• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Endoskeleton TC Interbody System with Titan nanoLOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Endoskeleton TC Interbody System with Titan nanoLOCKsee related information
Date Initiated by FirmMarch 17, 2021
Create DateApril 20, 2021
Recall Status1 Terminated 3 on May 03, 2023
Recall NumberZ-1439-2021
Recall Event ID 87626
510(K)NumberK192018 
Product Classification Intervertebral fusion device with bone graft, lumbar - Product Code MAX
ProductEndoskeleton TC Interbody System with Titan nanoLOCK Surface Technology, 6 Degree, Small, 8 mm, Product Number 5146-1208-N - Product Usage: indicated to be used with supplemental fixation cleared by the FDA for use in the cervical spine and autograft bone, allograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft with bone marrow aspirate, or a combination thereof.
Code Information GTIN 00191375013006, Lot Number TM0126527
FEI Number 3003120897
Recalling Firm/
Manufacturer
Medtronic Sofamor Danek USA, Inc
4340 Swinnea Rd
Memphis TN 38118-6603
For Additional Information ContactVictor Rocha
901-399-2401
Manufacturer Reason
for Recall
The shelf box identifies a small device actually contains a large device, and the shelf box identifying a large device actually contains a small device.
FDA Determined
Cause 2
Packaging process control
ActionThe firm initiated the recall by letter on 03/19/2021, The letter requested that the consignee locate and isolate the impacted product and contact the Medtronic Sales Representative to coordinate the return and replacement of the device.
Quantity in Commerce11 devices
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = MAX
-
-