| | Class 2 Device Recall Medtronic StimLoc Burr Hole Cover |  |
| Date Initiated by Firm | June 02, 2021 |
| Create Date | August 06, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2228-2021 |
| Recall Event ID |
88160 |
| 510(K)Number | K001096 |
| Product Classification |
Cover, burr hole - Product Code GXR
|
| Product | Medtronic StimLoc Burr Hole Cover, Part Numbers:
a) 3387S-40
b) 3389S-28
c) 3389S-40
d) 3391S-40
e) 3550S-01
f) 924256
g) DB-5000 |
| Code Information |
a) PART NUMBER: 3387S-40 - DBS Lead with StimLoc
LOTS # : VA21YUS, VA24RVV, VA25SDK, VA26Q2S, VA27X8H, VA2980G, VA29Z9A, VA2DA2Z,
VA21ZJU, VA24S65, VA25WK1, VA26XP2, VA27ZRF, VA29814, VA2A5MJ, VA2DA5Y, VA21ZYP, VA24S66, VA2613V, VA26ZFU, VA281L5, VA2985H, VA2AK84, VA2DA9E, VA222L5, VA24SRF, VA261EL, VA275SC, VA2820Y, VA29BSM, VA2APV8, VA2DEKH, VA227E2, VA24UVS, VA263GC, VA275SD, VA282CS, VA29E09, VA2B3PB, VA2DVM9, VA22LDZ, VA24ZFG, VA26460, VA276GS, VA283SF, VA29LP7, VA2B3PN, VA2DVME, VA236DC, VA254MA, VA2696J, VA277EB, VA28C1Z, VA29LPC, VA2B3PQ, VA2DVNX, VA23BWB, VA254MC, VA2696L, VA27JJV, VA28C36, VA29PPY, VA2BFSP, VA23BXM, VA254MD, VA269LV, VA27MQ3, VA28DVB, VA29S0E, VA2BHV6, VA23J1U, VA2562G, VA26AGU, VA27RMG, VA28LVR, VA29S0F, VA2CEVY, VA23J4X, VA257LH, VA26E5P, VA27SW9, VA28NFL, VA29SB3, VA2CPP0, VA23L21, VA25QDQ, VA26E5R, VA27V4Y, VA28NG1, VA29UNG, VA2CXDR, VA23S64, VA25RHC, VA26K1N, VA27WDK, VA28SC8, VA29UPA, VA2D9X8, VA24R3X, VA25S3A, VA26K1R, VA27WDP, VA28XL0, VA29Z98, VA2DA0S,
VA23LXN, VA23WY0, VA2446L, VA24AFN, VA24FA2, VA25DKB, VA23WXZ, VA23X94, VA249A6, VA24EYN, VA2562K, VA23LXN, VA23WY0, VA2446L, VA24AFN, VA24FA2, VA25DKB, VA23WXZ, VA23X94, VA249A6, VA24EYN and VA2562K
b) PART NUMBER: 3389S-28 DBS Lead with StimLoc
LOTS #: VA25K4W, VA25CMZ and VA25DKD
c) PART NUMBER: 3389S-40 - DBS Lead with StimLoc
LOTS #: VA21X0Y, VA24WXJ, VA260NN, VA27QE4, VA28NFM, VA29ZU0, VA2BE8A, VA2CQET, VA21ZJV, VA25101, VA26233, VA27RET, VA28SCE, VA2A3HK, VA2BEXC, VA2CQFM, VA21ZL1, VA259WX, VA2670U, VA27REU, VA28URD, VA2A6K7, VA2BF1G, VA2CQH2, VA227NQ, VA25B79, VA2672S, VA27SWB, VA28URE, VA2ACW4, VA2BFG0, VA2CRKW, VA228PU, VA25HSZ, VA26GGZ, VA27V8S, VA28XHX, VA2ACW5, VA2BHHW, VA2D6WD, VA22C6L, VA25N4Y, VA26GPA, VA27WDE, VA28ZQ0, VA2AD57, VA2BJEH, VA2D71G, VA22KF4, VA25N50, VA26KJ3, VA27ZA2, VA28ZQ2, VA2AD5G, VA2C8BG, VA2D730, VA22MXM, VA25N6D, VA26NWG, VA27ZRE, VA291YL, VA2AH9C, VA2CABX, VA2D739, VA22TDL, VA25RSZ, VA26Q16, VA280V2, VA291YM, VA2AHHR , VA2CBQ0, VA2D73E, VA22V9R, VA25TBA, VA26Q6E, VA283SH, VA2924E, VA2AHNG, VA2CFBH, VA2DAZA, VA22VZB, VA25UBK, VA26R4H, VA2865T, VA2924F, VA2AHNM, VA2CJ4Z , VA2DAZT, VA22Y79, VA25UG1, VA26TK5, VA2893X, VA29LPD, VA2AHT7, VA2CK66, VA2DB2C, VA237AP, VA25VKR, VA26XGF, VA289WG, VA29LUC, VA2AR36, VA2CK8L, VA2DB2L, VA2382J, VA25VQD, VA26ZFV, VA289WH, VA29MUW, VA2B3SB, VA2CLRF, VA2DB2V, VA2382Y, VA25XC7, VA272ZP, VA28DN6, VA29PLU, VA2B3SC, VA2CN0F, VA2DBAU, VA2384E, VA25YHU, VA275D6, VA28E06, VA29PM8, VA2B7G5, VA2CN0H, VA2DUR8, VA239JK, VA25YHW, VA276MJ, VA28KWJ, VA29UB3, VA2BDS1, VA2CN7Z, VA2DUR9, VA23P6D, VA25YR5, VA27E9P, VA28LHR, VA29UPD, VA2BDSE, VA2CN86, VA2DURH, VA23S5Z, VA25ZJ3, VA27JJS, VA28LHU, VA29UPH, VA2BDSX, VA2CN8A, VA24PD4, VA25ZJ6, VA27K8D, VA28M21, VA29ZES, VA2BE44, VA2CNX0, VA24RVR, VA25ZUB, VA27NE2, VA28M22, VA29ZM2, VA2BE7W, VA2CNXG, VA24WXH, VA25ZWB, VA27PDB, VA28M9B, VA29ZSC, VA2BE86, VA2CPPV, VA23MM7, VA23TVY, VA240PB, VA24237, VA2493U, VA24EYM, VA24L0S VA25LHE, VA23S65, VA23WXW, VA240YK, VA2440E, VA2493V, VA24G61 VA24M5P, VA25PZU, VA23TFY, VA23X95, VA241H1, VA244QV, VA24A58, VA24G7S, VA25B78, VA25QSW, VA23TVT, VA23XYL, VA241NE, VA245W8, VA24AYZ, VA24GWZ and VA25CF3
d) PART NUMBER: 3391S-40 - DBS Lead with StimLoc
LOTS #: VA2391V, VA25B91, VA2AD6N, VA2BJWK, VA2DN50 and VA2DR23
e) PART NUMBER: 3550S-01 - Spare Parts Kit for StimLoc¿ Burr Hole Cover
LOTS #: 082300620, 082307020, 082307620, 082329819, 082332919, 082335119, 082335320, 82316820 and 82326919
f) PART NUMBER: 924256 - StimLoc Burr Hole Cover
LOTS #: 082202020A, 082203820A, 082211120A, 082228819A, 082234520A, 082202121A, 082207920A, 082214120A, 082232319A, 082235019A and 082224219A
g) PART NUMBER: DB-5000 -StimLoc Support Clips
LOTS #: 082104420
|
| FEI Number |
2182207
|
Recalling Firm/ Manufacturer |
Medtronic Neuromodulation
7000 Central Ave Ne
Minneapolis
MN
55432-3568
|
| For Additional Information Contact | Medtronic Technical Services 800-707-0933 |
Manufacturer Reason for Recall | Excess nylon flash on the clip. |
FDA Determined Cause 2 | Process control |
| Action | On June 2,2021 the firm sent letter with the following information:
Dear Healthcare Professional,
Medtronic is voluntarily recalling specific lots of the StimLoc" Burr Hole Cover due to a manufacturing issue resulting in excess nylon flash on the clip. Please review the information contained in this letter, quarantine any affected product (see Attachment A for list of affected lot numbers) in your inventory for return to Medtronic, and sign and return the customer confirmation form included with this letter.
Medtronic email at RS.NeuAdverseEventReporting@Medtronic.com or Medtronic Technical Services at 1-800-707-0933 weekdays 7am - 6pm CST. |
| Quantity in Commerce | 15,192 |
| Distribution | AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI,
WV and Hawaii
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = GXR
|
|
|
|