4188436); 5454001-240, 00000396125HM7 (A4192117); 5454001-240, 00000397411HM0 (X41944005); 5454001-240, 00000410649HM8 (B4195740); 5454001-240, 00000378361HM0 (M25030109); 5454001-240, 00000411638HM0 (M2637323); 5454001-240, 00000377033HM6 (A4461318); 5454001-240, 00000375268HM0 (E8375416); 5454001-240, 00000372043HM0 (X970562018); 5454001-240, 00000406453HM1 (M4460821); 5454001-240, 00000381120HM5 (C4193351); 5454001-240, 00000414007HM5 (M1020820); 5454001-40, 00000066734YC5 (M4467041); 5454001-40, 00000058708YC9 (A5485268); 5454001-240, 00000382285HM5 (GP1046CT02); 5454001-240, 00000411632HM3 (F5367808); 5454001-240, 00000411633HM1 (F5367806); 5454001-240, 00000383053HM6 (B5603949); 5454001-240, 00000402441HM0 (C208415201); 5454001-240, 00000395532HM5 (A5138519); 5454001-240, 00000394187HM9 (A51773120); 5454001-240, 00000401668HM9 (A51773122); 5454001-240, 00000386835HM3 (A51014121); 5454001-240, 00000375266HM4 (LV4000CT05); 5454001-240, 00000403532HM5 (LV4025CT01); 5454001-240, 00000385660HM6 (EMM01127); 5454001-240, 00000389991HM1 (PL2693CT01); 5454001-240, 00000397582HM8 (PL1620CT03); 5454001-240, 00000385354HM6 (PL3349CT01); 5454001-240, 00000389994HM5 (600034CT02); 5454001-240, 00000390617HM9 (PL1180CT01); 5454001-240, 00000367714HM3 (PL3254CT02); 5454001-240, 00000382219HM4 (RE1047CT01); 5454001-240, 00000403047HM4 (O069CT01); 5454001-240, 00000409897HM6 (00555CTS03); 5454001-240, 00000394180HM4 (00916CTS01); 5454001-240, 00000405807HM9 (00260CTS02); 5454001-240, 00000414008HM3 (01211CTS03); 5454001-240, 00000406017HM4 (00680CTS01); 5454001-40, 00000061773YC8 (83119339107); 5454001-40, 00000060871YC1 (83119339252); 5454001-40, 00000059011YC7 (83111338230); 5454001-240, 00000401671HM3 (M28027105); 5454001-240, 00000394149HM9 (M4494710); 5454001-240, 00000389552HM1 (X59007328); 5454001-240, 00000406454HM9 (B1020805); 5454001-240, 00000422272HM5 (M45052123); 5454001-240, 00000396126HM5 (A5189734); 5454001-40, 00000065205YC7 (C5294657); 5454001-40, 00000067499YC4 (B5805826); 5454001-40, 00000064636YC4 (C5123654); 5454001-40, 00000066654YC5 (C5123656); 5454001-40, 00000062055YC9 (KW1067CT01); 5454001-40, 00000067502YC5 (KW1012CT03); 5454001-40, 00000061248YC1 (LB1027CT03); 5454001-40, 00000067498YC6 (1553CT0004); 5454001-40, 00000060450YC4 (1553CT0001); 5454001-40, 00000063385YC9 (OM1073CT01); 5454001-40, 00000064813YC9 (PK1124CT02); 5454001-40, 00000059545YC4 (PL3007CT03); 5454001-40, 00000068719YC4 (RU2751CT02); 5454001-40, 00000070630YC9 (SA2459CT01); 5454001-40, 00000055610YC0 (SA2299CT01); 5454001-40, 00000060445YC4 (SA2332CT01); 5454001-40, 00000059868YC0 (SA2426CT01); 5454001-40, 00000070843YC8 (SA2458CT01); 5454001-40, 00000069393YC7 (SA2462CT01); 5454001-40, 00000066737YC8 (SA1239CT02); 5454001-40, 00000062752YC1 (SA2268CT01); 5454001-40, 00000060449YC6 (AE1485CT01); 5454001-40, 00000059870YC6 (AE1476CT01); 5454001-40, 00000071619YC1 (AE1546CT01); 5454001-40, 00000059546YC2 (859210138); 5454001-40, 00000070848YC7 (859210154); 5454001-140, 443306CN6 (480425CT1); 5454001-140, 00000438837CN7 (KPFO660CT2); 5454001-40, 67557YC9. (410535OP660); 5454001-140, 441138CN5 (910450CT3); 5454001-140, 441597CN2 (419407CT64); 5454001-140, 439141CN3 (864261ERCT); 5454001-140, 442828CN0 (605256660CT); 5454001-140, CJRAX1600031CN (615867VACT); 5454001-40, 00000067495YC2 (361SACT6603); 5454001-140, 00000438920CN1 (509486NVCT3); 5454001-40, 00000067558YC7 (262687OPT660); 5454001-140, CJRMX1600001CN (CJRMX1600001CN); 5454001-140, CJRMX1600004CN (CJRMX1600004CN); 5454001-140, CJRMX1600007CN (CJRMX1600007CN); 5454001-140, 00000441554CN3 (CT441554CN3); 5454001-240, 00000405578HM6 (10583CTM01); 5454001-240, 00000409896HM8 (10583CTM02); 5454001-40, 60595YC6 (843782CT); 5454001-40, 00000074500YC0 (DZ2014CT01); 5454001-40, 00000058110YC8 (853210274); 5454001-40, 00000076302YC9 (853210306); 5454001-40, 00000067560YC3 (853210291); 5454001-40, 00000066081YC1 (853210289); 5454001-40, 00000066082YC9 (853210290); 5454001-40, 00000059867YC2 (853210272); 5454001-40, 00000061246YC5 (853210277); 5454001-40, 00000063727YC2 (853210284); 5454001-40, 00000069667YC4 (853210297); 5454001-40, 00000074067YC0 (853210299); 5454001-40, 00000062053YC4 (853210280); 5454001-40, 00000065208YC1 (853210288); 5454001-40, 40GAN1700004YC (409868CT06); 5454001-240, 00000382286HM3 (PC3383CT02); 5454001-40, 00000058396YC3 (409868CT04); 5454001-240, 00000432092HM5 (O005CT12); 5454001-240, 00000436247HM1 (00150CTS01); 5454001-140, CJRAX1600073CN (CT1555CL); 5454001-240, 00000369364HM5 (ZA2608CT01); 5454001-240, 00000461074HM7 (83021604891917); 5454001-240, 00000397410HM2 (83021864230915); 