| | Class 2 Device Recall Abbott Alinity i |  |
| Date Initiated by Firm | February 26, 2019 |
| Create Date | April 18, 2019 |
| Recall Status1 |
Terminated 3 on September 28, 2022 |
| Recall Number | Z-1196-2019 |
| Recall Event ID |
82511 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Abbott Alinity i Processing Module LN 03R65-01 - Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples. |
| Code Information |
AI01015 AI01100 AI01149 AI01180 AI01246 AI01292 AI01335 AI01376 AI01020 AI01101 AI01150 AI01181 AI01249 AI01293 AI01336 AI01377 AI01025 AI01102 AI01151 Ai01182 AI01252 AI01294 AI01337 AI01378 AI01027 AI01103 AI01152 AI01183 AI01253 AI01295 AI01338 AI01379 AI01032 AI01104 AI01153 AI01184 AI01254 AI01296 AI01339 AI01380 AI01036 AI01105 AI01154 AI01185 AI01255 AI01297 AI01340 AI01381 AI01054 AI01109 AI01155 AI01187 AI01256 AI01298 AI01341 AI01382 AI01055 AI01110 AI01156 AI01188 AI01257 AI01299 AI01343 AI01383 AI01056 AI01111 AI01157 AI01189 AI01258 AI01300 AI01344 AI01384 AI01057 AI01112 AI01158 AI01190 AI01259 AI01301 AI01345 AI01385 AI01061 AI01114 AI01159 AI01193 AI01260 AI01302 AI01346 AI01386 AI01064 AI01115 AI01160 Ai01194 AI01261 AI01303 AI01347 AI01387 AI01065 AI01116 AI01161 AI01196 AI01262 AI01304 AI01348 AI01388 AI01066 Ai01117 Ai01162 AI01197 AI01263 AI01305 AI01349 AI01389 AI01068 AI01120 AI01164 AI01198 AI01264 AI01306 AI01350 AI01390 AI01069 AI01122 AI01165 AI01200 AI01265 AI01307 AI01351 AI01391 AI01071 AI01123 AI01166 AI01201 AI01266 AI01308 AI01352 AI01392 Ai01072 AI01124 AI01167 AI01203 AI01267 AI01309 AI01353 AI01393 AI01073 AI01125 AI01168 AI01204 AI01268 AI01310 AI01354 AI01394 AI01074 AI01126 AI01169 AI01205 AI01269 AI01311 AI01355 AI01395 AI01075 AI01127 AI01170 AI01206 AI01270 AI01312 AI01356 AI01396 AI01079 AI01128 AI01171 AI01207 AI01271 AI01313 AI01357 AI01397 AI01081 AI01129 AI01172 AI01208 AI01272 AI01314 AI01358 AI01398 AI01082 AI01130 AI01173 AI01209 AI01274 Ai01315 AI01359 AI01399 AI01083 AI01131 AI01174 AI01210 AI01275 AI01316 AI01360 AI01400 AI01084 Ai01132 AI01175 AI01211 AI01276 AI01317 AI01361 AI01401 Ai01085 AI01133 AI01176 AI01212 AI01277 AI01319 AI01362 AI01402 AI01086 AI01136 AI01177 AI01213 AI01279 AI01320 AI01363 AI01403 AI01087 AI01137 AI01178 AI01214 AI01280 AI01321 AI01364 AI01404 AI01088 AI01138 AI01179 AI01216 AI01281 Ai01322 AI01365 AI01405 AI01089 AI01139 AI01180 AI01217 AI01282 AI01323 AI01366 AI01406 AI01090 AI01140 AI01181 AI01218 AI01283 AI01324 AI01367 AI01407 AI01091 AI01141 Ai01182 AI01219 AI01284 AI01326 AI01368 AI01408 AI01092 AI01142 AI01183 AI01220 AI01285 AI01327 AI01369 AI01409 AI01093 AI01143 AI01184 AI01221 AI01286 AI01328 AI01370 AI01410 AI01095 AI01144 AI01185 AI01222 AI01287 AI01329 AI01371 AI01411 AI01096 