| Date Initiated by Firm | April 18, 2019 |
| Create Date | June 13, 2019 |
| Recall Status1 |
Terminated 3 on April 17, 2020 |
| Recall Number | Z-1779-2019 |
| Recall Event ID |
82650 |
| 510(K)Number | K132439 |
| Product Classification |
Sterilizer, steam - Product Code FLE
|
| Product | Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam
Product Usage:
The SKYTRON Integrity 270 Steam Sterilizer for use in health care facilities. The Integrity Sterilizer incorporates high-pressure steam to sterilize nonporous or porous items and materials used in the health care industry |
| Code Information |
Serial Numbers: 14092619, 14092620, 14102621, 17123520, 18113521 |
| FEI Number |
1825014
|
Recalling Firm/ Manufacturer |
Skytron, Div. The KMW Group, Inc
5085 Corporate Exchange Blvd SE
Grand Rapids
MI
49512-5515
|
| For Additional Information Contact | 616-656-2900 |
Manufacturer Reason for Recall | Weld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure |
FDA Determined Cause 2 | Device Design |
| Action | Skytron issued Medical Device Correction letter on 4/18/19 via certified mail stating: reason for recall, action to be taken and will replace the air heater assembly on the affected devices.
Contact Larry Perez, Vice President of Technical Services, at 616-656-2900 or www.lperez@skytron.us, or Bob Vreeland, at 616-656-1189 or www.bvreeland@skytron.us with any questions. For after-hours support, please call 1-800-759-8766 (SKYTRON). |
| Quantity in Commerce | 5 units |
| Distribution | US Nationwide Distribution - AK, IA, MS |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = FLE
|