| | Class 2 Device Recall The CapsoCam Plus |  |
| Date Initiated by Firm | February 26, 2020 |
| Create Date | July 08, 2020 |
| Recall Status1 |
Terminated 3 on January 28, 2021 |
| Recall Number | Z-2536-2020 |
| Recall Event ID |
85540 |
| 510(K)Number | K192662 |
| Product Classification |
System, imaging, gastrointestinal, wireless, capsule - Product Code NEZ
|
| Product | The CapsoCam Plus (SV-3); Device Common Name: System, Imaging, Gastrointestinal, Wireless, Capsule; Catalog Number 2795 - Product Usage: intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel. |
| Code Information |
Lot number 01-19-0070 with the following S/N and associated UDIs: S/N A00Y2K.700, UDI 010086777000020913200204172107011001-19-007021A00Y2K.700; S/N A00Z9J.700, UDI 010086777000020913200204172107011001-19-007021A00Z9J.700; S/N A011EP.700, UDI 010086777000020913200204172107011001-19-007021A011EP.700; S/N A011F3.700, UDI 010086777000020913200204172107011001-19-007021A011F3.700; S/N A013BM.700, UDI 010086777000020913200204172107011001-19-007021A013BM.700; S/N A013D8.700, UDI 010086777000020913200204172107011001-19-007021A013D8.700; S/N A013EH.700, UDI 010086777000020913200204172107011001-19-007021A013EH.700; S/N A013F9.700, UDI 010086777000020913200204172107011001-19-007021A013F9.700; S/N A013FP.700, UDI 010086777000020913200204172107011001-19-007021A013FP.700; S/N A014ZB.700, UDI 010086777000020913200204172107011001-19-007021A014ZB.700; S/N A01535.700, UDI 010086777000020913200204172107011001-19-007021A01535.700; S/N A01536.700, UDI 010086777000020913200204172107011001-19-007021A01536.700; S/N A015AJ.700, UDI 010086777000020913200204172107011001-19-007021A015AJ.700; S/N A015DD.700, UDI 010086777000020913200204172107011001-19-007021A015DD.700; S/N A015FB.700, UDI 010086777000020913200204172107011001-19-007021A015FB.700; S/N A015MK.700, UDI 010086777000020913200204172107011001-19-007021A015MK.700; S/N A015S1.700, UDI 010086777000020913200204172107011001-19-007021A015S1.700; S/N A015S2.700, UDI 010086777000020913200204172107011001-19-007021A015S2.700; S/N A015S9.700, UDI 010086777000020913200204172107011001-19-007021A015S9.700; S/N A015T3.700, UDI 010086777000020913200204172107011001-19-007021A015T3.700; S/N A015V7.700, UDI 010086777000020913200204172107011001-19-007021A015V7.700; S/N A015WH.700, UDI 010086777000020913200204172107011001-19-007021A015WH.700; S/N A015YC.700, UDI 010086777000020913200204172107011001-19-007021A015YC.700; S/N A01605.700, UDI 010086777000020913200204172107011001-19-007021A01605.700; S/N A016B1.700, UDI 010086777000020913200204172107011001-19-007021A016B1.700; S/N A016EN.700, UDI 010086777000020913200204172107011001-19-007021A016EN.700; S/N A016G4.700, UDI 010086777000020913200204172107011001-19-007021A016G4.700; S/N A016G7.700, UDI 010086777000020913200204172107011001-19-007021A016G7.700; S/N A016JB.700, UDI 010086777000020913200204172107011001-19-007021A016JB.700; S/N A01878.700, UDI 010086777000020913200204172107011001-19-007021A01878.700; S/N A01880.700, UDI 010086777000020913200204172107011001-19-007021A01880.700; S/N A0188Z.700, UDI 010086777000020913200204172107011001-19-007021A0188Z.700; S/N A0189M.700, UDI 010086777000020913200204172107011001-19-007021A0189M.700; S/N A018BZ.700, UDI 010086777000020913200204172107011001-19-007021A018BZ.700; S/N A018D2.700, UDI 010086777000020913200204172107011001-19-007021A018D2.700; S/N A018DY.700, UDI 010086777000020913200204172107011001-19-007021A018DY.700; S/N A018T7.700, UDI 010086777000020913200204172107011001-19-007021A018T7.700; S/N A018TA.700, UDI 010086777000020913200204172107011001-19-007021A018TA.700; S/N A018TD.700, UDI 010086777000020913200204172107011001-19-007021A018TD.700; S/N A018TF.700, UDI 010086777000020913200204172107011001-19-007021A018TF.700; S/N A018TN.700, UDI 010086777000020913200204172107011001-19-007021A018TN.700; S/N A018UC.700, UDI 010086777000020913200204172107011001-19-007021A018UC.700; S/N A018UL.700, UDI 010086777000020913200204172107011001-19-007021A018UL.700; S/N A018VF.700, UDI 010086777000020913200204172107011001-19-007021A018VF.700; S/N A018VG.700, UDI 