AI01145 AI01187 AI01223 AI01288 AI01330 AI01372 AI01412 AI01097 AI01146 AI01188 AI01224 AI01289 AI01331 AI01373 AI01413 AI01098 AI01147 AI01189 AI01225 AI01290 AI01332 AI01374 Ai01414 AI01099 AI01148 AI01190 AI01226 AI01291 AI01333 AI01375 AI01416 AI01418 AI01459 AI01509 AI01553 AI01600 AI01641 AI01686 AI01729 AI01419 AI01460 AI01510 AI01554 AI01601 AI01642 AI01687 AI01730 AI01420 AI01461 AI01511 AI01555 AI01602 AI01643 AI01688 AI01731 AI01421 AI01462 AI01512 AI01556 AI01603 AI01644 AI01689 AI01732 AI01422 AI01463 AI01513 AI01557 AI01604 AI01645 AI01690 AI01733 AI01423 AI01464 AI01514 AI01558 AI01605 AI01646 AI01691 AI01734 AI01424 AI01465 AI01515 AI01560 AI01607 AI01647 AI01692 AI01735 AI01425 AI01466 AI01516 AI01561 AI01608 AI01648 AI01693 AI01736 AI01426 AI01467 AI01517 AI01562 AI01609 AI01649 AI01694 AI01737 AI01427 AI01468 AI01518 AI01563 AI01610 AI01650 AI01695 AI01738 AI01428 AI01471 AI01519 AI01564 AI01611 AI01651 AI01698 AI01739 AI01430 AI01472 AI01520 AI01565 AI01612 AI01652 AI01699 AI01740 AI01431 AI01473 AI01523 AI01566 AI01613 AI01653 AI01700 AI01741 AI01432 AI01474 AI01524 AI01567 AI01614 AI01654 AI01702 AI01742 AI01433 AI01475 AI01525 AI01568 AI01615 AI01655 AI01703 AI01743 AI01434 AI01476 AI01526 AI01569 AI01616 AI01656 AI01704 AI01744 AI01435 AI01479 AI01527 AI01570 AI01617 AI01657 AI01705 AI01745 AI01436 AI01480 AI01528 AI01571 AI01618 AI01658 AI01706 AI01746 AI01437 AI01483 AI01529 AI01572 AI01619 AI01659 AI01707 AI01747 AI01438 AI01484 AI01530 AI01573 AI01620 AI01660 AI01708 AI01748 AI01439 AI01485 AI01531 AI01574 AI01621 AI01661 AI01709 AI01749 AI01440 AI01486 AI01532 AI01575 AI01622 AI01662 AI01710 AI01750 AI01441 AI01487 AI01533 AI01576 AI01623 AI01663 AI01711 AI01751 AI01442 AI01488 AI01534 AI01579 AI01624 AI01664 AI01712 AI01752 AI01443 AI01489 AI01535 AI01580 AI01625 AI01665 AI01713 AI01753 AI01444 AI01490 AI01537 AI01581 AI01626 AI01670 AI01714 AI01754 AI01445 AI01491 AI01539 AI01584 AI01627 AI01671 AI01715 AI01755 AI01446 AI01492 AI01540 AI01585 AI01628 AI01672 AI01716 AI01756 AI01447 AI01493 AI01541 AI01586 AI01629 AI01673 AI01717 AI01757 AI01448 AI01494 AI01542 AI01587 AI01630 AI01675 AI01718 AI01758 AI01449 AI01495 AI01543 AI01588 AI01631 AI01676 AI01719 AI01759 AI01450 AI01496 AI01544 AI01590 AI01632 AI01677 AI01720 AI01760 AI01451 AI01497 AI01545 AI01591 AI01633 AI01678 AI01721 AI01761 AI01452 AI01498 AI01546 AI01592 AI01634 AI01679 AI01722 AI01762 AI01453 AI01499 AI01547 AI01593 AI01635 AI01680 AI01723 AI01763 AI01454 AI01502 AI01548 AI01594 AI01636 AI01681 AI01724 AI01764 AI01455 AI01503 AI01549 AI01595 AI01637 AI01682 AI01725 AI01766 AI01456 AI01505 AI01550 AI01597 AI01638 AI01683 AI01726 AI01768 AI01457 AI01507 AI01551 AI01598 AI01639 AI01684 AI01727 