• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Universal security cuffs

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Universal security cuffssee related information
Date Initiated by FirmFebruary 27, 2020
Create DateMarch 30, 2020
Recall Status1 Terminated 3 on May 10, 2022
Recall NumberZ-1604-2020
Recall Event ID 85102
Product Classification Restraint, protective - Product Code FMQ
ProductDeRoyal SECURITY CUFFS, Single Strap, QR Buckle Cuff and Bed, UNIVERSAL, REF M8138 - Product Usage: patient restraint.
Code Information GTIN: 0749756946865 Lot Code: 50192186
FEI Number 1043214
Recalling Firm/
Manufacturer
DeRoyal Industries Inc
200 Debusk Ln
Powell TN 37849-4703
For Additional Information Contact
865-938-7828
Manufacturer Reason
for Recall
One lot of product was inadvertently manufactured with a buckle which does not meet the mechanical hold force required for this device.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionDeRoyal issued their recall on 2/27/2020 to the end user level by emailing its direct accounts and sales representatives. The customers were asked to identify the affected products, quarantine them, return the affected product and the provided notice of return form back to DeRoyal indicating if replacement or credit is requested. Distributors were asked to further notify their customers. They were asked to do this directly or to provide the recalling firm with a customer listing and the recalling firm would notify them.
Quantity in Commerce94 pairs
DistributionVA, CA, FL, AZ
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-