| | Class 2 Device Recall Bronchial Microbiology Brush |  |
| Date Initiated by Firm | December 28, 2021 |
| Date Posted | March 02, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0711-2022 |
| Recall Event ID |
89488 |
| 510(K)Number | K893729 |
| Product Classification |
endoscopic cytology brush - Product Code FDX
|
| Product | Bronchial Microbiology Brush, Catalog Number 4310 and 4320. Microbial sampling is performed by compressing a spring as the base of brush wire to expose the brush, scraping the desired surface, and releasing the compression on the spring which retracts the brush back into the sheath. A protector tip on the end of the brush prevent contamination of the sample during the endoscopy procedure. The force required to dislodge the tip from the bristle is greater than to break the brush wire at the handle body. |
| Code Information |
Affected lots were distributed in the United States from 1/18/2017 to 8/26/2021. These lots have expiration dates of 1/01/2022 through 12/01/2026.
Catalog Number 4310, Package DI M84943100, Package DI Number M84943101;
Catalog Number 4320, Package DI M84943200, Package DI Number M84943201;
Lot Numbers: H12-16-112
H12-16-113
H12-16-114
H01-17-002
H01-17-003
H01-17-004
H01-17-005
H01-17-101
H01-17-134
H01-17-135
H01-17-136
H02-17-044
H02-17-045
H02-17-046
H02-17-047
H02-17-048
H03-17-113
H03-17-114
H03-17-115
H03-17-116
H03-17-117
H03-17-233
H03-17-234
H03-17-235
H03-17-236
H03-17-237
H03-17-238
H05-17-119
H05-17-120
H05-17-121
H05-17-122
H05-17-123
H05-17-124
H06-17-142
H06-17-143
H06-17-144
H06-17-145
H07-17-112
H07-17-113
H07-17-114
H07-17-115
H08-17-127
H08-17-128
H08-17-129
H08-17-130
H08-17-131
H08-17-269
H08-17-270
H08-17-271
H08-17-272
H08-17-273
H09-17-102
H09-17-103
H09-17-104
H10-17-053
H10-17-056
H10-17-057
H10-17-058
H10-17-059
H10-17-060
H10-17-061
H11-17-146
H11-17-147
H11-17-148
H11-17-149
H11-17-150
H11-17-151
H12-17-120
H12-17-121
H12-17-122
H12-17-123
H12-17-124
H02-18-001
H02-18-002
H02-18-003
H02-18-004
H02-18-005
H02-18-209
H02-18-210
H02-18-211
H02-18-212
H02-18-219
H02-18-220
H02-18-221
H03-18-176
H03-18-177
H03-18-178
H03-18-179
H03-18-180
H04-18-038
H04-18-039
H04-18-040
H04-18-151
H04-18-152
H04-18-153
H04-18-154
H04-18-155
H04-18-156
H04-18-157
H05-18-177
H05-18-178
H05-18-179
H05-18-180
H06-18-111
H06-18-112
H06-18-113
H06-18-114
H07-18-017
H07-18-018
H07-18-019
H07-18-021
H07-18-020
H08-18-070
H08-18-071
H08-18-073
H08-18-072
H08-18-206
H08-18-207
H08-18-208
H10-18-117
H10-18-118
H10-18-120
H10-18-121
H10-18-119
H11-18-133
H11-18-134
H11-18-135
H11-18-136
H11-18-137
H01-19-023
H01-19-024
H01-19-025
H01-19-026
H01-19-027
H12-18-086
H12-18-087
H12-18-088
H12-18-089
H12-18-090
H01-19-178
H01-19-179
H01-19-180
H01-19-181
H01-19-182
H01-19-183
H02-19-184
H02-19-185
H02-19-186
H02-19-187
H02-19-188
H02-19-189
H02-19-190
H04-19-067
H04-19-068
H04-19-069
H04-19-070
H04-19-071
H04-19-072
H04-19-073
H05-19-102
H05-19-103
H05-19-104
H05-19-160
H05-19-161
H05-19-162
H05-19-163
H05-19-164
H06-19-001
H06-19-003
H06-19-004
H06-19-002
H06-19-310
H06-19-311
H06-19-312
H06-19-313
H06-19-314
H06-19-315
H06-19-316
H06-19-317
H07-19-129
H07-19-130
H07-19-131
H07-19-132
H07-19-133
H07-19-134
H08-19-176
H08-19-177
H08-19-178
H08-19-179
H08-19-180
H09-19-113
H09-19-114
H09-19-115
H09-19-116
H10-19-104
H10-19-105
H10-19-107
H10-19-106
H10-19-108
H11-19-120
H11-19-121
H11-19-122
H01-20-008
H01-20-009
H01-20-010
H01-20-011
H01-20-012
H01-20-013
H01-20-207
H01-20-208
H01-20-209
H01-20-210
H01-20-211
H01-20-212
H01-20-213
H02-20-125
H02-20-126
H02-20-127
H02-20-128
H02-20-129
H03-20-169
H03-20-170
H03-20-171
H03-20-173
H03-20-172
H04-20-097
H04-20-098
H04-20-099
H04-20-100
H04-20-101
H04-20-120
H04-20-121
H04-20-122
H04-20-123
H04-20-124
H05-20-109
H05-20-110
H05-20-111
H05-20-112
H05-20-113
H05-20-114
H05-20-115
H05-20-116
H05-20-117
H05-20-118
H06-20-033
H06-20-034
H06-20-035
H06-20-036
H06-20-037
H07-20-060
H07-20-061
H07-20-062
H07-20-063
H08-20-102
H08-20-103
H08-20-104