5454001-240, 00000378363HM6 (83021860024814); 5454001-240, 00000376506HM2 (B208415203); 5454001-240, 00000368078HM2 (530332CT660); 5454001-240, 410651HM4 (510437ATR2); 5454001-240, 403538HM2 (760446CT2); 5454001-40, 00000057671YC0 (541426CT660); 5454001-240, 00000420998HM7 (PL1580CT03); 5454001-140, CJRAX1600075CN (CT2970CL); 5454001-240, 00000413611HM5 (409868CT05); 5454001-240, 00000411636HM4 (371961CT01); 5454001-240, 00000394181HM2 (HC1806CT02); 5454001-240, 00000398386HM3 (259130CT04); 5454001-240, 00000373268HM2 (HJJH01CT01); 5454001-240, 00000404915HM1 (CC7614CT04); 5454001-240, 00000390618HM7 (IL1191CT01); 5454001-40, 00000063141YC6 (HC1770CT04); 5454001-240, 00000368927HM0 (IE1166CT01); 5454001-40, 00000063381YC8 (83124805328); 5454001-240, 00000388471HM5 (83021806885314); 5454001-140, 438898CN9 (765747ER660); 5454001-140, Not Available (605234660CT); 5454001-140, CJRAX1600071CN (CT9939CL); 5454001-240, 00000376509HM6 (00314CTS04); 5454001-40, 00000057988YC8 (83113600922); 5454001-40, 00000057987YC0 (6073CT0006); 5454001-40, 00000057389YC9 (856210182); 5454001-40, 00000059547YC0 (856210169); 5454001-240, 00000378362HM8 (A8488413); 5454001-240, 00000375267HM2 (P003CT03); 5454001-240, 00000409895HM0 (M1400935); 5454001-240, 00000383052HM8 (A9190172); 5454001-240, 00000370113HM3 (479394M660); 5454001-240, 00000368928HM8 (970858CT); 5454001-140, 88636818 (LARGOEDCT660); 5454001-40, 00000057387YC3 (904337MKCT); 5454001-40, 00000057388YC1 (765832UCCT); 5454001-40, 00000059871YC4 (810376CT1); 5454001-240, 00000370112HM5 (702990CT660); 5454001-40, 63899YC9 (504712OPT); 5454001-140, 442885CN0 (508828ERCT); 5454001-140, 438545CN6 (607274ED660); 5454001-230, 00000416330HM9 (82421110129); 5454001-230, 00000456736HM8 (82421111119); 5454001-230, 00000448388HM9 (82421111124); 5454001-230, 00000458651HM7 (82421111108); 5454001-230, 00000464977HM8 (82421111145); 5454001-230, 00000432924HM9 (82421111102); 5454001-230, 00000393824HM8 (82421040343); 5454001-230, 00000407929HM9 (82421040360); 5454001-230, 00000408420HM8 (82421040357); 5454001-230, 00000409408HM2 (82421040365); 5454001-230, 00000415075HM1 (82421040367); 5454001-230, 00000418608HM6 (82421040353); 5454001-230, 00000426309HM1 (82421040368); 5454001-230, 00000442964HM3 (82421040385); 5454001-230, 00000447669HM3 (82421040388); 5454001-230, 00000418247HM3 (82421220129); 5454001-230, 00000426312HM5 (82421220131); 5454001-230, 00000452013HM6 (82421220150); 5454001-230, 00000456734HM3 (82421220151); 5454001-230, 00000427492HM4 (82421150168); 5454001-230, 00000418249HM9 (82421150161); 5454001-230, 00000466269HM8 (82421150179); 5454001-230, 00000393825HM5 (82421150154); 5454001-230, 00000416332HM5 (82421290096); 5454001-230, 00000461822HM9 (82421290113); 5454001-230, 00000423287HM2 (82421290097); 5454001-230, 00000465932HM2 (82421290114); 5454001-230, 00000427491HM6 (82421290102); 5454001-230, 00000445935HM0 (82421290107); 5454001-230, 00000451286HM9 (82421290108); 5454001-230, 00000401146HM6 (82421160390); 5454001-230, 00000447668HM5 (82421160455); 5454001-230, 00000456737HM6 (82421160465); 5454001-230, 00000464320HM1 (82421160467); 5454001-230, 00000440636HM9 (82421160439); 5454001-230, 00000391826HM5 (82421160391); 5454001-230, 00000416331HM7 (82421160411); 5454001-230, 00000417376HM1 (82421160412); 5454001-230, 00000423727HM7 (82421160408); 5454001-230, 00000434594HM8 (82421160433); 5454001-230, 00000427493HM2 (82421160426); 5454001-230, 00000415727HM7 (82421160414); 5454001-230, 00000439083HM7 (82421240186); 5454001-230, 00000458650HM9 (82421240193); 5454001-230, 00000463102HM4 (82421240198); 5454001-230, 00000399208HM8 (82421240160); 5454001-230, 00000404924HM3 (82421240165); 5454001-230, 00000415073HM6 (82421240166); 5454001-230, 00000438245HM3 (82421240184); 5454001-230, 00000463099HM2 (82421240195); 5454001-230, 00000434593HM0 (82421170049); 5454001-230, 00000464976HM0 (82421170061); 5454001-230, 00000442965HM0 (82421070442); 5454001-230, 00000392177HM2 (82421070393); 5454001-230, 00000383095HM7 (82421070382); 5454001-230, 00000442122HM8 (82421070443); 5454001-230, 00000367774HM7 (82421070348); 5454001-230, 00000427494HM0 (82421070427); 5454001-230, 00000432925HM6 (82421070387); 5454001-230, 00000435152HM4 (82421070401); 5454001-230, 00000461666HM0 (82421070460); 5454001-230, 00000464975HM2 (82421070454); 5454001-230, 00000451284HM4 (82421070447); 5454001-230, 00000383097HM3 (82421070364); 5454001-230, 00000423729HM3 (82421070416); 5454001-230, 00000459918HM9 (82421010376); 5454001-230, 00000463100HM8 (82421010380); 5454001-230, 00000456733HM5 (82421010366); 5454001-230, 00000400583HM1 (82421010330); 5454001-230, 00000468129HM2 (82421090609); 5454001-230, 00000458063HM5 (82421090588); 5454001-230, 00000461045HM7 (82421090598); 5454001-230, 00000434596HM3 (82421090563); 5454001-230, 00000383096HM5 (82421090508); 5454001-230, 00000439738HM6 (82421090569); 5454001-230, 00000426310HM9 (82421090553); 5454001-230, 00000440637HM7 (82421210364); 5454001-230, 00000407926HM5 (82421210326); 5454001-230, 00000392137HM6 (82421210315); 5454001-230, 00000380001HM8 (82421210306); 5454001-230, 00000442118HM6 (82421210354); 5454001-230, 00000444847HM8 (82421210365); 5454001-230, 00000463620HM5 (82421210380); 5454001-230, 00000418607HM8 (82421210341); 5454001-230, 00000461048HM1 (82421190361); 5454001-230, 00000458062HM7 (82421190365); 5454001-230, 00000417377HM9 (82421190320); 5454001-230, 00000435153HM2 (82421190336); 5454001-230, 00000456735HM0 (82421190362); 5454001-230, 00000423730HM1 (82421190323); 5454001-230, 00000458685HM5 (82421190345); 5454001-230, 00000440387HM9 (82421190342); 5454001-230, 00000441932HM1 (82421190348); 5454001-230, 00000451287HM7 (82421190353); 5454001-230, 00000459917HM1 (82421190360); 5454001-230, 00000426313HM3 (82421190322); 5454001-230, 00000468388HM4 (82421190370); 5454001-230, 00000433663HM2 (82421120515); 5454001-230, 00000401144HM1 (82421120480); 5454001-230, 00000440640HM1 (82421120544); 5454001-230, 00000461044HM0 (82421120494); 5454001-230, 00000440639HM3 (82421120559); 5454001-230, 00000449096HM7 (82421120551); 5454001-230, 00000459916HM3 (82421120545); 5454001-230, 00000392176HM4 (82421120463); 5454001-230, 00000415074HM4 (82421120500); 5454001-230, 00000450867HM7 (82421120567); 5454001-230, 00000420265HM1 (82421120531); 5454001-230, 00000408419HM0 (82421200258); 5454001-230, 00000439082HM9 (82421200278); 5454001-230, 00000442120HM2 (82421200284); 5454001-230, 00000467723HM3 (82421200300); 5454001-230, 00000400584HM9 (82421200250); 5454001-230, 00000433661HM6 (82421200276); 5454001-230, 00000448389HM7 (82421020219); 5454001-230, 00000428865HM0 (82421020212); 5454001-230, 00000401145HM8 (82421030412); 5454001-230, 00000461047HM3 (82421030459); 5454001-230, 00000448392HM1 (82421030454); 5454001-230, 00000435746HM3 (82421030195); 5454001-230, 00000416333HM3 (82421030205); 5454001-230, 00000392178HM0 (82421030427); 5454001-230, 00000448390HM5 (82421030453); 5454001-230, 00000426308HM3 (82421030209); 5454001-230, 00000452010HM2 (82421060165); 5454001-230, 00000367773HM9 (82421060149); 5454001-230, 00000461824HM5 (82421060176); 5454001-230, 00000423731HM9 (82421060162); 5454001-230, 00000442121HM0 (82421060167); 5454001-230, 00000463621HM3 (82421300076); 5454001-230, 00000439737HM8 (82421280059); 5454001-230, 00000429012HM8 (82421280058); 5454001-230, 00000398857HM3 (82421100765); 5454001-230, 00000452012HM8 (82421100924); 5454001-230, 00000461825HM2 (82421100918); 5454001-230, 00000461826HM0 (82421100917); 5454001-230, 00000468128HM4 (82421100935); 5454001-230, 00000426314HM1 (82421100799); 5454001-230, 00000391824HM0 (82421100755); 5454001-230, 00000406493HM7 (82421100778); 5454001-230, 00000426315HM8 (82421100832); 5454001-230, 00000452009HM4 (82421100902); 5454001-230, 00000461823HM7 (82421100923); 5454001-230, 00000403512HM7 (82421100780); 5454001-230, 00000437614HM1 (82421100837); 5454001-230, 00000442963HM5 (82421100872); 5454001-230, 00000383098HM1 (82421100748); 5454001-230, 00000401750HM5 (82421100771); 5454001-230, 00000409409HM0 (82421100789); 5454001-230, 00000444715HM7 (82421100881); 5454001-230, 00000433662HM4 (82421100845); 5454001-230, 00000436775HM1 (82421100861); 5454001-230, 00000461049HM9 (82421100919); 5454001-230, 00000468304HM1 (82421100932); 5454001-230, 00000461046HM5 (82421100907); 5454001-230, 00000400581HM5 (82421100776); 5454001-230, 00000440638HM5 (82421100854); 5454001-230, 00000417375HM3 (82421100795); 5454001-230, 00000461667HM8 (82421130195); 5454001-230, 00000448391HM3 (82421310294); 5454001-230, 00000401147HM4 (82421080149); 5454001-230, 00000445934HM3 (82421080165); 5454001-230, 00000451999HM7 (82421080168); 5454001-230, 00000440484HM4 (82421310290); 5454001-230, 00000441933HM9 (82421310292); 5454001-230, 00000463103HM2 (82421310306); 5454001-230, 00000406042HM2 (82421310270); 5454001-230, 00000420264HM4 (82421310278); 5454001-230, 00000447670HM1 (82421310300); 5454001-230, 00000452011HM0 (82421310298); 5454001-230, 00000459915HM5 (82421310304); 5454001-230, 00000468387HM6 (82421310309); 5454001-230, 00000423728HM5 (82421310281); 5454001-230, 00000463619HM7 (82421310303); 5454001-230, 00000441935HM4 (82421080161); 5454001-230, 00000466270HM6 (82421080172); 5454001-230, 