| | Class 2 Device Recall E.cam or Symbia systems |  |
| Date Initiated by Firm | March 31, 2016 |
| Create Date | August 27, 2020 |
| Recall Status1 |
Terminated 3 on October 29, 2020 |
| Recall Number | Z-2898-2020 |
| Recall Event ID |
86230 |
| 510(K)Number | K142006 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | E.cam or Symbia systems that use foresight detectors - Product Usage: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. |
| Code Information |
Model Numbers: 8717733, 8717741, 10275007, 10275009, 10275010, 10607649, 10764801, 10764802, 10764803, 10764804, 10275879, 10413009, 4380213, 4380221, 5242826, 5977066, 5977074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7324143, 7760932, 7761161, 7823920, 7823938, 7823946, 7823953, 7823979, 10151531, 10151532, 10520745, 10607648. E. Cam series serial numbers: 8015, 8061, RC493, 8005, 8660, 8669, 8689, 8687, 8723, 8708, 1078, 1093, 1094, 1104, 1108, 1112, 1207, 11751, 1065, 11623, 1007, 1003, 9965, 8149, 1203, 5130, 1005, 1069, 1073, 1074,1007, 1076, 1080, 1081, 1082, 1095, 1009, 1098, 1010, 1099, 1011, 1109, 1016, 1136, 1137, 1139, 1150, 1151, 1152, 1019, 1155, 1154, 1163, 1026, 1167, 1174, 1176, 1186, 1187, 1188, 1194, 1192, 1211, 1210, 1032, 1220, 1222, 1223, 1233, 1039, 1248, 1247, 1252,1257, 1260, 1259, 1270, 1050, 1280, 1053,1291,1296, 1301, 1321, 1068, 1069, 1347, 1350, 1353, 1078, 1364, 1079, 70001, 70001, 70005, 70006, 70007, 70011, 70010, 70012, 70013, 70015, 70003, 10016, 70017, 70019, 70030, 70031, 70034, 70035, 70037, 70038, 70004, 70040, 70041, 70043, 70044, 70045, 70046, 70047, 70048, 70049, 1014, 1023, 1033, 1037, 1044,1045, 1039, 1051,1054, 1056, 1057, 1003, 1063, 1062, 9995, 1062, 1065, 1096, 1097, 1013, 1124, 1130, 1129, 1147, 1142, 1149, 1156, 1157, 1166, 1185, 1184, 1190, 1203, 1221, 1035, 1271, 1272, 1297, 1304, 1313, 1316, 1320, 1064, 1330, 1357, 1361,1360, 1017, 1027, 1029, 1034, 1036, 1040, 1041,1060, 1059, 1061,10319, 1162, 11070. Symbia series serial numbers: Serial No 1001, 1002, 1008, 1001, 1007, 1348, 1012, 1001, 1014, 1009, 1433, 1011, 1015, 1482, 1018, 1007, 1003, 1004, 1008, 1023, 1011, 1024, 1026, 1028, 1029, 1013, 1038, 1008, 1015,1009, 1011, 1012, 1013, 1015, 1016, 1040, 1045, 1017,1019, 1018,1049, 1022, 1034, 1025, 1056, 1026, 1038, 1027, 1058, 1028, 1029, 1030, 1025, 1027, 1028, 1031, 1032, 1039, 1034, 1038, 1039, 1041, 1042, 1043, 1040, 1044, 1051, 1045, 1005, 1052
