Class 2 Device Recall Flexor CheckFlo Introducer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Flexor CheckFlo Introducersee related information
Date Initiated by FirmSeptember 28, 2021
Create DateOctober 29, 2021
Recall Status1 Terminated 3 on February 28, 2024
Recall NumberZ-0185-2022
Recall Event ID 88787
510(K)NumberK142829 
Product Classification Introducer, catheter - Product Code DYB
ProductFlexor Check-Flo Introducer, Introducer Set, RPN KCFW-6.0-35-45-RB, GPN G09908 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator.
Code Information lot 13861362
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Inc.
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Department
812-339-2235
Manufacturer Reason
for Recall
Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.
FDA Determined
Cause 2
Employee error
ActionOn September 28, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers should take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return to the firm. 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Cook Medical is removing potentially impacted devices from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue.
Quantity in Commerce48 (OUS)
DistributionDomestic distribution to AR AZ CT FL GA IL LA MA ME MN ND NJ NY OH PA TX WA. Foreign distribution to Germany, Finland, Italy, Netherlands, and Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = DYB
-
-