010086777000020913200204172107011001-19-007021A018VG.700; S/N A018X5.700, UDI 010086777000020913200204172107011001-19-007021A018X5.700; S/N A018XP.700, UDI 010086777000020913200204172107011001-19-007021A018XP.700; S/N A018YR.700, UDI 010086777000020913200204172107011001-19-007021A018YR.700; S/N A01901.700, UDI 010086777000020913200204172107011001-19-007021A01901.700; S/N A0190B.700, UDI 010086777000020913200204172107011001-19-007021A0190B.700; S/N A0191F.700, UDI 010086777000020913200204172107011001-19-007021A0191F.700; S/N A0191L.700, UDI 010086777000020913200204172107011001-19-007021A0191L.700; S/N A0192C.700, UDI 010086777000020913200204172107011001-19-007021A0192C.700; S/N A0192G.700, UDI 010086777000020913200204172107011001-19-007021A0192G.700; S/N A0192J.700, UDI 010086777000020913200204172107011001-19-007021A0192J.700; S/N A0192X.700, UDI 010086777000020913200204172107011001-19-007021A0192X.700; S/N A0192Y.700, UDI 010086777000020913200204172107011001-19-007021A0192Y.700; S/N A01936.700, UDI 010086777000020913200204172107011001-19-007021A01936.700; S/N A0193A.700, UDI 010086777000020913200204172107011001-19-007021A0193A.700; S/N A0193S.700, UDI 010086777000020913200204172107011001-19-007021A0193S.700; S/N A0194A.700, UDI 010086777000020913200204172107011001-19-007021A0194A.700; S/N A0194D.700, UDI 010086777000020913200204172107011001-19-007021A0194D.700; S/N A0194K.700, UDI 010086777000020913200204172107011001-19-007021A0194K.700; S/N A0194P.700, UDI 010086777000020913200204172107011001-19-007021A0194P.700; S/N A0194X.700, UDI 010086777000020913200204172107011001-19-007021A0194X.700; S/N A01950.700, UDI 010086777000020913200204172107011001-19-007021A01950.700; S/N A0195S.700, UDI 010086777000020913200204172107011001-19-007021A0195S.700; S/N A0195Y.700, UDI 010086777000020913200204172107011001-19-007021A0195Y.700; S/N A01968.700, UDI 010086777000020913200204172107011001-19-007021A01968.700; S/N A0196W.700, UDI 010086777000020913200204172107011001-19-007021A0196W.700; S/N A0196Y.700, UDI 010086777000020913200204172107011001-19-007021A0196Y.700; S/N A0196Z.700, UDI 010086777000020913200204172107011001-19-007021A0196Z.700; S/N A01970.700, UDI 010086777000020913200204172107011001-19-007021A01970.700; S/N A0197C.700, UDI 010086777000020913200204172107011001-19-007021A0197C.700; S/N A0197W.700, UDI 010086777000020913200204172107011001-19-007021A0197W.700; S/N A01986.700, UDI 010086777000020913200204172107011001-19-007021A01986.700; S/N A0198Z.700, UDI 010086777000020913200204172107011001-19-007021A0198Z.700; S/N A01994.700, UDI 010086777000020913200204172107011001-19-007021A01994.700; S/N A0199V.700, UDI 010086777000020913200204172107011001-19-007021A0199V.700; S/N A0199Y.700, UDI 010086777000020913200204172107011001-19-007021A0199Y.700; S/N A019A1.700, UDI 010086777000020913200204172107011001-19-007021A019A1.700; S/N A019AN.700, UDI 010086777000020913200204172107011001-19-007021A019AN.700. |
| FEI Number |
3008062894
|
Recalling Firm/ Manufacturer |
Capso Vision, Inc.
18805 Cox Ave Ste 250
Saratoga
CA
95070-6615
|
| For Additional Information Contact | 408-866-6358 |
Manufacturer Reason for Recall | Ingestible video capsule system has capsule with incorrect capture mode which will not record data when the capsule is ingested by patients. If used, there is a risk of delay of diagnosis and the patient may need to repeat the exam. |
FDA Determined Cause 2 | Employee error |
| Action | Beginning 02/26/2020, the firm notified affected customers via email of "URGENT: Medical Device Voluntary Recall Letter". The letter indicated, as soon as the customer receives the notice, please provide acknowledgment of receipt. The letter requested actions to be taken by the customer/user: We request that you ship back the affected capsules using the provided RMA form and return shipping label. For any concerns or question email or contact the firm at (408) 866-6358. |
| Quantity in Commerce | 82 |
| Distribution | US Nationwide distribution including in the states of AZ, CA, DE, MD, NJ, NY, SC, and VA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = NEZ
|
|
|
|