AI01769 AI01458 AI01508 AI01552 AI01599 AI01640 AI01685 AI01728 AI01770 AI01772 AI01818 AI01859 AI01902 AI01948 AI01990 AI02032 AI02083 AI01773 AI01819 AI01860 AI01903 AI01949 AI01992 AI02033 AI02085 AI01774 AI01820 AI01861 AI01904 AI01950 AI01993 AI02034 AI02086 AI01776 AI01821 AI01862 AI01905 AI01951 AI01995 AI02035 AI02087 AI01779 AI01822 AI01863 AI01907 AI01952 AI01996 AI02036 AI02088 AI01781 AI01823 AI01864 AI01908 AI01953 AI01997 AI02037 AI02089 AI01782 AI01824 AI01865 AI01909 AI01954 AI01998 AI02038 AI02090 AI01783 AI01825 AI01866 AI01910 AI01955 AI01999 AI02039 AI02091 AI01784 AI01826 AI01867 AI01911 AI01956 AI02000 AI02040 AI02092 AI01786 AI01827 AI01868 AI01912 AI01957 AI02001 AI02041 AI02093 AI01787 AI01828 AI01869 AI01913 AI01958 AI02002 AI02042 AI02094 AI01788 AI01829 AI01870 AI01914 AI01959 AI02003 AI02043 AI02095 AI01789 AI01830 AI01871 AI01915 AI01960 AI02004 AI02044 AI02096 AI01790 AI01831 AI01872 AI01916 AI01961 AI02005 AI02045 AI02097 AI01791 AI01832 AI01873 AI01917 AI01962 AI02006 AI02046 AI02098 AI01792 AI01833 AI01875 AI01918 AI01963 AI02007 AI02047 AI02099 AI01793 AI01834 AI01877 AI01919 AI01964 AI02008 AI02048 AI02100 AI01794 AI01835 AI01878 AI01920 AI01966 AI02009 AI02049 AI02101 AI01795 AI01836 AI01879 AI01924 AI01967 AI02010 AI02050 AI02102 AI01797 AI01837 AI01880 AI01926 AI01968 AI02011 AI02053 AI02103 AI01798 AI01838 AI01881 AI01927 AI01969 AI02012 AI02054 AI02104 AI01799 AI01840 AI01882 AI01928 AI01970 AI02013 AI02060 AI02106 AI01800 AI01841 AI01884 AI01929 AI01971 AI02014 AI02061 AI02107 AI01801 AI01842 AI01885 AI01930 AI01972 AI02015 AI02062 AI02108 AI01802 AI01843 AI01886 AI01931 AI01973 AI02016 AI02063 AI02111 AI01803 AI01844 AI01887 AI01932 AI01974 AI02017 AI02064 AI02113 AI01804 AI01845 AI01888 AI01933 AI01975 AI02018 AI02065 AI02114 AI01805 AI01846 AI01889 AI01934 AI01976 AI02019 AI02066 AI02115 AI01806 AI01847 AI01890 AI01935 AI01977 AI02020 AI02067 AI02116 AI01807 AI01848 AI01891 AI01936 AI01978 AI02021 AI02068 AI02117 AI01808 AI01849 AI01892 AI01938 AI01979 AI02022 AI02069 AI02118 AI01809 AI01850 AI01893 AI01939 AI01980 AI02023 AI02071 AI02119 AI01810 AI01851 AI01894 AI01940 AI01981 AI02024 AI02072 AI02120 AI01811 AI01852 AI01895 AI01941 AI01982 AI02025 AI02073 AI02121 AI01812 AI01853 AI01896 AI01942 AI01984 AI02026 AI02074 AI02122 AI01813 AI01854 AI01897 AI01943 AI01985 AI02027 AI02075 AI02123 AI01814 AI01855 AI01898 AI01944 AI01986 AI02028 AI02076 AI02124 AI01815 AI01856 AI01899 AI01945 AI01987 AI02029 AI02078 AI02125 AI01816 AI01857 AI01900 AI01946 AI01988 AI02030 AI02079 AI02126 AI01817 AI01858 AI01901 AI01947 AI01989 AI02031 AI02081 AI02128 AI02131 AI02149 AI02166 AI02194 AI02230 AI02321 AI02362 AI02424 AI02132 AI02150 AI02169 AI02196 AI02233 