| Date Initiated by Firm | May 26, 2021 |
| Date Posted | May 03, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0886-2022 |
| Recall Event ID |
89964 |
| 510(K)Number | K172822 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Azurion systems with software release R1.x |
| Code Information |
software release R1.x |
| FEI Number |
3002807880
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 4-6
Best
Netherlands
|
| For Additional Information Contact | Dusty Leppert, CHP 978-2280190 |
Manufacturer Reason for Recall | In the Azurion system, the user can add a new study to a patient by selecting the option Add
Study . The Add Study dialogue box is then displayed where the Patient Type is selected to
perform the study. Due to a software defect, when the study is initiated by pressing Start
Procedure , the Patient Type changes inadvertently to a Patient Type different than the one
selected as shown in the Table below. Patient type is one of the factors involved in the dose
control process. The incorrect patient type changes the technique factors to be used by the
system without notification to the user. |
FDA Determined Cause 2 | Software design |
| Action | Philips notified customers of the issue via Electronic Product Radiation Defect notification letter dated March 4, 2022. The letter identified the affected product, problem and actions to be taken. Customers were informed that they will be contacted by a Philips Field Service Engineers to schedule the software update at no cost. For questions contact your local Philips representative: 1-800-722-9377. |
| Quantity in Commerce | 722 systems |
| Distribution | US Nationwide and Worldwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|