00000391827HM3 (82421310248); 5454001-230, 00000434595HM5 (82421080159); 5454001-230, 00000461668HM6 (82421230418); 5454001-230, 00000407927HM3 (82421230359); 5454001-230, 00000451288HM5 (82421230407); 5454001-230, 00000464321HM9 (82421230425); 5454001-230, 00000418248HM1 (82421230365); 5454001-230, 00000412426HM9 (82421230357); 5454001-230, 00000468305HM8 (82421050124); 5454001-230, 00000392136HM8 (82421270153); 5454001-230, 00000399209HM6 (82421270155); 5454001-230, 00000467722HM5 (82421270178); 5454001-230, 00000380399HM6 (82421270144); 5454001-230, 00000438244HM6 (82421260021); 5454001-230, 00000442119HM4 (82421250185); 5454001-230, 00000463101HM6 (82421250195); 5454001-230, 00000449097HM5 (82421250188); 5454001-230, 00000458061HM9 (82421140402); 5454001-230, 00000383094HM0 (82421140317); 5454001-230, 00000430969HM6 (82421140369); 5454001-230, 00000413448HM2 (82421140353); 5454001-230, 00000397527HM3 (82421140332); 5454001-230, 00000418250HM7 (82421140347); 5454001-230, 00000400582HM3 (82421140335); 5454001-230, 00000451289HM3 (82421140389); 5454001-230, 00000462645HM3 (82421140413); 5454001-230, 00000464978HM6 (82421190381); 5454001-230, 00000510240HM5 (082421080193); 5454001-230, 00000489389HM7 (082421310323); 5454001-60, 00000063929YC4 (83129073471); 5454001-60, 00000071236YC4 (83144741048); 5454001-60, 00000066951YC5 (83136988535); 5454001-60, RE36A1700008YC (HC7137CT03); 5454001-60, RE36A1600020YC (828210101); 5454001-60, RE36A1600133YC (828210106); 5454001-60, RE36A1600016YC (83021600420416); 5454001-260, CBCFG1600055HM (83021865200517); 5454001-60, RE36A1700005YC (EV0054); 5454001-60, 00000067352YC5 (EV0014); 5454001-60, 00000067579YC3 (EV0015); 5454001-60, 00000066663YC6 (EV0004); 5454001-60, RE36A1600156YC (EV0047); 5454001-60, RE36A1600160YC (EV0099); 5454001-60, 00000068586YC7 (EV0019); 5454001-60, 00000068587YC5 (EV0018); 5454001-60, RE36A1700025YC (EV0052); 5454001-60, RE36A1700010YC (EV0073); 5454001-60, RE36A1700045YC (EV0049); 5454001-60, RE36A1600046YC (EV0036); 5454001-60, 00000072035YC9 (EV0026); 5454001-60, 00000062892YC5 (EV0001); 5454001-60, 00000070661YC4 (EV0021); 5454001-60, RE36A1600085YC (EV0029); 5454001-60, RE36A1600093YC (EV0038); 5454001-60, RE36A1700027YC (EV0057); 5454001-60, 00000066759YC2 (EV0012); 5454001-60, 00000067195YC8 (EV0016); 5454001-60, 00000070264YC7 (EV0020); 5454001-60, RE36A1700041YC (EV0066); 5454001-60, RE36A1600059YC (EV0027); 5454001-60, RE36A1700032YC (EV0053); 5454001-60, 00000070069YC0 (EV0022); 5454001-60, RE36A1700004YC (EV0060); 5454001-60, 00000069692YC2 (EV0025); 5454001-60, RE36A1600144YC (EV0045); 5454001-60, 00000063738YC9 (EV0007); 5454001-60, 00000072032YC6 (EV0024); 5454001-60, RE36A1600078YC (EV0033); 5454001-60, RE36A1700006YC (EV0044); 5454001-60, 00000062547YC5 (EV0002); 5454001-60, 00000067542YC1 (EV0017); 5454001-60, 00000065276YC8 (EV0008); 5454001-60, RE36A1600084YC (EV0035); 5454001-60, RE36A1700039YC (EV0058); 5454001-60, RE36A1600028YC (EV0030); 5454001-60, RE36A1600077YC (EV0034); 5454001-60, RE36A1600097YC (EV0039); 5454001-60, RE36A1600126YC (EV0050); 5454001-60, RE36A1600142YC (EV0046); 5454001-60, RE36A1700046YC (EV0059); 5454001-60, 00000062548YC3 (EV0005); 5454001-60, 00000062943YC6 (EV0006); 5454001-60, 00000071098YC8 (EV0023); 5454001-60, RE36A1600116YC (EV0040); 5454001-60, RE36A1700007YC (EV0051); 5454001-60, RE36A1600068YC (EV0032); 5454001-60, RE36A1600072YC (EV0028); 5454001-60, RE36A1600082YC (EV0031); 5454001-60, RE36A1700018YC (EV0048); 5454001-60, RE36A1600081YC (EV0037); 5454001-60, RE36A1600122YC (EV0041); 5454001-60, RE36A1600143YC (EV0055); 5454001-60, RE36A1700044YC (EV0090); 5454001-60, 00000062893YC3 (EV0003); 5454001-60, 00000063239YC8 (EV0009); 5454001-60, 00000065964YC9 (EV0010); 5454001-60, 00000066110YC8 (EV0011); 5454001-60, RE36A1600127YC (EV0043); 5454001-60, 00000069693YC0 (850210887); 5454001-60, 00000071100YC2 (850210891); 5454001-60, 00000072028YC4 (850210896); 5454001-60, 00000072029YC2 (850210897); 5454001-60, RE36A1600075YC (850210906); 5454001-60, 00000071799YC1 (850210895); 5454001-60, 00000072547YC3 (850210893); 5454001-60, 00000072030YC0 (850210892); 5454001-60, RE36A1600001YC (850210913); 5454001-60, RE36A1600092YC (850210911); 5454001-60, 00000068131YC2 (850210883); 5454001-60, 00000070072YC4 (850210886); 5454001-260, 00000422153HM7 (PL1029CT03); 5454001-60, RE36A1600163YC (10894CTS01); 5454001-260, 00000422151HM1 (913MISHEVO); 5454001-260, 00000422152HM9 (207283REV1); 5454001-260, 00000425961HM0 (401845CT1); 5454001-60, RE36A1600050YC (C5237834); 5454001-60, 00000072031YC8 (00463CTS02); 5454001-60, 00000071800YC7 (00440CTS09); 5454001-160, CJRBX1500002CN (480882CTER); 5454001-160, CJRBX1600001CN (614GMCEVO); 5454001-160, CJRBX1600013CN (281338CT1); 5454001-260, 00000425959HM4 (303GSCT1); 5454001-260, Not Available (816347EVO); 5454001-260, 414951HM4 (212255CT1); 5454001-260, CBAFG1600002HM (IL1041CT05); 5454001-60, 00000068739YC2 (83140390372); 5454001-60, 00000071615YC9 (83147877398); 5454001-60, RE36A1600003YC (M4164352); 5454001-60, 00000072548YC1 (M4234243); 5454001-60, 00000071597YC9 (M4488948); 5454001-60, RE36A1600006YC (M5484843); 5454001-60, 00000071614YC2 (M4489347); 5454001-60, 00000071441YC0 (477494CT02); 5454001-60, RE36A1600053YC (HC1709CT04); 5454001-60, RE36A1600002YC (A343134111); 5454001-60, 00000070070YC8 (A5348013); 5454001-60, 00000072382YC5 (A5115180); 5454001-60, 00000071613YC4 (A5201222); 5454001-60, RE36A1600058YC (MQ1000CT04); 5454001-60, 00000071099YC6 (NO1049CT02); 5454001-260, 00000420289HM1 (NO1052CT03); 5454001-60, 00000071097YC0 (PL1673CT02); 5454001-60, RE36A1700001YC (O012CT01); 5454001-60, RE36A1600040YC (266120CT01); 5454001-60, RE36A1600071YC (01278CTS01); 5454001-160, CJREX1600001CN (614DHWEVO); 5454001-160, CJRBX1600002CN (614WMCEVO1); 5454001-260, 00000424356HM4 (219932EVO); 5454001-260, 00000425960HM2 (NO1201CT01); 5454001-60, 00000071599YC5 (83143725811); 5454001-60, RE36A1600011YC (83148554598); 5454001-60, 00000066757YC6 (83141198489); 5454001-60, 00000072247YC0 (A5102805); 5345001, CLCJX1500165CN (A5109238); 5454001-260, 00000426390HM1 (NO1024CT01); 5454001-260, 00000425962HM8 (NO1221CT01); 5454001-60, 00000072248YC8 (RO1324CT01); 5454001-60, 00000071598YC7 (RO1184CT02); 5454001-60, 00000071237YC2 (GP178346); 5454001-160, 442576CN5 (574772EVO); 5454001-60, 00000072033YC4 (ZA2812CT01); 5454001-61, EVOCN1800002YC (082421250226); 5454001-260, CBCFG1600003HM (203276REVEVO); 5454001-260, CBCGG1600004HM (SHI94701417CT1); 5454001-260, CBCGG1700019HM (918494REVO5E); 5454001-260, CBCFG1600017HM (512816EVOCT); 5454001-260, CBCGG1600038HM (608417EVO); 5454001-260, CBCGG1600020HM (304388EVO1); 5454001-60, RE36A1600009YC (719776EVO); 5454001-60, RE36A1600060YC (631351EVO); 5454001-60, RE36A1600121YC (414649OPTCT1); 5454001-60, RE36A1600090YC (030036CT01); 5454001-60, RE36A1600045YC (M40480235); 5454001-160, CJRBX1600045CN (970CPMCT1); 5454001-160, CJRBX1600068CN (303LAED660CT); 5454001-160, CJRBX1700015CN (770824HREV); 5454001-160, CJRBX1600071CN (312996CTEVO); 5454001-160, CJRBX1600046CN (317582REVO); 5454001-160, CJRBX1600007CN (CEEVOCT1); 5454001-160, CJRBX1600017CN (603332EVO); 5454001-160, CJRBX1600073CN (580252REV1); 5454001-160, CJRBX1600011CN (830308VVEVO); 5454001-160, CJRBX1700014CN (254PROVCT1); 5454001-160, CJRBX1600060CN (703369REVO); 5454001-60, RE36A1600008YC (123285CT01); 5454001-60, RE36A1600087YC (779205CT01); 5454001-260, CBCFG1600012HM (334288REV); 5454001-260, CBCFG1600042HM (480425REV1); 5454001-260, CBCFG1600046HM (818885EVO); 5454001-260, CBCGG1600003HM (813916EVOER); 5454001-260, CBCFG1600053HM (352596CTEVO); 5454001-260, CBCFG1600015HM (850863REVO); 5454001-260, CBCGG1600032HM (305251REV); 5454001-260, CBCGG1600042HM (765649REVO); 5454001-260, CBCFG1600031HM (877285REV); 5454001-260, CBCFG1600013HM (765453REV); 5454001-260, CBCGG1600012HM (765747REVO); 5454001-260, CBCFG1600035HM (317204REVO); 5454001-260, CBCFG1600018HM (574224EVO); 5454001-260, CBCFG1600010HM (585922REVO1); 5454001-260, CBCFG1600016HM (440695CTEVO); 5454001-260, CBCFG1600030HM (216363CTEVO); 5454001-260, CBCFG1600036HM (216692CTEVO); 5454001-260, CBCFG1600026HM (216581CTEVO); 5454001-260, CBCFG1600023HM (541382EVO); 5454001-260, CBCGG1600007HM (814877EVO); 5454001-260, CBCGG1600008HM (215710RBOROCT); 5454001-260, CBCGG1600001HM (864530EVO); 5454001-260, CBCFG1600022HM (605338CTEVO); 5454001-260, CBCFG1600043HM (214633REV); 5454001-260, CBCGG1600002HM (613432EVO); 5454001-260, CBCFG1600020HM (M2748226); 5454001-60, RE36A1600141YC (CT6007AR); 5454001-60, RE36A1600057YC (808891CT01); 5454001-160, CJRBX1700009CN (CT7686PE); 5454001-60, RE36A1600123YC (480882REV); 5454001-60, 72034YC2 (310900EVO1); 5454001-60, RE36A1600067YC (765359REVO); 5454001-60, RE36A1600105YC (765448EVO1); 5454001-60, RE36A1600032YC (913345EVO); 5454001-60, RE36A1600021YC (913676EVO); 5454001-60, RE36A1600015YC (603421REV); 5454001-60, RE36A1600018YC (603433REV); 5454001-60, RE36A1600153YC (RADNETCT1562); 5454001-60, RE36A1600131YC (505262REVO); 5454001-60, RE36A1600111YC (212523REV); 5454001-60, RE36A1600145YC (212523REVO); 5454001-60, RE36A1600025YC (330896EVO); 5454001-60, RE36A1600118YC (614WMCEVO2); 5454001-60, RE36A1600041YC (570421ERCT); 