| | Class 2 Device Recall Alinity c Cuvette Segments |  |
| Date Initiated by Firm | April 15, 2019 |
| Create Date | June 02, 2019 |
| Recall Status1 |
Terminated 3 on October 13, 2020 |
| Recall Number | Z-1709-2019 |
| Recall Event ID |
82644 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Alinity c Cuvette Segments, List Number 04S47-01
Product Usage:
The Alinity c Cuvettes are used on the Alinity c system processing module. The Alinity c processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. |
| Code Information |
All serial numbers |
| FEI Number |
3002809144
|
Recalling Firm/ Manufacturer |
Abbott Gmbh & Co. KG
Max-Planck-Ring 2
65205
Wiesbaden
Germany
|
| For Additional Information Contact | Albert Chianello 224-668-1923 |
Manufacturer Reason for Recall | Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On April 15, 2019, Abbott sent Product Correction Urgent - Immediate Action Required letters to all impacted Customers. US customers were sent these letters via FedEx priority overnight express while customers outside the US were notified via mailings or visits. The letters sent to US consignees were accompanied by Customer Reply form which were to be completed and sent back to Abbott. |
| Quantity in Commerce | 19,184 |
| Distribution | US Nationwide Distribution- AL, AR,CA, FL, GA, IL, LA, MA, ME, MN, MO, NJ, NY, OK, SC, TN, TX, UT, & VA
WORLDWIDE - Albania, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, Norway, Pakistan, Philippines, Poland, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Turkey, United Kingdom, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|