AI02325 AI02363 AI02425 AI02133 AI02151 AI02175 AI02197 AI02238 AI02328 AI02364 AI02429 AI02134 AI02152 AI02176 AI02198 AI02248 AI02329 AI02379 AI02437 AI02136 AI02153 AI02178 AI02199 AI02250 AI02330 AI02380 AI02438 AI02137 AI02154 AI02180 AI02201 AI02267 AI02331 AI02381 AI02442 AI02138 AI02155 AI02181 AI02202 AI02278 AI02334 AI02383 AI02445 AI02139 AI02156 AI02182 AI02212 AI02282 AI02342 AI02385 AI02461 AI02140 AI02157 AI02183 AI02213 AI02283 AI02343 AI02386 AI02462 AI02141 AI02158 AI02184 AI02217 AI02288 AI02344 AI02389 AI02467 AI02142 AI02159 AI02185 AI02218 AI02297 AI02346 AI02394 AI02473 AI02143 AI02160 AI02186 AI02219 AI02298 AI02347 AI02397 AI02474 AI02144 AI02161 AI02187 AI02222 AI02302 AI02348 AI02398 AI02475 AI02145 AI02162 AI02188 AI02223 AI02303 AI02352 AI02399 AI02486 AI02146 AI02163 AI02189 AI02224 AI02306 AI02353 AI02406 AI02147 AI02164 AI02190 AI02225 AI02317 AI02354 AI02421 AI02148 AI02165 AI02193 AI02228 AI02319 AI02361 AI02422 AI01001 AI01047 AI01237 AI01538 AI02110 AI02240 AI02300 AI02408 AI02494 AI01002 AI01048 AI01238 AI01577 AI02112 AI02241 AI02304 AI02412 AI02499 AI01003 AI01049 AI01239 AI01578 AI02127 AI02243 AI02305 AI02415 AI02500 AI01004 AI01050 AI01242 AI01582 AI02129 AI02244 AI02307 AI02416 AI02502 AI01005 AI01051 AI01243 AI01583 AI02135 AI02245 AI02308 AI02417 AI02503 AI01006 AI01052 AI01247 AI01589 AI02167 AI02246 AI02309 AI02418 AI02504 AI01007 AI01053 AI01248 AI01596 AI02168 AI02247 AI02310 AI02419 AI02505 AI01008 AI01058 AI01250 AI01606 AI02170 AI02249 AI02313 AI02420 AI02506 AI01009 AI01059 AI01251 AI01666 AI02171 AI02251 AI02314 AI02423 AI02508 AI01010 AI01060 AI01273 AI01667 AI02172 AI02252 AI02320 AI02426 AI02510 AI01011 AI01062 AI01278 AI01668 AI02173 AI02253 AI02324 AI02427 AI02511 AI01012 AI01063 AI01318 AI01669 AI02174 AI02254 AI02326 AI02428 AI02514 AI01013 AI01067 AI01325 AI01674 AI02177 AI02255 AI02333 AI02431 AI02516 AI01014 AI01070 AI01334 AI01696 AI02191 AI02256 AI02335 AI02432 AI02517 AI01016 AI01076 AI01342 AI01697 AI02192 AI02257 AI02340 AI02433 AI02518 AI01017 AI01077 AI01415 AI01701 AI02195 AI02258 AI02341 AI02434 AI02540 AI01018 AI01078 AI01417 AI01765 AI02200 AI02259 AI02345 AI02435 AI02541 AI01019 AI01080 AI01429 AI01767 AI02203 AI02260 AI02349 AI02439 AI02544 AI01021 AI01094 AI01470 AI01771 AI02204 AI02261 AI02350 AI02440 AI02556 AI01022 AI01106 AI01477 AI01775 AI02205 AI02262 AI02351 AI02441 AI02564 AI01023 AI01107 AI01478 AI01777 AI02206 AI02263 AI02360 AI02444 AI02565 AI01024 AI01108 AI01481 AI01778 AI02207 AI02264 AI02368 AI02446 AI02579 AI01026 AI01113 AI01482 AI01780 AI02208 AI02268 AI02369 AI02447 AI02602 AI01028 AI01118 AI01500 AI01785 AI02209 AI02269 AI02373 AI02448 AI02609 AI01029 AI01119 AI01501 AI01796 AI02210 AI02271 