5454001-60, RE36A1600094YC (214828REVO); 5454001-60, RE36A1600149YC (435893EVO); 5454001-60, RE36A1600022YC (804320CT660); 5454001-60, RE36A1600044YC (020144CT01); 5454001-60, RE36A1600147YC (705435EVO); 5454001-60, RE36A1600136YC (905895EVO1); 5454001-60, RE36A1600157YC (905895EVO2); 5454001-60, RE36A1600140YC (613283EVO1); 5454001-60, RE36A1600155YC (418862EVO); 5454001-60, RE36A1600062YC (M4476535); 5454001-60, RE36A1700020YC (M4409443); 5454001-60, RE36A1600019YC (X118874111); 5454001-60, RE36A1700028YC (M4499411); 5454001-60, RE36A1600063YC (M4182834); 5454001-60, RE36A1600056YC (M4199027); 5454001-60, RE36A1700036YC (M4145034); 5454001-60, RE36A1600161YC (A4495715); 5454001-60, RE36A1600007YC (M4472026); 5454001-60, RE36A1600034YC (M1481112); 5454001-60, RE36A1600091YC (HC7048CT03); 5454001-60, RE36A1600073YC (631485CT02); 5454001-60, RE36A1600096YC (TP0575CT01); 5454001-60, RE36A1600079YC (83021864008816); 5454001-60, RE36A1600036YC (A5292321); 5454001-60, RE36A1600088YC (A5165722); 5454001-60, RE36A1600029YC (A5691227); 5454001-60, RE36A1600055YC (A5196358); 5454001-60, RE36A1600124YC (PL3955CT01); 5454001-60, RE36A1700003YC (ZA2029CT04); 5454001-60, RE36A1600061YC (UE260530); 5454001-160, CJRBX1600028CN (480324EVO); 5454001-160, CJRBX1600024CN (303SANCT3); 5454001-160, CJRBX1600036CN (561747CT1B); 5454001-160, CJRBX1600026CN (386326CT1); 5454001-160, CJRBX1600032CN (305284EVO); 5454001-160, CJRBX1600020CN (706291ERCT); 5454001-160, CJRBX1600003CN (208677EVO); 5454001-160, CJRBX1600027CN (260920EVOCT); 5454001-160, CJRBX1600069CN (260248EVO); 5454001-160, CJRBX1600050CN (913676SMBVCT); 5454001-160, CJRBX1600065CN (406363REV); 5454001-160, CJRBX1600033CN (704384REVO); 5454001-260, CBCGG1700018HM (LMEVOCT1); 5454001-160, CJRBX1600052CN (440997EVO); 5454001-160, CJRBX1600012CN (513686ERCT); 5454001-160, CJRBX1600056CN (440312CTEVO); 5454001-160, CJRBX1600006CN (843692GSCT3); 5454001-160, CJRBX1600048CN (423844EVO); 5454001-160, CJRBX1600025CN (972540REVO); 5454001-160, CJRBX1600063CN (210METROEVO); 5454001-160, CJRBX1600067CN (757261REV); 5454001-160, CJRBX1600070CN (705647EVO1); 5454001-260, CBCGG1600021HM (408358OPT660); 5454001-260, CBCFG1600033HM (530OROEVO); 5454001-260, CBCGG1600015HM (530626EVO); 5454001-260, CBCFG1600047HM (831476CT); 5454001-260, CBCFG1600040HM (805652REV); 5454001-260, CBCGG1600024HM (303UROCT); 5454001-260, CBCGG1700010HM (561965REV1); 5454001-260, CBCGG1600030HM (386TLREVOCT); 5454001-260, CBCFG1600006HM (352344CTEVO); 5454001-260, CBCFG1600048HM (352344OPEVO); 5454001-260, CBCFG1600014HM (407933ORHREV); 5454001-260, CBCGG1600022HM (561346REV); 5454001-260, CBCFG1600025HM (954721EVO); 5454001-260, CBCGG1600009HM (813873EVO); 5454001-260, CBCFG1600021HM (706278RCT); 5454001-260, CBCGG1600026HM (SMSLMS2); 5454001-260, CBCGG1700022HM (208557EVO); 5454001-260, CBCGG1700004HM (630946EVO); 5454001-260, CBCFG1600027HM (708349EVO); 5454001-260, CBCFG1600007HM (574546EVO); 5454001-260, CBCFG1600029HM (316962REV); 5454001-260, CBCFG1600050HM (810606EVO); 5454001-260, CBCGG1700009HM (517437EVO); 5454001-260, CBCFG1600044HM (601894REVO); 5454001-260, CBCGG1600037HM (662773REV); 5454001-260, CBCFG1600034HM (406283REV); 5454001-260, CBCGG1700003HM (406654EVO); 5454001-260, CBCGG1600017HM (406323CT); 5454001-260, CBCGG1700005HM (704994REVO); 5454001-260, CBCGG1600043HM (402887REVO); 5454001-260, CBCGG1700011HM (732671EVO); 5454001-260, CBCFG1600009HM (505863REVO); 5454001-260, CBCFG1600039HM (718375EVO); 5454001-260, CBCGG1600034HM (516632CTRM7); 5454001-260, CBCGG1700021HM (716592REVO); 5454001-260, CBCFG1600045HM (419824CT1); 5454001-260, CBCGG1600040HM (918542REVO); 5454001-260, CBCGG1600031HM (503652REVO); 5454001-260, CBCGG1600027HM (814889CTEVO); 5454001-260, CBCGG1600044HM (843832REVO); 5454001-260, CBCGG1600036HM (615769REV); 5454001-260, CBCGG1600019HM (254729REV); 5454001-260, CBCGG1600025HM (512SMMCEVO); 5454001-260, CBCGG1600039HM (254PROVCT2); 5454001-260, CBCFG1600005HM (703287REVEVO); 5454001-260, CBCFG1600008HM (757886CT1); 5454001-260, CBCGG1600018HM (920262EVO1); 5454001-260, CBCFG1600052HM (304256REV); 5454001-60, RE36A1600012YC (909558CTEVO); 5454001-60, RE36A1600049YC (510357EVO); 5454001-60, RE36A1600148YC (352333REVO); 5454001-60, RE36A1600066YC (904702SJCT); 5454001-60, RE36A1600043YC (561747CT2B); 5454001-60, 71235YC6 (407OVCT); 5454001-60, RE36A1600146YC (727521CTEVO); 5454001-60, RE36A1600120YC (770918EVO); 5454001-60, RE36A1700019YC (478757CTEVO); 5454001-60, RE36A1600017YC (770736EVO); 5454001-60, RE36A1700033YC (765584EVO); 5454001-60, RE36A1600083YC (504988REVO); 5454001-60, RE36A1700023YC (301609REVO); 