1054 ,1055, 1057, 1005, 1009, 1015, 1030, 1038, 1067, 1073, 1002, 1116, 1128, 1144, 1143, 1152, 1156, 1167, 1007, 1181, 1186, 1189, 1190, 1191, 1011, 1205, 1222, 1224, 1240, 1242, 1246, 1247, 1251, 1015, 1249, 1275, 1280, 1281, 1284, 1285, 1289, 1017, 1308, 1326, 1327, 1342, 1348, 1350, 1351, 1026, 1027, 1358, 1361, 1364, 1029, 1030, 1387, 1400, 1411, 1417,1421, 1427, 1437, 1445, 1482, 1484, 1485, 1045, 1501, 1503, 1509, 1517, 1518, 1528,1533, 1537, 1542, 1549, 1550, 1055, 70003, 1057, 1655, 1657, 1658, 1659, 1062, 1662, 70010,1065, 70011, 70012, 1676, 70013, 1677, 1067, 70015, 1068, 70018, 1150, 1306, 1001, 1011, 1012, 1065,1068, 1078, 1082,1001, 1083,1004,1103,1105,1109,1129,1147,1347,1384,1410, 1144, 1204, 1281, 1297, 1314,1050,1051,1052,1056,1057,1437,1058,1441,1090,1091,1119,1092,1252,1161,1120,1122,1121,1124, 1123, 1125, 1126 ,1131, 1170, 1128, 1129, 1132, 1133, 1134, 1175, 1135, 1136, 1137, 1138, 1139, 1142, 1143, 1180, 1182, 1145, 1146, 1147, 1187, 1148, 1150, 1149, 1152, 1191, 1189, 1154, 1155, 1159, 1158, 1162, 1161, 1160, 1163, 1198,1164, 1165,1169,1168, 1172, 1171, 1203, 1202, 1204, 1170, 1179, 1180,1177, 1178 ,1182, 1183, 1206, 1056, 1184, 1186, 1187, 1191, 1192, 1057, 1194, 1197, 1196, 1195, 1094, 1095, 1208, 1198, 1097, 1099, 1100, 1201, 1203, 1205, 1207, 1208, 1066, 1104,1212, 1210, 1214, 1211, 1107, 1215, 1064, 1218, 1217, 1216, 1220, 1219, 1222, 1108, 1111, 1225, 1069, 1113, 1074,1114,1229, 1230, 1231, 1232, 1071, 1117, 1235, 1233,1236, 1234, 1241, 1242, 1244, 1243, 1245, 1240, 1247, 1086, 1246, 1248, 1121, 1249, 1250,1076, 1123, 1253, 1077, 1087, 1255, 1254, 1078, 1257, 1090, 1256, 1259, 1262, 1258, 1264, 1263, 1266, 1091, 1267, 1270, 1271, 1082, 1269, 1083, 1084, 1276, 1284, 1283, 1286, 1285, 1287, 1288, 1291,1290, 1131, 1295, 1296, 1298, 1300, 1303, 1299, 1302, 1301, 1304, 1308, 1305, 1306, 1088, 1313, 1099, 1311, 1101, 1102, 1319, 1318, 1321, 1322, 1323, 1320, 1138, 1091, 1139, 1327, 1329,1328,1144, 1332, 1334, 1335, 1145, 1337, 1094, 1095, 1096,1341,1338, 1347,1346,1098 ,1099,1349 ,1101,1150, 1350, 1351, 1352, 1152, 1354, 1114, 1104, 1356, 1105, 1366, 1363, 1367, 1116, 1070, 1365, 1364, 1108, 1368
1372 ,1369, 1373, 1375, 1376, 1374, 1071, 1381, 1382, 1384, 1383, 1111, 1385, 1389,1075, 1390,1391, 1393,1165,1395,1394, 1128, 1398,1167, 1168, 1130, 1405, 1404, 1403, 1169, 1411, 1408, 1410, 1412, 1409, 1413, 1417, 1418, 1419, 1120, 1421, 1423, 1424, 1425, 1428, 1427, 1429, 1430, 1176, 1432, 1179, 1180, 1083, 1182,1440, 1184, 1185, 1442, 1443, 1445, 1448, 1449,1446, 1444, 1084, 1450, 1451, 1454, 1455, 1086, 1461, 1192, 1134, 1462, 1463, 1162,1135 ,1466 ,1468, 1195,1171, 1469, 1470, 1090, 1175,1471, 1139, 1176,1472,1094,1178, 1179,1479,1480, 1142, 1098, 1483,1144, 1145, 1185, 1186, 1486, 1146, 1487,1100, 1491,1489,1488,1492,1493,1192,1193,1102, 1194, 1195, 1151, 1203, 1152, 1495, 1496, 1497, 1104, 1204,1205,1105, 1500,1501,1504, 1502,1506,1511,1509,1508,1510, 1517, 1157, 1158, 1513, 1512, 1515, 1518, 1206, 1159, 1519, 1521, 1520, 1209, 1210, 1160, 1523, 1525, 1524,1526,1161,1110,1162, 1211,1530, 1164, 1211, 1212, 1216, 1217, 1535, 1536, 1115,1213, 1540, 1542,1229, 1541, 1544, 1234, 1546, 1236,1237, 1550, 1553, 1551, 1552, 1555, 1556, 1216, 1174, 1558, 1559, 