 Class 2 Device Recall AIRO Mobile CT Systemsee related information
Date Initiated by FirmAugust 01, 2018
Create DateSeptember 14, 2018
Recall Status1 Terminated 3 on December 13, 2019
Recall NumberZ-2991-2018
Recall Event ID 80867
510(K)NumberK180393 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductAIRO Mobile CT System Model # MobiCT-32
Code Information Model # MobiCT-32
FEI Number 3010151377
Recalling Firm/
Manufacturer
Mobius Imaging, LLC
2 Shaker Rd Ste F100
Shirley MA 01464-2535
For Additional Information Contact
978-796-5068
Manufacturer Reason
for Recall
The Tube Current Modulation feature (Modulated Scans) is not working in AIRO systems with software version 2.0.0.0, and operators would not be able to detect this fault until after a scan is completed.
FDA Determined
Cause 2
Radiation Control for Health and Safety Act
ActionMobius Imaging, LLC sent an URGENT MEDICAL DEVICE RECALL letter to customers advising that a voluntary recall related to the AIRO Mobile CT System was initiated. The letter included the reason for the recall and actions to be taken by the customers.
Quantity in Commerce147
DistributionUS and foreign distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = JAK
-
-