AI02374 AI02450 AI02610 AI01030 AI01121 AI01504 AI01839 AI02211 AI02277 AI02375 AI02451 AI02616 AI01031 AI01134 AI01506 AI01874 AI02214 AI02279 AI02384 AI02453 AI02617 AI01033 AI01135 AI01521 AI01876 AI02215 AI02280 AI02387 AI02458 AI02621 AI01034 AI01163 AI01522 AI01883 AI02216 AI02281 AI02390 AI02459 AI02622 AI01035 AI01186 AI01536 AI01906 AI02220 AI02285 AI02391 AI02464 AI02628 AI01037 AI01191 AI01538 AI01921 AI02221 AI02286 AI02392 AI02468 AI02632 AI01038 AI01192 AI01577 AI01922 AI02226 AI02287 AI02393 AI02470 AI02634 AI01039 AI01195 AI01578 AI01923 AI02227 AI02290 AI02395 AI02471 AI02636 AI01040 AI01199 AI01582 AI01925 AI02229 AI02291 AI02396 AI02482 AI02667 AI01041 AI01202 AI01583 AI01937 AI02231 AI02292 AI02400 AI02484 AI02668 AI01042 AI01215 AI01589 AI01965 AI02232 AI02293 AI02401 AI02488 AI01043 AI01232 AI01596 AI01983 AI02234 AI02294 AI02403 AI02489 AI01044 AI01234 AI01606 AI01991 AI02235 AI02295 AI02404 AI02491 AI01045 AI01235 AI01666 AI02055 AI02237 AI02296 AI02405 AI02492 AI01046 AI01236 AI01667 AI02056 AI02239 AI02299 AI02407 AI02493 |
| FEI Number |
3002809144
|
Recalling Firm/ Manufacturer |
Abbott Gmbh & Co. KG
Max-Planck-Ring 2
65205
Wiesbaden
Germany
|
| For Additional Information Contact | Albert Chianello 877-422-2688 |
Manufacturer Reason for Recall | The safety interlock covering the septum piercing probes within the bulk solution bottle holder may not deploy when a bulk solution bottle is removed from the Alinity ci series and Alinity s System. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On February 26, 2019, Abbott sent a correction notice to there customers stating to please continue to follow the instructions provided in the Alinity ci series Operations Manual when replacing the bulk solutions. Use caution when accessing bulk solution bottle holders or replacing bulk solution bottles as you may be exposed to a sharp hazard.
The Alinity ciseries Operations Manual will be updated in the future to add the appropriate hazard information to the procedure. If the safety interlock or bottle release does not function correctly, please contact your area customer support for assistance.
Your Abbott representative will contact you to schedule the placement of hazard labeling in the bulk solution area to increase awareness of this hazard.
If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service. |
| Quantity in Commerce | 1097 |
| Distribution | Worldwide distribution - US Nationwide in the states of AR, CA, GA, ME, MN, MO, NY, OK, SC, TN, TX, UT, and VA, and countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, LATVIA, LEBANON, MADAGASCAR, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, and VIETNAM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|