H08-20-105
H08-20-106
H08-20-107
H09-20-040
H09-20-041
H09-20-042
H09-20-043
H10-20-074
H10-20-075
H10-20-076
H10-20-077
H10-20-078
H11-20-096
H11-20-097
H11-20-098
H11-20-099
H11-20-100
H12-20-160
H12-20-161
H12-20-162
H12-20-163
H12-20-164
H01-21-105
H01-21-106
H01-21-107
H01-21-108
H01-21-110
H01-21-109
H02-21-047
H02-21-048
H02-21-049
H02-21-050
H02-21-051
H03-21-158
H03-21-159
H03-21-160
H03-21-161
H03-21-162
H05-21-137
H05-21-138
H05-21-139
H05-21-140
H06-21-148
H06-21-149
H06-21-150
H06-21-151
H06-21-152
H07-21-005
H07-21-006
H07-21-007
H07-21-008
H07-21-009
H07-21-133
H07-21-134
H07-21-135
H07-21-136
H07-21-137
H07-21-138
H12-16-115
H12-16-116
H02-17-131
H02-17-132
H02-17-133
H02-17-134
H02-17-135
H02-17-136
H03-17-118
H03-17-119
H03-17-120
H04-17-030
H04-17-031
H04-17-032
H04-17-033
H05-17-098
H05-17-099
H05-17-100
H05-17-101
H06-17-146
H06-17-147
H06-17-148
H06-17-149
H07-17-116
H07-17-117
H07-17-118
H07-17-119
H08-17-132
H08-17-133
H08-17-134
H08-17-135
H08-17-274
H08-17-275
H08-17-276
H08-17-277
H10-17-062
H11-17-109
H11-17-110
H11-17-111
H11-17-112
H12-17-125
H12-17-126
H12-17-127
H12-17-128
H12-17-129
H02-18-006
H02-18-007
H02-18-008
H02-18-009
H02-18-010
H02-18-213
H02-18-214
H02-18-215
H03-18-181
H03-18-182
H03-18-183
H04-18-158
H04-18-159
H05-18-181
H05-18-182
H05-18-183
H07-18-022
H07-18-023
H08-18-082
H08-18-083
H08-18-084
H08-18-204
H08-18-205
H10-18-114
H10-18-115
H10-18-116
H11-18-138
H11-18-139
H11-18-140
H01-19-184
H01-19-185
H01-19-186
H01-19-187
H01-19-188
H01-19-189
H01-19-190
H01-19-191
H01-19-192
H01-19-193
H01-19-194
H01-19-195
H01-19-196
H01-19-197
H01-19-198
H03-19-087
H03-19-088
H03-19-089
H04-19-124
H04-19-125
H04-19-126
H04-19-127
H06-19-318
H06-19-319
H06-19-320
H07-19-135
H07-19-136
H07-19-137
H08-19-173
H08-19-174
H08-19-175
H09-19-117
H09-19-118
H09-19-119
H10-19-109
H10-19-111
H01-20-014
H01-20-015
H01-20-016
H10-19-110
H01-20-214
H01-20-215
H01-20-216
H01-20-217
H03-20-174
H03-20-175
H03-20-176
H03-20-177
H05-20-119
H05-20-120
H05-20-121
H06-20-038
H06-20-039
H07-20-064
H07-20-065
H09-20-037
H09-20-038
H11-20-101
H11-20-102
H12-20-165
H12-20-166
H01-21-111
H01-21-112
H03-21-064
H03-21-065
H07-21-143
H07-21-144
|
| FEI Number |
1220592
|
Recalling Firm/ Manufacturer |
Hobbs Medical, Inc.
8 Spring St
Stafford Springs
CT
06076-1505
|
Manufacturer Reason for Recall | The labeling for the Microbiology Brush identifies the device as a Bronchial Microbiology Brush. The device is only cleared for gastrointestinal use and is not cleared for bronchial use. |
FDA Determined Cause 2 | No Marketing Application |
| Action | URGENT MEDICAL DEVICE RECALL notification letters were sent to customers beginning 12/28/21.
Actions requested on your part:
1. Read the Description of the problem section.
2. Immediately identify and set aside all product listed herein in a manner that ensures the
affected product will not be used and will be returned to Hobbs Medical.
3. Review, complete, and return the enclosed Acknowledgement Form directly to Hobbs
Medical Customer Service, return via Fax at 860-684-7574, or email to
CustomerService@HobbsMedical.com.
4. Pass on this notice to anyone in your facility that needs to be informed according to
your internal notification procedures.
5. If any product listed below has been forwarded to another facility, contact that facility
to arrange return.
6. Maintain awareness of this notice until all product listed above has been returned to
Hobbs Medical.
The following assistance is available:
-For questions related to the issue or product replacement, please contact Hobbs
Medical Customer Service.
-For questions related to product return please contact Hobbs Medical Customer Service
Return instructions are provided on the Acknowledgement Form.
-For questions related to the Acknowledgement Form and its return, please contact
Hobbs Customer Service at 1-800-344-6227 |
| Quantity in Commerce | 38,357 devices |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FDX
|
|
|
|