5454001-160, CJRBX1600064CN (269673EVO); 5454001-60, RE36A1600117YC (WMCOIEVO); 5454001-60, RE36A1700009YC (615344REVO); 5454001-60, RE36A1600098YC (575521EVO); 5454001-60, RE36A1600086YC (212312ER); 5454001-60, RE36A1600109YC (814837EVO); 5454001-60, RE36A1600112YC (931403OPT); 5454001-60, RE36A1600024YC (817465EVOCT); 5454001-60, RE36A1600023YC (972566EVO); 5454001-60, RE36A1600047YC (915577EVO); 5454001-60, RE36A1600151YC (817377EVO); 5454001-60, RE36A1600074YC (540994REV); 5454001-60, RE36A1600108YC (276596REV); 5454001-60, RE36A1600114YC (509962KREVCT); 5454001-60, RE36A1600070YC (83153613787); 5454001-60, RE36A1600030YC (83151144325); 5454001-60, RE36A1600051YC (83153230695); 5454001-60, RE36A1600089YC (83155767313); 5454001-60, RE36A1600026YC (X147408510); 5454001-60, RE36A1600080YC (M4489650); 5454001-60, RE36A1600027YC (M2663617); 5454001-60, RE36A1700022YC (M1230016); 5454001-60, RE36A1600052YC (X224352305); 5454001-60, RE36A1600039YC (M24582606); 5454001-60, RE36A1600099YC (126540CT01); 5454001-60, RE36A1600076YC (83021104241416); 5454001-60, RE36A1600069YC (5512CT0046); 5454001-60, RE36A1600065YC (5512CT0056); 5454001-60, RE36A1600125YC (RO1102CT03); 5454001-60, RE36A1600064YC (5973CT0063); 5454001-60, RE36A1600132YC (5973CT0069); 5454001-60, RE36A1600135YC (5973CT0071); 5454001-60, RE36A1600154YC (5973CT0079); 5454001-160, CJRBX1600054CN (205554REV); 5454001-160, CJRBX1700006CN (501450EVO); 5454001-160, CJRBX1600039CN (520WCT); 5454001-160, CJRBX1600055CN (310900EVO2); 5454001-260, CBCFG1600032HM (615812REVO); 5454001-260, CBCGG1600033HM (561965REV); 5454001-160, CJRBX1700017CN (863533CTEVO); 5454001-160, CJRBX1600008CN (813681CTEVO); 5454001-160, CJRBX1600061CN (850863REV); 5454001-160, CJRBX1700011CN (352333REV); 5454001-160, CJRBX1700008CN (386TCREVOCT); 5454001-160, CJRBX1600029CN (727834CTEVO); 5454001-160, CJRBX1600018CN (850CTEVO); 5454001-160, CJRBX1600031CN (352307EVO); 5454001-160, CJRBX1600010CN (515222RECT); 5454001-160, CJRBX1600030CN (217762REVO); 5454001-260, CBCGG1700015HM (708425EVO); 5454001-60, RE36A1600113YC (812283CT3); 5454001-160, CJRBX1600057CN (617421EVO); 5454001-160, CJRBX1600058CN (781624REVO); 5454001-160, CJRBX1600035CN (517353EVO); 5454001-160, CJRBX1600005CN (248652EVO); 5454001-160, CJRBX1600009CN (601288EVO); 5454001-160, CJRBX1600053CN (973322EVOCT); 5454001-160, CJRBX1600037CN (702880EVO); 5454001-60, RE36A1700043YC (702962ERCT); 5454001-160, CJRBX1700003CN (516632CTER); 5454001-160, CJRBX1600040CN (614293EVO1); 5454001-160, CJRBX1600021CN (513981CT); 5454001-160, CJRBX1600047CN (405636REVO); 5454001-260, CBCGG1600023HM (541247EVO); 5454001-160, CJRBX1600022CN (717272CTER); 5454001-160, CJRBX1600042CN (717854CT2); 5454001-160, CJRBX1700012CN (864512REVO); 5454001-160, CJRBX1600038CN (843228EVO); 5454001-160, CJRBX1600059CN (843355CT); 5454001-160, CJRBX1600023CN (843272CT); 5454001-160, CJRBX1600072CN (615777REVO); 5454001-60, RE36A1600138YC (979739EVO); 5454001-260, CBCFG1600038HM (254559REV); 5454001-260, CBCFG1600011HM (817639EVO); 5454001-160, CJRBX1600062CN (972334REV); 5454001-160, CJRBX1700004CN (409985REV); 5454001-160, CJRBX1600004CN (435462CT); 5454001-160, CJRBX1600014CN (435713CT); 5454001-160, CJRBX1600043CN (435676REV); 5454001-160, CJRBX1600074CN (801465REV); 5454001-260, CBCGG1600029HM (262289EVOCT); 5454001-60, RE36A1600150YC (304388EVO2); 5454001-160, CJRBX1700005CN (CT9287UY); 5454001-60, RE36A1700034YC (83021608240017); 5454001-60, RE36A1700037YC (560060CT02); 5454001-60, RE36A1700013YC (1553CT0010); 5454001-60, RE36A1700035YC (262010CT07); 5454001-60, RE36A1600162YC (725120CT01); 5454001-260, CBCGG1600016HM (NO1041CT01); 5454001-260, CBCGG1600006HM (PL1107CT04); 5454001-60, RE36A1700038YC (BH1047CT01); 5454001-60, RE36A1600104YC (110002CT05); 5454001-60, RE36A1700011YC (220826CT03); 5454001-60, RE36A1600095YC (M4171232); 5454001-60, RE36A1700280YC (HC7669CT06); 5454001-60, RE36A1700002YC (GH1078CT01); 5454001-60, RE36A1700030YC (A5333078); 5454001-60, RE36A1700021YC KW1088CT02); 5454001-60, RE36A1700029YC (KW1018CT04); 5454001-60, RE36A1600119YC (600040CT01); 5454001-60, RE36A1600137YC (PL1780CT03); 5454001-60, RE36A1600106YC (SA1010CT05); 5454001-60, RE36A1600107YC (SA2513CT02); 5454001-60, RE36A1600134YC (ZA2240CT02); 5454001-60, RE36A1700042YC (00521CTS02); 5454001-260, CBCGG1600035HM (M41981135); 5454001-60, RE36A1600054YC (83159000785); 5454001-60, RE36A1700040YC (HU1573CT01); 5454001-60, RE36A1700016YC (5512CT0070); 5454001-260, CBCGG1700007HM (A126322012); 5454001-60, RE36A1600130YC (A5196611); 5454001-60, RE36A1600158YC (A5320425); 5454001-60, RE36A1700012YC (A5359621); 5454001-60, RE36A1700014YC (B5320435); 5454001-60, RE36A1700017YC (A5138130); 5454001-60, RE36A1700024YC (A103421313); 5454001-60, RE36A1600110YC (PL0030CT01); 5454001-60, RE36A1700015YC (5973CT0083); 5454001-60, RE36A1600129YC (5973CT0068); 5454001-60, 00000062224YC1 (PC1455CT02); 5454001-60, 00000063575YC5 (CG2027CT04); 5454001-60, 00000064526YC7 (HC3295CT02); 5454001-60, 00000061001YC4 (M2753437); 5454001-60, 00000063576YC3 (D007CT08); 5454001-60, 00000061302YC6 (R002CT07); 5454001-60, 00000061547YC6 (AE1019CT02); 5454001-60, 00000059898YC7 (AE1445CT01); 5454001-60, 00000060310YC0 (AE1200CT01); 5454001-60, 00000061242YC4 (AE1122CT01); 5454001-60, 00000061548YC4 (AE1487CT01); 5454001-260, CBCFG1600004HM (83021247166016); 5454001-260, CBCFG1600037HM (83021242150416); 5454001-260, CBCFG1600054HM (83021246415317); 5454001-260, CBCGG1700001HM (83021827123117); 5454001-260, CBCGG1700012HM (83021828063217); 5454001-260, CBCGG1700017HM (83021868075017); 5454001-260, CBCFG1600056HM (83021284827817); 5454001-260, CBCGG1700020HM (83021608422417); 5454001-60, 69443YC0 (614RMHEVOCT); 5454001-160, CJREX1600004CN (CJREX1600004CN); 5454001-160, CJRBX1600049CN (CT8469CO); 5454001-260, CBCGG1700002HM (83021608239917); 5454001-160, Not Available (614GMOEVO); 5454001-160, CJREX1600003CN (514934EVO); 5454001-260, 418734HM0 (AM13RBW105); 5454001-260, 398107HM3 (604874EVO1); 5454001-260, 406001HM8 (416284EVO); 5454001-260, 396563HM9 (418748EVO); 5454001-260, 00000390008HM1 (514934GACT2); 5454001-260, 403057HM3 (450746CT1); 5454001-260, CBCGG1700013HM (83021849162417); 5454001-60, Not Available (DMHCT2); 5454001-60, 00000067354YC1 (83136988079); 5454001-60, 00000065277YC6 (83130826251); 5454001-60, 00000068130YC4 (DK1366CT01); 5454001-60, 00000063739YC7 (DK1025CT04); 5454001-60, 00000067540YC5 (DK1108CT03); 5454001-60, 00000070263YC9 (EG1196CT01); 5454001-60, 00000066661YC0 (M2510010); 5454001-60, 00000065963YC1 (M9861919); 5454001-60, 00000065962YC3 (M4165448); 5454001-60, 00000058952YC3 (M5484997); 5454001-60, 00000066756YC8 (A4201351); 5454001-60, 00000066760YC0 (M2850015); 5454001-60, RE36A1600010YC (GF1002CT04); 5454001-60, 00000066662YC8 (361441CT01); 5454001-60, 00000068129YC6 (HC1384CT01); 5454001-60, 00000065636YC3 (PC6066CT09); 5454001-60, 00000068588YC3 (157022CT02); 5454001-60, 00000068737YC6 (A5696313); 5454001-60, 00000067351YC7 (A5691225); 5454001-60, 00000069879YC5 (A409329101); 5454001-60, 00000059897YC9 (PL1476CT02); 5454001-60, 00000068996YC8 (PL3616CT01); 5454001-60, 00000066761YC8 (SI4031CT01); 5454001-60, 00000069880YC3 (ZA1247CT02); 5454001-60, 00000070071YC6 (ZA2103CT03); 5454001-60, 00000062944YC4 (BD002CT01); 5454001-60, 00000062945YC1 (R002CT08); 5454001-60, 00000065965YC6 (00468CTS02); 5454001-60, 00000067580YC1 (10717CTS01); 5454001-160, CJREX1600002CN (705645EVO); 5454001-160, 443690CN3 (819275CT); 5454001-160, 00000443690CN3 (AM13SNS17.DEL); 5454001-260, 419707HM5 (819537EVO); 5454001-160, 443746CN3 (AM141FK43); 5454001-260, 407415HM9 (832505CT2); 5454001-60, RE36A1600102YC (CT4422AR); 5454001-60, 00000067578YC5 (83137071341); 5454001-60, 00000068997YC6 (83136997603); 5454001-60, 61919YC7 (418248EVO); 5454001-60, 00000070262YC1 (CC0618CT10); 5454001-60, 00000063574YC8 (KW1001CT06); 5454001-60, 00000062942YC8 (AE1036CT01); 5454001-60, 00000061546YC8 (AE1140CT01); 5454001-60, 00000061241YC6 (AE1042CT01); 5454001-60, 00000061545YC0 (AE1022CT02); 5454001-160, CJREX1500001CN (708747EVO); 5454001-160, Not Available (DMHCT1); 5454001-160, 442551CM8 (514934ERCT); 5454001-260, 00000415732HM7 (M2148257); 5454001-60, 00000066952YC3 (M1118331); 5454001-60, 00000066664YC4 (M1807238); 5454001-60, 00000067575YC1 (B5483452); 5454001-260, 00000415733HM5 (D4200417); 5454001-60, 00000066111YC6 (M190136616) |
| FEI Number |
3002808529
|
Recalling Firm/ Manufacturer |
GE Healthcare Japan Corporation
Ge Herusukea.Japan ( Kabu -4 Chome 7-127
Asahigaoka
Hino
Japan
|
| For Additional Information Contact | 800-437-1171 |
Manufacturer Reason for Recall | GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry. |
FDA Determined Cause 2 | Employee error |
| Action | The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost.
Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative.
GE Healthcare expanded the recall to include an additional product and model number. Customers were notified on about 08/09/2018 via URGENT MEDICAL DEVICE CORRECTION letter. Instructions were the same as the original notification. |
| Quantity in Commerce | 4,007 devices total (products 1, 2, 3) |
| Distribution | Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|