1242, 1561,1560, 1563, 1566, 1568,1567, 1569, 1570, 1180, 1573, 1253, 1576, 1183, 1184, 1578, 1257, 1186, 1582, 1258, 1583, 1130, 1586, 1262,1585, 1584, 1588, 1566, 1593, 1592, 1266, 1190, 1595,1594, 1267, 1597, 1596, 1599, 1600, 1273, 1603, 1606, 1605, 1607,1608, 1137, 1138, 1139, 1613, 1278, 1614, 1280, 1617, 1196, 1615, 1619, 1143, 1282, 1622, 1624, 1626, 1625, 1628, 1201, 1632, 1290, 1291, 1149, 1634, 1151, 1636, 1631, 1637, 1638, 1639, 1158, 1641, 2006, 1643, 1645, 2008, 2009, 1650, 1652, 1654, 1655, 1656, 1660, 1661,1662, 1663, 1307, 1664, 1665, 1310, 1170, 1668, 1672, 1670,1671, 1673, 1181, 1212, 1676, 1677,1678, 1182, 1679, 1315, 1682, 1680, 1683, 1217, 1684, 1317, 1685, 1319 , 1690, 1689, 1692, 1691, 1694, 1693, 1695, 1321, 1697, 1322, 1223, 1702, 1701, 1703, 1704, 1198, 1705, 1202, 1706, 1707, 1708, 1334, 1225, 1709, 1710, 1711, 1712, 1713, 1227, 1336, 1715, 1228 , 1338, 1229, 1339, 1719, 1717, 1718, 1716, 1231, 1723, 1232, 1343, 1722, 1217, 1724, 1725 ,1234, 1727, 1728, 1348, 1729, 1349, 1732, 1731, 1733, 1735, 1737, 1354, 1736, 1738, 1739, 1740, 1225, 1743, 1741, 1742, 1744, 1745, 1236, 1237, 1356, 1229, 1748, 1358, 1359, 1750, 1751,1363 ,1752 ,1753 ,1754 ,1364, 1235, 1755, 1756, 1237, 1240, 1759, 1242, 1760, 1761, 1762, 1763, 1764, 1365, 1767, 1766, 1367, 1246, 1768,1770 , 1247, 1773,1774, 1771, 1250, 1249, 1251, 1371, 1783, 1786, 1252, 1784, 1785, 1253, 1790, 1789, 1791, 1792, 1373, 1374, 1794, 1793, 1795, 1798, 1797, 1799, 1378, 1801, 1796, 1805, 1381, 1804, 1382, 1806, 1807, 1809, 1383, 1808, 1810, 1264, 1261, 1812, 1384, 1262, 1263, 1814, 1815, 1813, 1818, 1816, 1817, 1386, 1819, 1820, 1823, 1822, 1824, 1826, 1825, 1827, 1828, 1267, 1829, 1830, 1831, 1390, 1832, 1833, 1835, 1394, 1284, 1840, 1837, 1396, 1839, 1842, 1841, 1397, 1844, 1398, 1845, 1399, 1400, 1846, 1847, 1850, 1851, 2003, 1853, 1856, 1855, 1854, 1858, 1860, 1859, 1861, 2007, 2003, 2009, 1862, 2004, 1863, 2005, 2006, 1865, 1405, 1864, 1866, 1868, 1872, 2010, 1408, 2016, 1869, 1871,1870, 2018,1873, 1410, 1874, 1876, 1411, 1877, 1412, 2011, 2012, 1413, 1414, 1878, 1416, 1879, 1880, 1881,1883, 1885, 1886, 1888, 1889, 1890, 2001, 1898, 1896, 2025, 1897, 2003,1902, 2004, 2030, 1903, 1904,1905, 2005, 1907, 2021, 1909, 2006, 1920, 2007, 1949, 1948, 1950, 1951, 1112, 1113,1150, 1153, 1152, 1114, 1154,1115, 1118, 1144, 1168, 1167, 1169, 1177, 1151, 1193, 1188, 1199, 2008, 1268, 1282, 1293, 1312, 1422, 1434, 1465, 1499, 2024, 2014, 2015, 2002, 1006, 2004, 1011, 1013,1010, 2001, 2003, 2002, 2005, 2007, 1005, 1006, 2012, 2013, 1008, 2012, 1012, 2004, 2005, 2007, 2008, 2010, 2018, 2012, 2019, 2014, 2015, 2021, 2022, 2020, 2016, 2020, 2022, 2018, 2031, 2019, 2020, 2021, 2023, 2024, 2025, 1012, 2026, 2033, 2027,1016, 1002, 2029, 2038, 2030, 2039, 1018, 2031, 2032, 2033, 1023, 1024, 2036, 1027, 2034, 1029, 1028, 2035, 1032, 1003, 2038, 1004, 1005, 2039, 2048, 1040, 2040, 1006, 2041, 2049, 1007, 2050, 1008, 1009, 2051, 1010, 1011, 1075, 1076, 2038, 1012, 2039, 2040, 2054, 2046, 2047, 1014, 2041, 2042, 1018, 2100, 1020, 1053, 1019, 2059, 1058, 2049, 2049, 2050, 2052, 2051, 1023, 1062, 1063, 1024, 2059, 2047, 1025, 2109, 1911, 1910, 1912, 1913, 1914, 1915, 2016, 1916, 2017, 1919, 1918, 1922, 1924, 1923, 2030, 2019, 1925, 1926, 1931, 1935, 1937, 1936, 1938, 1939, 1940, 1941, 1945, 1946, 2004, 2005, 2006, 2007, 2012, 2013, 2014, 2015,2016 , 2017,2018 ,2021, 2022, 2023, 2024, 2030, 2029, 2031, 2032, 2037, 2040, 2042, 2036, 2041, 2047, 2048, 2049, 2050, 2051, 2052, 2057, 2056, 2060, 2058, 2062, 2061,2063, 2064, 2065, 2067, 2068, 2066, 2069, 2071, 2072, 2075, 2073, 2078, 2079, 2080, 2082, 2083, 2084, 2085, 2086, 2087, 2089, 2091, 2092, 2094, 2093, 2096, 2095, 2097, 2099, 2102, 2103, 2104, 2105, 2110, 2111, 2113, 2115, 2117, 2119, 2118, 2120, 2121, 70017, 2063, 2080, 2067, 2068, 2069, 2070, 2071, 2072, 1110, 1027. |
| FEI Number |
1000514697
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc.
2501 Barrington Rd
Hoffman Estates
IL
60192-2061
|
| For Additional Information Contact | Laura Meyer 847-304-7700 |
Manufacturer Reason for Recall | E. CAM and Symbia system with foresight detectors performing gated or dynamic acquisition may lose some detector time information. Our risk analysis indicates that the probability of occurrence is remote.
SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.
CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles.
SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
|
FDA Determined Cause 2 | Software design |
| Action | Based on the results of our risk assessment indicating that the probability of occurrence is remote, you may continue to use your system. Please ensure that this advisory is placed in the system's instructions for use.
To reduce the likelihood of this issue, perform Daily Shutdown and Start-up as described in the operator manual.
If you have sold this equipment and it is no longer in your possession, we kindly ask that you forward this notice to the new owner of this equipment. Please inform us about the new owner of the equipment.
Adverse events or quality problems experienced with the use of this product should be reported to Siemens through the contact information provided below and may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mailor by fax.
If you have any questions regarding this important notice, please contact your local Service representative at the contact numbers provided below.
o America: 1-800-888-7436
o Europe, Middle East, and Afr ica: +49 9131 940 4000
o Asia and Australia: +86 (21) 3811 2121 |
| Quantity in Commerce | 5,066 systems. |
| Distribution | US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|